K252161 · Good Sleep CO Pte , Ltd. · LRK · Nov 25, 2025 · Dental
Device Facts
Record ID
K252161
Device Name
Hushd Avera
Applicant
Good Sleep CO Pte , Ltd.
Product Code
LRK · Dental
Decision Date
Nov 25, 2025
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hushd Avera device is intended to reduce or alleviate snoring in adults.
Device Story
Hushd Avera is a removable, non-patient-specific intraoral mandibular advancement device (MAD) for snoring treatment. It consists of moldable ethylene vinyl acetate (EVA) inner trays with rigid polypropylene (PP) outer shells, utilizing a 'boil and bite' fitting process. The device comprises one upper tray and two lower tray options (L1, L2) that interlock via integrated dorsal fins. The fins hold the mandible in a forward position during sleep, preventing tongue and soft tissue collapse into the airway. The user selects between L1 (1mm advancement) or L2 (3mm advancement) to achieve the desired mandibular protrusion. The device is intended for over-the-counter (OTC) use by adults. It is cleaned daily with non-abrasive toothpaste and periodic soaking. By maintaining airway patency, the device aims to reduce snoring, potentially improving sleep quality for the user.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including mechanical property testing (Charpy impact, flexural, shear, hardness, tear, water sorption/solubility, tensile strength) and biocompatibility testing (cytotoxicity, sensitization, systemic toxicity, irritation, and pyrogenicity) per ISO 10993 and USP standards.
Technological Characteristics
Materials: EVA (inner) and Polypropylene (outer). Principle: Mandibular advancement via interlocking dorsal fins. Form factor: Boil-and-bite intraoral trays. Connectivity: None. Sterilization: Non-sterile. Standards: ISO 179-1:2023, ISO 178:2019, ANSI/ADA No. 99-2001, ASTM D638-22, ISO 10993 series.
Indications for Use
Indicated for adults to reduce or alleviate snoring.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 25, 2025
Good Sleep Co Pte Ltd.
% Joseph Azary
Regulatory Consultant
Aztech Regulatory & Quality LLC
543 Long Hill Avenue
Shelton, Connecticut 06484
Re: K252161
Trade/Device Name: Hushd Avera
Regulation Number: 21 CFR 872.5570
Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea
Regulatory Class: Class II
Product Code: LRK
Dated: October 28, 2025
Received: October 28, 2025
Dear Joseph Azary:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252161 - Joseph Azary
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K252161 - Joseph Azary
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
Hushd Avera
Page 12 of 47
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252161 | ? |
| Please provide the device trade name(s). | | ? |
| Hushd Avera | | |
| Please provide your Indications for Use below. | | ? |
| The Hushd Avera device is intended to reduce or alleviate snoring. | | |
| Please select the types of uses (select one or both, as
applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
SUBSTANTIAL EQUIVALENCE COMPARISON
K252161
# 510(k) Summary
HUSHD PRO AVERA INTRAORAL APPLIANCE FOR SNORING
## 1. SUBMITTER/510(K) HOLDER
Good Sleep Co Pte Ltd.
105 Cecil Street,
#18-18, The Octagon
Singapore, 069534
Contact Name: Joseph Azary (Aztech Regulatory and Quality LLC.)
Email: aztechregulatory@gmail.com
Telephone: (203) 242-6670
Prepared by: Joseph Azary / Gayathri Manoj
Date Prepared: October 28, 2025
## 2. DEVICE NAME
Proprietary Name: Hushd Avera
Common/Usual Name: Intraoral device for snoring
Classification Name: Device Anti-Snoring
Classification Regulation: 21 CFR 872.5570
Classification Product code: LRK
Classification: Class 2
Medical Specialty (Panel): Dental
## 3. PREDICATE DEVICES
- Primary Predicate Device – SnoreRx (K170285)
- Reference Device – Dr. Greenburg’s Hybrid (K111680)
## 4. DEVICE DESCRIPTION
The Hushd Avera device is a removable intraoral non-patient specific device used for treating snoring. The device consists of multiple trays that fit over the upper and lower teeth. When the upper and lower tray engages, the device functions as a mandibular repositioner holding the mandible forward during sleep thus preventing the tongue and soft tissues of the throat from collapsing into the airway.
The Hushd Avera device is a “Boil and Bite” device where both the upper and lower trays are constructed out of moldable soft ethylene vinyl acetate (EVA) material with rigid polypropylene (PP) outer shells.
Hushd Avera consists of one upper and two lower trays. Each upper tray contains a trapezoidal block on each side of the posterior to mate with the lower tray. An identification mark is placed on the blocks. The blocks and the outer shell of the upper tray are molded in one piece.
{6}
SUBSTANTIAL EQUIVALENCE COMPARISON
Each lower tray contains a vertical dorsal fin on each side of the posterior to serve as mating mechanism. The fins are triangular shaped at approximately $12.0\mathrm{mm(H)}$ x $12.0\mathrm{mm(W)}$ x $2.5\mathrm{mm(D)}$. The mating surface of the fins are flat to ensure stable positioning of the upper and lower trays against each other. An identification mark is placed on the dorsal fins. The dorsal fins and the outer shell of the lower tray are molded in one piece.
There are no additional clip-on mechanisms connecting the upper and lower trays.
Relative to the upper tray (U), the first lower tray (L1) is positioned $1.0\mathrm{mm}$ forward from an edge-to-edge position, while the upper tray (U) and the second lower tray (L2) have an additional $2.0\mathrm{mm}$ protrusion. The user can select between the two positions as needed.
Therefore, the maximum protrusion is $+3\mathrm{mm}$ from edge to edge while using L2.
## Variants/Configurations
There are no variants for this device and it is available only in one configuration.
## 5. INTENDED USE
The Hushd Avera device is intended to reduce or alleviate snoring in adults.
## 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The subject device has the same intended use, materials and operating principles as the predicate devices. Additionally, the differences in technological characteristics do not raise new questions of safety or effectiveness. Therefore, based on the substantial equivalence evaluation represented below, the company concludes that the subject device is equivalent to the predicate devices.
{7}
SUBSTANTIAL EQUIVALENCE COMPARISON
| | Hushd Pro Avera | Reference Device: SnoreRx (K170285) PRIMARY PREDICTATE | Predicate Device: Dr. Greenburg's Hybrid (K111680) REFERENCE PREDICTATE | Comments |
| --- | --- | --- | --- | --- |
| Regulation | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | Same |
| Class | Class II | Class II | Class II | Same |
| Product Code | LRK | LRK | LRK | Same |
| Indications for Use | Reduce or alleviate snoring | Reduce or alleviate snoring | Reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) | Same for both subject and primary predicate device. While reference device is also intended for mild to moderate obstructive sleep apnea (OSA). |
| Usage frequency | Removable intraoral device for single patient multiple use, can be used as OTC | Removable intraoral device for single patient multiple use, can be used as OTC | Removable intraoral device for single patient multiple use, can be used as Rx only | The subject device and primary predicate device are for OTC use while the reference predicate device is for prescription use only. |
| Target Population | Adults | Adults | Adults | Same |
| Mechanism of Action | The device repositions the lower jaw by holding mandible forward during sleep, preventing the tongue and soft tissues of the throat from collapsing into the airway. The mandibular advancement is achieved using a series of custom-fit upper (maxillary) and lower (mandibular) splints, which interlock via dorsal fins to hold the lower jaw forward. | The device repositions the lower jaw by adding an elastic component connected between the posterior sections of the trays which impedes the tongue from falling back into the airway. The tray system is designed to posture the lower jaw into several positions, with the intent to increase the airway passage opening. | The device repositions the lower jaw by adding an elastic component connected between the posterior sections of the trays which impedes the tongue from falling back into the airway. The tray system is designed to posture the lower jaw into several positions, with the intent to increase the airway passage opening. | All the devices function as Mandibular Advancement Devices (MAD). The subject device uses dorsal fins in each splint to keep the lower jaw forward while the predicate device and reference device uses tray system. |
| Mandibular advancement range | Up to 3mm, not incremental advancement | Up to 6mm | Up to 2mm, not incremental advancement | The subject device provides a 2mm advancement up to 3mm. While the reference predicate device provides a fixed advancement of 2mm and the primary predicate device provides an advancement up to 6mm. |
{8}
SUBSTANTIAL EQUIVALENCE COMPARISON
| Fit | Boil-and-bite fit | Boil-and-bite fit | Boil-and-bite fit | Same |
| --- | --- | --- | --- | --- |
| Manufacturing process | Commercial injection molding | Commercial injection molding | Commercial injection molding | Same |
| Cleaning and Maintenance | After each use daily, use a non-abrasive toothpaste to gently clean the device. Periodically, soak the device in a mild denture cleaner or a mixture of water and soap for a more thorough cleanse. Do not use hot water, as it may warp the appliance. Store dry in the case provided. | After each use, use a non-abrasive toothpaste to gently clean the device. Periodically, soak the device in a mild denture cleaner or a mixture of water and soap for a more thorough cleanse. | After each use, use a non-abrasive toothpaste to gently clean the device. Periodically, soak the device in a mild denture cleaner or a mixture of water and soap for a more thorough cleanse. | Same |
| Material composition | Thermoplastics (Outer – Polypropylene, Inner - EVA) | Thermoplastics (Polycarbonate resin, EVA copolymer) | Thermoplastics (Outer – Polypropylene, Inner - EVA) | All of the devices are composed of thermoplastics. The subject device, primary predicate and reference predicate are all composed of EVA. The primary predicate has polycarbonate whereas the reference predicate has polypropylene which is the same as the subject device. |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Same |
{9}
SUBSTANTIAL EQUIVALENCE COMPARISON
## 7. PERFORMANCE TESTING AND BIOCOMPATIBILITY
The following standards were either used as a reference and/or for testing of the subject device:
- Charpy Unnotched Impact Strength Testing as per ISO 179-1: 2023
- Flexural Testing as per ISO 178: 2019 Method A
- Shear Testing as per In-house method
- Hardness Testing as per ANSI/ADA specification No. 99-2001 (reaffirmed May 2013) clause 4.2
- Tear Strength Testing as per ANSI/ADA specification No. 99-2001 (reaffirmed May 2013) clause 4.3
- Water Sorption Testing as per ANSI/ADA specification No. 99-2001 (reaffirmed May 2013) clause 4.5
- Water Solubility Testing as per ISO 20795-2: 2013 clause 5.2.13
- Tensile Strength Testing as per ASTM D638-22 Type IV
- Cytotoxicity Testing per ISO 10993-5 Biological Evaluation of Medical Devices – Part 5: Tests for in vitro Cytotoxicity.
- Sensitization Testing per ISO 10993-10 Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization.
- Systemic Toxicity Testing per ISO 10993-11 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity.
- Irritation Testing per ISO 10993-23 Biological Evaluation of Medical Devices – Part 23: Test for irritation
- Pyrogen Testing per USP 40, National Formulary 35, General Chapter <151> (2017).
{10}
SUBSTANTIAL EQUIVALENCE COMPARISON
| Summary of Testing | | |
| --- | --- | --- |
| Standard | Description of Test | Results |
| ISO 179-1 (2023) | Plastics – Determination of Charpy impact properties Part 1: Non-instrumented impact testCharpy Unnotched Impact Strength Testing | Virgin Material samples (A&B) demonstrated higher impact resistance compared to Mixed samples (C & D). |
| ISO 178 (2019) Method A | Plastics — Determination of flexural propertiesFlexural Testing | Mixed material samples (C & D) exhibited higher flexural strength and flexural modulus compared to Virgin samples (A & B). |
| In-house method | Shear Testing | Samples successfully withstood the applied forces respectively, achieving displacements without premature failure. |
| ANSI/ADA specification No. 99-2001 (reaffirmed May 2013) clause 4.2 | Athletic Mouth Protectors and MaterialsHardness Testing | Polypropylene showed higher Shore A hardness, while EVA components provided softer inner surfaces for patient comfort. |
| ANSI/ADA specification No. 99-2001 (reaffirmed May 2013) clause 4.3 | Athletic Mouth Protectors and MaterialsTear Strength Testing | Polypropylene demonstrated approximately 4x greater tear resistance than EVA, enhancing long-term durability. |
| ANSI/ADA specification No. 99-2001 (reaffirmed May 2013) clause 4.5 | Athletic Mouth Protectors and MaterialsWater Sorption Testing | Both materials exhibited low water absorption, with polypropylene showing negligible sorption. |
| ISO 20795-2: 2013 clause 5.2.13 | Dentistry — Base polymersPart 2: Orthodontic base polymersWater Solubility Testing | Solubility values for both polypropylene and EVA were well below the allowable limit. |
| ASTM D638-22 Type IV | Standard Test Method for Tensile Properties of PlasticsTensile Strength Testing | Polypropylene samples demonstrated significantly higher tensile strength than EVA. |
| ISO 10993-5 (2009) Cytotoxicity | Biological Evaluation of Medical Devices – Part 5: Tests for in vitro cytotoxicityMEM Elution | Non-cytotoxic |
{11}
SUBSTANTIAL EQUIVALENCE COMPARISON
| ISO 10993-10 (2010) Sensitization | Biological Evaluation of Medical Devices – Part 10: Tests for irritation and skin sensitization
Guinea Pig Maximization | Non-sensitizer |
| --- | --- | --- |
| ISO 10993-11 (2017) Systemic Toxicity | Biological Evaluation of Medical Devices – Part 11: Test for systemic toxicity
0.9% Sodium chloride injection and Sesame oil extract | Non-toxic |
| ISO 10993-23 (2021) Irritation | Biological Evaluation of Medical Devices – Part 23: Test for irritation
Mucosal Irritation Test | Non-irritant |
| USP 40, National Formulary 35, General Chapter <151> (2017) | Pyrogen Test
0.9% Sodium chloride injection extract | Absence of Pyrogens |
## 8. SAFETY AND EFFICACY
**Biocompatibility:** A biocompatibility assessment of the device was performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible, based on the identical use of materials of construction of the predicate devices.
**Clinical Testing:** Human clinical studies were not deemed necessary to evaluate the safety or effectiveness of the subject device. The subject device uses similar manufacturing process as the predicate devices, uses identical materials and has same indications for use.
**Risks:** As with the predicate devices, the subject device can have risks for dental or soft tissue soreness, TMJ dysfunction, obstruction to oral breathing, dental movement or changes in dental occlusion. The potential hazards, including allergy/irritation, poor fit, and incorrect mandibular protrusion have been systematically evaluated as per ISO 14971: 2019. The Instructions for Use include warnings, precautions and contraindications to address such risks. The device is manufactured using the same equipment and same materials as the predicate devices.
## 9. CONCLUSION
Information presented supports substantial equivalence of the Hushd Avera Intraoral Appliance for Snoring to the predicate devices based on similarities in intended use, design, principles of operation, material composition and performance specifications. The conclusions drawn from the nonclinical tests and device comparative analysis demonstrate that the Hushd Avera Intraoral Appliance for Snoring device is as safe, as effective, and performs as well as the identified legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.