K252110 · Guangzhou Ajax Medical Equipment Co., Ltd. · EHD · Mar 12, 2026 · Dental
Device Facts
Record ID
K252110
Device Name
Dental X-RAY Unit (AJX200)
Applicant
Guangzhou Ajax Medical Equipment Co., Ltd.
Product Code
EHD · Dental
Decision Date
Mar 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Unit is an extraoral source dental radiographic x-ray unit. This unit works as diagnostic purpose x-ray source for human teeth with resultant image recorded on intraoral dental x-ray film or image receptor. The unit is intended to be used in a dental hospital or general hospital with a dental office and should have a separate space to install and use this product. The device is intended to be used for adult and pediatric patients.
Device Story
Extraoral dental X-ray unit; emits X-rays via high-voltage tube assembly; produces diagnostic images on intraoral film or receptors. Used in dental or general hospitals; operated by dental professionals. Features wall-mounted design; jog dial interface for mode selection; preset exposure times based on patient size and tooth type. Output assists clinicians in dental diagnosis. Benefits include standardized diagnostic imaging for human teeth.
Clinical Evidence
Bench testing only. Performance data includes electrical safety and electromagnetic compatibility (IEC 60601-1, 1-2, 1-3, 1-6, 2-28, 2-65), software verification/validation, and bench testing of X-ray module voltage, current, loading accuracy, and imaging performance. Service lifetime verified to 10,000 exposures.
Technological Characteristics
Extraoral X-ray system; 65kV tube voltage; 2.5mA tube current; constant potential (DC) waveform; 200mm source-to-skin distance; 58mm circular field size. Wall-mounted form factor. Energy source: 100-240V AC. Complies with IEC 60601-1, 60601-1-3, 60601-2-65, 60601-2-28, 60601-1-2, and 60601-1-6.
Indications for Use
Indicated for adult and pediatric patients requiring extraoral dental radiographic imaging for diagnostic purposes.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 12, 2026
Guangzhou Ajax Medical Equipment Co., Ltd.
% Jett Lee
Regulation Manager
SGS-CSTC Standards Technical Services Co., Ltd
198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology Development District, Tianhe district
GUANGZHOU, GUANGDONG 510000
CHINA
Re: K252110
Trade/Device Name: Dental X-RAY Unit (AJX200)
Regulation Number: 21 CFR 872.1800
Regulation Name: Extraoral Source X-Ray System
Regulatory Class: Class II
Product Code: EHD
Dated: February 2, 2026
Received: February 2, 2026
Dear Jett Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252110 - Jett Lee
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252110 - Jett Lee
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Dental X-RAY Unit
Page 8 of 36
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252110 | ? |
| Please provide the device trade name(s). | | ? |
| Dental X-RAY Unit (AJX200) | | |
| Please provide your Indications for Use below. | | ? |
| The Unit is an extraoral source dental radiographic x-ray unit. This unit works as diagnostic purpose x-ray
source for human teeth with resultant image recorded on intraoral dental x-ray film or image receptor. The
unit is intended to be used in a dental hospital or general hospital with a dental office and should have a
separate space to install and use this product. The device is intended to be used for adult and pediatric
patients. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Date of the summary prepared: November 9, 2025
# 510(k) Summary: K252110
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a traditional 510(K) submission with no previous application.
# 1. Submitter's and Subject Device Information
# Sponsor
- Company Name: Guangzhou Ajax Medical Equipment Co., Ltd.
Address: Building No.2, Dagang Industrial Zone, Shilou Town, Panyu District, Guangzhou City, Guangdong Province, 511447, P.R. China
Phone: +8618776169232
- Contact Person (including title): Sophia Xu (Registered Engineer)
E-mail: xuefang.xu@ajaxdent.cn
# Application Correspondent
- Company Name: Guangzhou Ajax Medical Equipment Co., Ltd.
Address: Building No.2, Dagang Industrial Zone, Shilou Town, Panyu District, Guangzhou City, Guangdong Province, 511447, P.R. China
Tel: 86-13512755282
- Contact Person: Jett Lee
Email: jianda-lee@foxmail.com
# Subject Device Information
- Type of 510(k) submission: Traditional
Common Name: Unit, X-Ray, Extraoral With Timer
Trade Name: Dental X-RAY Unit (AJX200)
Model: AJX200
- Classification Name: Extraoral source x-ray system
Review Panel: Radiology
Product Code: EHD
Regulation Number: 21 CFR 872.1800
Regulation Class: II
# 2. Predicate Device Information
| | Predicate Device |
| --- | --- |
| Sponsor | Vatech Co., Ltd |
| Device Name | EzRay Air Portable |
| Model | VEX-P300 |
{5}
2
| 510(k) Number | K200182 |
| --- | --- |
| Product Code | EHD |
| Regulation Number | 21 CFR 872.1800 |
| Regulation Class | II |
## 3. Product Description
### 3.1 Description of the device and its operating principle
#### 3.1.1 Working Principle
The equipment acquires images by emitting X-rays continuously on a human tooth. X-rays are emitted when high voltage is supplied to the X-ray tube assembly, which frees electrons from the cathode. They hit the anode to produce X-rays.
#### 3.1.2 Indication for Use
The Unit is an extraoral source dental radiographic x-ray unit. This unit works as diagnostic purpose x-ray source for human teeth with resultant image recorded on intraoral dental x-ray film or image receptor. The unit is intended to be used in a dental hospital or general hospital with a dental office and should have a separate space to install and use this product. The device is intended to be used for adult and pediatric patients.
#### 3.1.3 Use Environment
To ensure proper function and prevent damage, Ajax equipment should be stored, transported, and operated within the following temperature, humidity, and air pressure ranges:
| Temperature/Humidity/Air Pressure | Specifications |
| --- | --- |
| Storage/Transportation | • Temperature: -4° F to 131° F (-20° C to 55° C)
• Relative Humidity: 5% to 95%
• Air Pressure: 700hPa to 1060hPa |
| Operation | • Room Ambient Temperature: 40°F to 104°F (4°C to 40°C)
• Relative Humidity: 20% to 80%
• Air Pressure: 700hPa to 1060hPa |
## 4. Substantial Equivalence Discussion
| Elements of Comparison | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Trade Name | Dental X-RAY Unit (AJX200) | EzRay Air Portable (Model: VEX-P300) | - |
| Submitter | Guangzhou Ajax Medical Equipment Co., Ltd. | Vatech Co., Ltd | - |
| Product Code | Applying | K200182 | - |
| Regulation Number | CFR872.1800 | CFR872.1800 | - |
| Product Code | EHD | EHD | - |
| Prescription / Over - The - Counter Use | Prescription Use | Prescription Use | - |
{6}
| Medical Specialty | Dental | Dental | - |
| --- | --- | --- | --- |
| Intended Use | The Unit is an extraoral source dental radiographic x-ray unit. This unit works as diagnostic purpose x-ray source for human teeth with resultant image recorded on intraoral dental x-ray film or image receptor. The unit is intended to be used in a dental hospital or general hospital with a dental office and should have a separate space to install and use this product. The device is intended to be used for adult and pediatric patients. | EzRay Air Portable (Model: VEX-P300) is an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors. It is indicated for use by a dentist or a dental technician for both adult and pediatric patients. | SE |
| Principle of Use | X-ray tube | X-ray tube | SE |
| Source to Skin Distance | 200mm | 200 mm | SE |
| X-ray Field Size | 58mm dia., circular | 6cm round | SE |
| User Interface | Jog dial for operating mode selection. Additionally, several user-selectable preset times with patient size and tooth selection icons on a display module. | Jog dial for operating mode selection. Additionally, several user-selectable preset times with patient size and tooth selection icons on a display module. | Same |
| Exposure Switch | Exposure switch on the handset | Exposure button on the handset | SE |
| Tube Head Mounting | Wall mounted | Handheld | SE Note 1 |
| Energy Source | Single-phase AC 100-240V, ±10% | Rechargeable 21.6 V DC Li-ion polymer battery pack (Nominal Capacity: 2,500 mAh) | SE Note 2 |
| Exposure Time | 0.02~2.00s, regulated according to the R10 coefficient | 0.05 - 1.0 seconds in 0.01 increments | SE Note 3 |
| Tube Current (mA) | 2.5mA | 2.5 mA fixed | SE |
| Tube Voltage (kVp) | 65kV | 60 or 65 kV fixed | SE |
| Waveform | Constant Potential (DC) | Constant Potential (DC) | SE |
| Applied Standard | IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-6, IEC 60601-2-28 | IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-2, 21 CFR1020.30, 1020.31 | SE |
## Comparison in Detail(s):
Note 1(Tube Head Mounting):
The subject device is a wired digital detector, while the predict device is portable. These minor differences are related to the product design of each manufacturer, but do not affect its function. And the subject device comply with IEC 60601-1 and will not raise any safety or effectiveness issue.
Note 2(Energy Source):
The AC energy source of subject device will be converted to DC by main control board. And the subject device comply with IEC 60601-1 and will not raise any safety or effectiveness issue.
Note 3(Exposure Time):
The exposure time is adjustable to the optimum value according to the tooth type. These minor differences do not affect its function. And the subject device comply with IEC 60601-1 and will not raise any safety or effectiveness issue.
{7}
5 Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
1) Electrical safety and electromagnetic compatibility
The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2005 A1: 2012+A2: 2020 Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-3:2008+AMD1:2013+AMD2:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
IEC 60601-2-28:2017 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis.
IEC 60601-2-65:2012+AMD1:2017+AMD2:2021 Medical electrical equipment - Part 2-65: Particular requirements for the basic safety and essential performance of dental intra-oral X-ray equipment.
IEC 60601-1-2:2014 A1:2020 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests.
IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
2) Software verification and validation testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's 2023 Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". The software for this device was developed and reviewed under an Enhanced Documentation Level.
3) Bench Testing
The performance test had been conducted to verify the X ray module voltage and current accuracy, loading accuracy, accuracy of loading factor and imaging performance testing.
The service lifetime of had been conducted, it is verified that the exposure life of the wall-mounted dental X-ray device reach 10000 times.
4
{8}
5
# 6 Comparison to predicate device and conclusion
The Intended Use, Principle of Use, Source to Skin Distance, X-ray Field Size, Tube Current and Tube Voltage of Dental X-RAY Unit, model: AJX200 is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.