K252005 · Liviwell, Inc. · HEB · Mar 20, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K252005
Device Name
Livi Device
Applicant
Liviwell, Inc.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Mar 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The Livi device is intended for insertion into the vagina following sexual intercourse for the absorption of semen or other vaginal discharge. The device does not prevent pregnancy or protect against sexually transmitted infections (STIs).
Device Story
Livi device is a single-use, non-sterile, flower-shaped polyurethane foam device with a removal string and polyethylene applicator. Designed for OTC use by patients to absorb semen or vaginal discharge post-intercourse. Device is inserted into the vagina and worn for up to 15 minutes. It does not provide contraception or STI protection. Healthcare providers or patients use the device to manage post-coital discharge; clinical decision-making is not affected as the device is for hygiene/comfort. Benefits include absorption of post-coital fluids. Device integrity and mucosal safety are maintained during use.
Clinical Evidence
Clinical investigation of 60 participants (180 cycles) evaluated acute vaginal mucosal safety via colposcopy. Results: No Grade ≥2 mucosal trauma, fragmentation, or infections. 6.7% transient Grade 1 findings; 17.8% reported mild/moderate insertion discomfort. Self-selection/labeling comprehension study (n=50) showed >85% success in understanding product purpose and safety information, confirming effective communication for intended/non-intended users.
Technological Characteristics
Materials: Polyurethane foam (device), 100% cotton/polyester (removal string), polyethylene (applicator). Form factor: Flower/tulip shape, 74-78mm length, 18-22mm diameter. Non-sterile. Single-use. Biocompatibility per ISO 10993-1, -5, -10, -11, -23. Bench testing per 21 CFR 801.430 (Syngyna) and FDA guidance for menstrual tampons.
Indications for Use
Indicated for vaginal insertion post-intercourse to absorb semen or vaginal discharge in adult females. Not for use as a contraceptive, proceptive, or for menstrual fluid absorption.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 20, 2026
LiviWell Inc.
% Chris Staab
Regulatory Consultant
Regulatory and Technical Associates
Contact Address
Re: K252005
Trade/Device Name: Livi Device
Regulation Number: 21 CFR§ 884.5470
Regulation Name: Unscented Menstrual Tampon
Regulatory Class: II
Product Code: HEB
Dated: June 27, 2025
Received: June 27, 2025
Dear Chris Staab:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252005 - Chris Staab
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252005 - Chris Staab
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JASON ROBERTS -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252005 | |
| Device Name Livi Device | |
| Indications for Use (Describe) The Livi device is intended for insertion into the vagina following sexual intercourse for the absorption of semen or other vaginal discharge. The device does not prevent pregnancy or protect against sexually transmitted infections (STIs). | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252005
Page 1 of 5
510(k) SUMMARY
I. SUBMITTER
510(k) Holder: LiviWell Inc.
333 US Highway 46, Suite 201
Mountain Lakes, NJ 07046
E-mail: mike@mylivi.com
Contact Person: Chris Staab
Regulatory Consultant
Regulatory and Technical Associates
chris.staab@regulatoryandtechnicalassociates.com
Date Prepared: March 20, 2026
II. DEVICES
Names of Devices: Livi Device
Regulation Name: Unscented Menstrual Tampon
Regulatory Class: Class II
Product Codes: HEB – Unscented Tampon
III. PREDICATE DEVICE
K241064 – Shandong Intco Hygiene Products Co., LTD. Unscented Tampon
The predicate device has not been subject to a design-related recall.
IV. DEVICE DESCRIPTION
The Livi device is an over the counter (OTC), single-use, non-sterile fluid absorptive device designed for insertion into the vagina for the absorption of semen or other vaginal discharge. The Livi device is intended to be worn for up to 15 minutes, and is not intended as a contraceptive, proceptive, or for use to absorb menstrual fluid during menstruation.
{5}
K252005
Page 2 of 5
# V. INDICATIONS FOR USE
The Livi device is intended for insertion into the vagina following sexual intercourse for the absorption of semen or other vaginal discharge. The device does not prevent pregnancy or protect against sexually transmitted infections (STIs).
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
In accordance with the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] issued July 28, 2014, the comparison between the predicate device and the subject device is shown to be substantially equivalent by comparing the indications for use, principles of operation, technological characteristics, and performance testing similarities and differences (Table 1). The technological characteristic differences do not raise different questions of safety and effectiveness than the predicate device.
Table 1. Substantial Equivalence Subject Device Livi Compared to the Predicate Device Unscented Tampon
| | K252005
LiviWell Inc.
Livi Device | | K241064
SHANDONG INTCO HYGIENE PRODUCTS CO., LTD.
Unscented Tampon | |
| --- | --- | --- | --- | --- |
| Product Code and Regulation Number | HEB (21 CFR 884.5470) | | HEB (21 CFR 884.5470) | |
| Classification | II | | II | |
| Indications for Use | The Livi device is intended for insertion into the vagina following sexual intercourse for the absorption of semen or other vaginal discharge. The device does not prevent pregnancy or protect against sexually transmitted infections (STIs). | | The Unscented Tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge. | |
| Single Use | Yes | | Yes | |
| Sterility | No | | No | |
| Design features | Livi is flower (tulip) shaped, with 3 petals
Assembled device with smooth, rounded tip. | | Tampon with cylindrical shape and bullet-like tip.
Applicator with smooth, rounded tip. | |
| Component Materials | Livi | Polyurethane foam | Pledglet | Cotton / Viscose |
| | Removal String | 100% cotton / polyester | Removal String | 100% Cotton / Polyester and cotton |
| | Applicator | Polyethylene (PE) | Applicator | Cardboard |
| Packaging Materials | Overwrap | Multi-layer laminate including:
Outer wrapper:
Polyethylene terephthalate (PET) | Overwrap | Polyethylene and Polyethylene terephthalate (PET) |
{6}
K252005
Page 3 of 5
| | | Middle layer: Aluminum Inner wrapper: Cast Polypropylene (CPP) | | |
| --- | --- | --- | --- | --- |
| Absorbency | Light (<6g*) | | Light: ≤6g , Regular: 6~9g , Super: 9~12g, and Super Plus: 12~15g | |
| Product Dimensions | Foam length 74-78mm
Foam diameter 18-22mm
Removal string length 198-202mm
Applicator length 90-94mm
Applicator diameter 19-22mm | | (For Light Absorbency Tampon):
Pledget length 45 – 50 mm
Pledget diameter 12.2–13.7 mm
Removal string length 115 -175 mm
Applicator length 120–125 mm
Applicator diameter 13.8–14.2 mm | |
*Note: There is only one Livi device; an absorbency range for each device is not applicable to this device as it is not intended for use to absorb menstrual fluid.
## VII. Summary of Non-Clinical Testing
Non-clinical testing was conducted to verify that the subject devices meet all design and performance specifications similarly to the predicate device. The following tests were conducted:
### Biocompatibility
- Biocompatibility studies were performed on the Livi device in accordance with the FDA guidance document “Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO 10993-1, ‘Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process’ issued on September 4, 2020 as follows:
- In vitro cytotoxicity test per ISO 10993-5:2009
- Skin sensitization test per ISO 10993-10:2021
- Vaginal irritation test per ISO 10993-23:2021
- Acute systemic toxicity test per ISO 10993-11:2017 (foam only)
- All the above tests were performed on the applicator and plunger, except acute systemic toxicity. The results demonstrated that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.
### Bench Performance Testing
- The following performance characteristics were assessed on the Livi in accordance with the FDA guidance document “Guidance for Industry and FDA Staff - Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)” issued on July 27, 2005.
- Dimensional verification
- String pull strength
- Absorbency per Syngyna Testing (21 CFR 801.430)
- Applicator expulsion force
- Foam (device) integrity
- Fiber/material shedding
{7}
K252005
Page 4 of 5
- Preclinical Microbiology
- Bioburden
- Effect on vaginal microflora
- S. aureus growth
- TSST-1 production
- The results demonstrated that the subject device performs as intended, maintains integrity, and that the Livi device does not enhance the growth of any microbes tested and yields less TSST-1 when compared to a legally marketed tampon.
# VIII. Clinical Tests
## Acute vaginal mucosal safety
- A clinical investigation to evaluate the acute vaginal mucosal safety of repetitive, consecutive use of the Livi post-coital absorptive device under controlled in-clinic conditions, enrolling 60 participants who completed 180 observed device-use cycles with digital colposcopic examinations and structured follow-up was conducted. No Grade ≥2 mucosal trauma, device fragmentation, retained material, infections, allergic reactions, or unanticipated adverse events were observed. 6.7%, (12/180) were noted to have transient Grade 1 mucosal findings. Subject-reported symptoms included mild to moderate pain or discomfort (32/180 or 17.8%) with insertion but were self-limited, and both 24-hour follow-up and 7-day surveillance identified no serious adverse events or fragmentation. Overall, the clinical study confirmed device integrity and acceptable mucosal safety with repetitive consecutive use representative of the device's intended use under the conditions evaluated.
## Self-selection and labeling comprehension study
- A self-selection and labeling comprehension study was conducted to evaluate whether both intended and non-intended users clearly understood the product's intended use and associated safety information. A total of 50 participants (25 intended users and 25 non-intended users) completed the study, which demonstrated that participants consistently understood the product's purpose, accurately identified whether they were appropriate users, and comprehended the labeling as presented. The results exceeded the predefined ≥85% success criteria for self-selection and all comprehension endpoints, supporting the conclusion that the labeling effectively communicates critical information for the intended and non-intended user population.
# IX. CONCLUSIONS
The Livi device has similar technological characteristics to the predicate device (K241064), including intravaginal placement and absorbent material composition. While the predicate is intended for menstrual use, the Livi device is intended for post-coital absorption of semen or vaginal discharge and is explicitly not intended for contraceptive, proceptive, or menstrual use. These differences in intended use do not raise new
{8}
K252005
Page 5 of 5
questions of safety and effectiveness, as the associated risks, including potential misuse as a tampon or misunderstanding of product purpose, are adequately mitigated through labeling and demonstrated user comprehension. Based on the totality of the evidence, including clinical and non-clinical performance data, the Livi device is as safe and as effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.