SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System

K251743 · Surgical Principals, Inc. · GCJ · Sep 22, 2025 · Gastroenterology, Urology

Device Facts

Record IDK251743
Device NameSPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System
ApplicantSurgical Principals, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 22, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery.

Device Story

The LaproSac™ is a sterile, single-use laparoscopic specimen retrieval system. It consists of a flexible rip-stop nylon bag, optionally paired with a deployment mechanism comprising an introducer, actuation rod, and biasing arms. The device is inserted through a standard laparoscopic port (5-15mm). Upon deployment, the biasing arms automatically open the bag mouth to facilitate tissue or organ collection. A closure suture is used to secure the bag after specimen collection. The device is operated by surgeons in professional healthcare facilities (operating rooms). It enables safe containment and removal of tissue, reducing the risk of contamination or spillage during extraction. The device is non-latex.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993 series), accelerated aging (ASTM F1980-21), sterilization validation (ISO 11135:2014), and performance testing (dimensional measurements, deployment/closure forces, seam strength per ASTM F88/F88M, and puncture resistance per MIL-STD-3010C).

Technological Characteristics

Materials: Rip-stop nylon. Dimensions: 50-6000 mL volume; 5-15mm introducer/trocar diameter. Sensing/Actuation: Mechanical biasing arms for bag opening. Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide (ISO 11135) to SAL 10^-6. Standards: ISO 10993 (biocompatibility), ASTM F1980-21 (aging), ASTM F88/F88M (seam strength), MIL-STD-3010C (puncture resistance).

Indications for Use

Indicated for use in surgical procedures to capture organs or tissue for removal from the body cavity during laparoscopic surgery.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 22, 2025 Surgical Principals, Inc. Timothy Wynne President and CEO 1625 South Tacoma Way Pierce, Washington 98409 Re: K251743 Trade/Device Name: SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: August 27, 2025 Received: August 27, 2025 Dear Timothy Wynne: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251743 - Timothy Wynne Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251743 - Timothy Wynne Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2025.09.22 16:33:44 -04'00' Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K251743 Device Name SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System Indications for Use (Describe) The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} Surgical Principals, Inc. 1625 South Tacoma Way Tacoma, WA 98409 Office: 888-801-9251 customerservice@surgicalprincipals.com 510(k) K251743 510(k) SUMMARY 1. SUBMITTER'S CONTACT INFORMATION Company: Surgical Principals Inc. Address: 1625 South Tacoma Way, Tacoma, WA 98409 Contact Person: Timothy Wynne Phone: 888-801-9251 Date Prepared: September 16, 2025 2. DEVICE NAME Trade Name – SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems Common Name – Retrieval Bag, Specimen Retrieval Bag Regulation Number – 21 CFR 876.1500 Classification Name – Laparoscope, General & Plastic Surgery Product Code – GCJ Device Classification – Class II Classification Panel – General and Plastic Surgery 3. PREDICATE DEVICE The SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems claim Substantial Equivalence to the Espiner Tissue Retrieval System (Sac or E-Sac) under 510(k) K111845. 4. DEVICE DESCRIPTION The SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems are sterile single patient use devices, which comprise of a flexible nylon bag with and without a deployment mechanism. The bag consists of a large, easily accessible opening and a closure suture that facilitates closure of the specimen bag after the specimen(s) have been collected. The deployment mechanism consists for an introducer and actuation rod with biasing arms, which allows for insertion through an appropriately sized port and automatic opening of the bag upon deployment. These devices are not made with natural rubber latex or synthetic derivatives of natural rubber latex. 5. INDICATIONS FOR USE The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery. Surgical Principals, Inc. www.SurgicalPrincipals.com Fax: 866-333-9524 {5} Surgical Principals, Inc. 1625 South Tacoma Way Tacoma, WA 98409 Office: 888-801-9251 customerservice@surgicalprincipals.com # 6. SUBSTANTIAL EQUIVALENCE TABLE | Category | Subject Device | Predicate Device | | --- | --- | --- | | Device | SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems (510(k) K251743) | The Espiner Tissue Retrieval System (K111845) | | Classification | Endoscope and Accessories (GCJ) | Same | | Intended Use | The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery. | The Espiner Tissue Retrieval System consists of a family of impervious sacs which are sterile single use device that can be used alone of with a dedicated introducer system for encapture and removal of an organ, tissue or fluid from the bag cavity during laparoscopic surgery. | | Product Picture | | | | Design | The LaproSac™ Specimen Retrieval Bags are sterile single patient use devices, which comprise of a flexible nylon bag with and without a deployment mechanism. The bag consists of a large, easily accessible opening and a closure suture that facilitates closure of the specimen bag after the specimen(s) have been collected. The deployment mechanism consists for an introducer and actuation rod with biasing arms, which allows for insertion through an appropriately sized port and automatic opening of the bag upon deployment. These devices are not made with natural rubber latex or synthetic derivatives of natural rubber latex. | The sac includes introduction tab(s) or tether and a drawstring (closure suture). The introduction tab(s) are used as a contact points for the instrument (tether for attachment to the dedicated introducer) for introduction through an appropriately sized cannula. The drawstring facilitates a secure closure of the sac. The dedicated introducer comprises an introducer tube with handle, and push rod. The introducer allows for the bag to be inserted through an appropriately sized port. The push rod assembly comprises a loop on one end and biasing arms on the other, which allow the bag mouth to open upon deployment. | | Bag Volume | 50 to 6000 mL | Unknown | | Introducer / Trocar Diameter | 5 to 15mm | Same | | Biocompatibility | Conforms to ISO 10993 series of FDA recognized consensus standards and associated FDA Guidance documents on application. | Unknown | | Sterilization | Sterilized using Ethylene Oxide for single patient use in accordance with ISO 11135 to an SAL of 10^{-6}. | Sterilized using Ethylene Oxide | | Prescription Use | Yes | Yes | | Intended Environment | Professional Healthcare Facility (Surgical Room or Operating theatre) | Same | Surgical Principals, Inc. www.SurgicalPrincipals.com Fax: 866-333-9524 {6} Surgical Principals, Inc. 1625 South Tacoma Way Tacoma, WA 98409 Office: 888-801-9251 customerservice@surgicalprincipals.com # 7. NONCLINICAL TESTS Nonclinical testing has been conducted to verify that these devices met all design specifications and are substantially equivalent to the predicate device. Testing included the following: - Biocompatibility Testing performed in accordance with the following: - ISO 10993-1: 2018 – Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (Recognition No. 2-258) - ISO 10993-5:2009 – Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (Recognition No. 2-245) - ISO 10993-7:2008 – Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (Recognition No. 14-408) - ISO 10993-10:2021 – Biological evaluation of medical devices - Part 10: Tests for skin sensitization (Recognition No. 2-296) - ISO 10993-11:2017 – Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (Recognition No. 2-255) - ISO 10993-23:2021 – Biological evaluation of medical devices - Part 10: Tests for irritation (Recognition No. 2-291) - Aging Study per ASTM F1980-21 – Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices (Recognition No. 14-575) - Ethylene Oxide Sterilization Validation per ISO 11135:2014 – Sterilization of health-care products - Ethylene Oxide (Recognition No. 14-529) In addition, these devices have been compared to the predicate device through various performance studies designed to test the general device operation, function and integrity. Testing included the following: - Dimensions: Dimensional measurements of the bag length, width/opening and volume. - Deployment Operation: Measurement of the bag retraction, deployment, and introduction forces. - Bag Operation: Measurement of the bag closure force, resulting reduction in the bag opening and force required to reopen the bag. - Device Integrity (Destruction): Measurement of the tensile strength of the tether, buttonhole, and introduction tabs. Measurement of the bag seam strength per ASTM F88/F88M and puncture resistance per MIL-STD-3010C Method 2065. # 8. CLINICAL TESTS There were no clinical trials performed on these devices. # 9. CONCLUSIONS The subject device has equivalent indications for use as the predicate device. There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device. Surgical Principals, Inc. www.SurgicalPrincipals.com Fax: 866-333-9524
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%