K251802 · Merit Medical Systems, Inc. · GXI · Feb 13, 2026 · Neurology
Device Facts
Record ID
K251802
Device Name
STAR RF Ablation System
Applicant
Merit Medical Systems, Inc.
Product Code
GXI · Neurology
Decision Date
Feb 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STAR RF Ablation System is intended for ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least 6 months duration that has not responded to at least six months of conservative care, and is also accompanied by either Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change).
Device Story
Bipolar, high-frequency electrosurgical system; includes MetaSTAR RF Generator, SpineSTAR Ablation Instrument, StabiliT Introducer, and PowerCURVE Navigational Osteotome. Used in clinical settings by physicians to treat chronic low back pain. Procedure involves creating access to the vertebral body via the introducer and osteotome; placing the RF probe at the target site; and delivering RF energy to induce thermal coagulation (cellular necrosis) of the basivertebral nerve. The generator monitors temperature at the ablation zone periphery via integrated thermocouples to enable controlled ablation. Output includes real-time display of cycle timer, temperature, power level, and tissue impedance. Benefits include targeted nerve ablation for pain relief. Differences from predicate include articulation capability of the probe shaft and specific temperature monitoring configurations.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993), mechanical performance (insertion/retraction force, torque, tensile strength), electrical safety/EMC (IEC 60601 series), sterilization validation, and ex vivo thermal effect comparisons measuring lesion size. Software verification performed per IEC 62304.
Technological Characteristics
Bipolar RF electrosurgical system; 480 kHz frequency; 3-10W power output. Components: MetaSTAR generator, SpineSTAR probe, StabiliT introducer, PowerCURVE osteotome. Features integrated thermocouples for temperature monitoring. Articulating distal shaft design. Materials tested per ISO 10993. Sterilization per ISO 11135/11137. Complies with IEC 60601-1, 60601-1-2, and 60601-2-2.
Indications for Use
Indicated for adult patients with chronic low back pain (≥6 months) unresponsive to ≥6 months of conservative care, with Type 1 or Type 2 Modic changes on MRI, for basivertebral nerve ablation of L3-S1 vertebrae.
Regulatory Classification
Identification
A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.
Predicate Devices
Intracept Intraosseous Nerve Ablation System (K153272)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 13, 2026
Merit Medical Systems, Inc.
Sandeep Saboo
Principal Regulatory Affairs Specialist
1600 West Merit Parkway
South Jordan, Utah 84095
Re: K251802
Trade/Device Name: STAR RF Ablation System
Regulation Number: 21 CFR 882.4725
Regulation Name: Radiofrequency Lesion Probe
Regulatory Class: Class II
Product Code: GXI
Dated: January 15, 2026
Received: January 16, 2026
Dear Sandeep Saboo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251802 - Sandeep Saboo
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K251802 - Sandeep Saboo
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**JAIME RABEN -S**
Jaime Raben, Ph.D.
Division Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251802 | |
| Device Name STAR RF Ablation System | |
| Indications for Use (Describe) The STAR RF Ablation System is intended for ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least 6 months duration that has not responded to at least six months of conservative care, and is also accompanied by either Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251802 Page 1 of 6
| 510(k)#: | K251802 |
| --- | --- |
| 510(k) Summary | | |
| --- | --- | --- |
| General Provisions | Submitter Name: | Merit Medical Systems, Inc. |
| | Address: | 1600 West Merit Parkway South Jordan, UT 84095 |
| | Telephone Number: | (510) 468-9995 |
| | Email: | sandeep.saboo@merit.com |
| | Contact Person: | Sandeep Saboo |
| | Date of Preparation: | 12 February 2026 |
| | Registration Number: | 1721504 |
| Subject Device | Trade Name: | STAR RF Ablation System |
| | Common/Usual Name: | Probe, Radiofrequency Lesion |
| | Classification Name: | Probe, Radiofrequency Lesion |
| PREDICATE DEVICE | Trade Name: | Intracept Intraosseous Nerve Ablation System |
| | Classification Name: | Probe, Radiofrequency Lesion |
| | Premarket Notification: | K153272 |
| | Manufacturer: | Relievant Medsystems, Inc. |
| | Predicate has not been a subject of a design related recall. | |
| Classification | Class: | Class II |
| | Regulation: | 21 CFR 882.4725 |
| | FDA Product Code: | GXI |
| | Review Panel: | Neurology |
| Intended Use/Indications For Use | The STAR RF Ablation System is intended for ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least 6 months duration that has not responded to at least six months of conservative care, and is also accompanied by either Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change). | |
{5}
K251802 Page 2 of 6
## Device Description
The subject device (STAR RF Ablation System) is a bipolar, high frequency electrosurgical system comprising the SpineSTAR Ablation Instrument, MetaSTAR RF Generator, AE Cable, Hand Switch Cable, StabiliT Introducer and PowerCURVE Navigational Osteotome. The subject device is intended to produce lesions by direct application of radiofrequency currents for the relief of chronic low back pain by basivertebral nerve ablation. The MetaSTAR RF Generator applies radiofrequency (RF) energy into the SpineSTAR. StabiliT Introducer is used to create access to the vertebral body and PowerCURVE Navigational Osteotome is used to create a channel to the target site for ablation. During lesion creation, targeted tissue in vertebral body is exposed to RF energy. The application of RF energy causes a thermal reaction at the targeted tissue site to ablate the basivertebral nerve. When used as directed, the SpineSTAR Ablation Instrument is connected to the AE Cable, which is connected to MetaSTAR. A Hand Switch Cable is connected to MetaSTAR to control delivery of RF energy to the target tissue. MetaSTAR monitors the temperature at the periphery of the ablation zone to enable controlled ablation.
## Comparison to Predicate
When compared to the predicate device (Relievant Medsystems, Inc. Intracept Intraosseous Nerve Ablation System, the STAR RF Ablation System:
- Has the same indications for use;
- Has the same intended use;
- Has the same mechanism of action and principles of operation; and
- Has similar technological characteristics.
Results of verification and validation demonstrate that minor differences that may exist do not introduce new questions of safety and effectiveness. The subject device is at least as safe and effective as the predicate devices.
Provided in the tables below is a comparison of the subject and the predicate devices. Based on the comparison, the intended/indications for use and technological characteristics of the STAR RF Ablation System are substantially equivalent to the predicate (Relievant Medsystems, Inc. Intracept Intraosseous Nerve Ablation System cleared per K153272).
Intended Use/Indications For Use Comparison:
| Device Type→ | K251802
SUBJECT Device | K153272
PREDICATE Device |
| --- | --- | --- |
| Manufacturer→ | Merit Medical Systems, Inc. | Relievant Medsystems, Inc. |
| Device Names → | STAR RF Ablation System | Intracept Intraosseous Nerve Ablation System
(Relievant Intracept System) |
| Class | II | II |
| Regulation Name | Radiofrequency Lesion Probe | Radiofrequency Lesion Probe |
| Regulation Number | 882.4725 | 882.4725 |
| Product Code | GXI | GXI |
| Product Code Description | Probe, Radiofrequency Lesion | Probe, Radiofrequency Lesion |
| Intended Use | Delivery of RF energy into tissue to ablate tissue and relieve pain. | Delivery of RF energy into tissue to ablate tissue and relieve pain. |
{6}
K251802 Page 3 of 6
| Device Type→ | K251802
SUBJECT Device | K153272
PREDICATE Device |
| --- | --- | --- |
| Manufacturer→ | Merit Medical Systems, Inc. | Relievant Medsystems, Inc. |
| Device Names → | STAR RF Ablation System | Intracept Intraosseous Nerve Ablation System
(Relievant Intracept System) |
| Indications for Use: | The STAR RF Ablation System is intended for ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least 6 months duration that has not responded to at least six months of conservative care, and is also accompanied by either Type 1 or Type 2 Modic changes on an MRI such as inflammation, edema, vertebral endplate changes, disruption and fissuring of the endplate, vascularized fibrous tissues within the adjacent marrow, hypointensive signals (Type 1 Modic change), and changes to the vertebral body marrow including replacement of normal bone marrow by fat, and hyperintensive signals (Type 2 Modic change). | The Intracept Intraosseous Nerve Ablation System is intended to be used in conjunction with radiofrequency (RF) generators for the ablation of basivertebral nerves of the L3 through S1 vertebrae for the relief of chronic low back pain of at least 6 months duration that has not responded to at least six months of conservative care, and is also accompanied by either Type 1 or Type 2 Modic changes on an MRI. |
Technological Comparison
| Device Type → | K251802
SUBJECT DEVICE | K153272
PREDICATE Device |
| --- | --- | --- |
| Device Name → | STAR RF Ablation System | Intracept Intraosseous Nerve Ablation System
(Relievant Intracept System) |
| Anatomical site where used | Vertebral body | Vertebral body |
| Patient population | Adult patient with chronic lower back pain. | Adult patient with chronic lower back pain. |
| System components | • MetaSTAR RF Generator (RF Generator)
• SpineSTAR Ablation Instrument (i.e., RF Probe)
• AE Cable
• Hand Switch Cable
• StabiliT Introducer
• PowerCURVE Navigational Osteotome | • RF Generator (Stockert Neuro N50)
• Intracept Flexible Bi-Polar Probe (i.e., RF Probe)
• Interconnect Cable
• Intracept Easy Access Instrument Set (Instrument Set) |
| RF Probe access to Target location | The RF Probe is used with the StabiliT Introducer and the PowerCURVE for the placement of the RF probe into the target site. | The RF Probe is used with the Access Instrument Set, which allows for the placement of the RF probe into the target site. |
| Method of Use | Placement of the RF probe at the target tissue; delivery of RF energy into the tissue to achieve tissue ablation (i.e., Cellular necrosis through thermal coagulation). | Placement of the RF probe at the target tissue; delivery of RF energy into the tissue to achieve tissue ablation (i.e., Cellular necrosis through thermal coagulation). |
| Capability to Monitor Ablation Zone Temperature? | YES | YES |
| Max Temp Measured At Center of Abl. Zone | • 71°C to 89°C (depending on power setting and device TC used for monitoring the ablation) | • 75°C (Parameter Set #1)
• 85°C (Parameter Set #2) |
| Temp Rise Rate | • 0.2 to 2.9°C/sec (depending on power setting and device TC used for monitoring the ablation) | • 0.5 to 1°C/sec |
| Ablation Duration | • 18 to 178sec (depending on power setting used for monitoring the ablation) | • 420 to 900secs |
| Energy Type | Radiofrequency | Radiofrequency |
{7}
K251802 Page 4 of 6
| Device Type → | K251802
SUBJECT DEVICE | K153272
PREDICATE Device |
| --- | --- | --- |
| Device Name → | STAR RF Ablation System | Intracept Intraosseous Nerve Ablation System
(Relievant Intracept System) |
| Output Power Levels | 3, 5, 7.5 and 10W. | Up to 20W |
| Output Frequency | 480 kHz | 475 kHz |
| Principle of operation | Provide bipolar RF energy to the tissue between and around the electrodes to achieve tissue ablation (i.e., cellular necrosis through thermal ablation) | Provide bipolar RF energy to the tissue between and around the electrodes to achieve tissue ablation (i.e., cellular necrosis through thermal ablation) |
| Electrical Safety EMC | Conforms to IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-2 | Conforms to IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-2 |
| Information Displayed on Device Screen | • RF Cycle count-down timer
• Number of Cycles
• TOTAL RF Time (cumulative total)
• Distal TC temperature reading
• Proximal TC temperature reading
• Power Level
• Tissue Impedance | • RF Cycle count-down timer
• TC reading
• Power Delivered
• Tissue Impedance
• A chart with impedance and power during the RF Cycle. |
| Shaft OD | 3.0 mm | 1.98 mm |
| # of TCs and their position for monitoring ablation zone temperature | Incorporates 1 TC along the shaft at 5 mm from the center of the ablation zone. | Incorporates 1 TC at the center of the ablation zone. |
| Distal Shaft Design | Can be articulated (i.e., curled/bent for steering using device handle) by the user | Pre-Curved (not articulated by the user) |
| Shaft Marks | Yes (Circumferential) | Yes (Circumferential) |
| Cannula/Introducer | 10G Cannula with Bevel and Diamond Tip Stylets | 8G Cannula with Bevel and Diamond Tip Stylets |
| Access Conduit | 11G Osteotome with stainless steel shaft | 10G Curved Introducer Conduit Peek |
# Biocompatibility:
Patient contacting materials of the components of STAR RF Ablation System are classified as external communicating devices in contact with tissue/bone/dentin < 24 hours and tested for compliance with applicable ISO 10993 standards and in accordance with FDA's biocompatibility guidance and that all acceptance criteria were met.
| TEST | TEST METHOD | ACCEPTANCE CRITERIA | RESULT |
| --- | --- | --- | --- |
| Cytotoxicity-MEM | ISO 10993-5:2009 | Test article to be ≤ Grade 2. | PASS |
| Sensitization | ISO 10993-10:2021 | Test article causes no sensitization reaction based on scoring and comparison to control | PASS |
| Irritation | ISO 10993-23:2021 | Difference between the test article and control is less than or equal to 1.0 | PASS |
| Acute Systemic Toxicity | ISO 10993-11:2017 | Test article shows no greater biological reaction than animals treated with control | PASS |
| Material Mediated Pyrogenicity | ISO 10993-11:2017 | If no single animal showed a rise of 0.5°C or more above its baseline temperature, then the extract is judged nonpyrogenic. If the total maximum temperature rise of all three animals exceeded 3.3°C, the extract is judged pyrogenic. | PASS |
{8}
K251802 Page 5 of 6
Performance, Sterilization, Software, and EMC/Electrical Safety Testing:
| TEST | TEST METHOD | RESULT |
| --- | --- | --- |
| Distribution Testing, Shelf-life (Accelerated Aging) | • ASTM D4169: Standard Practice for Performance Testing of Shipping Containers and Systems
• ASTM F1980: Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices
• ISO11607-1: Requirements for materials, sterile barrier systems and packaging systems
• ASTM F2096: Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
• ASTM F88: Standard Test Method for Seal Strength of Flexible Barrier Materials | PASS |
| Dimensional | Meet dimensional specifications per product specifications (OD, ID, length, shaft marks, protrusion, tip angle, handle orientation) | PASS |
| Visual Inspection | Ensure test samples are free from surface defects and material degradation | PASS |
| Temp Accuracy | Temperature measurement accuracy using integrated thermocouples covering full range of use | PASS |
| Mechanical Testing | • Device insertion in Bone
• Device insertion and retraction force through Cannula
• Tensile strength
• Torque strength (articulation, handle, knob)
• Articulation angle
• Articulation cycles
• Impact force
• Lift force
• Hammer impulse
• Ratchet impulse
• Device lifetime
• Radiopacity
• Luer
• Torque limiter | PASS |
| Electrical Test | • Electrical continuity
• Electrical resistance
• Hand switch functionality | PASS |
| Corrosion Test | ASTM F1089: Standard Test Method for Corrosion of Surgical Instruments | PASS |
| Design Validation | Simulated use testing | PASS |
| Thermal Effect and comparison to predicate | Measured RF lesion size in ex vivo tissue model. | PASS |
| Sterilization validation | Sterilization Validation per ISO 11137-1/-2/-3 and 11135 | PASS |
| Software Verification | Testing performed per IEC 62304 and documentation provided per FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” | PASS |
| Electromagnetic compatibility and Electrical Safety | STAR RF Ablation System was evaluated for compliance with the following standards:
• IEC 60601-1: Medical electrical equipment; Part 1: General requirements for basic safety and essential performance
• IEC 60601-1-2: Medical electrical equipment; Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility
• IEC 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories | PASS |
{9}
K251802 Page 6 of 6
Summary of Substantial Equivalence
Based on the comparison of intended/indications for use and technological characteristics, the subject device is substantially equivalent to the predicate device. The performance tests for each system component and the hardware and software verification and validation testing demonstrate that the subject device meets its performance specifications and will perform as intended in the specified use conditions and that any differences between the subject device and predicate devices do not raise new questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device (Relievant Intracept Intraosseous Nerve Ablation System per K153272).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.