Li-Battery Powered OTC TENS/EMS Combination Stimulator, Model Z4
K251856 · Everyway Medical Instruments Co.,Ltd · NUH · Nov 21, 2025 · Neurology
Device Facts
Record ID
K251856
Device Name
Li-Battery Powered OTC TENS/EMS Combination Stimulator, Model Z4
Applicant
Everyway Medical Instruments Co.,Ltd
Product Code
NUH · Neurology
Decision Date
Nov 21, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator is intended for over-the-counter (OTC) use. It integrates both TENS and EMS functions in a single unit. However, the device can only operate in one mode (either TENS or EMS) at a time; simultaneous operation of both modes is not possible. TENS is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS Modes P1 to P19 programs) EMS is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS Modes P20 to P30 programs)
Device Story
Four-channel electrical stimulation device; delivers electrical current via adhesive electrodes on user's skin. Operates in TENS mode for pain relief or EMS mode for muscle performance; cannot operate both modes simultaneously. User selects from 30 preset programs (P1-P19 TENS; P20-P30 EMS); adjusts intensity and treatment duration. Powered by rechargeable Li-Polymer battery via USB-C. Used in home/OTC setting; operated by patient. Device produces fixed, pre-configured waveforms (asymmetrical bi-phasic square pulse for TENS; pulsed monophasic for EMS). Output affects muscle/nerve tissue to provide therapeutic benefit. Healthcare provider not involved in routine operation; patient self-manages based on device interface. Benefits include non-pharmacological pain management and muscle conditioning.
Clinical Evidence
No clinical data. Bench testing only, including waveform analysis, power/current density calculations, and compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10.
Indicated for OTC use by adults for temporary relief of pain in low back, arms, and legs due to exercise or household/work strain (TENS) and for stimulation of healthy muscles to improve/facilitate muscle performance (EMS).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 21, 2025
Everyway Medical Instruments Co.,LTD
Jimmy Cheng
Vice President R&D Department of Everyway Medical Instruments Co., Ltd.
8Fl. No.5, Lane 155, Sec.3 Beishen Rd., Shenkeng Dist.
New Taipei City, 22203
Taiwan
Re: K251856
Trade/Device Name: Li-Battery Powered OTC TENS/EMS Combination Stimulator, Model Z4
Regulation Number: 21 CFR 882.5890 and 21 CFR 890.5850
Regulation Name: Transcutaneous electrical nerve stimulator for pain relief; and Powered muscle stimulator
Regulatory Class: Class II
Product Code: NUH, NGX
Dated: May 2, 2025
Received: June 17, 2025
Dear Jimmy Cheng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251856 - Jimmy Cheng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251856 - Jimmy Cheng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**XIAORUI TANG -S**
For: CDR Jitendra Virani, MS, MBA
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251856 | |
| Device Name Li-Battery Powered OTC TENS/EMS Combination Stimulator, Model Z4 | |
| Indications for Use (Describe) 1. The Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator is intended for over-the-counter (OTC) use. It integrates both TENS and EMS functions in a single unit. However, the device can only operate in one mode (either TENS or EMS) at a time; simultaneous operation of both modes is not possible. 2. TENS is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS Modes P1 to P19 programs) 3. EMS is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS Modes P20 to P30 programs) | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(K) Summary: K251856
# I. SUBMITTER
Applicant: Everyway Medical Instruments Co., LTD.
Address: 8Fl. No.5, Lane 155, Sec.3 Beishen Rd. ,Shenkeng Dist., New Taipei City, Taiwan 22203
Applicant Contact: Jimmy Cheng
Email: jimmy@everyway-medical.com
Phone Number: +886 2 26620038
Fax Number: +886 2 26645566
Date Prepared: Nov 20th, 2025
# II. DEVICE
Name of Device: Li-Battery Powered OTC TENS/EMS Combination Stimulator, Model Z4
Common or Usual Name: Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator
Regulation Names: Transcutaneous electrical nerve stimulator for pain relief; and
Powered muscle stimulator
Regulation Numbers: 21 CFR 882.5890 and 21 CFR 890.5850
Product Codes: NUH and NGX
Regulatory Class: II
# III. PREDICATE DEVICE
510(k) Number: K201335
Name of Device: Li-Battery powered OTC TENS/EMS Combination Stimulator, Models EV-906/EV-906A
Regulation Names: Transcutaneous electrical nerve stimulator for pain relief; and Powered muscle stimulator
Regulation Numbers: 21 CFR 882.5890 and 21 CFR 890.5850
Product Codes: NUH and NGX
Regulatory Class: II
# IV. DEVICE DESCRIPTION
The Li-Battery powered OTC TENS/EMS Combination Stimulator, model Z4, is a four-channel electrical stimulation device intended for pain relief and muscle training. It delivers electrical current through electrodes placed on the user's skin. Each operating mode delivers a fixed, pre-
{5}
510(K) Summary: K251856
configured waveform (e.g., asymmetrical bi-phasic square pulse) and output characteristics; users may only adjust intensity and treatment duration within predefined limits.
The device primarily consists of two components: the stimulation generator and adhesive electrodes. The generator produces the stimulation current, which is delivered to the body via lead wires connected to the electrodes. Depending on the selected mode—TENS or EMS—the device helps achieve either temporary pain relief or improved muscle performance.
The Z4 includes the following preset programs:
- TENS Mode (Programs P1–P19): Intended for temporary relief of pain in the lower back, arms, and legs due to exercise or household/work strain. Users are advised to begin with the lowest intensity and gradually increase to a comfortable “tingling” sensation. Each program differs in waveform parameters and may produce varied sensations.
- EMS Mode (Programs P20–P30): Designed to stimulate healthy muscles to enhance or support muscle performance. These programs cause the muscles to contract and relax. Users should start at low intensity for warm-up and increase gradually as needed.
The Z4 device package includes the following accessories:
- Adhesive Electrode Pads × 24
- Lead Wires × 4
- USB Cable × 1
- Instruction Manual × 1
- Carrying Case × 1
V. INDICATIONS FOR USE
1. The Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator is intended for over-the-counter (OTC) use. It integrates both TENS and EMS functions in a single unit. However, the device can only operate in one mode (either TENS or EMS) at a time; simultaneous operation of both modes is not possible.
2. TENS is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS Modes P1 to P19 programs)
3. EMS is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS Modes P20 to P30 programs)
{6}
510(K) Summary
# VI. COMPARISON WITH PREDICATE DEVICE
The subject device, Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator, is substantially equivalent to the predicate device EV-906/EV-906A (K201335) in terms of intended use, basic design, technological characteristics, and performance. Both devices are classified as Class II under product codes NUH (TENS) and NGX (EMS), and conform to the applicable regulations 21 CFR 882.5890 and 21 CFR 890.5850.
# VI.1 Intended Use Comparison
Both the subject and predicate devices combine TENS and EMS functions in a single unit. They are intended for over-the-counter (OTC) use and can only operate in one mode at a time (TENS or EMS), not simultaneously.
- TENS function: Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities.
- EMS function: Stimulation of healthy muscles to improve or facilitate muscle performance.
# VI.2 Substantial Equivalence Table
A. Basic Unit Characteristics
| Feature / Characteristic | New Device | Predicate Device | Same or Equivalent? | Explanation of Differences |
| --- | --- | --- | --- | --- |
| 510(k) Number | K251856 | K201335 | — | — |
| Device Name, Model | Li-Battery Powered OTC TENS/EMS Combination Stimulator, Model Z4 | Li-Battery Powered OTC TENS/EMS Combination Stimulator, models EV-906/EV-906A | Same | -- |
| Manufacturer | Everyway Medical Instruments Co., Ltd. | Everyway Medical Instruments Co., Ltd. | Same | -- |
{7}
510(K) Summary
| Indications for Use | 1. The Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator is intended for over-the-counter (OTC) use. It integrates both TENS and EMS functions in a single unit. However, the device can only operate in one mode (either TENS or EMS) at a time; simultaneous operation of both modes is not possible.
2. TENS is intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS Modes P1 to P19 programs)
3. EMS is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS Modes P20 to P30 programs) | •TENS: intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS Modes B/N/M/SD1/SD2 as adjustable mode and P1 to P12 preset programs only for EV-906A).
•EMS: intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS Modes C/S/A as adjustable mode and P13 to P24 preset programs only for EV-906A). | Equivalent | The differences are limited to mode naming and program numbering (e.g., EV-906A uses B/N/M/SD1/SD2 and C/S/A modes with 24 programs, while Z4 uses 30 preset programs P1–P30). These changes result from software and interface optimization for OTC use and do not affect the device’s intended use, safety, or effectiveness. |
| --- | --- | --- | --- | --- |
| Power Source(s) | Rechargeable Li-Polymer battery via USB-C | Rechargeable Li-Polymer battery via USB charging cable | Equivalent | The difference in the new device is limited to the connector type and does not impact the device’s safety and effectiveness. |
| - Method of Line Current Isolation | Type BF | Type BF | Same | -- |
| - Patient Leakage Current(Normal) | Under 0.1 | Under 0.1 | Same | -- |
| - Patient Leakage Current (Single fault) | Under 0.5 | Under 0.5 | Same | -- |
{8}
510(K) Summary
| Number of Output Modes | 2 output modes: TENS mode and EMS mode | 2 output modes: TENS mode and EMS mode | Same | -- |
| --- | --- | --- | --- | --- |
| Number of Output Channels | 4 | 4 | Same | -- |
| - Synchronous or Alternating? | Synchronous | Synchronous | Same | -- |
| - Method of Channel Isolation | Transformer-based or equivalent circuit isolation | Transformer-based or equivalent circuit isolation | Same | -- |
| Regulated Current or Regulated Voltage? | Constant Current | Constant Current | Same | -- |
| Software/Firmware/Microprocessor Control? | Yes | Yes | Same | -- |
| Automatic Overload Trip? | No | No | Same | -- |
| Automatic No-Load Trip? | Yes | Yes | Same | -- |
| Automatic Shut Off? | Yes | Yes | Same | -- |
| Patient Override Control? | No | No | Same | -- |
| Indicator Display: - On/Off Status? | Yes | Yes | Same | -- |
| - Low Battery? | Yes | No | Equivalent | The new device includes a low battery indicator. This difference does not impact the device's safety and effectiveness. |
| - Voltage/Current Level? | Yes (Intensity) | Yes (Intensity) | Same | -- |
| Timer Range (minutes) | 5–60 minutes or continuous mode | 5–60 minutes or continuous mode | Same | -- |
| Compliance with Voluntary Standards? | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 | Same | -- |
| Compliance with 21 CFR 898? | Yes | Yes | Same | -- |
{9}
510(K) Summary
B. TENS Mode Output Specifications
| Feature / Characteristic | New Device: Z4 | Predicate Device: K201335 (EV-906/EV-906A) | Same or Equivalent? | Explanation of Differences |
| --- | --- | --- | --- | --- |
| Waveform (e.g., pulsed monophasic, biphasic) | Asymmetrical Bi-Phasic Square Pulse | Pulsed monophasic | Equivalent | The new device uses an asymmetrical biphasic square pulsed waveform, while the predicate device uses a pulsed monophasic waveform. This difference does not impact the device's safety and effectiveness. |
{10}
510(K) Summary
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular | Same | -- |
| --- | --- | --- | --- | --- |
| Maximum Output Voltage (Volts) (+/-20 %) | 51.6 @500Ω
128.0V @2KΩ
218.0V @10KΩ | 51.6 @500Ω
128.0V @2KΩ
218.0V @10KΩ | Same | -- |
| Maximum Output Current (mA) (+/-20 %) | 103.2mA @500Ω
64.0mA @2KΩ
21.8mA @10KΩ | 103.2mA @500Ω
64.0mA @2KΩ
21.8mA @10KΩ | Same | -- |
| Duration of primary phase (uS) | 300 max.
(Single phase) | 300 max.
(Single phase) | Same | -- |
| Pulse Duration (mS) | 45 max. | 45 max. | Same | -- |
| Frequency (Hz) [or Rate (pps)] | 150 max. | 150 max. | Same | -- |
| For multiphasic waveforms only: | Symmetrical phases? | No
Asymmetrical phases | No
Asymmetrical phases | -- |
| Maximum Charge (Microcoulombs per pulse)
(if zero, state method of achieving zero net charge for net charge/pulse) | 14.97 | 14.97 | Same | -- |
| - Maximum Phase Charge (uC) | 6.57 | 6.57 | Same | -- |
{11}
510(K) Summary
C. EMS Mode Output Specifications
| Feature / Characteristic | New Device: Z4 | Predicate Device: K201335 (EV-906/EV-906A) | Same or Equivalent? | Explanation of Differences |
| --- | --- | --- | --- | --- |
| Waveform (e.g., pulsed monophasic, biphasic) | Pulsed monophasic | Pulsed monophasic | Same | -- |
| - Maximum Current Density(mA/cm^{2}) | 0.09 | 0.09 | Same | -- |
| --- | --- | --- | --- | --- |
| - Maximum Average Current (mA) | 2.25 | 2.25 | Same | -- |
| - Maximum Average Power Density (mW/cm^{2}) | 0.98 | 0.98 | Same | -- |
| Average DC current through electrodes when device is on but no pulses are being applied (uA) | Not applicable | Not applicable | Same | -- |
| Pulse per burst | 9 | 9 | Same | -- |
| Burst per second (Hz) | 0.5~5 | 0.5~5 | Same | -- |
| Bust duration (uS) | 300 | 300 | Same | -- |
| Duty Cycle | Same for each program | Same for each program | Same | -- |
| Method of achieving zero net charge for net charge/pulse | Monophasic asymmetric wave for each pulse | Monophasic asymmetric wave for each pulse | Same | -- |
{12}
510(K) Summary
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular | Same | -- |
| --- | --- | --- | --- | --- |
| Maximum Output Voltage (Volts) (+/-20 %) | 50.8@500Ω
127.0V @2KΩ
212.0V @10KΩ | 52.0 @500Ω
128.0V @2KΩ
216.0V @10KΩ | Equivalent | Minor variation within ±20% tolerance; no impact on safety or effectiveness |
| Maximum Output Current (mA) (+/-20 %) | 101.6mA @500Ω
63.5mA @2KΩ
21.2mA @10KΩ | 104.0mA @500Ω
64.0mA @2KΩ
21.6mA @10KΩ | Equivalent | There is a minor difference in the maximum output current for the new device. This difference does not impact safety and effectiveness. |
| Duration of primary phase (uS) | 300 max.
(Single phase) | 300 max.
(Single phase) | Same | -- |
| Pulse Duration (mS) | 45 max. | 45 max. | Same | -- |
| Frequency (Hz) [or Rate (pps)] | 150 max. | 150 max. | Same | -- |
| For multiphasic waveforms only: | Symmetrical phases? | No
Asymmetrical phases | No
Asymmetrical phases | -- |
| Maximum Charge (Microcoulombs per pulse)
(if zero, state method of achieving zero net charge for net charge/pulse) | 14.87 | 15.02 | Equivalent | Difference within acceptable range; does not affect safety or effectiveness |
{13}
510(K) Summary
| - Maximum Phase Charge (uC) | | 6.47 | 6.62 | Equivalent | This difference does not impact safety and effectiveness |
| --- | --- | --- | --- | --- | --- |
| - Maximum Current Density(mA/cm²) | | 0.09 | 0.09 | Same | -- |
| - Maximum Average Current (mA) | | 2.23 | 2.25 | Equivalent | This difference does not impact safety and effectiveness |
| - Maximum Average Power Density (mW/cm²) | | 0.96 | 0.99 | Equivalent | This difference does not impact safety and effectiveness |
| Burst Mode (i.e. pulse trains) | Pulse per burst | No burst mode | No burst mode | Same | -- |
| | Burst per second | | | | |
| | Bust duration | | | | |
| | Duty Cycle | | | | |
| On Time (Second) | | 2~99 | 2~99 | Same | -- |
| Off Time (second) | | 0~99 | 0~99 | Same | -- |
| Ramp Time (second) | | 1~8 must less than On time | 1~8 must less than On time | Same | -- |
VI.3 Output Safety & Performance Comparison
Based on waveform and Vrms/Arms analysis across different load impedances (500Ω, 2kΩ, 10kΩ), the subject device meets IEC 60601 safety limits and demonstrates equivalent.
- Maximum average current
- Maximum current density
- Maximum power density
Despite minor variations in waveform parameters and program configurations, performance testing and waveform comparisons confirm that these differences do not raise new safety or effectiveness concerns.
{14}
510(K) Summary
## VI.4 Conclusion
The subject device Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator is substantially equivalent to the cited predicate device (K201335) in design, function, and safety. All differences, including waveform shape, pulse duration, output values, and interface features, were evaluated through bench testing, waveform analysis, and power/current density calculations. Results confirm that the subject device performs comparably and meets safety standards under normal and worst-case operating conditions.
## VII. PERFORMANCE DATA
### VII.1 Non-Clinical Tests Performed
Compliance to applicable voluntary standards include: IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices.
### VII.2 Clinical Testing
No clinical studies were required or conducted.
## VIII. CONCLUSION
The subject device, Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator, has been demonstrated to be substantially equivalent to the cited predicate device (K201335), in terms of intended use, design, technological characteristics, and performance.
From a clinical perspective, both devices are intended for the same over-the-counter (OTC) applications: the temporary relief of pain using TENS programs and the stimulation of healthy muscles using EMS programs. They are intended for use by the same user group, under similar conditions and environments.
From a technical perspective, both devices operate based on similar principles and deliver electrical stimulation via electrodes through pre-programmed waveforms. The differences in waveform configuration, pulse duration, output voltage/current, timer settings, device dimensions, weight, and number of control buttons were carefully evaluated and validated through applicable safety and performance testing, including compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10. These differences were determined to have no impact on the safety or effectiveness of the device.
{15}
510(K) Summary
From a biological perspective, the materials used in patient-contacting components, their surface characteristics, contact duration, and location of contact are identical between the subject and predicate devices. Therefore, no additional biological risk is introduced by the subject device.
In conclusion, the differences in waveform parameters, device dimensions, user interface features, and output specifications do not affect the product's safety or performance. Performance testing has been completed to demonstrate that the subject device, Z4 Li-Battery Powered OTC TENS/EMS Combination Stimulator, is substantially equivalent to the cited predicate device (K201335).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.