The Aura Glide is an over-the-counter handheld device for aesthetic purposes. (1) The microcurrent mode is indicated for facial stimulation. (2) The LED mode is intended for the treatment of periorbital wrinkles (red light) and for the treatment of mild to moderate inflammatory acne (blue light).
Device Story
Aura Glide is a lemon-shaped, handheld, battery-powered cosmetic device for home use. It features a main control body with two detachable attachments: an LED attachment for light therapy and a microcurrent attachment for facial stimulation. Users apply a conductive primer gel for microcurrent treatments, which offers three intensity levels (Gentle, Standard, Strong). The device uses pulsed biphasic, modulated square waveforms for stimulation and specific wavelengths (633nm red, 415nm blue) for light therapy. Operation is controlled by an internal microprocessor with preset treatment times (8 minutes for microcurrent, 10 minutes for LED) and automatic shut-off. The device is charged via USB-C. It is intended for consumer self-use to improve facial appearance; the healthcare provider does not interact with the device output, as it is an OTC aesthetic product.
Clinical Evidence
Bench testing only. No clinical data was conducted or required to support substantial equivalence. Safety and performance were validated via IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-10, 60601-2-57, 62366-1, 62133-2, and 62471. Biocompatibility was confirmed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-23 (irritation).
Technological Characteristics
Handheld device; ABS plastic and chrome-plated metal housing. Power: 3.7V 600mAh rechargeable Li-ion battery. Modalities: Microcurrent (pulsed biphasic, modulated square waveform) and LED (633nm red, 415nm blue). Connectivity: USB-C charging. Safety: Type BF isolation. Standards: IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-10, 60601-2-57, 62471, 62366-1, 62133-2; ISO 10993-5, 10, 23.
Indications for Use
Indicated for OTC aesthetic facial stimulation (microcurrent mode), treatment of periorbital wrinkles (red LED), and treatment of mild to moderate inflammatory acne (blue LED).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2025
Aura Medical LLC
% Gina Walljasper
Principal Consultant
GPW Enterprises LLC
47 Gatehouse Rd
Bedminster, New Jersey 07921
Re: K252187
Trade/Device Name: Aura Glide (FC40)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO, OHS, OLP
Dated: November 24, 2025
Received: November 24, 2025
Dear Gina Walljasper:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252187 - Gina Walljasper
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252187 - Gina Walljasper
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Aura Glide
Page 11 of 54
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252187 | ? |
| Please provide the device trade name(s). | | ? |
| Aura Glide (FC40) | | |
| Please provide your Indications for Use below. | | ? |
| The Aura Glide is an over-the-counter handheld device for aesthetic purposes.
(1) The microcurrent mode is indicated for facial stimulation.
(2) The LED mode is intended for the treatment of periorbital wrinkles (red light) and for the treatment of mild to moderate inflammatory acne (blue light). | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
# 1. SUBMITTER/OWNER:
ALC Enterprises Inc (DBA Aura Medical LLC)
134 Stockton St, 1C
Brooklyn, NY 11206
Contact: Connie Lefkowitz, Managing Partner of Aura Medical LLC
2 Skillman St., Suite 512
Brooklyn, NY 11205
FEI Number: 3011721071
Telephone: (347) 831-2215
Email: connie@aura-medical.com
# Application Correspondent:
Gina Walljasper, Principal Consultant
GPW Enterprises LLC
47 Gatehouse Rd
Bedminster, NJ 07921
Tel: (908) 507-6503
Email: gina@walljasperpc.com
Date prepared: December 22, 2025
# 2. DEVICE NAME:
| Trade Name | Aura Glide |
| --- | --- |
| Common Name | Stimulator, Transcutaneous Electrical, Aesthetic Purposes; Light Based Over-the-Counter Wrinkle Reduction; Over-the-Counter Powered light based laser for acne. |
| Classification Name | Stimulator, Transcutaneous Electrical, Aesthetic Purposes; Light Based Over-the-Counter Wrinkle Reduction; Over-the-Counter Powered light based laser for acne. |
| Device Classification | Class II |
| Review Panel | Neurology; General & Plastic Surgery |
| 510(k) Number | K252187 |
| Product Code(s) | NFO, OHS, OLP |
| Regulation Number | 21CFR882.5890, 21CFR878.4810 |
{5}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
3. PREDICATE DEVICE(S): Substantial equivalence is based on following legally marketed device.
| Sponsor | Heat In Click |
| --- | --- |
| Device Name and Model | 2 Face/Face Evolution |
| 510(k) Number | K171821 |
| Regulation Name | Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| Product Code(s) | NFO, OHS, OLP |
| Regulation Number | 21CFR 882.5890, 21CFR 878.4810 |
| Review Panel | Neurology, General & Plastic Surgery |
| Regulation Class | Class II |
4. DESCRIPTION OF THE DEVICE:
Aura Glide is a lemon shaped handheld device that consists of:
- A main control body that is programmed with preset treatment modes depending on the application at the time of use.
- Two (2) essential attachments that are detachable. Each attachment is specifically designed to sit securely on the main body to ensure proper functionality and stability during use.
- LED (Light Emitting Diode) Attachment – intended for the treatment of periorbital wrinkles (red light) and for the treatment of mild to moderate acne (blue light).
- Microcurrent Attachment – indicated for facial stimulation. Microcurrent can be delivered with three (3) levels of intensity; L1 (Gentle; Default), L2 (Standard), and L3 (Strong) depending on user preference.
- Conductive gel (Glide On Primer gel) must be applied prior to starting microcurrent treatment. It enhances conductivity ensuring safe and effective microcurrent treatment.
- The Aura Glide is equipped with a USB-C charging port located on the bottom of the device and an internal rechargeable battery. A USB-C charging cable is provided with every device.
The Aura Glide is an over-the-counter cosmetic device for home use.
5. INDICATIONS FOR USE:
The Aura Glide is an over-the-counter handheld device for aesthetic purposes.
(1) The microcurrent mode is indicated for facial stimulation.
(2) The LED mode is intended for the treatment of periorbital wrinkles (red light) and for the treatment of mild to moderate inflammatory acne (blue light).
Page | 2 of 8
{6}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
## 6. TESTING SUMMARY:
Aura Glide has been evaluated for safety and performance has been conducted by laboratory testing for the following:
- IEC 60601-1:2005 + A1:2012 +A2: 2020
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 + A1:2020
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]
- IEC 60601-1-11:2015 + A1:2020
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-10:2012 +A1:2016
Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- IEC 60601-2-57:2023
Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- IEC TR 60601-4-2:2016
Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical system
- IEC 62366-1:2020
Medical devices - Part 1: Application of usability engineering to medical devices
- IEC 62133-2:2017
Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- IEC 62471 First edition 2006-07
Photobiological safety of lamps and lamp systems
- ISO 10993-5:2009
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021
Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021
Biological evaluation of medical devices - Part 23: Tests for irritation
Page | 3 of 8
{7}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
Animal Testing – Animal testing is not necessary to support Substantial Equivalence and therefore no animal testing was conducted.
Clinical Testing – Clinical testing is not needed for this premarket notification to support substantial equivalence.
# 7. COMPARISON WITH PREDICATE DEVICE:
Comparison in Detail(s):
| Element of Comparison | Subject Device (Aura Glide - K252187) | Predicate Device (2 Face/Face Evolution - K171821) | Remark |
| --- | --- | --- | --- |
| Manufacturer | Aura Medical | Heat in a Click | – |
| 510Knumber | K252187 | K171821 | – |
| ProductName | Aura Glide | 2 Face/Face Evolution | – |
| Classification Name | Stimulator, Transcutaneous Electrical, Aesthetic Purposes; Light Based Over-the-Counter Wrinkle Reduction, Over-the-Counter Powered light-based laser for acne | Stimulator, Transcutaneous Electrical, Aesthetic Purposes; Light Based Over-the-Counter Wrinkle Reduction, Over-the-Counter Powered light-based laser for acne | SE |
| Regulation Class | II | II | SE |
| Regulation Number | 21CFR882.5890, 21CFR 878.4810 | 21CFR882.5890, 21CFR 878.4810 | SE |
| Review Panel | Neurology; General & Plastic Surgery | Neurology; General & Plastic Surgery | SE |
| Product Code | NFO, OHS, OLP | NFO, OHS, OLP | SE |
| OTC & Rx | OTC | OTC | SE |
| Indications for Use | | | |
| Indications for Use | The Aura Glide is an over-the-counter handheld device for aesthetic purposes. (1) The microcurrent mode is indicated for facial stimulation. (2) The LED mode is intended for the treatment of periorbital wrinkles (red light) and for the treatment of mild to moderate inflammatory acne (blue light). | 2 Face/Face Evolution is a handheld device for over-the-counter aesthetic purposes. (1) The EMS mode is indicated for facial stimulation. (2) The Photon mode: The red light is intended for the treatment of periorbital wrinkles and the blue light is intended for the treatment of the mild to moderate inflammatory acne | SE |
| Treatment Area | Face | Face | SE |
{8}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
| Element of Comparison | Subject Device (Aura Glide - K252187) | Predicate Device (2 Face/Face Evolution - K171821) | Remark |
| --- | --- | --- | --- |
| Device Design | | | |
| Design | Hand-held | Hand-held | SE |
| Environment of Use | Home | Home | SE |
| Method of Line Current Isolation | Type BF | Type BF | SE |
| Main Unit Weight | 198 g | 200 g | Similar See Note a. |
| Dimensions of Device (mm)[W x Lx D] | 60mm x 58mm x 98mm | 158mm x 56mm x 51.5mm | Similar See Note a. |
| Housing Materials of the Main Unit | ABS Plastic and Chrome plated metal | ABS Plastic and Stainless Steel | Similar See Note b. |
| Power Source | Internal rechargeable Lithium battery DC 3.7V 600mAh | Internal rechargeable Lithium battery DC 3.7V 2200mAh | Similar See Note c. |
| Software/Firmware Microprocessor Control? | Yes | Yes | SE |
| Automatic Shut Off | Yes | Yes | SE |
| User Override Control | Yes | Yes | SE |
| Indicator | Microcurrent intensity indicator lights. LED mode working indicator lights, Charging Indicator | EMS Mode intensity indicator lights. Photon mode working indicator lights, Charging Indicator | SE |
| Treatment Time | Microcurrent: 8 minutes (preset) LED: 10 minutes (preset) | EMS Mode: 5 minutes Photon Mode: 5 – 7 minutes | Similar See Note d. |
| Wavelengths | Red Light: 633 nm ± 10nm Blue Light: 415 nm ± 10nm | Red Light: 630 nm ± 3nm Blue Light: 415 nm ± 3nm | Similar See Note e. |
| LED Power | 0.1W | 0.1W | SE |
| Energy emitted during treatment | 1050J | 1050J | SE |
| Distance between LEDs to treatment surface | 2-3cm | 2-3cm | SE |
| Irradiation Area | 22.5cm² ±5% | 26cm² ±5% | Similar See Note e. |
| Waveform and Shape | Pulsed Biphasic, Modulated Square | Pulsed Biphasic, Modulated Square | SE |
Page | 5 of 8
{9}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
| Element of Comparison | Subject Device (Aura Glide - K252187) | Predicate Device (2 Face/Face Evolution - K171821) | Remark |
| --- | --- | --- | --- |
| Maximum Output Voltage (+/- 10%) | 252 mV @ 500Ω
940 mV @ 2000Ω
3.92 V @ 10,000Ω | 310 mV @ 500Ω
1.16 V @ 2000Ω
5.56 V @ 10,000Ω | Similar
See Note f. |
| Maximum Output Current (+/- 10%) | 504 μA @ 500Ω
470 μA @ 2000Ω
392 μA @ 10,000Ω | 620 μA @ 500Ω
580 μA @ 2000Ω
556 μA @ 10,000Ω | Similar
See Note f. |
| Maximum Phase Charge | 34.89 μC @ 500Ω | 26.31 μC @ 500Ω | Similar
See Note f. |
| Maximum Current Density | 0.315 mA.cm²@500Ω (The minimum electrode size 1.6cm²) | 0.330 mA.cm²@500Ω (The minimum electrode size 2.91cm²) | |
| Maximum Power Density | 7.94μW/cm² @ 500Ω (The minimum electrode size 1.6cm²) | 4.34μW/cm² @ 500Ω (The minimum electrode size 2.91cm²) | Similar
See Note f. |
| Pulse Duration | 60ms | 60ms | SE |
| Irradiance Source | LED | LED | SE |
| Visible Light LEDs | Yes | Yes | SE |
| Compliance with 21 CFR 898 | Yes | Yes | SE |
| Testing Summary | | | |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | SE |
| Safety | IEC 60601-1
IEC 60601-1-11
IEC 60601-2-10
IEC 60601-2-57
IEC 62471 | IEC 60601-1
IEC 60601-2-10
IEC 60601-2-57 | SE |
| Biocompatibility | ISO 10993-5 (Cytotoxicity)
ISO 10993-10 (Skin Sensitization)
ISO 10993-23 (Irritation) | ISO 10993-5 (Cytotoxicity)
ISO 10993-10 (Irritation and Skin Sensitization) | SE |
| Battery | IEC 62133-2 | IEC 62133-2 | SE |
| Usability | IEC 60601-1-6
IEC 62366 -1
Light Power Density | IEC 60601-1-6
IEC 62366 | SE |
Page | 6 of 8
{10}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
# 8. NOTES:
a. The minor differences in weight and dimensions are due to the difference in design, appearance, and shape of the main body for both the subject and predicate devices. Both devices are designed to be handheld and used in a home environment. The minor differences in weight and dimensions pose no safety or effectiveness concerns or issues as demonstrated by the safety and performance testing conducted.
b. Stainless steel is used in the manufacture of the Predicate device and the microcurrent spheres are chrome plated on the Subject device, Aura Glide, this minor difference poses no concern for safety or effectiveness. The difference between stainless steel and chrome plating does not affect the safety and/or effectiveness of the Subject device as provided in the ISO 10993-5, ISO 10993-10, and ISO 10993-23 Biocompatibility reports.
The Subject device, Aura Glide, also has a plastic shield covering the LED lights in the LED attachment. This information was not available for the Predicate device. The plastic used on the subject device is among the list of synthetic polymers that pose low biocompatibility risk because of the long history of safe use in other legally marketed devices that contact intact skin.
c. The Subject device, Aura Glide, has been evaluated and found to comply with the safety standards as required per IEC 60601-1 and IEC 62133-2. Therefore, the minor difference in Power Source poses no effect on safety and/or effectiveness.
d. The preset treatment time for microcurrent and LED are slightly different between the Subject and Predicate devices. The preset treatment time for the Subject device is based on current user practice. Based on user preference, the user can also manually turn off the device at any time before the treatment ends. This minor difference poses no safety and/or effectiveness concerns.
e. The differences in Wavelength and Irradiation Area between the Subject and Predicate devices are very minor and well within their tolerance that the differences do not pose any concerns with regards to safety and/or effectiveness.
f. The Subject device, Aura Glide, has been evaluated extensively and found to comply with the safety standards set forth in IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10, and IES 60601-2-57. The minor differences in maximum output voltage, maximum output current, maximum phase charge and maximum power density poses no adverse effect on safety and/or effectiveness.
Page | 7 of 8
{11}
Sponsor: Aura Medical LLC
Subject Device: Aura Glide (K252187)
510(k) Summary (as required by 807.92)
## 9. CONCLUSION:
The Aura Glide is substantially equivalent to the predicate device with regards to the indications for use, and overall technological characteristics. Any minor differences that may exist between the subject and predicate devices do not raise questions of safety and effectiveness based on satisfactory test results obtained from safety and performance testing conducted. Thus, the Aura Glide is as safe, and as effective, as the predicate device.
Page | 8 of 8
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.