← Product Code [MBI](/productcode/MBI) · K251809

# Arthrex FiberTak Suture Anchor (K251809)

_Arthrex, Inc. · MBI · Jul 18, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K251809

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jul 18, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the elbow, shoulder, hand/wrist, foot/ankle, knee, and hip the following procedures: • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction • Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty) • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-Foot Reconstruction • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis • Hip: Acetabular Labral Repair and Reconstruction

## Device Story

All-suture soft-tissue fixation device; push-in design. Consists of anchor with 1.3 mm SutureTape (polyester sheath) and polyester/UHMWPE repair suture. Preloaded on disposable inserter. Impacted into pilot hole; manual tensioning causes suture sleeve to bulge/bunch, securing anchor in bone. Suture then passed around soft tissue for fixation. Used by surgeons in clinical/OR settings. Provides mechanical fixation of soft tissue to bone to facilitate healing. Single-use; sterile.

## Clinical Evidence

Bench testing only. Straight pull testing performed per ASTM F3690. Results demonstrate performance statistically equivalent to predicate and reference devices.

## Technological Characteristics

All-suture anchor; polyester sheath and polyester/UHMWPE suture. Push-in deployment mechanism. Sterile, single-use. Testing conducted per ASTM F3690.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex FiberTak Suture Anchor ([K181769](/device/K181769.md))

## Reference Devices

- Smith & Nephew Microraptor Knotless Suture Anchor ([K181746](/device/K181746.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

July 18, 2025

Arthrex, Inc.
Alex Underberg
Regulatory Affairs Specialist, Sr.
1370 Creekside Blvd.
Naples, Florida 34108

Re: K251809
Trade/Device Name: Arthrex FiberTak Suture Anchor
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: MBI
Dated: June 6, 2025
Received: June 12, 2025

Dear Alex Underberg:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251809 - Alex Underberg
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K251809 - Alex Underberg
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CHRISTOPHER FERREIRA -S

Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

Arthrex FiberTak Suture Anchor
Page 7 of 30

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251809 | ?  |
|  Please provide the device trade name(s). |  | ?  |
|  Arthrex FiberTak Suture Anchor  |   |   |
|  Please provide your Indications for Use below. |  | ?  |
|  The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the elbow, shoulder, hand/wrist, foot/ankle, knee, and hip the following procedures:  |   |   |
|  • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction  |   |   |
|  • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction  |   |   |
|  • Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty)  |   |   |
|  • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-Foot Reconstruction  |   |   |
|  • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis  |   |   |
|  • Hip: Acetabular Labral Repair and Reconstruction  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

Arthrex

K251809

# 510(k) Summary

|  Date Prepared | 6/6/2025  |
| --- | --- |
|  Submitter | Arthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945  |
|  Contact Person | Name: Alex Underberg
Title: Regulatory Affairs Specialist, Senior
Phone: 901-606-4046
Email: Alex.Underberg@Arthrex.com  |
|  Trade Name | Arthrex FiberTak Suture Anchor  |
|  Common Name | Suture Anchor  |
|  Product Code | MBI  |
|  Classification Name | 21 CFR 888.3040: Fastener, Fixation, Nondegradable, Soft-Tissue  |
|  Regulatory Class | Class II  |
|  Primary Predicate Device | K181769: Arthrex FiberTak Suture Anchor  |
|  Additional Predicate Device | N/A  |
|  Reference Device | K181746: Smith & Nephew Microraptor Knotless Suture Anchor  |
|  Purpose of Submission | This Traditional 510(k) premarket notification is submitted to obtain clearance for the expanded indication of an Arthrex FiberTak Suture Anchor  |
|  Device Description | The fundamental technology of the Arthrex FiberTak Suture Anchors is that the Arthrex FiberTak Anchors are “all-suture” soft-tissue fixation devices with a push-in design. The subject device is FiberTak Anchor with 1.3 mm SutureTape (White/Blue) and is a knotted FiberTak with a Polyester Sheath and a Polyester/UHMWPE repair suture.

The subject device is preloaded on a disposable inserter, which facilities deployment. The subject anchor and connected sutures are impacted into a pilot hole. The suture is then manually tensioned to set the anchor by “bulging/bunching” the suture sleeve within the pilot hole. Once the anchor is set, the suture is passed around the soft tissue and is fixated via its surgical technique. The device is provided sterile and is intended for single-use.  |
|  Indications for Use | The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the elbow, shoulder,  |

Page 1 of 3

{5}

Arthrex
K251809

|   | hand/wrist, foot/ankle, knee, and hip in the following procedures:  |
| --- | --- |
|   | Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction  |
|   | Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction  |
|   | Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty)  |
|   | Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-Foot Reconstruction  |
|   | Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis  |
|   | Hip: Acetabular Labral Repair and Reconstruction  |
|  Performance Data | Straight Pull testing (method based on ASTM F3690) was conducted on the subject device. The test data demonstrates that the subject device performs statistically equivalent to the predicate and reference devices for the intended indications for use.  |
|  Technological Comparison | The subject Arthrex FiberTak Suture Anchor is identical to the primary predicate Arthrex FiberTak Suture Anchor except that it is seeking the additional Acetabular Labral Reconstruction indication. The subject Arthrex FiberTak Suture Anchor and reference device Smith & Nephew Microraptor Knotless Suture Anchor have similar technologies. Any differences between the subject and predicate devices are considered minor and do not raise any  |

Page 2 of 3

{6}

Arthrex
K251809

|  Conclusion | new or different questions concerning safety or effectiveness.
The subject Arthrex FiberTak Suture Anchor is substantially equivalent to the predicate and reference devices. The subject device and primary predicate are of identical designs. Per this 510(k) submission, Arthrex is seeking to add Acetabular Labral Reconstruction as an additional indication. The reference device, Smith & Nephew Microraptor Knotless Suture Anchor, has Acetabular Labrum Reconstruction indication, which is equivalent to the indication Arthrex is seeking. Any differences between the subject device and the predicate devices are considered minor and do not result in new or different questions concerning safety or effectiveness. Based on the indications for use, technological characteristics, and the summary of data submitted, Arthrex has determined that the subject device is substantially equivalent to the currently marketed predicate and reference devices.  |
| --- | --- |

Page 3 of 3

---

**Source:** [https://fda.innolitics.com/device/K251809](https://fda.innolitics.com/device/K251809)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com