K252260 · Li Medical Corporation , Ltd. · LKF · Nov 26, 2025 · Obstetrics/Gynecology
Device Facts
Record ID
K252260
Device Name
RELIEEV HSG Catheter (HSG7FA1)
Applicant
Li Medical Corporation , Ltd.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Nov 26, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The RELIEEV HSG Catheter (HSG7FA1) is for the delivery of contrast media or saline into the uterine cavity during hysterosalpingography or sonohysterography for examination of the uterus and fallopian tubes.
Device Story
The RELIEEV HSG Catheter is a 7 French dual-lumen polyurethane catheter used for hysterosalpingography or sonohysterography. It features a silicone balloon (3.5 mL capacity) and a proximal double-lumen hub with a polycarbonate stopcock. The device is inserted manually by a physician; the balloon is inflated via a syringe to seal the cervix, and contrast media or saline is injected through the center lumen into the uterine cavity. The device is single-use and supplied sterile. It facilitates the visualization of the uterus and fallopian tubes, aiding in the diagnosis of uterine pathology. The catheter is designed for limited mucosal contact (≤24 hours).
Clinical Evidence
Bench testing only. Testing included balloon diameter/volume, failure testing, lumen patency, liquid leakage, and tensile strength. Biocompatibility testing (cytotoxicity, sensitization, vaginal irritation) per ISO 10993 confirmed non-cytotoxic, non-sensitizing, and non-irritant status. Sterilization validated per ISO 11135. Shelf-life and packaging integrity validated per ASTM D4332, D4169, F1980, and ISO 11607.
Technological Characteristics
7 Fr dual-lumen catheter; materials: TPU 75D tube, silicone balloon, polypropylene guide sleeve. Manual operation via syringe inflation. Sterilized via Ethylene Oxide (EO) to SAL 10⁻⁶. Shelf life: 3 years. Packaging: medical grade paper pouch and PET film.
Indications for Use
Indicated for patients undergoing hysterosalpingography or sonohysterography for examination of the uterus and fallopian tubes. Contraindicated in cases of pregnancy, suspected pregnancy, active pelvic infections, recent pelvic infections, severe uterine bleeding, gynecologic malignancies, or allergy to contrast media.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 26, 2025
LI Medical Corporation Ltd.
Jago Chen
CEO
2F., No. 43, Zhongxing Rd.,
Xizhi Dist., New Taipei City 221012
TAIWAN
Re: K252260
Trade/Device Name: RELIEEV HSG Catheter (HSG7FA1)
Regulation Number: 21 CFR 884.4530
Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II
Product Code: LKF
Dated: July 18, 2025
Received: October 27, 2025
Dear Jago Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252260 - Jago Chen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K252260 - Jago Chen
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K252260
Device Name
RELIEEV HSG Catheter (HSG7FA1)
Indications for Use (Describe)
The RELIEEV HSG Catheter (HSG7FA1) is for the delivery of contrast media or saline into the uterine cavity during hysterosalpingography or sonohysterography for examination of the uterus and fallopian tubes.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{5}
K252260
Page 1 of 8
510(k) Summary – K252260
I. SUBMITTER INFORMATION
Submitter: LI Medical Corporation LTD.
2F., No. 43, Zhongxing Rd., Xizhi Dist.,
New Taipei City 221012, Taiwan
Submission: Jago Chen, CEO
Correspondent: 2F., No. 43, Zhongxing Rd., Xizhi Dist.,
New Taipei City 221012, Taiwan
Phone: +886-2-86461999
Fax: +886-2-26471999
Email: jago.chen@li-med.com.tw
Date Summary Prepared: November 25, 2025
II. DEVICE INFORMATION
Device Trade Name: RELIEEV HSG Catheter (HSG7FA1)
Common Name: HSG Catheter
Regulation Number: 21 CFR 884.4530
Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument
Regulatory Class: Class II
Product Code: LKF (Cannula, Manipulator/Injector, Uterine)
III. PREDICATE DEVICE INFORMATION
| Predicate Device: | PANPAC HSG Catheter Set |
| --- | --- |
| Company Name: | PANPAC |
| Submission number: | K092983 |
The predicate device has not been subject to a design-related recall.
{6}
K252260
Page 2 of 8
# IV. DEVICE DESCRIPTION
The RELIEEV HSG Catheter (HSG7FA1) is a 7 French (Fr) dual lumen polyurethane catheter with a working length of 30 cm. RELIEEV HSG Catheter (HSG7FA1) is available with a closed-end tip with two side ports. The catheter consists of two lumens: an open lumen used to inflate the balloon and an open lumen used to inject diagnostic media into the uterus. The silicone balloon is designed with a balloon volume of 3.5 mL. When the balloon is inflated inside the uterus, the subject device seals the cervix. The catheter features a proximal fitting of a double-lumen polyurethane hub with a polycarbonate stopcock attached. Both the hub and stopcock accept a standard Luer lock. A stopcock is provided at the proximal end of the device to allow inflation of the balloon with a syringe. The center lumen is open through the device to the distal end, and the device has a luer connector at the proximal end for injecting fluids. An introducing sheath is placed over the tube as a guide and into the endocervix to the level of the internal os (opening of the cervix) with no dilatation required; therefore, allowing the catheter to pass easily through the cervix and into the uterine cavity.
The RELIEEV HSG Catheter (HSG7FA1) consists of a balloon catheter and a 5 mL syringe. The subject device is supplied sterile and is single-use.
# V. INDICATIONS FOR USE
The RELIEEV HSG Catheter (HSG7FA1) is for the delivery of contrast media or saline into the uterine cavity during hysterosalpingography or sonohysterography for examination of the uterus and fallopian tubes.
# VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE AND PREDICATE DEVICE
The table below compares the intended use and technological characteristics of the subject and predicate device.
Table 1. Comparison to Predicate Device
| Device Attribute | Proposed Device
RELIEEV HSG Catheter (HSG7FA1)
(Subject Device: K252260) | Predicate Device
PANPAC HSG Catheter Set
(Predicate Device: K092983) | Comparison |
| --- | --- | --- | --- |
| Classification | Intrauterine Catheter | Intrauterine Catheter | Same |
| Regulation | Class II, 21 CFR | Class II, 21 CFR | Same |
{7}
K252260
Page 3 of 8
| Product code | 884.4530 | 884.4530 | |
| --- | --- | --- | --- |
| Indications for Use | The RELIEEV HSG Catheter (HSG7FA1) is for the delivery of contrast media or saline into the uterine cavity during hysterosalpingography or sonohysterography for examination of the uterus and fallopian tubes. | PANPAC HSG Catheter Sets are indicated to evaluate the causes of abnormal uterine bleeding, menstrual disorders, recurring pregnancy loss, or unexplained infertility. They can also be used to assess uterine pathology and patients on tamoxifen therapy.
HSG type catheter sets are used to infuse a fluid (either a contrast media or a sterile saline) into the uterine cavity while blocking the external cervical as to retain the fluid in the uterus during the procedure. | *Different*: The differences in the Indications for Use statement between the subject device and predicate device do not alter the intended use of the subject device. |
| Manufacturer | LI Medical Corporation LTD. | PANPAC Medical Corporation | - |
| Device name | RELIEEV HSG Catheter (HSG7FA1) | PANPAC HSG Catheter Set | - |
| Prescription medical device | YES | YES | Same |
| Single patient use | YES | YES | Same |
| Reusable | NO | NO | Same |
| Contraindications | HSG procedures are contraindicated in cases of pregnancy, suspected pregnancy, | HSG procedures are contraindicated in cases of pregnancy, suspected pregnancy, | Similar |
{8}
K252260
Page 4 of 8
| | active pelvic infections, recent pelvic infections, severe uterine bleeding, gynecologic malignancies, or if the patient is allergic to the contrast media. This device should not be used for intrafallopian procedures. | active pelvic infections, recent pelvic infections, severe uterine bleeding. | |
| --- | --- | --- | --- |
| Technical Characteristics | | | |
| Lumen Number | 2, dual lumen (for contrast and air inflated balloon) | 2, dual lumen (for contrast and air inflated balloon) | Same |
| Catheter Tip Diameter | 7 Fr | 5 Fr
7 Fr | *Different*: The differences in the catheter tip diameters are supported by testing (e.g., performance testing). |
| Size of Catheter Tube | O.D. (7.0 French (Fr)) | O.D. (7.0 French (Fr)) | Same |
| ID of Tube (Air) | Ø0.8mm | Ø0.8mm | Same |
| ID of Tube (Liquid) | 1.3*0.8mm | 1.3*0.8mm | Same |
| Catheter Tip Shape | Closed and round | Closed and round | Same |
| Working Length | 30 cm | 30 cm | Same |
| TPU Tube flexible | YES | YES | Same |
| Capacity of Inflated Balloon | 3.5 mL | 5 Fr 1.5 ml
7 Fr 3.0 ml | *Different*: The differences in the balloon volume are |
{9}
K252260
Page 5 of 8
| | | | supported by testing (e.g., performance testing). |
| --- | --- | --- | --- |
| Balloon Shape | Spherical | Spherical | Same |
| Operation Principle | Manual insert and inflate the balloon with a sterile syringe by a trained physician. | Manual insert and inflate the balloon with a sterile syringe by a trained physician. | Same |
| Removal Mechanism | Deflate the balloon with a syringe and remove the catheter. | Deflate the balloon with a syringe and remove the catheter. | Same |
| Media Expelled (Contrast or Saline Injection) | Via catheter side port | Via catheter side port | Same |
| Materials | Catheter tube: Thermoplastic polyurethane (TPU) 75D; Balloon material: Silicone; Guide Sleeve: Polypropylene (PP) | Catheter tube: Polyurethane (PU) 75D; Balloon material: Silicone; Guide Sleeve: Polypropylene (PP) | *Different*: The differences in the materials are supported by testing (e.g., biocompatibility testing, shelf-life testing, performance testing, etc.). |
| Biocompatibility | Surface device, in contact with mucosal tissue, with limited contact (≤24h) | Surface device, in contact with mucosal tissue, with limited contact (≤24h) | Same |
| Single Patient Use | YES | YES | Same |
| Reusable | NO (Single use) | NO (Single use) | Same |
| Anatomical Site | Uterine cavity | Uterine cavity | Same |
| Target Patient | Patient undergoing SHG or HSG | Patient undergoing SHG or HSG | Same |
{10}
K252260
Page 6 of 8
| | procedure | procedure | |
| --- | --- | --- | --- |
| Clinical purpose on Clinical environment | The catheter is to be used for sonohysterography procedure that aid in the analysis of uterine pathology. | The catheter is to be used for sonohysterography procedure that aid in the analysis of uterine pathology. | Same |
| Sterilized Method | Ethylene Oxide (EO) gas | Ethylene Oxide (EO) gas | Same |
| Packaging | Medical grade paper pouch and PET film | Dupont Tyvek | **Different:** The differences in the packaging material are supported by testing (e.g., biocompatibility testing, shelf-life testing, etc.). |
| Shelf life | 3 years | 3 years | Same |
The indications for use of the subject and predicate device are not identical. However, the intended use of the subject and predicate is the same (i.e., delivery of contrast medium or saline to the uterine cavity for HSG and SHG procedures). The subject and predicate device have the same designs and dimensions, shelf life, and sterilization method. The predicate and subject device differ in catheter tip sizes, tube material, maximum balloon inflation, and packaging materials. The differences identified do not raise different questions of safety and effectiveness as compared to the predicate device.
## VIII. PERFORMANCE DATA
The following testing was performed to demonstrate that the RELIEEV HSG Catheter (HSG7FA1) met applicable design and performance requirements. All predetermined acceptance criteria were met in the following tests:
### Bench Testing
Physical bench testing was performed to confirm that the device performs according to the product specifications. Device evaluation consisted of physical and functional testing including:
- Balloon Diameter and Maximum Balloon Volume Testing,
{11}
K252260
Page 7 of 8
- Balloon Failure Test,
- Lumen Patency and Liquid Leakage Test:
- Tensile Testing (luer to tube, stopcock to tube, balloon adhesive catheter tube)
The subject device test results were compared with the predicate device and passed all functional bench testing.
## Biocompatibility Testing
Per the indications for use, the RELIEEV HSG Catheter (HSG7FA1) is categorized as a surface device contacting mucosal membranes for a limited duration (≤ 24 hours). According to the 2023 FDA Guidance Document, Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", the following endpoints were testing for the subject device:
| Test Description | Method | Results |
| --- | --- | --- |
| Cytotoxicity | ISO 10993-5: 2009, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity | Non-cytotoxic |
| Sensitization | ISO 10993-10: 2021, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization | Non-sensitizer |
| Vaginal Irritation | ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation | Non-irritant |
## Sterilization Validation
The subject device is sterilized via Ethylene Oxide (EO) to achieve a Sterility Assurance Level (SAL) of 10⁻⁶. Sterilization validation was conducted as per ISO 11135:2014 Sterilization of health-care products – Ethylene oxide requirement for the development, validation, and routine control of a sterilization process for medical devices. Sterility testing and documentation were provided as outlined in the 2024 FDA Guidance Document, Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile.
## Shelf Life
The RELIEEV HSG Catheter (HSG7FA1) has a shelf life of 3 years (36-months). Shelf life studies have been conducted to demonstrate that the subject device can maintain functionality and
{12}
K252260
Page 1 of 8
the packaging in will maintain its sterile barrier over the entirety of the intended shelf life.
The devices were subjected to environmental conditioning per ASTM D4332-22:2022, Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing, and simulated shipping distribution per ASTM D4169-22: 2022, Distribution Cycle 2, Standard Practice for Performance Testing of Shipping Containers and Systems. Sterile barrier packaging testing was performed per ISO 11607-1:2019, Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile barrier systems and packaging systems, and ISO 11607-2:2019, Packaging for Terminally Sterilized Medical Devices -Part 2: Validation requirements for forming, sealing and assembly processes. Following accelerated aging, per ASTM F1980-21: 2023, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, and simulated shipping distribution, the following package integrity tests were completed:
- Visual Inspection per ASTM F1886/F1886M-16: 2017, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- Seal Strength per ASTM F88/F88M-23: 2023, Standard Test Method for Seal Strength of Flexible Barrier Materials
- Dye Penetration per ASTM F1929-23: 2024, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
## IX. CONCLUSION
The results of the testing described above demonstrate that the RELIEEV HSG Catheter (HSG7FA1) is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.