K252137 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Apr 2, 2026 · General Hospital
Device Facts
Record ID
K252137
Device Name
BD Insyte™ IV Catheter
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FOZ · General Hospital
Decision Date
Apr 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
BD Insyte™ IV Catheter is intended to be inserted into a patient’s peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
Device Story
BD Insyte™ IV Catheter is a conventional, non-safety, over-the-needle peripheral intravascular catheter. Device consists of a radiopaque BD Vialon™ polyurethane catheter, stainless steel needle, flash chamber with vent plug, and needle cover. BD Insyte-N™ variant features BD Instaflash™ technology for immediate blood visualization. Available in various gauges (14-24 GA) and configurations (winged/non-winged). Used in professional healthcare, transport, and MR environments for vascular access. Clinicians insert the device; blood flashback in the chamber confirms vessel entry. Device supports blood sampling, pressure monitoring, and fluid/power-injected contrast administration. Modifications include updated performance specs for blood fill time, frequency response, kink resistance, and Instaflash visualization, alongside packaging material changes and compliance with ISO 80369-7.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 10555-1/5, ISO 80369-7, ISO 11607, and ISO 10993-1. Testing included catheter-to-adapter separation, needle-to-hub pull force, kink resistance, frequency response, blood fill time, Instaflash visualization, and power injection compatibility. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, genotoxicity, implantation, and hemocompatibility. All predetermined acceptance criteria were met.
Technological Characteristics
Materials: BD Vialon™ polyurethane catheter, 304 stainless steel needle, polypropylene needle cover/adapter/vent plug, epoxy/polyamide adhesive. Dimensions: 14-24 GA. Energy: User-operated. Connectivity: None. Sterilization: Ethylene oxide (EO). Standards: ISO 80369-7 (connectors), ISO 10555-1/5 (catheters), ISO 11607 (packaging), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for patients requiring peripheral vascular access for short-term blood sampling, blood pressure monitoring, or fluid administration. Suitable for all patient populations based on vascular anatomy, procedure, and therapy duration. 22-18 GA models are compatible with power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
BD Angiocath™ and BD Insyte™ IV Catheters (K151698)
BD Insyte Autoguard BC Shielded IV Catheter (K251654)
BD Insyte Autoguard BC Pro Shielded IV Catheter (K251654)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 2, 2026
Becton Dickinson Infusion Therapy Systems Inc.
Rupali Gupta
Associate Director, Regulatory Affairs
9450 South State Street
Sandy, Utah 84070
Re: K252137
Trade/Device Name: BD Insyte™ IV Catheter
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular Catheter
Regulatory Class: Class II
Product Code: FOZ
Dated: February 26, 2026
Received: March 2, 2026
Dear Rupali Gupta:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252137 – Rupali Gupta
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252137 – Rupali Gupta
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
DAVID WOLLOSCHECK -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252137 | |
| Device Name BD Insyte™ IV Catheter | |
| Indications for Use (Describe) BD Insyte™ IV Catheter is intended to be inserted into a patient’s peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
K252137 - 510(k) Summary (21 CFR §807.92)
BD Insytem IV Catheter
| Submitter Information | Submitter Name: Submitter Address: Contact Person: Email Address: Phone Number: Date of Preparation: | Becton Dickinson Infusion Therapy Systems Inc. 9450 South State Street Sandy, Utah 84070 Rupali Gupta Associate Director, Regulatory Affairs rupali.gupta@bd.com (801) 522-5000 April 01, 2026 |
| --- | --- | --- |
| Subject Device | Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel: | BD Insytem IV Catheter Short-Term Less than 30 Days Therapeutic, Intravascular Catheter 21 CFR §880.5200 Intravascular Catheter II FOZ General Hospital |
| Predicate Device | Trade Name: Common Name: Regulation Number: Regulation Name: Regulatory Class: Product Code: Classification Panel: | BD Angiocath™ and BD Insytem IV Catheters K151698 Short-Term Less than 30 Days Therapeutic, Intravascular Catheter 21 CFR §880.5200 Intravascular Catheter II FOZ General Hospital |
Page 1 of 8
{5}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
# Reason for Submission
The purpose of this submission is to notify CDRH that BD is updating the 510(k) baseline for BD Insyte™ IV Catheter devices (BD Insyte™, BD Insyte-W™, BD Insyte-N™) to the current state-of-the-art, which represents modifications to Becton Dickinson’s own BD Insyte™ IV Catheter (K151698) as follows:
- Creation of new performance specifications for Blood Fill Time to support blood sampling; and Frequency Response and Catheter Kink Resistance to support blood pressure monitoring, as noted in the current BD Insyte™ IV Catheter indications for use statement.
- Creation of new performance specifications for existing Instaflash claim to strengthen supporting data:
- Force to remove needle cover
- Time to visualize flashback in flash chamber
- Instaflash time of product with patent needles
- Vent Plug does not fall off
- Modifications to product labeling, including the Instructions for Use, to comply with international requirements, EU MDR requirements, regulations, standards, and FDA labeling guidelines.
- Compliance with ISO 80369-7 (in place of ISO 594-1 and ISO 594-2).
- Modification to unit package top web and bottom web material formulation, as well as secondary level packaging box formulation & thickness.
- Provide design verification evidence supporting the regrind process in injection molding of the needle cover and needle hub component.
- Additional minor changes to the BD Insyte™ IV Catheter product materials, design, performance specifications, and primary packaging that were previously implemented via letter-to-file have been consolidated in this submission. This 510(k) provides FDA with information on the currently marketed device (subject device) along with supporting testing needed to demonstrate continued substantial equivalence.
Page 2 of 8
{6}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
| Device Description | BD Insyte™ IV Catheter is a conventional (non-safety) over-the-needle intravascular (IV) catheter. BD Insyte™ IV Catheter has a radiopaque BD Vialon™ Catheter Material. This device includes a needle and flash chamber with removable vent plug. The needle and catheter are protected by a needle cover.
BD Insyte-N™ IV Catheter has BD Instaflash™ Needle Technology, allowing for immediate visualization of blood along the catheter. The flash chamber on all devices provides confirmation that the catheter has entered the vessel.
BD Insyte™ IV Catheter is available without wings. BD Insyte-W™ IV Catheter is only available with wings. BD Insyte-N™ IV Catheter is available with or without wings. The catheter hub and wings are color coded to indicate catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grey, 14 GA (2.1 mm) = Orange).
BD Insyte™ IV Catheter 22-18 GA (0.9-1.3 mm) are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). |
| --- | --- |
| Indications for Use (21 CFR § 807.92(a)(5)) | BD Insyte™ IV Catheter is intended to be inserted into a patient’s peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). |
| Technological Characteristics | Technological characteristics of the subject BD Insyte™ IV Catheter device is substantially equivalent to the predicate device. The subject BD Insyte™ IV Catheter (Insyte™, Insyte-W™, and Insyte-N™) achieves its intended use based on the same technology and principles of operation as the predicate device.
Comparisons of the subject and predicate devices technological characteristics are provided in Tables 1-1, 1-2, and 1-3 below. |
Page 3 of 8
{7}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
# bd.com
Table 1-1 Comparison of Subject / Predicate Device
| Attribute | SUBJECT
Insytem IV Catheters | PREDICATE
(K151698)
Insytem IV Catheters | Substantial Equivalent? |
| --- | --- | --- | --- |
| Classification | 21 CFR 880.5200
Class II
FOZ - Intravascular Catheter | 21 CFR 880.5200
Class II
FOZ - Intravascular Catheter | Yes
Same as predicate |
| Indications for Use | BD Insytem IV Catheter is intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. This device may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). | An intravascular catheter is a device that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure. | Yes
The indications for use statements have been revised to include word clarification, information on patient population as well as power injection to align with current requirements. These revisions serve to clarify current product use in clinical practice and align with the indications for use of other BD PIVC devices (e.g., BD Nexiva Diffusics (K250682), BD Insyte Autoguard Shielded IV Catheter, BD Insyte Autoguard BC Shielded IV Catheter and BD Insyte Autoguard BC Pro Shielded IV Catheter (K251654).
These updates do not change the intended use of the product or introduce new features or functionality; there are no new questions of safety or effectiveness. |
| Intended Use | Intravascular access | Intravascular access | Yes
Same as predicate |
| Fundamental Scientific Technology | Peripheral intravascular catheter | Peripheral intravascular catheter | Yes
Same as predicate |
| Catheter Dimensions | Catheter Diameter and lengths
14G x 1.75 IN (3.1mm)
16G x 1.16 IN (1.7mm)
16G x 1.77 IN (1.7mm)
18G x 1.16 IN (1.3mm)
18G x 1.88 IN (1.3mm) | Catheter Diameter and lengths
14G x 1.75 IN (3.1mm)
16G x 1.16 IN (1.7mm)
16G x 1.77 IN (1.7mm)
18G x 1.16 IN (1.3mm)
18G x 1.88 IN (1.3mm) | Yes
Same as predicate |
Page 4 of 8
{8}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
| Attribute | SUBJECT
Insytet™ IV Catheters | PREDICATE
(K151698)
Insytet™ IV Catheters | Substantial Equivalent? |
| --- | --- | --- | --- |
| | 20G x 1.00 IN (1.1 mm)
20G x 1.16 IN (1.1 mm)
20G x 1.88 IN (1.1 mm)
22G x 1.00 IN (0.9 mm)
24G x 0.56 IN (0.7 mm)
24G x 0.75 IN (0.7 mm) | 20G x 1.00 IN (1.1 mm)
20G x 1.16 IN (1.1 mm)
20G x 1.88 IN (1.1 mm)
22G x 1.00 IN (0.9 mm)
24G x 0.56 IN (0.7 mm)
24G x 0.75 IN (0.7 mm) | |
| Product
Configurations | Winged
Non-Winged | Winged
Non-Winged | Yes
Same as predicate |
| Sterilization
Modality | Ethylene oxide (EO) | Ethylene oxide (EO) | Yes
Same as predicate |
| Sterilization
Sites | BD Curitiba
BD Tuas or BD Sterile
Services | BD Curitiba
BD Tuas or BD Sterile
Services | Yes
Same as predicate |
| Minimum SAL
(Sterility) | 1 x 10⁻⁶ | 1 x 10⁻⁶ | Yes
Same as predicate |
| Shelf Life | 5 years | 5 years | Yes
Same as predicate |
| Energy Source | User operated | User operated | Yes
Same as predicate |
| Disposable or
Reusable | Disposable | Disposable | Yes
Same as predicate |
| Use
Environment | • Professional Healthcare Facility
• Magnetic Resonance (MR) Environment
• Transport (Ambulatory) Environment | • Professional Healthcare Facility
• Magnetic Resonance (MR) Environment
• Transport (Ambulatory) Environment | Yes
Same as predicate |
| Performance
Testing | **Performance**
ISO 80369-7
ISO 10555-1/ISO 10555-5
ISO 11607
ISO 10993-1
ISO 11135-1
**Shipping**
ASTM D4169
**Aging**
ASTM F1980
BD Internal Requirement(s) | **Performance**
ISO 80369-7
ISO 10555-1/ISO 10555-5
ISO 11607
ISO 10993-1
ISO 11135-1
**Shipping**
ASTM D4169
**Aging**
ASTM F1980
BD Internal Requirement(s) | Dimension and performance testing per ISO 80369-7:2021.
ISO 594-1 and ISO 594-2 have been withdrawn and replaced by ISO 80369-7 which is now recognized by the FDA. |
Page 5 of 8
{9}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
Table 1-2 Comparison of Subject / Predicate Device & Packaging Materials
| Attribute | SUBJECT
Insytem™ IV Catheters | PREDICATE
(K151698)
Insytem™ IV Catheters | Substantial Equivalence? |
| --- | --- | --- | --- |
| Needle (Cannula) | 304 Stainless Steel | Stainless Steel | Yes
Same as predicate |
| Needle Lubricant | Polydimethylsiloxane | Polydimethylsiloxane | Yes
Same as predicate |
| Catheter (Catheter Tubing) | BD Vialon™ Polyurethane | BD Vialon™ Polyurethane | Yes
Same as predicate |
| Catheter Lubricant | Polydimethylsiloxane | Polydimethylsiloxane | Yes
Same as predicate |
| Catheter Tipping Lubricant | Polydimethylsiloxane | Polydimethylsiloxane | Yes
Same as predicate |
| Needle Hub | Propionate | Propionate | Yes
Same as predicate |
| Needle Cover | Polypropylene | Polypropylene | Yes
Same as predicate |
| Metal Wedge | Stainless Steel | Stainless Steel | Yes
Same as predicate |
| Wedge Lubricant | Polydimethylsiloxane | Polydimethylsiloxane | Yes
Same as predicate |
| Vent Plug | Polypropylene (plug)
Acrylic copolymer (filter) | Polypropylene (plug)
Acrylic copolymer (filter) | Yes
Same as predicate |
| Catheter Adapter
(with and without wings) | Polypropylene | Polypropylene | Yes
Same as predicate |
| Catheter Adapter
Colorants | Polypropylene + Colorant
24 GA (Yellow)
22 GA (Blue)
20 GA (Pink)
18 GA (Green)
16 GA (Gray)
14 GA (Orange) | Polypropylene + Colorant
24 GA (Yellow)
22 GA (Blue)
20 GA (Pink)
18 GA (Green)
16 GA (Gray)
14 GA (Orange) | Biocompatibility Testing
per ISO 10993-1:2018,
10993-3, 10993-11,
10993-12, 10993-18 |
| Adhesive (Glue) | Epoxy Resin + Polyamide
curing agent | Epoxy Resin + Polyamide
curing agent | Yes
Same as predicate |
Page 6 of 8
{10}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
| Attribute | SUBJECT
Insytem™ IV Catheters | PREDICATE
(K151698)
Insytem™ IV Catheters | Substantial Equivalence? |
| --- | --- | --- | --- |
| Top Web | Medical Grade Paper | Medical Grade Paper | Unit Packaging Testing per ISO 11607-1:2019
Biocompatibility Testing per ISO 10993-1:2018 |
| Bottom Web | Nylon Thermoforming Film | Copolymer Resin | Unit Packaging Testing per ISO 11607-1
Biocompatibility Testing per ISO 10993-1:2018 |
## Summary of Performance Tests
The following performance tests were completed on the subject device to support the determination of substantial equivalence to the predicate device.
A risk analysis was conducted in accordance with ISO 14971:2019 Medical Devices – Application of Risk Management to Medical Devices to assess the impact of the proposed modifications to the predicate device. BD has determined the technological characteristics between the subject and predicate devices were found to be substantially equivalent. The results of performance testing current and new were found to be applicable to the subject device.
The following guidance documents and standards in conjunction with in-house protocols were used to determine appropriate methods for evaluating the performance of the device:
1. Compliance Testing
- Testing per ISO 10555-1:2013 + Amd 1:2017, ISO 10555-5:2013, ISO 80369-7:2021
- Packaging validation per ISO 11607-1:2019
- Biological evaluation per ISO 10993-1:2018
- EO residuals per ISO 10993-7:2008
- Sterilization validation per ISO 11135-1:2014
2. BD Internal Requirements
- Package Peal Force
- Package Integrity Test
- Catheter to Adapter Separation Force
- Needle-to-Hub Pull Force
Page 7 of 8
{11}
Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 USA
bd.com
- Needle Cover remains in place
- Vent Plug does not fall off
- Frequency Response
- Catheter tubing shall not kink or create a crease in the tubing wall
- Blood Fill Time shall be measured
- Force to Remove Needle Cover
- Chamber Flashback Visible at Needle Hub
- Instaflash Time
- Power Injection
A biocompatibility evaluation, in accordance with ISO 10993-1:2018, Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process and FDA guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (issued September 2023), was conducted. The following testing was undertaken to support the biocompatibility of the subject devices:
- Cytotoxicity per ISO 10993-5:2009
- Sensitization per ISO 10993-10:2021
- Intracutaneous Reactivity per ISO 10993-23:2021
- Acute Systemic Toxicity per ISO 10993-11:2017
- Material-mediated Pyrogenicity per ISO 10993-11:2017 & USP <151>
- Subacute/Subchronic Toxicity per ISO 10993-11:2017
- Genotoxicity per ISO 10993-3:2014
- Implantation per ISO 10993-6:2016
- Hemocompatibility per ISO 10993-4:2017 & ASTM F756-17
- Particulate Matter per USP <788> & per acceptance criteria in ISO 8536-4:2019
Per design control requirements specified in 21 CFR §820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate device.
Summary of Substantial Equivalence
Based on the risk management activities, identical intended use, technological characteristics, and results of performance testing, the subject BD Insyte™ IV Catheter is considered substantially equivalent to the predicate device.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.