UE BRONCHO Single-Use Bronchoscope: The UE BRONCHO Single-Use Bronchoscopes has been designed to be used with the UE Display, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The UE BRONCHO Single-Use Bronchoscopes is for use in a hospital environment. The UE BRONCHO Single-Use Bronchoscopes is single-use device designed for use in adults. UE Display: The UE Display is reusable digital monitor, intended to display live imaging data from UE Medical visualization devices.
Device Story
System comprises sterile, single-use flexible video bronchoscopes (EBS-380C, EBS-500C, EBS-600C) and reusable UE-M10S digital display monitor. Bronchoscopes feature distal CMOS imaging sensor and LED light source; transmit live anatomical video to monitor. Used by clinicians in hospital settings for airway/tracheobronchial tree visualization and endotherapy. Provides real-time imaging to assist clinical decision-making during bronchoscopy. Single-use design eliminates reprocessing requirements, reducing cross-contamination risk.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including optical performance (ISO 8600), electrical safety, EMC, biocompatibility, and software verification/validation.
Technological Characteristics
Flexible video bronchoscope with CMOS sensor and LED illumination. Three sizes (Slim, Regular, Large). Reusable 10.1" TFT LCD monitor. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993. Electrical/EMC: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2. Shelf life: 3 years (ASTM F1980). Packaging: ISO 11607.
Indications for Use
Indicated for use in adults for endoscopy within the airways and tracheobronchial tree in a hospital environment.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 11, 2025
Zhejiang UE Medical Corp.
% Jie Yang
Consultant
Chonconn Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong
China
Re: K251823
Trade/Device Name: UE BRONCHO Single-Use Bronchoscopes (EBS-380C); UE BRONCHO Single-Use Bronchoscopes (EBS-500C); UE BRONCHO Single-Use Bronchoscopes (EBS-600C); UE Display (UE-M10S)
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: June 13, 2025
Received: June 13, 2025
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251823 - Jie Yang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251823 - Jie Yang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
UE BRONCHO Single-Use Bronchoscopes
Page 11 of 50
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251823 | ? |
| Please provide the device trade name(s). | | ? |
| UE BRONCHO Single-Use Bronchoscopes (EBS-380C);
UE BRONCHO Single-Use Bronchoscopes (EBS-500C);
UE BRONCHO Single-Use Bronchoscopes (EBS-600C);
UE Display (UE-M10S) | | |
| Please provide your Indications for Use below. | | ? |
| UE BRONCHO Single-Use Bronchoscope:
The UE BRONCHO Single-Use Bronchoscopes has been designed to be used with the UE Display, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The UE BRONCHO Single-Use Bronchoscopes is for use in a hospital environment.
The UE BRONCHO Single-Use Bronchoscopes is single-use device designed for use in adults. | | |
| UE Display:
The UE Display is reusable digital monitor, intended to display live imaging data from UE Medical visualization devices. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2025/6/13
1. Submission sponsor
Name: Zhejiang UE Medical Corp.
Address: No.8, Youyi Road, Baita Economic Development Zone, Xianju, Zhejiang 317300, China
Contact person: Carol He
Title: Regulatory Affairs Supervisor
E-mail: hejy@ueworld.com
Tel: +86 755 8977-4077
2. Submission correspondent
Name: Chonconn Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160
3. Subject Device Information
| Trade/Device Name | UE BRONCHO Single-Use Bronchoscopes, UE Display |
| --- | --- |
| Model | UE BRONCHO Single-Use Bronchoscopes (EBS-380C);
UE BRONCHO Single-Use Bronchoscopes (EBS-500C);
UE BRONCHO Single-Use Bronchoscopes (EBS-600C);
UE Display (UE-M10S) |
| Common Name | Endoscope and accessories |
| Regulatory Class | Class II |
| Classification | 21CFR 874.4680 / Bronchoscope (Flexible Or Rigid) / EOQ |
| Submission type | Traditional 510(K) |
4. Predicate Device
510(k) number: K173727
Product name: Ambu® aScope™ 4 Broncho, Ambu® aView™ Monitor
Product code: EOQ
5. Device Description
The bronchoscope system consists of UE BRONCHO Single-Use Bronchoscopes and UE Display. The UE BRONCHO Single-Use Bronchoscopes (the bronchoscopes) are sterile,
{5}
single-use flexible video bronchoscopes available in three sizes (Slim, Regular, Large). The bronchoscopes have been designed to be used with the UE Display (reusable, non-sterile), endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The bronchoscope system is designed for use by adults in a hospital environment.
# 6. Intended use & Indication for use
UE BRONCHO Single-Use Bronchoscope:
The UE BRONCHO Single-Use Bronchoscopes has been designed to be used with the UE Display, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The UE BRONCHO Single-Use Bronchoscopes is for use in a hospital environment.
The UE BRONCHO Single-Use Bronchoscopes is single-use device designed for use in adults.
UE Display:
The UE Display is reusable digital monitor, intended to display live imaging data from UE Medical visualization devices.
7. Comparison to the Predicate Device
| Features | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| K number | K251823 | K173727 | / |
| Manufacturer | Zhejiang UE Medical Corp. | Ambu Inc. | / |
| Model | UE BRONCHO Single-Use Bronchoscopes, model EBS-380C, EBS-500C, EBS-600C UE Display, model UE-M10S. | Ambu ® aScope™ 4 Broncho System: Ambu® aScope™ 4 Broncho Slim 3.8/1.2 Ambu® aScope™ 4 Broncho Regular 5.0/2.2 Ambu® aScope™ 4 Broncho Large 5.8/2.8 Ambu® aView™ Monitor | / |
| Classification Name | Bronchoscope (flexible or rigid) and accessories | Bronchoscope (flexible or rigid) and accessories | / |
| Device trade name | UE BRONCHO Single-Use Bronchoscopes, model EBS-380C, EBS-500C, | Ambu® aScope™ 4 Broncho Slim 3.8/1.2; Ambu® | / |
{6}
| Features | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| | EBS-600C
UE Display, model UE-M10S. | aScope™ 4 Broncho Regular 5.0/2.2; Ambu® aScope™ 4 Broncho Large 5.8/2.8; Ambu® aView Monitor | |
| Product Code | EOQ | EOQ | Same |
| Indication for use | The UE BRONCHO Single-Use Bronchoscopes has been designed to be used with the UE Display, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The UE BRONCHO Single-Use Bronchoscopes is for use in a hospital environment.
The UE BRONCHO Single-Use Bronchoscopes is single-use device designed for use in adults.
UE Display:
The UE Display is reusable digital monitor, intended to display live imaging data from UE Medical visualization devices. | The aScope 4 Broncho endoscopes have been designed to be used with the aView monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The aScope 4 Broncho system are for use in a hospital environment.
The aScope 4 Broncho are single-use devices designed for use in adults. | Same |
| Population | Adults | Adults | Same |
| Anatomic sites | airways and tracheobronchial tree | airways and tracheobronchial tree | Same |
| Rx only | Yes | Yes | Same |
| Technology | The Flexible bronchoscope is inserted through the airways and tracheobronchial tree during Bronchoscopy. Anatomical | The Flexible bronchoscope is inserted through the airways and tracheobronchial tree during Bronchoscopy. | Same |
3 / 6
{7}
| Features | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| | images are transmitted to the user by the video processor with a CMOS chip at the distal end of the endoscope and the images showing on a monitor. | Anatomical images are transmitted to the user by the video processor with a CMOS chip at the distal end of the endoscope and the images showing on a monitor. | |
| Performance | Complies with: ISO 8600 | Complies with: ISO 8600 | Same |
| Field of view (degree) | 90° | 85° | Similar |
| Direction of view (degree) | 0° | 0° | Same |
| Depth of view | 6-50 mm | 6-50 mm | Same |
| Working length (mm) | 600 mm | 600mm | Same |
| Deflection angle | 180° up, 180° down | Slim: 180° up, 180° down
Regular: 180° up, 180° down
Large: 180° up, 160° down | Same |
| Insertion Tube Outer Diameter | EBS-380C: 3.8 mm
EBS-500C: 5.0 mm
EBS-600C: 5.8 mm | Slim: 3.8 mm
Regular: 5.0 mm
Large: 5.8 mm | Same |
| Insertion Portion Maximum Diameter | EBS-380C: 4.4 mm
EBS-500C: 5.5 mm
EBS-600C: 6.5 mm | Slim: 4.3 mm
Regular: 5.5 mm
Large: 6.3 mm | Equivalent |
| Minimum Working Channel Diameter | EBS-380C: 1.15 mm
EBS-500C: 2.15 mm
EBS-600C: 2.75 mm | Slim: 1.2 mm
Regular: 2.0 mm
Large: 2.6 mm | Equivalent |
| Average Working Channel Diameter | EBS-380C: 1.2 mm
EBS-500C: 2.2 mm
EBS-600C: 2.8 mm | Slim: 1.2 mm
Regular: 2.2 mm
Large: 2.8 mm | Same |
{8}
| Features | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Minimum | EBS-380C: 5.0 mm | Slim: 5.0 mm | Same |
| ETT Inner | EBS-500C: 6.0 mm | Regular: 6.0 mm | |
| Diameter Size | EBS-600C: 7.0 mm | Large: 7.0 mm | |
| Minimum | EBS-380C: 35 Fr | Slim: 35 Fr | Similar |
| DLT Inner | EBS-500C: N/A | Regular: 41 Fr | |
| Diameter Size | EBS-600C: N/A | Large: N/A | |
| Light Source | LED | LED | Same |
| Sterilization | EO | EO | Same |
| Imaging Sensor | CMOS | CMOS | Same |
| Single-Use | Yes | Yes | Same |
| Image Display | Displays image on a reusable video monitor (UE Display) | Displays image on a reusable video monitor (aView™ Monitor) | Same |
| Maximum resolution | 1920*1200 | 800 x 480 | Different |
| Display Size | 10.1" | 8.5" | Similar |
| Display Type | 10.1" color TFT LCD | 8.5" color TFT LCD | Similar |
| Brightness control | Yes | Yes | Same |
| Contrast control | Yes | Yes | Same |
| Storage Capacity | 128 GB | 8 GB | Different |
| Display-Operating Environment | 5 - 40°C | 10 - 40°C | Similar |
| | ≤85% | 30 - 85% | Similar |
| | 860hPa ~ 1060hPa | 800 - 1090 hPa | Similar |
| Dimensions (W x H x D) | 257 mm x 195 mm x 36.2 mm | 241 mm x 175 mm x 33.5 mm | Similar |
| Weight | 1.50 kg | 1.50 kg | Same |
| Battery Type | Lithium battery | Lithium battery | Same |
# 8. Non-clinical Data
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
Biocompatibility of the UE BRONCHO Single-Use Bronchoscopes was evaluated in
{9}
accordance with the FDA guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA.
## Sterilization and shelf life testing
The UE BRONCHO Single-Use Bronchoscopes are provided sterile, and its shelf-life is 3 years.
Sterilization Process has been validated accordance with ISO 11135:2014.
EO/ECH residual test was performed according to ISO 10993-7.
The shelf life is determined based on optical testing and product performance testing after accelerated aging test according to ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
Package validation was conducted according to ISO 11607-1 and ISO 11607-2.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Bronchoscope system. The system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.
## Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”.
## Bench performance testing
The following bench tests were performed:
1. Optical performance testing according to ISO 8600 series.
2. Color performance (color reproduction), geometric distortion, optical performance (resolution, depth of field and image intensity uniformity), Noise and dynamic range, Frame rate compared with the predicate device.
## 9. Clinical study
Not applicable.
## 10. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed Bronchoscope system is substantially equivalent to the predicate device.
6 / 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.