← Product Code [PMC](/productcode/PMC) · K252289

# Eva Scalp Cooling System (K252289)

_Stemtech Medical Devices Private Limited · PMC · Nov 25, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252289

## Device Facts

- **Applicant:** Stemtech Medical Devices Private Limited
- **Product Code:** [PMC](/productcode/PMC.md)
- **Decision Date:** Nov 25, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4360
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Eva Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.

## Device Story

Eva Scalp Cooling System is a cooling unit with a built-in control system and touch-screen interface used in hospitals and healthcare facilities. The device circulates temperature-controlled fluid (propylene glycol and water mixture) between a refrigeration unit and a cooling cap worn by the patient. The system maintains a steady scalp temperature between -3°C and 1°C. Treatment time is adjustable by a clinician based on cancer and drug types, covering periods before, during, and after chemotherapy. The device features two cooling lines and caps. By cooling the scalp, the device aims to reduce chemotherapy-induced alopecia. Healthcare providers operate the system via an integrated touch-screen display to monitor and control the cooling process.

## Clinical Evidence

No clinical data was necessary to determine substantial equivalence. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC 60601-4-2), software verification/validation (IEC 62304), shelf-life testing, transportation testing (ASTM D4169), and performance testing (thermal stability, flow rate, leakage, and cleaning validation).

## Technological Characteristics

System includes a refrigeration unit, touch-screen controller, coolant tank, and two cooling caps. Cooling cap outer cover is neoprene; inner components use thermoplastic polyurethane (TPU). Cooling principle: circulating temperature-controlled fluid (propylene glycol/water mixture) between -3°C and 1°C. Sensors located in the coolant reservoir regulate fluid temperature. Dimensions: 1140x460x480mm; Weight: 80kg. Power: 230V, 50/60Hz. Connectivity: touch-screen interface. Sterilization: non-sterile. Software-controlled operation.

## Regulatory Identification

A scalp cooling system to reduce the likelihood of chemotherapy-induced alopecia is a prescription device intended to reduce the frequency and severity of alopecia during chemotherapy in which alopecia-inducing chemotherapeutic agents are used.

## Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use. This information must include testing to demonstrate accuracy of the temperature control mechanism.
(2) Performance testing must demonstrate the electromagnetic compatibility and electrical safety of the device.
(3) Software verification, validation, and hazard analysis must be performed.
(4) The patient contacting components of the device must be demonstrated to be biocompatible. Material names must be provided.
(5) Labeling must include the following:
(i) A statement describing the potential risk of developing scalp metastasis.
(ii) Information on the patient population and chemotherapeutic agents/regimen for which the device has been demonstrated to be effective.
(iii) A summary of the non-clinical and/or clinical testing pertinent to use of the device.
(iv) A summary of the device technical parameters, including temperature cooling range and duration of cooling.
(v) A summary of the device- and procedure-related adverse events pertinent to use of the device.
(vi) Information on how the device operates and the typical course of treatment.
(6) Patient labeling must be provided and must include:
(i) Relevant contraindications, warnings, precautions, and adverse effects/complications.
(ii) Information on how the device operates and the typical course of treatment.
(iii) Information on the patient population for which there is clinical evidence of effectiveness.
(iv) The potential risks and benefits associated with use of the device.
(v) Postoperative care instructions.
(vi) A statement describing the potential risk of developing scalp metastasis.

## Predicate Devices

- DigniCap Delta Scalp Cooling System ([K191166](/device/K191166.md))

## Reference Devices

- DigniCap Scalp Cooling System ([K170871](/device/K170871.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 25, 2025

Stemtech Medical Devices Private Limited
% Giselle Zhang
Lead QA&amp;RA Consultant
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, Texas 78746

Re: K252289
Trade/Device Name: Eva Scalp Cooling System
Regulation Number: 21 CFR 878.4360
Regulation Name: Scalp Cooling System To Reduce The Likelihood Of Chemotherapy-Induced Alopecia
Regulatory Class: Class II
Product Code: PMC
Dated: July 23, 2025
Received: July 23, 2025

Dear Giselle Zhang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252289 - Giselle Zhang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252289 - Giselle Zhang
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252289  |   |
|  Device Name Eva Scalp Cooling System  |   |
|  Indications for Use (Describe) The Eva Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

Page 1 of 5

# 510(k) Summary

# Eva Scalp Cooling System

## 1. Submission Sponsor

Stemtech Medical Devices Private Limited

411, 4th Floor, Tesla Industrial Building No-1 Tesla, Plot No Gen-2/1/c (Part), Of Trans Thane Creek Industrial Area Shiravane-Juinagar, Navi Mumbai, Tal. &amp; District. Thane, Thane, Maharashtra 400706

India

Subrai Pai, Chief Technology Officer

Email: subrai@stemtech.in

Phone number: +91 7977012055

## 2. Submission Correspondent

Emergo by UL

2500 Bee Cave Road

Building 1, Suite 300

Austin, TX 78746

Office Phone: (512) 327-9997

Email: LST.US.EmergoFDASubmissions@ul.com

Contact: Giselle Zhang

Title: Lead QA&amp;RA Consultant

## 3. Date Prepared

04/15/2025

## 4. Device Identification

Trade/Proprietary Name: Eva Scalp Cooling System

Common/Usual Name: Scalp Cooling System

Regulation Number(s): 878.4360

Product Code(s): PMC, Scalp cooling system to reduce the likelihood of chemotherapy-induced alopecia.

Class: II

Classification Panel: General &amp; Plastic Surgery

## 5. Legally Marketed Predicate Device(s)

Page 1 of 5

{5}

Page 2 of 5

Device name: DigniCap Delta Scalp Cooling System

510(k) number: K191166

Manufacturer: Dignitana, Inc.

The DigniCap Delta Scalp Cooling System ("DigniCap Delta") is a device that is intended to function as a cooling system to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The proposed therapy of the new device is comparable to the therapy of the predicate device, DigniCap Scalp Cooling System (K170871).

DigniCap Delta cools fluid to a prescribed set temperature and circulates that cooled fluid through a cooling wrap and then back to the device. This mode of operation is equivalent to other scalp cooling devices. User operation of the DigniCap Delta system is carried out on the illuminated graphics display with the integrated touch control, located on the front of the unit. The navigation and selection buttons on the display are used to interface with the device.

## 6. Device Description

The Eva Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. The Eva Scalp Cooling System is a cooling unit with a built-in control system. It can be operated using a touch-screen display. The device maintains a cooling temperature between -3°C and 1°C, and the treatment time can be adjusted by the doctor for before, during, and after chemotherapy. The treatment time also can be adjusted by doctor for different cancer types and drug types. The system features a computerized control unit that keeps the scalp temperature steady during treatment.

## 7. Indication for Use Statement

The Eva Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors.

## 8. Substantial Equivalence Discussion

The following table compares the Eva Scalp Cooling System to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

Comparison of Characteristics

|  Attribute | Subject: Eva Scalp Cooling System | Predicate: DigniCap Delta Scalp Cooling System | Comparison  |
| --- | --- | --- | --- |
|  510(k) Number | / | K191166 |   |
|  Product Code | PMC | PMC | Same  |
|  Regulation Number | §878.4360 | §878.4360 | Same  |
|  Regulation Name | Scalp Cooling System | Scalp Cooling System | Same  |

Page 2 of 5

{6}

Page 3 of 5

|  Attribute | Subject: Eva Scalp Cooling System | Predicate: DigniCap Delta Scalp Cooling System | Comparison  |
| --- | --- | --- | --- |
|  Indications for Use | The Eva Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. | The DigniCap Delta Scalp Cooling System is indicated to reduce the likelihood of chemotherapy-induced alopecia in cancer patients with solid tumors. | Same  |
|  Designed for Use | Hospitals and healthcare facilities | Hospitals and healthcare facilities | Same  |
|  Environment of Use | Ambient temperature 10 - 30°C, relative humidity 30 - 75% | Temperature: 10°C – 25°C
Relative Humidity: 30% to 60%, non-condensing | Different  |
|  Mechanism of Action | The Eva Scalp Cooling System transports temperature-controlled cooled fluid from the device to a cooling cap to cool the patient’s scalp, thereby reducing chemotherapy-induced alopecia. | The DigniCap Delta Scalp Cooling System transports temperature-controlled cooled fluid from the device to a cooling cap to cool the patient’s scalp, thereby reducing chemotherapy-induced alopecia. | Same  |
|  Technology Overview | The unit is composed of the main unit that contains the refrigeration components, touch screen controller, and coolant tank. There are detachable coolant lines with covers, detachable cooling cap with covers, and proprietary coolant. | The DigniCap Delta Scalp Cooling System consists of a solid-state thermoelectric cooling unit with integral control system operated via a touchscreen monitor, capable of precisely controlling the temperature of one cooling cap. | Different  |
|  Patient Population | Chemotherapy Patients with Solid Tumors. | Chemotherapy Patients with Solid Tumors. | Same  |
|  Set Cooling Time | Yes | Yes | Same  |
|  Pre/Post Cooling Time | Yes | Yes | Same  |
|  Material of Cooling Cap | Neoprene (Outer Cover) | Neoprene (Outer Cover) | Same  |
|   |  Thermoplastic polyurethane (TPU) | Polyacrylic, nylon | Different  |
|  Size of Cooling Cap | Adjustable | Adjustable | Same  |
|  Inner Cooling Cap | With adjustable tabs
Single Patient Use | With adjustable tabs
Single Patient Use | Same  |
|  Outer Cooling Cap | Single Patient Use | Single Patient Use | Same  |
|  Number of Cooling Caps/Lines | 2 | 1 | Different  |

Page 3 of 5

{7}

|  Attribute | Subject: Eva Scalp Cooling System | Predicate: DigniCap Delta Scalp Cooling System | Comparison  |
| --- | --- | --- | --- |
|  Quick Disconnect | Yes | Yes | Same  |
|  Coolant Temperature Range | -3 to 1°C | -7 to 1.5°C | Different  |
|  Sensors | The temperature of the fluid is regulated by temperature sensors located in the coolant reservoir. | The temperature of the fluid is regulated by temperature sensors located in the fluid lines. | Different  |
|  Refrigerant Type | Proprietary mixture (Propylene glycol +water) | Isopropyl alcohol/water mixture | Different  |
|  Coolant Refilling | Yes | Yes | Same  |
|  Sterile | No | No | Same  |
|  Single-Use | No | No | Same  |
|  Main Unit Dimensions | 1140(L)x460(H)x480(W)mm | 952.5(L)x 508 (D) x 381 (W) mm | Different  |
|  Weight | 80 kg (Eva Cooling Unit) | 65kg | Different  |
|  AC Powered | 230 V, 50/60 Hz | 100-240 V, 50/60 Hz | Different  |
|  Touch Screen Interface | Yes | Yes | Same  |
|  Software Controlled | Yes | Yes | Same  |
|  Use Life | 12 sessions | NA | Different  |
|  Battery Operated | NA | Yes | Different  |
|  Complies with ISO 10993-1 | Pass | Pass | Same  |
|  Electrical Safety Testing | Pass | Pass | Same  |
|  Electromagnetic Compatibility Testing | Pass | Pass | Same  |

# 9. Non-Clinical Performance Data

To demonstrate safety and effectiveness of Eva Scalp Cooling Cap and to show substantial equivalence to the predicate device, Stemtech completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The Eva Scalp Cooling passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:

Cytotoxicity testing per ISO 10993-5 - Passed
- Sensitization testing per ISO 10993-10 - Passed
- Irritation testing per ISO 10993-23 - Passed
- Electrical safety testing per IEC 60601-1 - Passed
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2 and IEC 60601-4-2 - Passed

{8}

Page 5 of 5

- Software verification and validation per IEC 62304/FDA Guidance – results /conclusion
- Shelf-Life Testing – Supports shelf life of 1 years
- Transportation Testing per ASTM D4169 – Demonstrates package integrity maintained
- Thermal Performance
- Stable Fluid Temperature Verification
- Flow Rate Verification
- Fluid Containment and Leakage Test
- System Integration Test
- Cleaning Validation Test

## 10. Clinical Performance Data

No clinical data was necessary to determine the substantial equivalence of this device.

## 11. Statement of Substantial Equivalence

The Eva Scalp Cooling Cap has the same intended use as the predicate device and similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Based on the results of the risk assessment and all applicable testing, the Eva Scalp Cooling Cap is determined to perform as intended during normal anticipated usage and is at least as safe, as effective, and performs as well as or better than the legally marketed predicate device and therefore can be determined to be substantially equivalent to the predicate device.

Page 5 of 5

---

**Source:** [https://fda.innolitics.com/device/K252289](https://fda.innolitics.com/device/K252289)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
