K251818 · Hudens Bio Co., Ltd. · LYC · Mar 6, 2026 · Dental
Device Facts
Record ID
K251818
Device Name
BONTREE PLUS
Applicant
Hudens Bio Co., Ltd.
Product Code
LYC · Dental
Decision Date
Mar 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BONTREE PLUS is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects, including - Periodontal/infrabony defects - Ridge augmentation - Extraction sites (implant preparation/placement) - Sinus lifts - Cystic cavities
Device Story
BONTREE PLUS is a synthetic bone graft substitute; composed of hydroxyapatite (HA) and octacalcium phosphate (OCP). Supplied as granules in vials; sterile; single-use; prescription-only. Used by dental surgeons in clinical settings to fill or reconstruct oral/maxillofacial bone defects. Operates as an osteoconductive scaffold; provides structural support for cell migration and bone ingrowth; gradually resorbs over time. Healthcare providers pack the granules into the surgical site; facilitates natural bone regeneration. Benefits patient by restoring bone volume in periodontal or maxillofacial defects.
Clinical Evidence
No human clinical data. Evidence includes biocompatibility testing (ISO 10993-1) and a canine mandibular 1-wall intrabony defect model. Study compared BONTREE PLUS to predicate OSTEON II at 4, 8, and 12 weeks. Endpoints included new bone formation (micro-CT, histomorphometry), material degradation, and inflammatory response. Results showed comparable bone regeneration to predicate with minimal inflammatory response.
Technological Characteristics
Synthetic bone graft; 30wt% HA and 70wt% OCP. Granule form; particle size 0.3-1.4mm; porosity ≥40%. Sterilized via gamma radiation (SAL 10^-6). Mechanical integrity verified via compression testing (ISO 13175-3). Packaging validated per ISO 11607. No software or electronic components.
Indications for Use
Indicated for patients requiring bone grafting to fill, augment, or reconstruct periodontal or oral/maxillofacial defects, including periodontal/infrabony defects, ridge augmentation, extraction sites, sinus lifts, and cystic cavities.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
HUDENS BIO Co., Ltd.
March 06, 2026
% Mikael Hellstrand
RA Manager
K-Bio Solutions
Suite 701,125 Seochojungang-ro, Seocho-gu
Seoul, 06644
KOREA, SOUTH
Re: K251818
Trade/Device Name: Bontree Plus
Regulation Number: 21 CFR 872.3930
Regulation Name: Bone Grafting Material
Regulatory Class: Class II
Product Code: LYC
Dated: February 4, 2026
Received: February 6, 2026
Dear Mikael Hellstrand:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K251818 - Mikael Hellstrand
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K251818 - Mikael Hellstrand
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Elitzer
for Andrew Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251818 | |
| Device Name BONTREE PLUS | |
| Indications for Use (Describe) BONTREE PLUS is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects, including - Periodontal/infrabony defects - Ridge augmentation - Extraction sites (implant preparation/placement) - Sinus lifts - Cystic cavities | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
510(k) Summary
Page 1 / 10
The assigned 510(k) Number: K251818
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
- Sponsor Name: HUDENS BIO Co., Ltd.
- Address: 318 Cheomdanyeonsin-ro, Buk-gu, Gwangju, Republic of Korea
- Postal Code: 61088
# Application Consulting Correspondent:
- Contact Person: Mr. Kyungyoon Kang
- Company: KBIO Solutions
- Address: 125 Seochojungang-ro, Seocho-gu, Seoul, Republic of Korea
- Tel: USA: 408-750-7843, KOR: +82-10-3648-7021
- E-mail: kyungyoon.kang@kbiotechsolutions.com
2. Date of the summary prepared: March 6, 2026
# 3. Subject Device Information
- Type of 510(k): Traditional
- Trade Name: BONTREE PLUS
- Classification Name: Bone grafting material, Synthetic
- Review Panel: Dental
- Product Code: LYC
- Regulation: 21 CFR 872.3930
- Regulatory Class: Class II
# 4. Predicate Device Information
- Sponsor: GENOSS Co., Ltd.
- Trade Name: OSTEON II
- Classification Name: Bone grafting material, Synthetic
- 510(k) Number: K112716
- Product Code: LYC
- Regulation Number: 21 CFR 872.3930
- Regulation Class: Class II
{6}
510(k) Summary
Page 2 / 10
# 5. Reference Device Information
- Sponsor: Tissue Regeneration Systems, Inc.
- Trade Name: TRS PCL Cranial Bone Void Filler
- Classification Name: Burr Hole Cover
- 510(k) Number: K123633
- Product Code: GXR
- Regulation Number: 21 CFR 882.5250
- Regulation Class: Class II
# 6. Device Description
BONTREE PLUS is a synthetic bone graft substitute composed of hydroxyapatite (HA) and octacalcium phosphate (OCP). BONTREE PLUS is available in granule type ranging in weight (g)/volume (cc) from 0.15g/0.3cc to 1.0g/2.0cc, and particle size ranges from 0.3mm to 1.4mm. BONTREE PLUS is supplied in vials as a single-use, prescription-only, sterile device (sterilization by gamma radiation).
# 7. Indications for Use
BONTREE PLUS is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects, including
- Periodontal/infrabony defects
- Ridge augmentation
- Extraction sites (implant preparation/placement)
- Sinus lifts
- Cystic cavities
# 8. Comparison to Predicate Device and Reference Device
The differences between the subject device, predicate device, and reference device do not raise new issues of safety or effectiveness.
| Comparison Item | Proposed Device, BONTREE PLUS | Predicate Device, OSTEON II, K112716 | Reference Device, TRS PCL Cranial Bone Void Filler, K123633 | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| 510(k) Number | K251818 | K112716 | K123633 | 510(k) Number |
| Trade/Device Name | BONTREE PLUS | OSTEON II | TRS PCL Cranial Bone Void Filler | Device Name |
{7}
510(k) Summary
Page 3 / 10
| Comparison Item | Proposed Device, BONTREE PLUS | Predicate Device, OSTEON II, K112716 | Reference Device, TRS PCL Cranial Bone Void Filler, K123633 | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| Regulation Number | 21 CFR 872.3930 | 21 CFR 872.3930 | 21 CFR 882.5250 | Regulation Number |
| Regulation Name | Bone grafting material | Bone grafting material | Burr hole cover | Same as the predicate, K112716 |
| Regulatory Class | Class II | Class II | Class II | Same as the predicate, K112716 |
| 510(k) Review Panel | Dental | Dental | Neurology | Same as the predicate, K112716 |
| Product Code | LYC | LYC | GXR | Same as the predicate, K112716 |
| Manufacturer | HUDENS BIO Co., Ltd. | GENOSS Co., Ltd. | Tissue Regeneration Systems, Inc. | Manufacturer |
| Indications for Use | Intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects.
• Periodontal/infrabony defects
• Ridge augmentation
• Extraction Sites (implant preparation/placement)
• Sinus lifts
• Cystic cavities | Intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects.
• Periodontal/infrabony defects
• Ridge augmentation
• Extraction Sites (implant preparation/placement)
• Sinus lifts
• Cystic cavities | Intended for use in the repair of 13mm neurosurgical cranial burr holes. It should be gently packed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. | Same as the predicate, K112716 |
| Surgical Site | Oral, periodontal | Oral, periodontal | Cranial, skeletal | Same as the predicate, K112716 |
| Components | • Hydroxyapatite (HA)
• Octacalciu | • Hydroxyapatite (HA)
• Beta- | • Hydroxyapat | BONTREE PLUS uses OCP instead of β-TCP, though, both materials are well-established, resorbable calcium phosphates with similar osteoconductive functions and safety profiles. In |
{8}
510(k) Summary
Page 4 / 10
| Comparison Item | Proposed Device, BONTREE PLUS | Predicate Device, OSTEON II, K112716 | Reference Device, TRS PCL Cranial Bone Void Filler, K123633 | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| | m phosphate (OCP) | tricalcium phosphate (β-TCP) | ite (HA)
• Octacalcium phosphate (OCP)
• Polycaprolactone (PCL) | addition, OCP is the same calcium phosphate component used in the cleared reference device (TRS PCL Cranial Bone Void Filler, K123633), further supporting its regulatory precedent and safety profile in bone repair applications. Biocompatibility testing and preclinical performance data of the finished device of BONTREE PLUS confirm that this material difference does neither alter the device’s safety or effectiveness profile nor affect substantial equivalence with the predicate, K112716.
Accordingly, the material difference does not raise different questions of safety and effectiveness, and BONTREE PLUS remains substantially equivalent to the predicate device under 21 CFR 872.3930. |
| Material Composition | • HA: 30wt%
• OCP: 70wt% | • HA: 25~45wt%
• β-TCP: 55~75wt% | N/A | With respect to the Material Composition Ratio Difference,
BONTREE PLUS contains 30 wt% HA and 70 wt% OCP, while the predicate contains 25~45 wt% HA and 55~75 wt% β-TCP. This difference does not raise new questions of safety or effectiveness:
The HA content (30 wt%) falls within the predicate’s HA range, and the resorbable phase is present at comparable proportions. OCP and β-TCP are well-established, resorbable calcium phosphates with similar osteoconductive functions. The biphasic calcium phosphate design and mechanism of action remain unchanged.
Both devices consist entirely of calcium phosphate without novel additives. ISO 10993 biocompatibility and preclinical data of BONTREE PLUS confirm no new risks compared to the predicate. Therefore, BONTREE PLUS remains substantially equivalent to OSTEON II (K112716) under 21 CFR 872.3930. |
| Porosity | 40% or more | 70% | N/A | Although BONTREE PLUS has a porosity of ≥40% compared to 70% for OSTEON II (K112716), this difference does not raise new questions of safety or effectiveness.
BONTREE PLUS maintains an interconnected porous structure adequate for cell migration, fluid exchange, and bone regeneration while preserving mechanical integrity. Preclinical data demonstrate comparable bone ingrowth and remodeling without |
{9}
510(k) Summary
Page 5 / 10
| Comparison Item | Proposed Device, BONTREE PLUS | Predicate Device, OSTEON II, K112716 | Reference Device, TRS PCL Cranial Bone Void Filler, K123633 | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| | | | | new risks. Therefore, intended use, technological characteristics, and safety/effectiveness remain unchanged, and substantial equivalence is maintained under 21 CFR 872.3930. |
| Pore Size | 0.1μm or more | 250μm | N/A | Although BONTREE PLUS specifies a pore size of ≥0.1 μm and OSTEON II (K112716) reports a macropore size of ~250 μm, this difference does not raise new safety or effectiveness concerns. The 0.1 μm value represents the minimum detectable pore structure, and BONTREE PLUS also contains an interconnected porous architecture that supports osteoconduction and fluid exchange.
The predicate’s value reflects its dominant macropore size; both devices function as porous calcium phosphate scaffolds for bone ingrowth and resorption. Preclinical data confirm comparable bone regeneration without new risks. Therefore, substantial equivalence is maintained under 21 CFR 872.3930. |
| Particle Size | 0.3~1.4mm | 0.2~2.0mm | N/A | BONTREE PLUS has a particle size range of 0.3–1.4 mm, fully within the predicate OSTEON II (K112716) range of 0.2–2.0 mm. Thus, it represents a subset of previously cleared dimensions.
This ensures comparable handling, packing, and osteoconductive surface area. Because the particle size falls within an established clinical range for calcium phosphate grafts, it raises no new safety or effectiveness concerns, and substantial equivalence is maintained under 21 CFR 872.3930. |
| Ca/P Ratio | 1.4 ± 0.1 | N/A | N/A | Although the predicate does not specify a Ca/P ratio, the 1.4 ± 0.1 ratio of BONTREE PLUS is consistent with the stoichiometric range of octacalcium phosphate and within the established spectrum of resorbable calcium phosphate grafts.
Such variations do not alter the osteoconductive mechanism based on scaffold-mediated bone ingrowth and gradual resorption. Biocompatibility and preclinical data confirm no adverse impact on safety or performance. Therefore, this specification does not raise new questions of safety or effectiveness compared to K112716. |
{10}
510(k) Summary
Page 6 / 10
| Comparison Item | Proposed Device, BONTREE PLUS | Predicate Device, OSTEON II, K112716 | Reference Device, TRS PCL Cranial Bone Void Filler, K123633 | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| Crystalline Phase Composition | ≥ 60% | N/A | N/A | Although the predicate does not specify crystalline phase composition, the ≥60% crystalline phase in BONTREE PLUS is consistent with sintered HA- and OCP-based calcium phosphate grafts. The crystalline structure supports mechanical integrity and controlled resorption without changing the osteoconductive mechanism.
While crystallinity may influence resorption rate, it does not introduce new biological risks. Biocompatibility and preclinical data confirm no adverse impact on safety, tissue response, or bone regeneration. Therefore, this characteristic does not raise new questions of safety or effectiveness. |
| Crystallinity | 60% or more | N/A | N/A | Although the predicate does not specify crystalline phase composition, the ≥60% crystalline phase of BONTREE PLUS is consistent with synthetic calcium phosphate graft materials. It provides structural stability and predictable resorption while maintaining the same osteoconductive mechanism as HA-based grafts.
Differences in crystallinity do not introduce new constituents or modes of action. Biocompatibility and preclinical studies confirm no adverse impact on safety or performance. Therefore, this specification does not raise new questions of safety or effectiveness compared to K112716. |
| pH | pH 0.41, pH difference shall be ≤ 1.5. | N/A | N/A | Although the predicate does not specify extraction pH, the ≤1.5 pH difference of BONTREE PLUS falls within the acceptable range for implantable calcium phosphate grafts and reflects normal material–fluid interaction. Such minor variations are not expected to cause tissue irritation due to physiological buffering at the implant site.
ISO 10993 biocompatibility testing, including cytotoxicity and irritation assessments, showed no adverse tissue response. Therefore, the pH specification does not introduce new risks or raise new questions of safety or effectiveness. |
| Supplied Sterile; | Supplied sterile; Gamma radiation | Supplied sterile; Gamma radiation | Supplied sterile; Ethylene | Same as the predicate, K112716 |
{11}
510(k) Summary
Page 7 / 10
| Comparison Item | Proposed Device, BONTREE PLUS | Predicate Device, OSTEON II, K112716 | Reference Device, TRS PCL Cranial Bone Void Filler, K123633 | Substantial Equivalence Discussion |
| --- | --- | --- | --- | --- |
| Sterilization Method | | | Oxide | |
| Packaging | Glass vial | Glass vial | N/A | Same as the predicate, K112716 |
| Shelf-life | 2 years | N/A | N/A | Labeled shelf life |
## 9. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## A. Biocompatibility Testing
BONTREE PLUS falls under the category of implant device in contact with tissue and/or bone with permanent contact duration (>30 days) (FDA 2023). Based on this categorization, BONTREE PLUS has been evaluated on a risk basis for biocompatibility endpoints according to ISO 10993-1 and FDA’s modified biocompatibility matrix Table A.1 as listed in Attachment A of the FDA guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
The battery of biocompatibility testing for BONTREE PLUS included the following.
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Irritation (ISO 10993-23; intracutaneous reactivity / irritation)
- Material-mediated pyrogenicity (ISO 10993-11)
- Acute systemic toxicity (ISO 10993-11)
- Subacute / subchronic systemic toxicity (ISO 10993-11)
- Hemocompatibility (Hemolysis) (ISO 10993-4)
- Genotoxicity – Bacterial Reverse Mutation (Ames) (ISO 10993-3)
- Genotoxicity – Micronucleus Assay (ISO 10993-3)
- Genotoxicity – Chromosomal Aberration Test (ISO 10993-3)
- Implantation- Animal Performance Study, Rabbit Bone Implantation Study 13- and 26-week, (ISO 10993-6:2016)
Chronic toxicity and carcinogenicity were addressed through the chemical characterization, toxicological risk assessment, and biological risk assessment.
{12}
510(k) Summary
Page 8 / 10
## B. Animal Study
Pre-clinical study was conducted in a canine dental defect model to evaluate the in vivo bone healing/regeneration ability. The study was comparative between the subject and predicate device. Radiography, histology, and histomorphometry were performed at 4, 8, and 12 weeks post-operation.
a. The Size and Type of Defect: A canine mandibular 1-wall intrabony defect model was used. Box-shaped critical-size defects measuring $4 \times 5 \mathrm{~mm}$ (length × height) were surgically created in the mandibular alveolar bone.
b. The Types of Histology Stains that Were Done: Histological evaluation was performed using Hematoxylin and Eosin (H&E) staining to assess cellular morphology and inflammatory response, and Masson's Trichrome staining to evaluate and quantify new bone formation within the defect area.
c. Endpoints Assessed (New Bone Formation, Degradation, and Inflammatory Response): Endpoints included new bone formation quantified by micro-CT total bone volume and histomorphometric analysis of new bone area, material degradation and graft integration observed through micro-CT and histology, and inflammatory response evaluated by histological assessment and gingival inflammation scoring.
d. Conclusion (Study Results): The animal study demonstrated that BONTREE PLUS supported significant new bone formation compared with the defect-only control and showed comparable bone regeneration to the predicate, OSTEON II (K112716). Minimal inflammatory response and no material-related adverse reactions were observed, confirming equivalent biocompatibility and bone regeneration performance.
## C. Sterilization Tests
The validation of gamma irradiation sterilization was performed following ISO 11137-1 and ISO 11137-2, achieving a sterility assurance level of $10^{-6}$.
BONTREE PLUS is provided sterile and is terminally sterilized by gamma irradiation to achieve a Sterility Assurance Level (SAL) of $1 \times 10^{-6}$.
- Sterilization validation was conducted in accordance with ISO 11137-1:2006/Amd.2:2018, ISO 11137-2:2013, and ISO 11137-3:2017.
- Bioburden determination and verification sterility testing were conducted in accordance with ISO 11737-1:2018 and ISO 11737-2:2019.
{13}
510(k) Summary
Page 9 / 10
- Bacterial endotoxin testing was performed using the Limulus Amebocyte Lysate (LAL) method in accordance with USP <85> Bacterial Endotoxins Test and met the established limit of <20 EU/device.
- Additionally, rabbit pyrogen testing was conducted under GLP conditions in accordance with ISO 10993-11 and applicable pharmacopeial requirements, demonstrating absence of pyrogenic response.
Shelf-life validation was conducted to demonstrate maintenance of sterility and package integrity over the labeled shelf life. Packaging validation was performed in accordance with ISO 11607-1:2019 and ISO 11607-2:2019.
Seal strength testing was conducted per ASTM F88, dye penetration testing per ASTM F1929, and visual seal integrity testing per ASTM F1886 to verify sterile barrier integrity. These tests confirmed that the packaging system maintains sterility and device integrity throughout the labeled shelf life.
## E. Bench Performance Testing
Bench-top performance testing was conducted to characterize the physicochemical and mechanical properties of BONTREE PLUS (synthetic bone graft material), as summarized below:
- Extraction – Buffering Capacity: Conducted in accordance with USP 38 <661> The test confirms buffering interaction with aqueous environments and demonstrates chemical stability consistent with calcium phosphate bone graft materials. These results support substantial equivalence to the predicate, OSTEON II (K112716), which uses similar calcium phosphate chemistry and osteoconductive function.
- Extractable Substances (pH): Performed in accordance with ASTM F619-20 The extraction pH difference (≤1.5) reflects normal interaction between calcium phosphate materials and aqueous solutions. Although not specified for the predicate device, this value is within the acceptable range for implantable calcium phosphate grafts and supports substantial equivalence with the predicate, OSTEON II (K112716).
- Physical Characterization: Evaluated in accordance with ISO 2591-1 Physical characterization testing confirms that the particle morphology, granule structure, and material consistency of BONTREE PLUS are comparable to those of cleared calcium phosphate bone graft materials. These characteristics are consistent with the technological properties of the predicate OSTEON II (K112716), and support substantial equivalence.
- Scanning Electron Microscopy (SEM) – Dimensional Analysis: Surface morphology and particle dimensions were evaluated using SEM in accordance with the laboratory’s validated test method for dimensional measurement SEM analysis confirmed a porous
{14}
510(k) Summary
Page 10 / 10
calcium phosphate microstructure consistent with osteoconductive bone graft materials. The observed morphology supports the same scaffold-based bone regeneration mechanism as the predicate OSTEON II (K112716).
- Elastic Modulus (Compression Testing): Determined using a Universal Testing Machine (UTM) in accordance with ISO 13175-3 Compression testing confirmed that the mechanical integrity of BONTREE PLUS is appropriate for bone graft handling and packing during surgical procedures. The mechanical characteristics are consistent with porous calcium phosphate graft materials and do not introduce different questions of safety or effectiveness relative to the predicate, OSTEON II (K112716),
- Particle Size Distribution (Sieving Method): Conducted in accordance with ISO 2591-1:1988 Testing confirmed that the particle size range of 0.3–1.4 mm falls entirely within the cleared predicate device range of 0.2–2.0 mm (OSTEON II, K112716). Because the particle size represents a subset of previously cleared dimensions, it ensures comparable handling, packing characteristics, and osteoconductive surface area, supporting substantial equivalence with the predicate, OSTEON II (K112716).
- Heavy Metal Study: Tested in accordance with the Korean Pharmacopoeia General Test Methods 66. Heavy Metal Test Methods, (the color of the test solution should not be darker than the control solution). Results demonstrated that heavy metal content remains below pharmacopeial limits and consistent with purified synthetic calcium phosphate materials used in bone grafts. These results confirm chemical purity comparable to other cleared calcium phosphate graft materials and support its substantial equivalence to the predicate OSTEON II (K112716).
All bench testing results demonstrated that the device meets its predetermined acceptance criteria and supports substantial equivalence to the predicate device.
## 10. Conclusion
The test results indicate that BONTREE PLUS is substantial equivalence to the predicate device OSTEON II.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.