DeGen Medical Patient Specific Implant (PSI) System
K251829 · Degen Medical · OVD · Dec 8, 2025 · Orthopedic
Device Facts
Record ID
K251829
Device Name
DeGen Medical Patient Specific Implant (PSI) System
Applicant
Degen Medical
Product Code
OVD · Orthopedic
Decision Date
Dec 8, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Solar-S™ PSI (Anterior, Standalone, With Integrated Fixation) and Solar-A™ PSI Spacers (Anterior, Non-Standalone, With Integrated Fixation) The Solar-S™ PSI (Standalone) and Solar-A™ PSI (Non-Standalone) are lumbar interbody fusion devices intended for use at one or more levels of the lumbosacral spine (L1-S1) in patients with degenerative disc disease (DDD), disc herniation (myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-S™ PSI and Solar-A™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. Solar-S™ PSI Used with Screws or Tusks When used with three (3) screws, interbody devices with a lordotic angle ≤ 20° can be used as standalone interbody fusion devices at 1 or 2 contiguous levels. When used with three (3) tusks, interbody devices must always be used with supplemental fixation and may be used at one or more levels. Hyperlordotic interbody devices (≥20° lordosis) used with screws or tusks, must always be used with supplemental fixation, and may be used at one or more levels. When used with supplemental fixation, all Solar-S™ PSI Spacers used with screws or tusks may be used at one or more levels. Solar-A™ PSI: These devices are intended to be used at one or more levels with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. Solar-M™ PSI Spacers (Anterior, Without Integrated Fixation) Solar-M™ PSI Spacers are lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-M™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. Impulse™ PSI Spacers (Posterior / Transforaminal) The Impulse™ PSI is an intervertebral body fusion device intended for use in skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or more levels in the lumbosacral spine (L1-S1) as an adjunct to fusion. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The Impulse™ PSI is indicated to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and are intended to be used with supplemental fixation systems cleared for use in the lumbar spine. The device is to be used in patients who have had six months of nonoperative treatment.
Device Story
Patient-specific lumbar/thoracic interbody spacers; additively manufactured from Puri-Ti unalloyed titanium. Inputs: patient imaging data (CT, MRI, X-ray). Transformation: custom design of dimensions (height, lordosis, width, length) and patient-matched endplates based on imaging. Output: final geometry implant; not intended for intraoperative alteration. Used by surgeons in spinal fusion procedures. Integrated fixation (screws/tusks) available for specific configurations. Porous endplates mitigate subsidence/expulsion. Central aperture holds bone graft. Benefits: patient-matched fit to anatomy.
Clinical Evidence
Bench testing and cadaver validation study performed. Results demonstrated that expanded CT scan parameters for patient-matched endplate design did not impact the ability to implement the intended surgical plan.
Technological Characteristics
Additively manufactured Puri-Ti unalloyed titanium; integrated screws (Ti-6AL-4V ELI per ASTM F136). Porous surface endplates. Patient-specific geometry (height, lordosis, width, length). Standalone or supplemental fixation configurations.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD), disc herniation, spondylolisthesis/retrolisthesis, deformity (scoliosis/kyphosis), spinal stenosis, or pseudarthrosis requiring lumbar (L1-S1) or thoracic (T1-T12) interbody fusion. Requires 6 months non-operative treatment. Contraindications include skeletal immaturity.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
DeGen Medical Patient-Specific Implant (PSI) System (K250667)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
December 8, 2025
DeGen Medical
% Justin Gracyalny, MSE
Regulatory Affairs Program Manager
Secure BioMed Evaluations
7828 Hickory Flat Highway Suite 120
Woodstock, Georgia 30188
Re: K251829
Trade/Device Name: DeGen Medical Patient Specific Implant (PSI) System
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: OVD, MAX, PHM
Dated: November 10, 2025
Received: November 10, 2025
Dear Justin Gracyalny:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K251829 - Justin Gracyalny
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251829 - Justin Gracyalny
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251829 | ? |
| Please provide the device trade name(s). | | ? |
| DeGen Medical Patient Specific Implant (PSI) System | | |
| Please provide your Indications for Use below. | | ? |
| Solar-S™ PSI (Anterior, Standalone, With Integrated Fixation) and Solar-A™ PSI Spacers (Anterior, Non-Standalone, With Integrated Fixation)
The Solar-S™ PSI (Standalone) and Solar-A™ PSI (Non-Standalone) are lumbar interbody fusion devices intended for use at one or more levels of the lumbosacral spine (L1-S1) in patients with degenerative disc disease (DDD), disc herniation (myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-S™ PSI and Solar-A™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. | | |
| Solar-S™ PSI Used with Screws or Tusks
When used with three (3) screws, interbody devices with a lordotic angle ≤ 20° can be used as standalone interbody fusion devices at 1 or 2 contiguous levels.
When used with three (3) tusks, interbody devices must always be used with supplemental fixation and may be used at one or more levels.
Hyperlordotic interbody devices (≥20° lordosis) used with screws or tusks, must always be used with supplemental fixation, and may be used at one or more levels.
When used with supplemental fixation, all Solar-S™ PSI Spacers used with screws or tusks may be used at one or more levels. | | |
| Solar-A™ PSI:
These devices are intended to be used at one or more levels with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. | | |
| Solar-M™ PSI Spacers (Anterior, Without Integrated Fixation)
Solar-M™ PSI Spacers are lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-M™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for | | |
{4}
use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
## Impulse™ PSI Spacers (Posterior / Transforaminal)
The Impulse™ PSI is an intervertebral body fusion device intended for use in skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or more levels in the lumbosacral spine (L1-S1) as an adjunct to fusion. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The Impulse™ PSI is indicated to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and are intended to be used with supplemental fixation systems cleared for use in the lumbar spine. The device is to be used in patients who have had six months of nonoperative treatment.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{5}
# Classification
Add a primary product code and any associated product codes below. You may type in the primary product code directly (only the product code field is required) or you may filter down by choosing first a medical specialty, regulation, then product code. Please verify whether a device specific guidance or special control is available for the product code at the Product Classification website linked to below. The regulation should be consulted to determine if any special controls need to be considered (e.g. PAE and 21 CFR 890.3450). Use this website to verify information for associated product codes as well.
| Medical Specialty | Orthopedic |
| --- | --- |
| Regulation | 888.3080 - Intervertebral body fusion device |
| Product Code | OVD (Class 2) - Intervertebral Fusion Device With Integrated Fixation, Lumbar |
Information about your primary product code, such as whether any applicable guidances or standards should be consulted, is available at the website below.
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPCD/classification.cfm?ID=OVD
| Associated Product Code(s) | MAX, PHM |
| --- | --- |
DeGen Medical Patient Specific Implant (PSI) System
Page 10 of 48
{6}
K251829
Page 1 of 3
| Date | November 7, 2025 |
| --- | --- |
| Sponsor | DeGen Medical
1321-C North Cashua Drive
Florence, SC 29501
Phone 877-240-7838
Fax 843-407-0545 |
| 510(k) Contact | Secure BioMed Evaluations
Justin Gracyalny, MSE
7828 Hickory Flat Highway, Suite 120
Woodstock, GA 30188
770-837-2681
Regulatory@SecureBME.com |
| Trade Name | DeGen Medical Patient-Specific Implant (PSI) System |
| Common Name | Intervertebral body fusion device |
| Code–Classification | MAX, OVD, PHM
21 CFR 888.3080 : Class II |
| Primary Predicate | K250667 DeGen Medical Patient-Specific Implant (PSI) System |
| Device Description | The DeGen Medical Patient-Specific Implant (PSI) System is a series of patient specific lumbar interbody spacers intended for use in anterior (Solar™ PSI) or posterior / transforaminal (Impulse™ PSI) lumbar fusion procedures. The DeGen Medical PSI devices are comprised of a single component that is additively manufactured. The superior and inferior endplates feature a porous surface to mitigate subsidence and expulsion. Superior and inferior faces feature a central aperture to constrain bone graft. The Impulse™ PSI spacers are available in two configurations: Impulse™ PSI and Impulse™ PSI Hyperlordotic. The Solar™ PSI patient specific spacers are available in three configurations; Solar-S™ (integrated fixation), Solar-A™ (anterolateral, with integrated fixation), and Solar-M™ (monolithic, without integrated fixation). The DeGen Medical PSI spacers are additively manufactured from Puri-Ti™ unalloyed titanium. The DeGen Medical Solar-S™ PSI and Solar-A™ PSI incorporate integrated fixation in the form of screws manufactured from Ti-6AL-4V ELI Alloy per ASTM F136 or tusks additively manufactured from Puri-Ti™ unalloyed titanium. The Solar-S™ PSI Lumbar Interbody Spacer must be used with three (3) integrated screws and the spacer must have ≤20° of lordosis to be considered for standalone use. All implants may include patient matched endplates when provided with an adequate CT scan.
The Impulse™ and Solar™ PSIs are designed individually for each patient using patient imaging data (X-Ray, MRI, CT). Variable dimensions include anterior and posterior heights, bone graft window, lordosis, coronal angle, length, and width. When provided with an adequate CT scan, implants may also include patient-matched endplates. All implants are provided in their final geometry and are not intended to be altered or reshaped at the time of surgery. |
{7}
K251829
Page 2 of 3
| Indications for Use | Solar-S™ PSI (Anterior, Standalone, With Integrated Fixation) and Solar-A™ PSI Spacers (Anterior, Non-Standalone, With Integrated Fixation)
The Solar-S™ PSI (Standalone) and Solar-A™ PSI (Non-Standalone, With Integrated Fixation) are lumbar interbody fusion devices intended for use at one or more levels of the lumbosacral spine (L1- S1) in patients with degenerative disc disease (DDD) disc herniation (myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-S™ PSI and Solar-A™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. |
| --- | --- |
| | Solar-S™ PSI Used with Screws
When used with three (3) screws, Solar-S™ PSI Spacers devices with a lordotic angle ≤ 20° can be used as standalone interbody fusion devices at 1 or 2 contiguous levels.
When used with three (3) tusks, interbody devices must always be used with supplemental fixation and may be used at one or more levels.
Hyperlordotic interbody devices (>20° lordosis) used with screws or tusks, must always be used with supplemental fixation, and may be used at one or more levels.
When used with supplemental fixation, all Solar-S™ PSI Spacers used with screws or tusks may be used at one or more levels. |
| | Solar-A™ PSI:
These devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
Solar-M™ PSI Spacers (Anterior, Without Integrated Fixation)
Solar-M™ PSI Spacers are lumbar interbody fusion devices indicated at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis or retrolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as |
{8}
K251829
Page 3 of 3
| | discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. Solar-M™ PSI Spacers are to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.
**Impulse™ PSI Spacers (Posterior / Transforaminal)**
The Impulse™ PSI is an intervertebral body fusion device intended for use in skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or more levels in the lumbosacral spine (L1-S1) as an adjunct to fusion. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The Impulse™ PSI is indicated to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and are intended to be used with supplemental fixation systems cleared for use in the lumbar spine. The device is to be used in patients who have had six months of nonoperative treatment. |
| --- | --- |
| Technological Characteristics | The purpose of this submission was to expand the CT scan parameters suitable for design of patient matched endplates. The subject device is otherwise identical in intended use, indications for use, design, and function to the predicate device.
In conclusion, the technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. |
| Performance Testing | A cadaver validation study was performed to support substantial equivalence of the predicate device. The results of this assessment show the change in CT scan parameters did not impact the ability of the subject device to implement the intended surgical plan. The subject DeGen Medical Patient-Specific Implant (PSI) System is substantially equivalent to the predicate device. |
| Conclusions | Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate device, the subject DeGen Medical Patient-Specific Implant (PSI) System is as safe and as effective as the legally marketed predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.