K251927 · Zeta Surgical, Inc. · SGE · Oct 10, 2025 · Neurology
Device Facts
Record ID
K251927
Device Name
Zeta Navigation System (ZNS131-US)
Applicant
Zeta Surgical, Inc.
Product Code
SGE · Neurology
Decision Date
Oct 10, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Zeta TMS Navigation System is a stereotaxic image guidance system intended for accurate positioning of the treatment coil of TMS systems with respect to target brain locations based on data obtained from 3D patient imaging. The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R30 stimulator, the R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: Cool-B65 RO.
Device Story
Zeta TMS Navigation System is a stereotaxic, image-guided planning and intraoperative guidance system for cranial interventions. It takes 3D preoperative medical scans (MRI/CT) as input. The system uses a structured light-based infrared 3D camera to register patient anatomy and track interventional instruments. It assists clinicians in planning and positioning TMS coils relative to target brain locations. The system displays the position of the navigated coil relative to the 3D scan on a monitor. It is used in clinical environments (offices/outpatient facilities) by trained clinical professionals. The output allows for precise coil placement, potentially improving the accuracy of TMS therapy delivery.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including accuracy assessments (single-point, orientation, point-to-point, and angular accuracy), electrical safety, and electromagnetic compatibility testing.
Technological Characteristics
Stereotaxic image guidance system; structured light-based infrared 3D camera for registration and tracking. Compatible with MagVenture R30/X100 stimulators and Cool-B65 RO coils. Supports 3D DICOM CT and MRI. Software-based planning and navigation. Tested to IEC 60601-1, IEC 62304, IEC 60601-1-2, and IEC 60601-1-6.
Indications for Use
Indicated for accurate positioning of TMS treatment coils relative to target brain locations in patients undergoing TMS therapy, using 3D imaging data. Intended for use by trained clinical professionals in clinical environments.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 10, 2025
Zeta Surgical Inc.
Hieu Le Mau
Chief Operating Officer
280 Summer Street, Floor 7
Boston, Massachusetts 02210
Re: K251927
Trade/Device Name: Zeta Navigation System (ZNS131-US)
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: SGE, QFF Dated: August 9, 2025
Received: August 11, 2025
Dear Hieu Le Mau:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}
K251927 - Hieu Le Mau
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K251927 - Hieu Le Mau
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jaime Raben -S
for Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251927 | |
| Device Name Zeta TMS Navigation System | |
| Indications for Use (Describe) The Zeta TMS Navigation System is a stereotaxic image guidance system intended for accurate positioning of the treatment coil of TMS systems with respect to target brain locations based on data obtained from 3D patient imaging. The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R30 stimulator, the R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: Cool-B65 RO. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251927, Page 1 of 4
510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary for the Zeta TMS Navigation System is provided below.
1. SUBMITTER
Applicant: Zeta Surgical Inc.
280 Summer Street, Floor 7
Boston, MA 02210
Contact: Hieu Le Mau
Chief Operating Officer
Zeta Surgical Inc.
280 Summer Street, Floor 7
Boston, MA 02210
+1 (857) 225-4138
hieu.lemau@zetasurgical.com
2. DEVICE
Device Trade Name: Zeta TMS Navigation System
Device Common Name: Neurological Stereotaxic Instrument
Classification Name: Stereotaxic instrument, 21 CFR 882.4560
Regulatory Class: Class II
Product Code: SGE, QFF
3. PREDICATE DEVICE
Localite TMS Navigator TS [K223577]
4. DEVICE DESCRIPTION
The Zeta TMS Navigation System is a stereotaxic, image guided planning and intraoperative guidance system enabling computer-assisted cranial interventional procedures. The system assists trained clinical professionals with the precise positioning of interventional instruments relative to patient anatomy by displaying the position of navigated interventional instruments relative to 3D preoperative medical scans.
5. INDICATIONS FOR USE
The Zeta TMS Navigation System is a stereotaxic image guidance system intended for accurate positioning of the treatment coil of TMS systems with respect to target brain locations based on data obtained from 3D patient imaging.
The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R30 stimulator, the R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: Cool-B65 RO.
{5}
K251927, Page 2 of 4
# 6. SUBSTANTIAL EQUIVALENCE
| | PREDICATE DEVICE | SUBJECT DEVICE |
| --- | --- | --- |
| 510(K) NUMBER | K223577 | K251927 |
| DEVICE NAME | Localite TMS Navigator TS | Zeta TMS Navigation System |
| CLASSIFICATION REGULATION | 21 CFR 882.4560 | 21 CFR 882.4560 |
| PRODUCT CODE | HAW | SGE, QFF |
| CLASSIFICATION | Class II | Class II |
| INDICATIONS FOR USE | The TMS Navigator TS helps users to plan, implement and document treatment involving TMS of the brain. The system provides planning and navigation functions using anatomical MR data. Regions of the brain to be stimulated can be determined on the basis of anatomy, functional areas or by entering previously calculated coordinates from brain atlases. The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R20 and R30 stimulators, the R30 TMS stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: C-100, C-B60, Cool-B65, Cool-B70, Cool D-B80, MC-125, MC-B70, MCF-75, MCF-125, MCF-B65 and MMC-140-II. | The Zeta TMS Navigation System is a stereotaxic image guidance system intended for accurate positioning of the treatment coil of TMS systems with respect to target brain locations based on data obtained from 3D patient imaging. The product is intended for use with the MagVenture therapy systems supplied by Tonica Elektronik A/S (Farum, Denmark), more precisely the R30 stimulator, the R30 stimulator with MagOption, the X100 stimulator and the X100 stimulator with MagOption only with following magnetic coils: Cool-B65 RO. |
| INTENDED USER | Trained clinical professionals | Trained clinical professionals |
| INTENDED USE ENVIRONMENT | Clinical environments such as medical offices and outpatient facilities, where TMS is administered under the supervision of qualified medical personnel. | Clinical environments such as medical offices and outpatient facilities, where TMS is administered under the supervision of qualified medical personnel. |
| PRESCRIPTION USE | Yes | Yes |
| ANATOMICAL SITE | Head | Head |
| NEURONAVIGATION PRINCIPLE | Anatomy registered to patient via physical point selection. Instrument tracking done by infrared 3D camera. | Anatomy registered to patient via structured light-based infrared 3D camera. Instrument tracking done by infrared 3D camera. |
| ACCEPTED IMAGING MODALITIES | MRI | 3D DICOM CT, MRI |
510(K) SUMMARY
{6}
K251927, Page 3 of 4
| | PREDICATE DEVICE | SUBJECT DEVICE |
| --- | --- | --- |
| MAJOR SYSTEM COMPONENTS | Cart, Sensor head, Sensor head positioning arm, Monitor, Monitor stand, Tracked TMS coil, Software | Cart, Sensor head, Sensor head positioning arm, Monitor, Monitor positioning arm, Tracked TMS coil, Software |
| WORKFLOW COMPONENTS | Upload, Segmentation, Planning, Staging/Positioning, Registration, Navigation | Upload, Segmentation, Planning, Staging/Positioning, Registration, Navigation |
| SELECTION OF TARGETS | Regions of the brain to be stimulated can be determined on the basis of anatomy, functional areas or by entering previously calculated coordinates from brain atlases. | Targets can be determined on the basis of anatomy and functional areas. |
| PLANNING FEATURES | Multiple target point selection, multiple instrument selection | Multiple target point selection, multiple instrument selection |
| 2D VIEWING | Yes: axial, coronal, sagittal slices through configurable cut planes in 3D scene | Yes: axial, coronal, sagittal slices through configurable cut planes in 3D scene |
| 3D VIEWING | Yes: 3D viewing of skin, brain surface and activation maps, using surface rendering techniques | Yes: 3D viewing of skin for segmentation and planning |
| SCANNER INTERFACE | DICOM import of MR images | DICOM import of MR/CT images |
| REGISTRATION FEATURES | Manual point selection registered to reference array | Automatic, pinless, and markerless |
| COMPATIBLE COILS | C-100, C-B60, Cool-B65, Cool-B70, Cool D-B80, MC-125, MC-B70, MCF-75, MCF-125, MCF-B65 and MMC-140-II | Cool-B65 RO |
| COMPATIBLE NAVIGATED POINTERS | Pointer ST, Pointer MR | IZI Disposable Passive Blunt Probe |
| SYSTEM ACCURACY FOR TMS COIL TRACKING | Average: 3.47 mm with a 95% CI of [3.40 mm, 3.53 mm]Best Case: 2.19 mm with a 95% CI of [2.11 mm, 2.26 mm] | Standard Case: 0.45mm mean with 99% CI of [0.40 mm, 0.50 mm]Worst Case: 0.93mm mean with 99% CI of [0.75 mm, 1.11 mm] |
# 7. PERFORMANCE DATA
# 7.1. Biocompatibility Testing
There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.
# 7.2. Electrical safety and electromagnetic compatibility (EMC)
{7}
K251927, Page 4 of 4
The subject device was tested in accordance with the following standards:
- IEC 60601-1:2005 Ed. 3 + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical electrical equipment: Part 1: General requirements for basic safety and essential performance including US deviations, except Clause 11.7 regarding biocompatibility. The device passed all tests.
- IEC 62304:2006+ Amd 1:2015, Medical device software - Software life cycle processes. The device passed all tests.
- IEC 60601-1-2:2014+A1:2021, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests. The device passed all tests.
- IEC 60601-1-6, Ed. 3.2 2020-07, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. The device passed all tests.
## 7.3. Software Verification and Validation Testing
Software verification and validation testing was conducted, and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device was considered a Moderate level of concern.
Cybersecurity documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices and Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.
## 7.4. Sterilization, Cleaning, and Shelf Life
The device is reusable, provided non-sterile, and is not sterile when used. Cleaning instructions are provided in the labeling. Shelf-life is not applicable due to the low likelihood of time-dependent product degradation.
## 7.5. Bench Testing
Bench testing was conducted to demonstrate substantial equivalence. Accuracy testing was included assessments of single-point accuracy across the TMS coil, accuracy at various coil orientations, relative point-to-point accuracy, accuracy at extreme positions within the measurement volume, and angular accuracy.
## 7.6. Clinical Testing
No clinical testing was required for the subject device.
## 8. CONCLUSION
The results of testing described above demonstrate that the Zeta TMS Navigation System is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.