The Adaptix Ortho350 is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained doctors or technicians on both adult and pediatric subjects. The Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of body extremities, upper/lower limbs, and adjacent anatomy in all routine radiography examinations. The device is not intended/indicated for use in imaging of the skull, chest, spinal column, hip, shoulder or in mammography applications.
Device Story
Portable 3D tomographic X-ray system; comprises C-Arm assembly and control PC. Operates by positioning patient anatomy against detector; X-ray source emits radiation from controlled sequence of locations; synchronized electronically with flat panel detector. Produces 3D tomosynthesis images and 2D radiographs. Used in hospitals/clinics by X-ray technologists or qualified healthcare professionals. Output displayed on workstation for physician review; aids in diagnosis of orthopedic conditions (fractures, arthritis). Benefits include high-resolution visualization of bone structures in extremities.
Clinical Evidence
Bench testing only. Image quality verified using geometrical phantoms, adult human cadavers, and anthropomorphic pediatric phantoms. Evaluations performed by independent radiologists and orthopedic clinicians confirmed images were clinically acceptable for diagnostic use. Biocompatibility, electrical safety (IEC 60601 series), and cybersecurity (IEC 81001-5-1) testing passed.
Technological Characteristics
Portable tomographic X-ray system. Monoblock X-ray source (70 kV). Flat panel detector (16 cm x 24 cm, 98 μm pixel spacing). Connectivity: Gigabit Ethernet, Wi-Fi/WLAN. Software: Meshless filtered back projection algorithm. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-54. Power: 100-230 V AC. Operating system: Windows 11.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic tomosynthesis imaging of body extremities, upper/lower limbs, and adjacent anatomy. Contraindicated for imaging of the skull, chest, spinal column, hip, shoulder, or mammography.
Regulatory Classification
Identification
A tomographic x-ray system is an x-ray device intended to be used to produce radiologic images of a specific cross-sectional plane of the body by blurring or eliminating detail from other planes. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 6, 2025
Adaptix Limited
% Prithul Bom
Regulatory Technology Services, LLC
1000 Westgate Drive,
Suite 510k
SAINT PAUL, MN 55114
Re: K252133
Trade/Device Name: Adaptix Ortho350
Regulation Number: 21 CFR 892.1740
Regulation Name: Tomographic x-ray system
Regulatory Class: Class II
Product Code: IZF
Dated: April 24, 2025
Received: September 23, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250318 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250318 - Prithul Bom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252133 | |
| Device Name Adaptix Ortho350 | |
| Indications for Use (Describe) The Adaptix Ortho350 is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained doctors or technicians on both adult and pediatric subjects. The Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of body extremities, upper/lower limbs, and adjacent anatomy in all routine radiography examinations. The device is not intended/indicated for use in imaging of the skull, chest, spinal column, hip, shoulder or in mammography applications. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
# I. SUBMITTER
Adaptix Limited
Centre for Innovation and Enterprise, University of Oxford, Begbroke Science Park
Woodstock Road, Oxford, OX5 1PF
United Kingdom
Phone: +44 1865 309619
Contact Person: Martin Stofanko
Date Prepared: November 3,
2025
# II. DEVICE
Name of Device: Adaptix Ortho350
Common or Usual Name: System, X-Ray, Tomographic
Classification Name: Tomographic x-ray system (21 CFR 892.1740)
Regulatory Class: II
Product Code: IZF
# III. PREDICATE DEVICE
Name of Device: ADAPTIX 3D Orthopedic Imaging System ("Ortho Device")
510(k) number: K221949
Manufacturer: Pausch Medical GmbH
Common or Usual Name: System, X-Ray, Tomographic
Classification Name: Tomographic x-ray system (21 CFR 892.1740),
Regulatory Class: II
Product Code: IZF
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
# IV. DEVICE DESCRIPTION
The Adaptix Ortho350 (K252133) is a 3D tomographic X-ray device intended to generate
tomosynthesis images of human anatomy for diagnostic purposes of body extremities, upper/lower limbs, and adjacent anatomy in all routine radiography examinations. The device comprises of an Ortho350 C-Arm Assembly and a Control PC. It is also possible to create 2D X-ray images of the desired anatomy.
The Adaptix Ortho350 device is a portable system that can be placed on a benchtop or the floor or mounted on an optional cart for added mobility. To obtain 3D tomosynthesis images, the relevant area of the patient is positioned against the detector cover then X-rays are emitted from a controlled sequence of locations. The system uses a low-power, monoblock X-ray source with integrated controls and feedback, synchronized electronically with the detector for tomosynthesis image acquisition. The system is operated by an X-ray technologist or suitably qualified healthcare professional using and designed with safety features to only allow X-ray emission when multiple controls are in place.
Page 1 of 9
{5}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
# V. INDICATIONS FOR USE
The Adaptix Ortho350 is intended to be used in hospitals, clinics, imaging centres, and/or other healthcare facilities by qualified/ trained doctors or technicians on both adult and pediatric subjects.
The Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of body extremities, upper/lower limbs, and adjacent anatomy in all routine radiography examinations.
The device is not intended/indicated for use in imaging of the skull, chest, spinal column, hip, shoulder or in mammography applications.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE
# Device Comparison Summary
The following Table 1 compares the Adaptix Ortho350 in a side-by-side manner to the predicate device with respect to indications for use, target population, principles of operation, technological characteristics, components and types of verification testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
Table 1 – Comparison of Main Characteristics between Subject Device and Predicate Device
| Features/Technology | New Device | Predicate Device | Comparison to Predicate, Comments to Differences |
| --- | --- | --- | --- |
| | | | |
| Product Codes | | | |
| Proprietary Name: | Adaptix Ortho350 | Adaptix 3D Orthopedic Imaging System | N/A |
| Classification Name | Tomographic x-ray system | Tomographic x-ray system | Same |
{6}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
| Classification Product Code | IZF | IZF | Same |
| --- | --- | --- | --- |
| Device Class | 2 | 2 | Same |
| Regulation Number | §892.1740 | §892.1740 | Same |
| Premarket Review Panel | Radiology | Radiology | Same |
| Registered Establishment Name | Adaptix Limited | Pausch Medical GmbH | N/A |
| Registered Establishment Number | TBD | 9610903 | N/A |
| Premarket Submission Number: | K252133 | K221949 | N/A |
| Intended use / indications for use | | | |
| | The Adaptix Ortho350 device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/ trained doctors or technicians on both adult and pediatric subjects.
The Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of body extremities, upper/lower limbs, and adjacent anatomy in all routine radiography examinations.
The device is not intended/indicated for use in imaging of the skull, chest, spinal column, hip, shoulder or in mammography applications. | The Ortho Device is intended to generate tomosynthesis images of human anatomy for diagnostic purposes of the hand, elbow, and foot in patients of all ages. The imaging will provide the physician visualized information about anatomical structures to facilitate assessment in orthopedic cases such as:
• Fractures of bones in finger, metacarpus or wrist
• Fractures of foot, ankle, or elbow joint Arthritis | Similar, with no new issues of safety or effectiveness.
The Adaptix Ortho350 device covers all the body parts explicitly stated in the Intended Use of the predicate device (finger, metacarpus, wrist, foot, ankle, elbow). In addition, the larger detector of the Adaptix Ortho350 device allows thicker parts of body extremities (e.g. knees) to also be imaged. Both devices are intended to be used in healthcare settings like hospitals, clinics, and imaging centers by qualified healthcare professionals. Their primary purpose is to provide physicians with visualized information to aid in the diagnosis and assessment of orthopedic conditions such as fractures, arthritis, and other bone-related issues. |
| Device Components | | | |
{7}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
| Workstation | Yes | Yes | Same |
| --- | --- | --- | --- |
| System Status Lights | Yes | Yes | Same |
| Flat panel x-ray detector | Yes | Yes | Same |
| Detector Specifications | | | |
| Digital Flat Panel Detector/Image Receptor | Flat Panel Detector | Flat Panel Detector | Same |
| Detector Size | 16 cm x 24 cm | 15 cm x 12 cm | Similar, with no new issues of safety or effectiveness. (Adaptix Ortho350 has a larger detector size, which means a larger area can be covered in a single exposure.) |
| Pixel Spacing | 98 μm | 99 μm | Similar, with no new issues of safety or effectiveness (Both devices have very similar pixel spacing) |
| Frame Rate | • 45 fps (1×1 binning, Full area)
• 90 fps (2×2 binning, Full area)
• 500 fps (1×1 binning, 6mm-width area)
* Frame Rate (Max.) depends on the ROI size | • Up to 34fps at full resolution,
• 86fps in 2x2 pixel binning mode
• 300fps in Panoramic Mode
Frame Rate (Max.) depends on the ROI size | Similar, with no new issues of safety or effectiveness (The frame rates are similar. Both sets offer similar options for binning modes and resolution, but the highest frame rates (500 fps vs. 300 fps) are different, with the first set offering more flexibility in areas like binning and ROI) |
| ROI Mode | Free ROI | Free ROI | Same |
| Color Resolution | Mono only | Mono only | Same |
| Number of pixels | 2,430 (w) x 1628 (h) pixels | 1488 (w) x 1148 (h) pixels | Similar, with no new issues of safety or effectiveness. (Adaptix Ortho350 has more pixels, which means a larger area can be covered in a single exposure) |
| Communication | Gigabit Ethernet | Gigabit Ethernet | Same |
| A/D Conversion | 16 bit | 14 bit | Similar, with no new issues of safety or effectiveness (Adaptix Ortho350 offers a higher bit depth in A/D conversion, which would result in more detailed image data) |
| Panel Interface | Wired (Ethernet) | Wired (Ethernet) | Same |
| Spatial resolution | >= 3.4 lp/mm | >= 2.8 lp/mm | Similar, with no new issues of safety or effectiveness |
Page 4 of 9
{8}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
| | | | (Adaptix Ortho350 offers a higher spatial resolution) |
| --- | --- | --- | --- |
| Contrast resolution | at least 10 (out of 17) copper steps can be seen in the images obtained of a Pehamed Fluorad 150 phantom | at least 10 (out of 17) copper steps can be seen in the images obtained of a Pehamed Fluorad 150 phantom | Same |
| **X-ray Source** | | | |
| kV Range | 70 kV (fixed) from a monoblock source emitting X-rays from a square array of positions | 60 kV (fixed) from a flat panel X-ray source emitting X-rays from a square array of positions | Similar, with no new issues of safety or effectiveness (Adaptix Ortho350 operates at slightly higher kV than Adaptix 3D Ortho Imaging System). The exact kV is less important for tomosynthesis imaging than it is for 2D X-ray. However, 70 kV makes penetrating thicker tissue easier. |
| Collimator | Fixed Aperture at fixed SID of 35 cm | Fixed Aperture at fixed SID of 20 cm | Similar, with no new issues of safety or effectiveness (Adaptix Ortho350 operates at larger SID than Adaptix 3D Ortho Imaging System) to cover the wider range of extremities in the intended use (to include knees). |
| **Image Documentation** | | | |
| DICOM 3 Compliant | DICOM 3.0 File Output | DICOM 3.0 File Output | Same |
| **Workstation** | | | |
| Wireless Communication (Wi-Fi)/(WLAN) | Both Wi-Fi and WLAN capable | Both Wi-Fi and WLAN capable | Same |
| USB Ports | Yes | Yes | Same |
| Printer option | Yes | Yes | Same |
| Display | 1920 x 1080 pixels as standard (all-in-one pc) | 1920 x 1080 pixels as standard (laptop) | Same |
| **Pediatric Features** | | | |
| Pediatric Dose Reduction (IDR) | Software controlled protocols | Software controlled protocols | Same |
| Adult Dose Reduction (IDR) | Software controlled protocols | Software controlled protocols | Same |
| **Software** | | | |
| Software Architecture | Adaptix Software Package | Adaptix Software Package (Acquisition & Control) | Similar, with no new issues of safety or effectiveness |
Page 5 of 9
{9}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
| | (Acquisition & Control) | | (the Adaptix Ortho350 has additional and upgraded features) |
| --- | --- | --- | --- |
| Reconstruction Algorithm | Meshless filtered back projection | Meshless filtered back projection | Same |
| Graphical User Interface (GUI) | Embedded owner control computer | Embedded owner control computer | Same |
| Operating system | Windows 11 and above | Windows 10 and above | Similar, with no new issues of safety or effectiveness (using currently supported OS release) |
| Image manipulation functions | Slice scrolling, zoom, windows level, markers | Slice scrolling, zoom, windows level, markers | Same |
| Measurement | ruler | ruler | Same |
| Physical Dimensions | | | |
| Height | 61 cm
24 in | 50 cm
20 in | Similar, with no new issues of safety or effectiveness |
| Footprint | 40 (w) cm x 42(l) cm
[15.7 in(w) x 16.5 in (w) | 50 (w) cm x 47 (l) cm [20 in(w) x 18.5 in (w) | Similar, with no new issues of safety or effectiveness (the Adaptix Ortho350 has a base plate rather than a stand) |
| Weight C-arm | 23.6 kg [52 lbs] | 15.0 kg [33 lbs] | Similar, with no new issues of safety or effectiveness (Adaptix Ortho350 is heavier than Adaptix 3D Ortho Imaging System by design. This difference has no impact on the intended use or safety of the product.) |
| Weight Computer | 7.8 kg [17.2 lbs] (all-in-one pc) | 2.7 kg [6 lbs] (laptop) | Similar, with no new issues of safety or effectiveness (Adaptix Ortho350 has a heavier computer than Adaptix 3D Ortho Imaging System by design. This difference has no impact on the intended use or safety of the product.) |
| Operating Conditions | | | |
| Conditions of Use | Temperature: +10°C to 35°C
(50°F to 95°F)
Humidity: 30% to 70%
Air Pressure: 79 to 106 kPa | Temperature: +10°C to 40°C
(50°F to 104°F)
Humidity: 20% to 80%
Air Pressure: 57 to 106 kPa | Similar, with no new issues of safety or effectiveness (Meeting requirements of typical Healthcare facility environment) |
| Conditions for storage / transportation | Temperature: -10°C to 55°C
(14°F to 131°F)
Humidity: 30% to 70 | Temperature: -10°C to 55°C
(14°F to 131°F)
Humidity: 20% to 80%
Air Pressure: 57 to 106 kPa | Similar, with no new issues of safety or effectiveness (Meeting requirements of typical transport |
Page 6 of 9
{10}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
| | %
Air Pressure: 57 to 106 kPa | | conditions. No specialized equipment required to control humidity for transport or storage). |
| --- | --- | --- | --- |
| Power System | | | |
| Input Power | AC: 100-230 V, 50-60 Hz | AC: 100-230 V, 50-60 Hz | Same |
| EMI Filter | IEC 60601-1-1 and 60601-1-2 compliant | IEC 60601-1-1 and 60601-1-2 compliant | Same |
| AC Power Cord | Hospital Grade NEMA 5-15 | Hospital Grade NEMA 5-15 | Same |
| Isolation Transformer | EN 60601-1, EN 60601-1-1, EN 60601-1-2, EN 61558-1, EN 61558-2-4, UL 60601-1, CAN / CSA 22.2 No. 601.1 / NRTL approval | EN 60601-1, EN 60601-1-1, EN 60601-1-2, EN 61558-1, EN 61558-2-4, UL 60601-1, CAN / CSA 22.2 No. 601.1 / NRTL approval | Similar, with no new issues of safety or effectiveness. The Adaptix 3D Ortho Imaging System had transformers located in its control box. Adaptix Ortho350 has transformers located in the C-arm and also a separate isolation transformer. |
| Protection type and level against electric shock | Class 1 | Class 1 | Same |
| Electrical Safety, Electromagnetic Compatibility standards | | | |
| Standards | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-3
IEC 60601-1-6
IEC 60601-2-28
IEC 60601-2-54 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-3
IEC 60601-1-6
IEC 60601-2-28
IEC 60601-2-54 | Same |
## Substantial Equivalence Summary
Based on the information provided above, the Adaptix Ortho350 system shares the same or similar intended use, modes of operation, design and overall technical and functional capabilities, and therefore is substantially equivalent to the selected predicate device Adaptix 3D Orthopedic Imaging System.
## VII. Performance Testing Summary
The following performance testing was carried out on the Adaptix Ortho350 device:
## Biocompatibility Testing
A Biological Evaluation was performed on the Ortho350 device in accordance with ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," and the FDA's Guidance for Industry and FDA Staff - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The evaluation concluded that the materials used in the Ortho350 device are biocompatible for its intended use and does not pose any biological safety risks to patients.
Page 7 of 9
{11}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the Adaptix Ortho350 device, consisting of the C-Arm, Isolation Transformer and the Control PC. The testing determined that the electric safety of the Ortho350 Device is acceptable for its intended use, and that the system complies with the IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601 -2-54 and the IEC 60601-1-2 standard for EMC.
# Software Verification and Validation Testing
Software Verification and Validation testing was performed on the Ortho350 device, in accordance with IEC 62304:2006 + A1:2015 Medical device software — Software life cycle processes and the FDA's Guidance for Industry and FDA Staff Content of Premarket Submissions for Device Software Functions (2023). Overall, the software for the Ortho350 system was determined as a Safety Class 'B' under IEC 62304. All design requirements and risk mitigations passed their planned verification and validation tests.
# Cybersecurity Testing
Security Requirements Testing, Threat Mitigation Testing, Vulnerability Testing, and Penetration Testing were performed on the Adaptix Ortho350 System in line with IEC 81001-5-1 (2022) and the FDA's Guidance for Industry and FDA Staff - Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (2023). The testing concluded that the Ortho350 system is safe to install and operate within a HDO environment from a cybersecurity perspective.
# Transport Testing
Transport verification was performed on the Ortho350 device within its packaging to ensure that the device could be reliably delivered to end users without sustaining damage that could affect its performance, safety, or usability. The verification was done in accordance with ASTM D4169-22, which included environmental conditioning, mechanical handling (e.g., incline impacts and drops), and vibration testing. The overall testing concluded that the Ortho350 system, along with its chosen packaging solution, can safely endure the rigors of transportation. The overall outcome of the study was a conclusive pass, validating that the product is fit for shipment and capable of immediate operation upon delivery.
# Image Verification Testing
Image Quality verification testing was conducted using geometrical phantoms, adult human cadavers and an anthropomorphic pediatric phantom. Digital tomosynthesis images and standard 2D radiographs evaluated by qualified independent radiologists and orthopedic clinicians from the USA and UK. The tests included assessments of the impact of typical image and motion artifacts, including an assessment of image artifacts in the presence of metal. The tests confirmed that images were clinically acceptable and within the intended use for diagnosis for both adult and pediatric patients.
# Hardware Verification Testing
Verification of the main hardware components of the Ortho350 device (Source and Stage Control Board and C-Arm Desktop Assembly) were performed against their corresponding requirements. The overall verification activities passed accordingly.
Page 8 of 9
{12}
Adaptix Transforming Radiology
Adaptix Ortho350 510(k) Summary K252133
# Usability Testing
Usability Engineering in accordance with IEC 62366 Part 1: Application of usability engineering to medical devices and the FDA's Guidance for Industry and FDA Staff - Applying Human Factors and Usability Engineering to Medical Devices was applied throughout the product development process of the Ortho350 system. During development, several rounds of formative evaluations were conducted, with a final summative evaluation documented in the subsequent report confirmed adequate tomosynthesis workflow for orthopedic settings.
# VIII. CONCLUSIONS
The subject device and the predicate device are substantially equivalent in the areas of technical characteristics, general function, application, and intended use and the subject device does not raise any new potential safety risks. The conclusions drawn from the nonclinical tests discussed above demonstrate that the device is as safe, as effective, and performs as well or better than the legally marketed device K221949.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.