K251860 · Copioumed International, Inc. · FYA · Feb 26, 2026 · General, Plastic Surgery
Device Facts
Record ID
K251860
Device Name
Copioumed AAMI 3 Surgical Gown
Applicant
Copioumed International, Inc.
Product Code
FYA · General, Plastic Surgery
Decision Date
Feb 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Copioumed AAMI 3 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate materials. The gown is single-use, disposable, and provided sterile, and is classified as AAMI Level 3.
Device Story
Surgical gown; single-use; disposable; sterile. Constructed from SMS (polypropylene laminated with polyethylene). Provides AAMI Level 3 barrier protection. Used by operating room personnel during surgical procedures. Protects patient and personnel from microorganisms, body fluids, and particulates. Features adjustable neck with Velcro closure; knit cuffs. Available in standard and high-performance models. Bench testing confirms liquid barrier performance, tensile strength, tear resistance, and biocompatibility.
Clinical Evidence
Bench testing only. Performance validated via AATCC 42 (water resistance), AATCC 127 (hydrostatic pressure), ANSI/AAMI PB70:2012 (liquid barrier), NWSP 080.8.R0 (alcohol repellency), ASTM D3776 (weight), ASTM D5034 (tensile), ASTM D5587 (tear), 16 CFR 1610 (flammability), ISO 9073-10 (linting), and ASTM D1683 (seam strength). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity), ISO 10993-23 (irritation), and ISO 10993-10 (sensitization). Sterility validated per ISO 11737-1 and EO residuals per ISO 10993-7.
Indicated for operating room personnel during surgical procedures to protect patient and personnel from transfer of microorganisms, body fluids, and particulate materials.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 26, 2026
Copioumed International, Inc.
% Ming-Yie Jan
Principal Consultant
Ruscert Technology Co., Ltd.
8f., # 187, Lequn 2nd Rd., Zhongshan Dist.
Taipei City, 10462
Taiwan
Re: K251860
Trade/Device Name: Copioumed AAMI 3 Surgical Gown
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Regulatory Class: Class II
Product Code: FYA
Dated: January 26, 2026
Received: January 26, 2026
Dear Ming-Yie Jan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K251860 - Ming-Yie Jan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K251860 - Ming-Yie Jan
Page 3
Sincerely,
ALLAN GUAN -S
For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251860 | |
| Device Name Copioumed AAMI 3 Surgical Gown | |
| Indications for Use (Describe) Copioumed AAMI 3 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate materials. The gown is single-use, disposable, and provided sterile, and is classified as AAMI Level 3. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
新萬盟有限公司
COPIOUMED INTERNATIONAL INC.
台北市復興北路 1號 16樓之 4
16F-4, NO.1, FUXING N.RD., SONGSHAN DIST., TAIPEI CITY 105, TAIWAN (R.O.C.)
TEL:886-2-2772-1815 FAX:886-2-2711-9232
# 510(k) SUMMARY
This 510 (K) summary is being submitted in accordance with requirements of Title 21,CFR Section 807.92.
A. 510(k) NUMBER K251860
B. DATE PREPARED February 17th, 2026
C. SUBMITTER COPIOUMED INTERNATIONAL INC.
16F-4, No.1, Fuxing N. Rd., Songshan Dist., Taipei City 105,
Taiwan (R.O.C.)
FEI Number: 9681151
Tel: +886 22772 1815
D. CONTACT PERSON
Primary Contact
Title: Chairman of the Board
Name: James Liu
Tel: +886 928 644 966
E-mail: james.liu@gcv.com.tw
Official Correspondent
Title: Principal Consultant
Name: Ming-Yie Jan
Tel: +886 2 8785 8989
E-mail: FDA.sub@ruscert.net
E. DEVICE
Proprietary Name: Copioumed AAMI 3 Surgical Gown
Product Code: FYA
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Device Class: Class II
Review Panel: General Hospital
F. INDICATION(S) FOR USE
Copioumed AAMI 3 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate materials. The gown is single-use, disposable, and provided sterile, and is classified as AAMI Level 3.
Page 1 of 13 pages
{6}
Page 2 of 13 pages
G. PRIMARY PREDICTATE DEVICE
Proprietary Name: Cardinal Health™ Non-Reinforced Surgical Gown
Product Code: FYA
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
510(k) Number: K170762
510(k) Submitter: Cardinal Health 200, LLC
Device Class: Class II
Review Panel: General Hospital
H. DEVICE DESCRIPTION
Copioumed AAMI 3 Surgical Gown provides protection with the materials of SMS. Copioumed AAMI 3 Surgical Gown has comfort and breathability and for a better fit on the user. The neck design of Copioumed Surgical Gown is adjustable and closes with a Velcro. Copioumed Surgical Gown is available in several sizes and lengths with two models (standard performance and high standard).
I. COMPARISON OF CHARACTERISTICS WITH THE PREDICTATE DEVICES
{7}
B. Substantial Equivalence Comparison Table-Basic Information
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown K251860 | Predicate Device: Cardinal HealthTM Non-Reinforced Surgical Gown K170762 | Comparisons |
| --- | --- | --- | --- |
| Product Code | FYA | FYA | N/A |
| Intended Use/Indications for Use | Copioumed AAMI 3 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both surgical patient and the operating room personnel from the transfer of microorganisms, body fluids, and particulate materials. The gown is single-use, disposable, and provided sterile, and is classified as AAMI Level 3. | The Cardinal HealthTM Non-Reinforced Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. In addition, this surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities. The Cardinal HealthTM Non-Reinforced Surgical Gowns are single use, disposable medical devices, provided sterile and non-sterile. | Similar |
| Materials | Polypropylene Gown laminated with polyethylene | Polyolefin (Polypropylene) SMS nonwoven | Same |
| Use | Single Use; Disposable | Single Use; Disposable | Same |
{8}
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown | Predicate Device: Cardinal Health™ Non-Reinforced Surgical Gown K170762 | Comparisons |
| --- | --- | --- | --- |
| Design Features | Knit cuff and Velcro in Standard
Knit cuff and Velcro in High Performance | Neck Closure: Hook and Loop Belt Ties
Knit Cuffs
Transfer Tab | Similar |
| Sterility | Sterile | Sterile and non-sterile | For sterile products: Same |
| Sterilization Modality | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Same |
| Contact Durations | Surface, Intact, < 24 hours | Surface, Intact, < 24 hours | Same |
| Color | light blue | Dark blue | Different |
| Size | Standard: Size M, L, XL, XXL
High Performance: M, L, XL, XXL | L, XL,2XL,3XL | Different |
Page 4 of 13 pages
{9}
C. Substantial Equivalence Comparison Table-Predicate Device-Properties
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown | Predicate Device1: Cardinal HealthTM Non-Reinforced Surgical Gown K170762 | | | Comparisons |
| --- | --- | --- | --- | --- | --- |
| Basis weight (oz/yd2) ASTM D3776 | 40G±4G | Test Results (3 Lots) | | Specification | Similar2 |
| | | Mean (min/max) | | | |
| | | Sleeve | Body | | |
| | | Unit4= oz/yd4 | | | |
| | | Mean = 1.32 | Mean = 1.32 | Target Mean= 1.30 | |
| | | Ind Min = 1.28 | Ind Min = 1.28 | Mean Min = 1.20 | |
| | | Ind Max = 1.36 | Ind Max = 1.36 | Ind Min = 1.07 | |
| Grab tensile, MD* (lbs) ASTM D5034 | MD≥55N | Test Results (3 Lots) | | Specification | Similar2 |
| | | Mean (min/max) | | | |
| | | Sleeve | Body | | |
| | | Unit = lbs | | | |
| | | Mean = 21.57 | Mean = 21.57 | N/A | |
| | | Ind Min = 18.60 | Ind Min = 18.60 | | |
| | | Ind Max = 23.60 | Ind Max = 23.60 | | |
{10}
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown | Predicate Device1: Cardinal HealthTM Non-Reinforced Surgical Gown K170762 | | | Comparisons |
| --- | --- | --- | --- | --- | --- |
| Grab tensile, CD* (lbs) ASTM D5034 | CD≥40N | Test Results (3 Lots) | | Specification | Similar2 |
| | | Mean (min/max) | | | |
| | | Sleeve | Body | | |
| | | Unit = lbs | | | |
| | | Mean = 13.60 | Mean = 13.60 | Target Mean = 15.00 | |
| | | Ind Min = 11.20 | Ind Min = 11.20 | Mean Min = 11.00 | |
| | | Ind Max = 16.0 | Ind Max = 16.0 | Ind Min = 9.50 | |
| Trap Tear, MD* (lbs) ASTM D5587-15 Highest Peak | MD≥10N | Test Results (3 Lots) | | Specification | Similar3 |
| | | Mean (min/max) | | | |
| | | Sleeve | Body | | |
| | | Unit = lbs | | | |
| | | Mean = 3.47 | Mean = 3.47 | Target Mean = 4.00 | |
| | | Ind Min = 2.73 | Ind Min = 2.73 | Mean Min = 2.40 | |
| | | Ind Max = 4.00 | Ind Max = 4.00 | Ind Min = 2.00 | |
{11}
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown | Predicate Device1: Cardinal HealthTM Non-Reinforced Surgical Gown K170762 | | | | Comparisons |
| --- | --- | --- | --- | --- | --- | --- |
| Trap Tear, CD* (lbs) ASTM D5587-15Highest Peak | CD≥10N | Test Results (3 Lots) | | Specification | | Similar2 |
| | | Mean (min/max) | | | | |
| | | Sleeve | Body | | | |
| | | Unit = lbs | | | | |
| | | Mean = 5.63 | Mean = 5.63 | N/A | | |
| | | Ind Min = 4.67 | Ind Min = 4.67 | | | |
| | | Ind Max = 7.90 | Ind Max = 7.90 | | | |
| | | | | | | |
| Flammability (sec) CPSC, Part 1610 | Class I | Class I | | | | Same |
| Alcohol Repellency (rating) NWSP 080.8.R0 (15) | >6 | Test Results (3 Lots) | | Specification | | Similar3 |
| | | Mean (min/max) | | | | |
| | | Sleeve | Body | | | |
| | | Mean = 9 | Mean = 9 | Target Mean = 7 | | |
| | | Ind Min = 8 | Ind Min = 8 | Mean Min = 6 | | |
| | | Ind Max = 9 | Ind Max = 9 | Ind Min = 5 | | |
| | | | | | | |
| Hydrostatic Head (cm) AATCC-127 | ≥50cm | Test Results (3 Lots) | | Specification | | Similar2 |
| | | Mean (min/max) | | | | |
| | | Sleeve | Body | | | |
{12}
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown | Predicate Device1: Cardinal HealthTM Non-Reinforced Surgical Gown K170762 | | | Comparisons |
| --- | --- | --- | --- | --- | --- |
| | | Unit = cm | | | |
| | | Mean = 89.90 | Mean = 89.90 | Target Mean = 70 | |
| | | Ind Min = 73.95 | Ind Min = 73.95 | Mean Min = 60 | |
| | | Ind Max = 100.98 | Ind Max = 100.98 | Ind Min = 53 | |
| | | | | | |
| Water Impact (g) AATCC-42 | ≤1.0g | Test Results (3 Lots) | | Specification | Similar2 |
| | | Mean (min/max) | | | |
| | | Sleeve | Body | | |
| | | Unit = g | | | |
| | | Mean = 0.10 | Mean = 0.10 | Target Mean = 0.10 | |
| | | Ind Min = 0.07 | Ind Min = 0.07 | Mean Max = 0.30 | |
| | | Ind Max = 0.13 | Ind Max = 0.13 | Ind Max = 1.0 | |
{13}
| lint generation | Standard Performance: log10 ≈ 3.5–3.6 High Performance log10 ≈ 3.2–3.3 | No available data | Different4 |
| --- | --- | --- | --- |
| Seam strength | ≥30N | No available data | Different5 |
| Performance per AAMI PB70:2012 | AAMI Level 3 | AAMI Level 3 | Same |
| Contact Durations | Surface, Intact, < 24 hours | Surface, Intact, < 24 hours | Same |
| Biocompatibility | Under the test conditions, | Under the conditions of each study, the Cardinal HealthTM non-reinforced | Same |
{14}
| Element of Comparison | Proposed Device: Copioumed AAMI 3 Surgical Gown | Predicate Device^{1}: Cardinal Health^{TM} Non-Reinforced Surgical Gown K170762 | Comparisons |
| --- | --- | --- | --- |
| | the subject device was shown to be noncytotoxic, non-irritating and non-sensitizing per ISO 10993-5, ISO 10993-23 & ISO 10993-10. | Surgical gown is non-cytotoxic, non-irritating, and non-sensitizing per ISO 10993-1. | |
Note 1: Individual Maximum (Ind. Max.): specification allows only one value at this level. Individual minimum (Ind. Min.): specification allows only one value at this level. The mean performance of the proposed device was compared to the mean specification values.
Note 2: The proposed device and predicate device passed the same criteria listed in the standard.
Note 3: We only compared the value of proposed device to the mean value of predicate device for the property. The proposed device and predicate device passed the same criteria listed in the standard.
Note 4: No publicly available lint generation data were identified for the predicate device. Lint generation testing was performed as supplementary characterization of the proposed device. This parameter is not required under AAMI PB70.
Note 5: No publicly available seam strength data were identified for the predicate device. Seam strength testing was performed as physical property characterization of the proposed device.
Page 10 of 13
{15}
J. SUMMARY OF NON-CLINICAL DATA
The following performance data were provided to demonstrate the safety and effectiveness:
| No. | Standards | Purpose | Acceptance | Results |
| --- | --- | --- | --- | --- |
| 1. | AATCC 42 | Water Resistance | ≤1.0G | Pass |
| 2. | AATCC 127 | Hydrostatic Pressure | ≥50CM H2O | Pass |
| 3. | ANSI/AAMI PB70:2012 AAMI Level 3 | Liquid Barrier Performance | Blotter weight gain ≤ 1.0 g Hydrostatic resistance ≥ 50 cm | Pass |
| 4. | NWSP 080.8.R0 | Alcohol Repellency (rating) | >6 | Pass |
| 5. | ASTM D3776/D3776M-17 | Weight of Woven Fabric | 40G±4G | Pass |
| 6. | ASTM D5034-9 2017 | Grab Tensile, Peak Stretch, and Peak Energy - Nonwovens | MD≥55N CD≥40N | Pass |
| 7. | ISO 10993-7:2008 | EO Sterilization Residuals Test | EtO < 4mg ECH < 9 mg | Pass |
| 8. | ASTM D5587-15 (2019) | Standard Test Method for Tearing Strength of Fabrics by the Trapezoid Procedure | MD≥10N CD≥10N | Pass |
| 10. | 16 CFR 1610 | Standard for the Flammability for Clothing Textiles | Class 1 | Pass |
| 11. | ISO 9073-10:2003 | Lint and other particles generation in the dry state | Standard Performance: log10 ≈ 3.5–3.6 High Performance log10 ≈ 3.2–3.3 | Pass |
| 12 | ASTM D1683 | Seam strength | ≥30N | Pass |
{16}
| 13. | ISO 10993-5:2009 | Cytotoxicity | The viability should be ≥ 70% of the blank. And the 50% extract of the test sample should have at least the same or a higher viability than the 100% extract. | Pass |
| --- | --- | --- | --- | --- |
| 14. | ISO 10993 23:2021 | Intracutaneous Irritation Test | Non-irritating | Pass |
| 15. | ISO 10993-10:2010 | Maximization Test | Non-sensitizing | Pass |
| 16. | EN ISO 11737-1 (2018/A1:2021) | Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products | ≤300 CFU/100cm2 | Pass |
K. The Conclusion drawn from the Non-Clinical tests demonstrates that the subject device Copioumed AAMI 3 Surgical Gown is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.