K251770 · Sleepres, Inc. · BZD · Dec 15, 2025 · Anesthesiology
Device Facts
Record ID
K251770
Device Name
SleepRes PAP System
Applicant
Sleepres, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Dec 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SleepRes PAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lbs. (30 kg.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings.
Device Story
SleepRes PAP system delivers pressurized air to patient airways via mask/nasal pillows to maintain patency during sleep. Device comprises firmware-controlled console (blower, motor, heating system), water tank, air filter, and heated breathing tube. Operates in CPAP, APAP (titrating pressure 4-20 cmH2O), and KPAP (pressure reduction during inspiration/expiration) modes. Used in home, hospital, or sleep center settings by patients under clinician prescription. Cloud connectivity via Wi-Fi enables data transfer for provider review and remote therapy adjustments. Healthcare providers use output data to monitor therapy efficacy and adjust settings. Benefits include reduced/eliminated sleep events and improved sleep quality for OSA patients.
Clinical Evidence
Clinical validation study with 50 OSA patients comparing SleepRes APAP/KPAP algorithms against predicate. Patients underwent full-night polysomnography (split-night: CPAP vs. APAP vs. APAP+KPAP). Results showed no significant difference in disordered breathing events (AHI3a), sleep quality, or sleep architecture metrics.
Technological Characteristics
Microprocessor-controlled blower and integrated humidifier. Materials: permanent contact gas pathways. Connectivity: Wi-Fi. Standards: ISO 5367, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-70, ISO 80601-2-74, ISO 10993 series, ISO 18562 series. Software-controlled pressure delivery (4-20 cmH2O).
Indications for Use
Indicated for treatment of obstructive sleep apnea (OSA) in patients weighing >66 lbs (30 kg). Intended for home, hospital/institutional, and sleep center use.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
December 15, 2025
SleepRes, Inc
% Paul Dryden
Consultant
ProMedic, LLC
131 Bay Point Dr NE
St. Petersburg, Florida 33704
Re: K251770
Trade/Device Name: SleepRes PAP System
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD, BZE
Dated: November 16, 2025
Received: November 17, 2025
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K251770 - Paul Dryden
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K251770 - Paul Dryden
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J. Mathews -S
Digitally signed by Binoy J. Mathews -S
Date: 2025.12.15 17:08:05
-05'00"
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251770 | |
| Device Name SleepRes PAP system | |
| Indications for Use (Describe) The SleepRes PAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lbs. (30 kg.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251770
# 510(k) Summary
Page 1 of 9
Date Prepared: December 15, 2025
Sponsor: SleepRes, Inc.
1508 Carl Adams Dr Ste 200
Murfreesboro, TN 37129
Tel: (615) 921-2406
Sponsor Contact: Craig Salazar - Chief Financial Officer
Submission Correspondent: Paul Dryden
ProMedic, LLC
Proprietary or Trade Name: SleepRes PAP system
Common/Usual Name: CPAP
Classification Name: BZD – 868.5905 - Noncontinuous Ventilator (IPPB) Secondary
BZE – 868.5270 - Breathing System Heater
Predicate Device: Philips DreamStation 2 - K200480
Common/Usual Name: CPAP
Classification Name: BZD – 868.5905 - Noncontinuous Ventilator (IPPB)
Reference Device: Exceleron Thermalite CPAP Heated Tubing - K241268
Common/Usual Name: Heated CPAP Hose
Classification Name: BZE – 868.5270 - Breathing System Heater
## Device Description
The SleepRes PAP system consists of a firmware-controlled console, power cable and power adapter, air filter, water tank, and heated breathing tube. The PAP console is composed of a control system, motor system, and heating system. The control system includes the main board, buttons, screen, and sensor. The motor system includes the motor, fan blade, motor outer shell, and silence box. The heating system includes a heating plate and temperature sensor. The PAP console of the subject device does not include foam.
The SleepRes PAP system also includes the following components of PAP therapy systems:
- Water tank
- Air filter
- Breathing Tube
The SleepRes PAP system offers several modes:
- CPAP
- Provides continuous positive airway pressure at the prescribed pressure
- APAP
- Functions to titrate the pressure up (and down as well) to balance reducing or eliminating sleep events while using the lowest effective pressure. Pressure limits are pre-set by the prescribing clinician.
- Kairos Positive Airway Pressure (KPAP)
- This algorithm reduces airway pressure during both inspiration and during most of expiration, only returning pressure to the prescribed treatment level towards the end of expiration.
Connectivity to a Cloud platform via Wi-Fi connection.
{5}
K251770
510(k) Summary
Page 2 of 9
## Principle of Operation
A PAP therapy system works by creating and delivering pressurized air into the airway through a patient interface, mask or nasal pillows, to keep the airway open while sleeping. The pressure is set to a therapy level, or it can change based on the patient's breathing needs. Typical therapy pressure averages 11-12 cmH₂O but can range from between 4 and 20 cmH₂O during APAP use.
## Indications for Use:
The SleepRes PAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lbs. (30 kg.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings.
{6}
K251770
510(k) Summary
Page 3 of 9
Table 1 - Comparison of Subject vs. Predicate
| | Subject device
SleepRes PAP System
K251770 | Predicate
Philips DreamStation 2
K200480 | Comparison |
| --- | --- | --- | --- |
| Product Code | BZD - 868.5905
Noncontinuous Ventilator (IPPB)
BZE - 868.5270
Breathing System Heater | BZD - 868.5905
Noncontinuous Ventilator (IPPB) | Similar |
| Indications for Use | The SleepRes PAP system is intended to deliver positive airway pressure therapy for the treatment of obstructive sleep apnea (OSA) in patients weighing over 30 kg. (66 lbs.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings. | The DreamStation 2 CPAP / DreamStation 2 Auto CPAP system delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing over 30 kg (66 lbs). It is for use in the home or hospital/institutional environment. | Similar to Predicate |
| Patient Population | Patients weighing over 30 kg (66 lbs). | Patients weighing over 30 kg (66 lbs). | Similar to Predicate |
| Environment of Use | Home, hospital/institutional and sleep center settings | Home, hospital/institutional environment | Similar to Predicate |
| Fundamental scientific technology | Microprocessor
Software controlled
Blower
Heater | Microprocessor
Software controlled
Blower
Heater | Similar to Predicate |
| Modes of Operation | CPAP
Auto CPAP (APAP)
KPAP | CPAP
Auto CPAP
- Auto-Trial
- CPAP-check
C-Flex | Both devices adjust pressure during inspiration and then increase pressure during expiration. Sleep comparison of algorithms demonstrated any difference does not raise different risks compared to predicate |
| Energy delivered | Continuous Positive Airway Pressure (CPAP) | Continuous Positive Airway Pressure (CPAP) | Similar to Predicate |
| Pressure Range | 4 to 20 cmH2O | 4 to 20 cmH2O | Similar to Predicate |
| Pressure Accuracy
Static
Dynamic | +/- 0.5 cmH2O
+/- 1.0 -1.5 cmH2O from 10-20 BPM | +/- 0.5 cmH2O for 15 and 22mm tubing
+/- 1.0 cmH2O for 12 mm tubing
+/- 1.0 cmH2O for 15 and 22mm tubing
+/- 2.0 cmH2O for 12 mm tubing | Similar to Predicate |
| Flow Range | 10-60 lpm | 10-60 lpm | Similar to Predicate |
{7}
K251770
# 510(k) Summary
Page 4 of 9
| | Subject device
SleepRes PAP System
K251770 | Predicate
Philips DreamStation 2
K200480 | Comparison |
| --- | --- | --- | --- |
| Accuracy | +/- 2 lpm or 10% of reading | | |
| Humidifier | Integrated | Integrated | Similar to Predicate |
| Water tank capacity | 250 ml | 325 ml | Similar to Predicate |
| Patient Interface | Mask | Mask | Similar to Predicate |
| User supplied | Nasal pillows | Nasal pillows | |
| Cleaning | Main unit – exterior only
Water tank – soap and water
Heated Tube – soap and water | Main unit – exterior only
Water tank – soap and water
Heated Tube – soap and water | Similar to Predicate |
| Specifications | | | |
| Dimensions | 243mm x 181mm x 110mm | 274mm x 158mm x 85mm | Similar to Predicate |
| Weight | 1370 grams | 1040 grams | |
| Operating temperature range | 5°C to 32°C (41°F to 95°F) | 5° to 35°C (41° to 95°F) | Similar to Predicate |
| Storage Temperature range | -20°C to 55°C (-13°F to 158°F) | -20° to 60°C (-4° to 140°F) | Similar to Predicate |
| Service Life | Water Tank – 6 months
Heated Tubing – 3 months | Water Tank – 12 months
Heated Tubing – 6 months | Similar to Predicate |
| Atmospheric Pressure | 700 to 1060 hPa (0-3000m / 0-9880ft) | 101 to 77 kPa (0 - 2286m / 0 - 7500ft) | Similar to Predicate |
| Electrical Safety and EMC | Input: AC100-240V, 50/60Hz
1.8Amax
Output: DC24V, 3.33A | Input: AC100-240 VAC, 50/60 Hz
2.0-1.0 A
Output: DC 12V | Similar to Predicate |
| Noise / Sound | <35 dB(A) | 27 dB(A) | higher than the predicate; Labeling mitigation added to address difference* |
| Therapy algorithm comparison | Comparative Sleep studies were substantially equivalent | Comparative Sleep studies were substantially equivalent | Similar to Predicate |
| Communications | Wi-Fi | Wi-Fi
BLE
Cellular | Similar to Predicate |
*Warnings have been included on the user manual to inform the users about the device sound pressure level as compared to the recommended level by the World Health Organization.
{8}
K251770
510(k) Summary
Page 5 of 9
Table 2 - Comparison of Heated Tubing Subject vs. Reference
| | Subject device SleepRes PAP System K251770 | Reference Exceleron Thermalite CPAP Heated Tubing K241268 | Comparison |
| --- | --- | --- | --- |
| Product Code | BZD - 868.5905 Noncontinuous Ventilator (IPPB) BZE - 868.5270 Breathing System Heater | BZE - 868.5270 Breathing System Heater | Similar |
| Indications for Use | The SleepRes PAP system is intended to deliver positive airway pressure therapy for the treatment of obstructive sleep apnea (OSA) in patients weighing over 30 kg. (66 lbs.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings. | The THERMALITE® CPAP Heated Tubing is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with noninvasive ventilation for adult patients. It is compatible with the Philips Respironics DreamStation 1 and DreamStation 2 heated humidifiers | Similar |
| Patient Population | Patients weighing over 30 kg (66 lbs.). | Adult patients | Similar |
| Environment of Use | Home, hospital/institutional and sleep center settings | Home, hospital/institutional environment | Similar |
| Operating Principle | During use a voltage is applied and a current flows through the heating wires. | During use a voltage is applied and a current flows through the heating wires, encapsulated in the tubing. | Similar |
| Technology | The power is generated by the humidifier and due to the wire resistance in the device. Main unit controller monitors temperature and regulates the current through the heated wires and thereby regulates the temperature of the breathed air. | The power is generated by the humidifier and due to the wire resistance in the device, heat is dissipated through the wall of the tube construction into the air flow in the lumen of the tubing. An in circuit integrated Negative Temperature Coefficient thermistor (NTC) at the mask-end senses the temperature of the passing air flow. The resistance characteristics of the NTC changes with temperature and regulates the current through the heated wires and thereby regulates the temperature of the breathed air. | Similar |
| Dimensions | ID - 22 mm Length - 1.83 m | ID - 22 mm Length - 1.83 m | Similar |
{9}
K251770
# 510(k) Summary
Page 6 of 9
| | Subject device
SleepRes PAP System | Reference
Exceleron
Thermalite CPAP Heated Tubing
K241268 | Comparison |
| --- | --- | --- | --- |
| Reusable | Single-patient, multi-use
Clean – soap and water | Single-patient, multi-use
Clean – soap and water | Similar |
| Power Source | Humidifier controller | Humidifier controller | Similar |
{10}
K251770
# 510(k) Summary
Page 7 of 9
## Discussion of Differences Between Subject and Predicates
We presented in the tables above a comparison of the proposed device and then the accessory. We have provided the rationale for the use of a predicate for the proposed PAP system. Separately, we compared the accessory heated wire hose to a reference device.
## Technological Characteristics
The subject device and the predicate, Philips DreamStation 2 utilize a blower to generate pressure and therefore flow to the patient which is software controlled. Each device has a series of algorithms which include APAP, CPAP and a feature referred to as KPAP and C-Flex for the subject device and predicate, respectively. The subject device and the predicate both include an integrated humidifier which provides heat and humidification to the patient via a heated wire hose. We compared the technology of the heated wire hose to a reference device, Exceleron Thermalite, which was found to be similar in all features and functions and is considered substantially equivalent. Both the subject device and the predicate include types of wireless connectivity which permits data transfer for review by the provider who may also provide changes to the PAP device.
## Non-clinical Testing
The proposed device has been tested to all applicable standards that include:
- ISO 5367:2023 Anaesthetic and respiratory equipment — Breathing sets and connectors
- IEC 60601-1: 2005 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2020 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility – Requirements and tests
- IEC 60601-1-11: 2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-70:2020 Medical electrical equipment – Part 2-70: Particular requirements for basic safety and essential performance of sleep apnoea breathing therapy equipment
- ISO 80601-2-74:2021 Medical electrical equipment – Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff:
- “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices”
- “Content of Premarket Submissions for Management of Cybersecurity in Medical Devices”
Human factors validation was conducted according to FDA’s Guidance for Industry and FDA Staff:
- “Applying Human Factors and Usability Engineering to Medical Devices”
Additional non-clinical performance testing included:
- Storage
- Transportation
- Service-life
- Altitude
- Cleaning
- Wireless coexistence
- Comparative waveform Testing
{11}
K251770
# 510(k) Summary
Page 8 of 9
## Biocompatibility Assessment
The subject device is categorized as permanent contact duration (>30 days) including dry and humidified gas pathways. The following Biocompatibility testing was performed.
- ISO 10993-1: 2025 Biological evaluation of medical devices Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:201 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- ISO 10993-23, 2021, Biological Evaluation of Medical Devices - Part 23: Tests for Irritation
- ISO 10993-11, 2017, Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity -Acute Systemic Toxicity and Material Mediated Pyrogenicity
- ISO/TR 10993-33, 2015, Biological Evaluation of Medical Devices - Part 33: Guidance on Tests to Evaluate Genotoxicity - Supplement to ISO 10993-3.
- ISO 10993-18:2020 – Biological evaluation of medical devices – part 18: Chemical characterization of medical device materials within a risk management process and TRA
- ISO 18562-1: 2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds (VOCs) and Inorganic gases – CO, CO2 and Ozone
- ISO 18562-4: 2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 4: Tests for leachables in condensate
Additionally, to verify the biocompatibility of the subject device at end of service-life following testing was conducted post accelerated aging.
- ISO 18562-2:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds (VOCs)
The proposed device demonstrated that it met the performance and safety requirements to be found safe and effective compared to the predicates.
## Clinical Testing
A clinical validation study was conducted to support the substantial equivalence of the SleepRes APAP device, a Philips APAP device and the SleepRes APAP + KPAP algorithms. The proposed device was compared in PSG sleep studies to the primary predicate for the treatment of OSA.
This study consisted of 50 OSA patients all of whom were on standard APAP in the home. Each patient underwent a full night polysomnogram. Half of the night was spent on CPAP and the other half of the night on APAP, and APAP + KPAP.
The results demonstrated no significant difference in disordered breathing events (AHI3a), sleep quality and metrics of sleep architecture between treatments.
## Substantial Equivalence Conclusion
The comparison of features and specifications, non-clinical performance testing and the comparative clinical studies demonstrate that the proposed device is substantially equivalent to the predicates.
{12}
K251770
# 510(k) Summary
Page 9 of 9
Any differences do not raise different concerns of safety or effectiveness when compared to the predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.