K252016 · Arthrex, Inc. · HSB · Mar 20, 2026 · Orthopedic
Device Facts
Record ID
K252016
Device Name
Arthrex Humeral Nails
Applicant
Arthrex, Inc.
Product Code
HSB · Orthopedic
Decision Date
Mar 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex Humeral Nail System is intended to treat stable and unstable proximal fractures of the humerus including two and three, and in some cases, four part humerus fractures. The Humeral Nail is also intended to treat proximal and distal one third fractures, midshaft fractures and pathological fractures.
Device Story
Titanium intramedullary nail system for humeral fracture fixation; enters humerus via greater tuberosity; includes nail, proximal/distal locking screws, compression screw, and end cap. Used by orthopedic surgeons in clinical settings to stabilize fractures. Provides mechanical support for bone healing. Device is MR Conditional; requires specific scanning conditions for patient safety.
Clinical Evidence
Bench testing only. Mechanical performance evaluated via static and dynamic four-point bend testing and static torsional testing per ASTM F1264-16. MR safety and compatibility validated via non-clinical testing and in-vivo electromagnetic simulation (displacement force, torque, RF heating, image artifact) per FDA guidance.
Technological Characteristics
Titanium intramedullary nail; proximal diameter 10.0mm; distal diameters 7.0mm, 8.0mm, 9.0mm; lengths 15-30cm. Mechanical testing per ASTM F1264-16. MR Conditional status established via non-clinical testing and simulation.
Indications for Use
Indicated for patients with stable/unstable proximal humerus fractures (2, 3, or 4-part), proximal/distal one-third fractures, midshaft fractures, and pathological fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 20, 2026
Arthrex Inc.
Tiffany Mentzel
Principal Regulatory Affairs Specialist
1370 Creekside Blvd
Naples, Florida 34108
Re: K252016
Trade/Device Name: Arthrex Humeral Nails
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB
Dated: February 17, 2026
Received: February 18, 2026
Dear Tiffany Mentzel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252016 - Tiffany Mentzel
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252016 - Tiffany Mentzel
Page 3
Sincerely,
Joseph P. Russell -S
for: Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Arthrex Humeral Nails
Page 8 of 47
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252016 | ? |
| Please provide the device trade name(s). | | ? |
| Arthrex Humeral Nails | | |
| Please provide your Indications for Use below. | | ? |
| The Arthrex Humeral Nail System is intended to treat stable and unstable proximal fractures of the humerus including two and three, and in some cases, four part humerus fractures. The Humeral Nail is also intended to treat proximal and distal one third fractures, midshaft fractures and pathological fractures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252016 (Page 1 of 3)
# 510(k) Summary
| Date Prepared | March 3, 2026 |
| --- | --- |
| Submitter | Arthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945 |
| Contact Person | Name: Tiffany Mentzel
Title: Principal Regulatory Affairs Specialist
Phone: 239-566-5833
Email: tiffany.mentzel@arthrex.com |
| Trade Name | Arthrex Humeral Nails |
| Common Name | Rod, Fixation, Intramedullary and Accessories |
| Product Code | HSB |
| Classification Name | 21 CFR 888.3020: Intramedullary Fixation Rod |
| Regulatory Class | II |
| Primary Predicate Device | K050241 AOS Humeral Nail |
| Additional Predicate Devices | K090478 AOS Proximal Humeral Nail
K103002 Synthes MultiLoc Humeral Nailing System |
| Reference Predicate | K210792 AUXEIIN Humeral Nail System |
| Purpose of Submission | This Traditional 510(k) premarket notification is submitted to obtain clearance for the Arthrex Humeral Nail System which was previously cleared as the AOS Humeral Nail System under K050241. |
| Device Description | The Arthrex Humeral Nails were previously cleared as the AOS Humeral Nail system under K050241. The Arthrex Humeral Nail is a titanium humeral intramedullary nail that is designed to enter the humerus through the greater tuberosity. It consists of an intramedullary nail, proximal and distal locking screws, compression screw and an end cap. |
| Indications for Use | The Arthrex Humeral Nail System is intended to treat stable and unstable proximal fractures of the humerus including two and three, and in some cases, four-part humerus fractures. The Humeral Nail is also intended to treat proximal and distal one third fractures, midshaft fractures and pathological fractures. |
| Performance Data | To address the potential risk of decreased mechanical strength of the proposed Arthrex Humeral Nail System compared to the reference predicate Synthes MultiLoc Humeral Nail System (K103002), Arthrex conducted mechanical testing (static and dynamic four-point bend testing) in accordance with ASTM F1264-16. The proposed |
Page 1 of 3
{5}
K252016 (Page 2 of 3)
| | Arthrex Humeral Nail System performed equivalent to the Synthes Nail test mean for bending yield moment, ultimate bending moment, and bending structural stiffness.
Fatigue four-point bend testing in accordance with ASTM F1264-16, Annex A3 was performed.
Arthrex also conducted static torsional testing to evaluate and measure the torsion stiffness inherent to the design and material for the Arthrex Humeral Nail per ASTM F1264-16, Annex A2. The results show the acceptance criterion was met.
To demonstrate that the components of the Arthrex Humeral Nail System are “MR Conditional”, Arthrex has conducted the recommended non-clinical testing and in-vivo electromagnetic simulation per the FDA Guidance, “Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment”, which includes Magnetically Induced Displacement Force, Magnetically Induced Torque, Heating by RF Fields, and Image Artifact. The results were used to develop safe scanning conditions and guidelines for the Arthrex Humeral Nail System for an hour-long scanning session. Simulated in-vivo results determined that patients with these devices can safely be scanned according to the conditions and guidelines listed in the provided Directions for Use (DFU). |
| --- | --- |
| Technological Comparison | Since the original clearance on March 14, 2005 (K050241), there have been modifications that are incorporated into this submission. The proposed Arthrex Humeral Nails are being offered with a smaller and larger distal diameter (7.0 and 9.0 mm) than the 8.0 distal diameter submitted via K050241. The proximal diameter of 10.0 mm remains unchanged for all three (3) sizes. The nail length is being offered in sizes 15 – 30 cm which is identical to the predicate nail length offering of 15 -30mm. The proposed Arthrex Humeral Nail System has been evaluated for MR Safety; however, it was not evaluated for MR safety when it was originally cleared under additional predicate devices AOS Humeral Nail System, K050241. The proposed Arthrex Humeral Nail System will be labeled as MR Conditional Safety. |
Page 2 of 3
{6}
K252016 (Page 3 of 3)
| | The overall design and configuration of the proposed Arthrex Humeral Nail System, surgical technique, fundamental scientific technology, sterility, materials and manufacturing processes remain identical to the predicate device AOS Humeral Nails (K050241). |
| --- | --- |
| **Conclusion** | Based on the intended use, fundamental scientific technology, and the data provided in this Traditional 510(k), Arthrex has determined that Arthrex Humeral Nail System is substantially equivalent to the predicate device. Any differences between the proposed and predicate device are considered minor and do not raise questions concerning safety and effectiveness. |
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.