Uterine Aspiration Set is used with the vacuum system to aspirate transcervically the products of conception or menstruation from the uterus.
Device Story
Uterine Aspiration Set comprises rigid/flexible curettes and a tubing set; used in hospitals or clinics by clinicians. Device connects to a vacuum system to generate negative pressure; suction force is manually adjusted via a sliding ring on the vacuum control hole. Curettes scrape and remove endometrial or uterine cavity tissue; tubing set facilitates tissue transport to a collection container. System enables efficient removal of uterine contents; provides stable, adjustable negative pressure. Disposable, sterile components; single-use. Benefits include effective, controlled evacuation of uterine cavity contents during gynecological procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including appearance, dimension, connection firmness, leak, print fastness, negative pressure, patency, and kink resistance. Biocompatibility testing performed per ISO 10993-5, ISO 10993-10, and ISO 10993-23.
Technological Characteristics
Materials: ABS, LDPE, SBC, PVC. Sensing/Actuation: Manual vacuum control via sliding ring. Energy: External vacuum source (ISO 10079-1 compliant, <98 kPa). Form factor: Rigid/flexible curettes (4-12mm) and tubing (1.8m-3m). Connectivity: Mechanical connection to vacuum system. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Shelf life: 5 years.
Indications for Use
Indicated for transcervical aspiration of products of conception or menstruation from the uterus in patients requiring uterine curettage or evacuation.
Regulatory Classification
Identification
An endometrial suction curette is a device used to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction. This device is used to obtain tissue for biopsy or for menstrual extraction. This generic type of device may include catheters, syringes, and tissue filters or traps.
Predicate Devices
Berkeley® VC-10 Vacuum Curettage System, Berkeley® Vacurette® Cannulas and Cannulas(F-tips), Collection Tubing Set (K030935)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 15, 2025
Gcmedica Enterprise Ltd.(wuxi)
Adam Jiang
Quality Manager
18 Xinjin Road, Xinwu Industrial Zone
Wuxi, Jiangsu 214000
CHINA
Re: K251882
Trade/Device Name: Uterine Aspiration Set
Regulation Number: 21 CFR 884.1175
Regulation Name: Endometrial Suction Curette and accessories
Regulatory Class: II
Product Code: HHK
Dated: June 19, 2025
Received: September 12, 2025
Dear Adam Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251882 - Adam Jiang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K251882 - Adam Jiang
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251882
Device Name
Uterine Aspiration Set
Indications for Use (Describe)
Uterine Aspiration Set is used with the vacuum system to aspirate transcervically the products of conception or menstruation from the uterus.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
K251882
Page 1 of 7
# 510(k) Summary
# Uterine Aspiration Set
# K251882
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
# 5.1 Submitter
| Submitted by: | GCMEDICA ENTERPRISE LTD.(WUXI) 18 Xinjin Road, Xinwu Industrial Zone, Wuxi, 214000, China |
| --- | --- |
| Contact Person: | Adam Jiang, Quality Manager 18 Xinjin road, Xinwu industrial zone, Wuxi, 214000, China Phone: +86 15206189652 E-mail: adam@gcmedica.com |
| Date Prepared: | October 15, 2025 |
# 5.2 Device
| Name of Device: | Uterine Aspiration Set |
| --- | --- |
| Common or Usual Name: | Uterine Aspiration Set |
| Name of Device: | Rigid Uterine Aspiration Curette |
| --- | --- |
| Common or Usual Name: | Intrauterine curette, suction |
| Classification Name: | Curette, Suction, Endometrial (And Accessories) (21 CFR 884.1175) |
| Regulatory Class: | II |
| Product Code: | HHK |
| Name of Device: | Flexible Uterine Aspiration Curette |
| --- | --- |
| Common or Usual Name: | Intrauterine curette, suction |
| Classification Name: | Curette, Suction, Endometrial (And Accessories) (21 CFR 884.1175) |
| Regulatory Class: | II |
| Product Code: | HHK |
| Name of Device: | Uterine Aspiration Tubing Set |
| --- | --- |
| Common or Usual Name: | Intrauterine curette, suction |
| Classification Name: | Curette, Suction, Endometrial (And |
{5}
K251882
Page 2 of 7
| | Accessories) (21 CFR 884.1175) |
| --- | --- |
| Regulatory Class: | II |
| Product Code: | HHK |
| Name of Device: | Uterine Aspiration Tubing |
| --- | --- |
| Common or Usual Name: | Intrauterine curette, suction |
| Classification Name: | Curette, Suction, Endometrial (And Accessories) (21 CFR 884.1175) |
| Regulatory Class: | II |
| Product Code: | HHK |
| Name of Device: | Swivel Handle |
| --- | --- |
| Classification Name: | Curette, Suction, Endometrial (And Accessories) (21 CFR 884.1175) |
| Regulatory Class: | II |
| Product Code: | HHK |
Note: Uterine Aspiration Tubing and Swivel Handle are components of the Uterine Aspiration Tubing Set.
## 5.3 Predicate Device
Berkeley® VC-10 Vacuum Curettage System, Berkeley® Vacurette® Cannulas and Cannulas(F-tips), Collection Tubing Set, K030935
The predicate device experienced one recall event due to a manufacturing process control issue. However, the recall was not related to the device design.
No reference devices were used in this submission.
## 5.4 General Device Description
The Uterine Aspiration Set consists of rigid/flexible uterine aspiration curette and uterine aspiration tubing set(Composed of uterine aspiration tubing, swivel handle and other components). Rigid/flexible uterine aspiration curette and uterine aspiration tubing set(including uterine aspiration tubing and swivel handle) are all disposable and supplied sterile.
### Rigid/Flexible Uterine Aspiration Curette
The rigid/flexible uterine aspiration curette is mainly used to scrape and remove tissues in the endometrium or uterine cavity. Its tip is suitable for entering the uterine cavity and performing scraping operations.
### Uterine Aspiration Tubing Set
The uterine aspiration tubing set is mainly used to generate negative pressure during the surgical
{6}
K251882
Page 3 of 7
process. It is usually connected at one end to a Rigid/Flexible Uterine Aspiration Curette for scraping operations, and at the other end to a suction device, and tissues are sucked out by negative pressure. The suction force can be adjusted by sliding the ring on the vacuum control hole.
## 5.5 Indications for Use
Uterine Aspiration Set is used with the vacuum system to aspirate transcervically the products of conception or menstruation from the uterus.
## 5.6 Comparison of Technological Characteristics with the Predicate Device
The proposed device is compared with some components (Berkeley® Vacurette® Cannulas and Cannulas(F-tips), Collection Tubing Set) of the predicate device. It can be effectively connected with the suction device to generate a stable and adjustable negative pressure, so as to efficiently remove the uterine tissue and adapt to the uterine conditions of different patients.
The following tables (Table 5.6.1, Table 5.6.2, Table 5.6.3 and Table 5.6.4) are summaries of the uterine aspiration set technological characteristics as compared to the predicated device from ACMI CORPORATION.
Table 5.6.1 Summary of technological characteristics between proposed device Rigid Uterine Aspiration Curette and the predicate device Berkeley® Vacurette® Cannulas from ACMI CORPORATION
| Item | Proposed Device | Predicated Device |
| --- | --- | --- |
| Device name | Rigid Uterine Aspiration Curette | Berkeley® Vacurette® Cannulas |
| 510(k) number | K251882 | K030935 |
| Manufacturer | GCMEDICA ENTERPRISE LTD.(WUXI) | ACMI CORPORATION |
| Product Code | HHK | HHK |
| Regulation No. | 884.1175 | 884.1175 |
| Classification | 2 | 2 |
| Intended use | The rigid or flexible Uterine Aspiration Curette is used for curettage and evacuation of the contents of the endometrium during induced termination of pregnancy or after miscarriage. | For rapid transcervical aspiration of the uterine cavity. (The intended use of the entire system) |
| Environment of Use | Hospital or clinics | Hospital or clinics |
| Supplied Sterile | Sterile | Sterile |
| Sterilization method | EO | EO |
| Compatibility | 1. Devices that comply with ISO 10079-1;
2. Devices with a claimed maximum vacuum range of less than 98 kPa. | Gyrus ACMI Berkeley vacuum curettage system |
{7}
K251882
Page 4 of 7
Table 5.6.2 Summary of technological characteristics between proposed device Flexible Uterine Aspiration Curette and the predicate device Berkeley® Vacurette® Cannulas(F-Tip) from ACMI CORPORATION
| Item | Proposed Device | Predicated Device |
| --- | --- | --- |
| Device name | Flexible Uterine Aspiration Curette | Berkeley® Vacurette® Cannulas(F-Tip) |
| 510(k) number | K251882 | K030935 |
| Manufacturer | GCMEDICA ENTERPRISE LTD.(WUXI) | ACMI CORPORATION |
| Product Code | HHK | HHK |
| Regulation No. | 884.1175 | 884.1175 |
| Classification | 2 | 2 |
| Intended use | The rigid or flexible Uterine Aspiration Curette is used for curettage and evacuation of the contents of the endometrium during induced termination of pregnancy or after miscarriage. | For rapid transcervical aspiration of the uterine cavity. (The intended use of the entire system) |
| Environment of Use | Hospital or clinics | Hospital or clinics |
| Supplied Sterile | Sterile | Sterile |
| Sterilization method | EO | EO |
| Compatibility | 1. Devices that comply with ISO 10079-1 ; 2. Devices with a claimed maximum vacuum range of less than 98 kPa. | Gyrus ACMI Berkeley vacuum curettage system |
| Single Use | Yes | Yes |
| Material | ABS, LDPE | Propionate, polyethylene |
| Sizes (mm) | 4-12 | 4-8 |
Table 5.6.3 Summary of technological characteristics between proposed device Uterine Aspiration Tubing Set and the predicate device Collection Tubing Set from ACMI CORPORATION
| Item | Proposed Device | Predicated Device |
| --- | --- | --- |
| Device name | Uterine Aspiration Tubing Set | Collection Tubing Set |
| 510(k) number | K251882 | K030935 |
| Manufacturer | GCMEDICA ENTERPRISE LTD.(WUXI) | ACMI CORPORATION |
| Product Code | HHK | HHK |
| Regulation No. | 884.1175 | 884.1175 |
| Classification | 2 | 2 |
| Intended use | The rigid or flexible Uterine Aspiration Curette is used for curettage and evacuation of the contents of the endometrium during induced termination of pregnancy or after miscarriage. | For rapid transcervical aspiration of the uterine cavity. (The intended use of the entire system) |
| Environment of Use | Hospital or clinics | Hospital or clinics |
| Supplied Sterile | Sterile | Sterile |
| Sterilization method | EO | EO |
| Compatibility | 1. Devices that comply with ISO 10079-1 ; 2. Devices with a claimed maximum vacuum range of less than 98 kPa. | Gyrus ACMI Berkeley vacuum curettage system |
| Single Use | Yes | Yes |
| Material | ABS, LDPE | Propionate, polyethylene |
| Sizes (mm) | 4-12 | 4-8 |
{8}
K251882
Page 5 of 7
| Regulation No. | 884.1175 | 884.1175 |
| --- | --- | --- |
| Classification | 2 | 2 |
| Intended use | The Uterine Aspiration Tubing Set is designed for use in conjunction with a Uterine Aspiration Curette to enable suction/aspiration during the process of gynecological operations and provide more convenience for easier operation. | For rapid transcervical aspiration of the uterine cavity. (The intended use of the entire system) |
| Environment of Use | Hospital or clinics | Hospital or clinics |
| Supplied Sterile | Sterile | Sterile |
| Sterilization method | EO | EO |
| Compatibility | 1. Devices that comply with ISO 10079-1 ; 2. Devices with a claimed maximum vacuum range of less than 98 kPa. | Gyrus ACMI Berkeley vacuum curettage system |
| Single Use | Yes | Yes |
| Material | SBC, PVC | PVC |
| tubing length | 1.8m,2m,3m | 6ft (1.8288m) |
## 5.7 Non-Clinical Performance Data
GCMEDICA performed bench testing to support substantial equivalence. The following testing was performed on GCMEDICA samples from production lots, including sterilization.
| Proposed Device | Bench Testing |
| --- | --- |
| Rigid Uterine Aspiration Curette | 1. Appearance
2. Dimension
3. Connection Firmness
4. Leak Test
5. Print Fastness
6. Negative Pressure Test
7. Patency |
| Flexible Uterine Aspiration Curette | 1. Appearance
2. Dimension
3. Connection Firmness
4. Leak Test
5. Print Fastness
6. Negative Pressure Test
7. Patency
8. Kink Resistance Testing |
| Uterine Aspiration Tubing Set | 1. Appearance |
{9}
K251882
Page 6 of 7
| | 2. Dimension
3. Negative Pressure Test
4. Leak Test
5. Connection Firmness |
| --- | --- |
| Uterine Aspiration Set | 1. Appearance
2. Dimension
3. Negative Pressure Test
4. Leak Test
5. Connection Firmness |
## 5.7.1 Sterilization
The Uterine Aspiration Set consists of rigid/flexible uterine aspiration curette, uterine aspiration tubing set (Composed of uterine aspiration tubing, swivel handle and other components). The products are all sold in sterile packages. The Uterine Aspiration Set has been sterilized in a validated EO sterilization cycle. The EO sterilization cycle has a Sterility Assurance Level (SAL) of $10^{-6}$. EO residuals on the Uterine Aspiration Set are below the maximum levels defined in ANSI/AAMI/ISO 10993-7:2008 Biological Evaluation of Medical Device - Part 7: Ethylene Oxide sterilization residuals. The Uterine Aspiration Set, and the predicate devices, are not labeled as pyrogen-free because they do not have any blood or cerebrospinal fluid contact. Rigid/flexible uterine aspiration curette, uterine aspiration tubing set and uterine aspiration set are packaged in a paper/film peel pouch like other sterile products GCMEDICA currently manufacture. These pouches have been tested by GCMEDICA and shown to provide and maintain a sterile barrier for at least five (5) years.
## 5.7.2 Shelf Life
The Uterine Aspiration Set has a five (5) years expiration date, based on the design and material equivalence to the predicated device and existing sterile barrier data from GCMEDICA existing packaging. The Uterine Aspiration Set is packaged in a paper/film peel pouch like other sterile products GCMEDICA currently manufacture. These pouches and products have been tested by GCMEDICA, including accelerated aging, seal strength, dye penetration, packaging persistence bacteria performance and shown to provide and maintain a sterile barrier for at least five (5) years. The Uterine Aspiration Set has been tested by GCMEDICA before and after accelerated aging test. The product's visual appearance, dimensions, and performance remained qualified for at least five (5) years.
## 5.8 Clinical Testing
Similar devices have been on the market for many years with proven safety and efficacy for the use of the device. These devices have no direct patient contact. Based on this history and the use of the device, clinical testing was not necessary to support substantial equivalence data. The non-clinical testing performed supports safety and efficacy of the device and provides data to show substantial equivalence to the predicated devices.
## 5.9 Biocompatibility
{10}
K251882
Page 7 of 7
Biocompatibility studies were performed in accordance with the 2020 FDA guidance document “Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and “testing” and current applicable standards, as follows:
- Cytotoxicity: ISO 10993-5:2009
- Sensitization: ISO 10993-10:2010
- Irritation: ISO 10993-23:2021
## 5.10 Conclusion
Based on the technological characteristics and bench testing performance, GCMEDICA believes there are no significant differences between the Uterine Aspiration Set and the predicate devices.
The proposed device does not raise new issues regarding safety or effectiveness and is considered substantially equivalent to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.