← Product Code [MBI](/productcode/MBI) · K251750

# Trax EX Anchor (K251750)

_Trax Surgical, Inc. · MBI · Mar 12, 2026 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K251750

## Device Facts

- **Applicant:** Trax Surgical, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Mar 12, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Trax EX Anchors are intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. Specific indications are listed below: Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty) Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

## Device Story

Trax EX Anchor system comprises 6AL-4V ELI titanium screw-in and push-in anchors preloaded with UHMWPE sutures; used for soft tissue-to-bone fixation in orthopedic procedures (foot/ankle, knee, hand/wrist, elbow, shoulder). Screw-in anchors utilize threaded engagement; push-in anchors utilize press-fit engagement. System includes delivery instruments for site preparation and anchor placement. Used by surgeons in clinical/OR settings. Provides mechanical fixation of tendons/ligaments to bone to facilitate healing. Sterilized via ethylene oxide.

## Clinical Evidence

No clinical or animal testing performed. Substantial equivalence supported by bench testing including cyclic pullout, static pullout, insertion/removal torque, and torsional yield strength testing in accordance with ASTM F3690-24 and FDA guidance.

## Technological Characteristics

Materials: 6AL-4V ELI titanium; UHMWPE sutures. Design: 2-component system (threaded screw-in, ribbed push-in). Sizes: 3.5mm to 5.0mm. Sterilization: Ethylene oxide (EO). Standards: ASTM F3690-24.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex DX SwiveLock SL with Forked Eyelet ([K150648](/device/K150648.md))

## Reference Devices

- Arthrex PushLock Anchor ([K173240](/device/K173240.md))
- Arthrex Corkscrew Anchor ([K050358](/device/K050358.md))
- Arthrex Corkscrew Anchor ([K003816](/device/K003816.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

March 12, 2026

Trax Surgical, Inc.
Shalain Siddiqui
QA/RA Engineer II
75 Mill St.
Stoughton, Massachusetts 02072

Re: K251750
Trade/Device Name: Trax EX Anchor
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: MBI
Dated: February 10, 2026
Received: February 10, 2026

Dear Shalain Siddiqui:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251750 - Shalain Siddiqui
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K251750 - Shalain Siddiqui
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CHRISTOPHER FERREIRA -S

Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number(if known) K251750  |   |
|  Device Name  |   |
|  Trax EX Anchor  |   |
|  Indications for Use(Describe) The Trax EX Anchors are intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. Specific indications are listed below: Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty) Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

Trax Surgical, Inc.
Trax EX Anchor
Traditional 510(k) – K251750

# 510(k) Summary

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the Trax EX Anchor.

Submitted By: Trax Surgical, Inc.
75 Mill Street
Stoughton, MA 02072

Date: 20 November 2025

Contact Person: Shalain Siddiqui
781.828.4400
ssiddiqui@primomedicalgroup.com

Proprietary Name: Trax EX Anchor

Common Name: Fastener, Fixation, Nondegradable, Soft Tissue

Classification Name and Reference: 21 CFR 888.3040, Smooth or threaded metallic bone fixation fastener, Class II

Device Product Code, Device Panel: MBI, Orthopedic

Predicate Device: Arthrex DX SwiveLock SL with Forked Eyelet (K150648)

Reference Devices: Arthrex PushLock Anchor (K173240)
Arthrex Corkscrew Anchor (K050358)
Arthrex Corkscrew Anchor (K003816)

# 1.0 Device Description

The Trax EX Anchors are intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. The anchors are manufactured from 6AL-4V ELI Titanium. The Trax EX Anchor is a 2-component anchor compromised of the Trax EX Screw-In and the Trax EX Push-In Anchor with diameter sizes ranging from 3.5mm to 5.0mm. The Trax EX Screw-In Anchor is a hollow 6AL-4V ELI titanium anchor that is threaded, preloaded with suture and designed to fixate into bone by way of threaded engagement. The Trax EX Push-In Anchor component is a hollow 6AL-4V ELI titanium anchor that is ribbed, preloaded with suture, and designed to fixate into bone by way of press-fit engagement. Both anchors will be offered in multiple diameters, lengths, and variations in configuration of the distal end of the implant to allow further functional options to the surgeon. The Trax EX Screw-In Anchor is designed to be used alone or in conjunction with the Trax EX Push-In Anchor. The Trax EX Push-In Anchor component is designed to be used only in conjunction with the Trax EX Screw-In Anchor and is not sold separately. The sutures utilized in all implants are high-strength, Ultra High Molecular Weight Polyethylene (UHMWPE). The Trax EX Anchor Delivery

Page 1 of 2

{5}

Trax Surgical, Inc.
Trax EX Anchor
Traditional 510(k) – K251750

Instruments are intended to prepare the site and fixate the anchors. All kits are intended to be sterilized by ethylene oxide (EO) and provided in a blister tray inside of a Tyvek pouch.

## 2.0 Indication for Use

The Trax EX Anchors are intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. Specific indications are listed below:

- Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
- Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

## 3.0 Performance Data

Performance of the Trax EX Anchors was evaluated in accordance with ASTM F3690-24 and the FDA guidance Bone Anchors – Premarket Notification (510(k)) Submissions. Cyclic pullout testing, static pullout testing, insertion / removal torque testing, and torsional yield strength testing were performed to demonstrate substantial equivalence to the predicate devices. No clinical testing or animal testing was performed.

## 4.0 Technological Characteristics

The subject Trax EX Anchors have similar technical characteristics as the predicate devices including indications for use, function, and range of sizes.

Trax Surgical has demonstrated that the Trax EX Anchors are safe and effective and perform in a substantially equivalent manner to the predicate devices.

## 5.0 Conclusion

The subject Trax EX Anchors has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. Any differences between the subject device and the predicate device are considered minor and do not raise questions concerning safety or effectiveness. The information provided supports equivalence to the predicate device.

Page 2 of 2

---

**Source:** [https://fda.innolitics.com/device/K251750](https://fda.innolitics.com/device/K251750)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
