K251724 · Grindguard, Inc. · SHI · Feb 25, 2026 · Dental
Device Facts
Record ID
K251724
Device Name
Remi Impression Material
Applicant
Grindguard, Inc.
Product Code
SHI · Dental
Decision Date
Feb 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Indications for Use
The Remi Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient’s teeth and gums to fabricate patient matched nightguards/ mouthguards.
Device Story
Remi Impression Material is an addition-cure silicone (Type 0 putty) used for dental impressions. The device consists of a base and catalyst (vinyl polysiloxane, fillers, platinum catalyst, dimeticone, white oil, silicon) mixed in a 1:1 ratio. It is intended for OTC home use by lay adults; users place the material into a preformed tray, seat it in the mouth, and maintain a bite for five minutes to capture dental anatomy. The resulting impression is used to fabricate custom nightguards. The device enables patients to obtain dental impressions without professional assistance, reducing access barriers. Clinical validation confirms that at-home impressions produce nightguards with dimensional accuracy and fit equivalent to those made from clinician-collected impressions.
Clinical Evidence
Prospective, single-arm, blinded study (n=30) compared at-home impressions to dentist-collected impressions. Primary endpoints: dimensional accuracy and nightguard fit. Results: Mean dimensional difference -0.001mm (95% CI: -0.255mm to +0.253mm), well within ±0.30mm tolerance. Pearson correlation 0.99998. 100% of nightguards from both methods rated 'Excellent Fit' by physicians. No adverse events reported. Human factors study (n=35) confirmed 100% successful self-selection and critical task completion.
Technological Characteristics
Addition-cure silicone (Type 0 putty consistency). Composition: Vinyl polysiloxane, Methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil, silicon, platinum catalyst. Mixing ratio 1:1. Non-sterile. Dimensional accuracy >99%. Linear dimensional change 0.69%. Hardness 58 +/- 2 Shore A. Testing performed per ISO 4823 and ISO 10993.
Indications for Use
Indicated for use by lay adults (age 18-71) to reproduce dental and gingival structures for the fabrication of custom nightguards/mouthguards. Contraindicated for individuals with specific restorative dental work as defined in the labeling.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
Zhengzhou Huaer Impression Material (putty) (K201483)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 25, 2026
GrindGuard Inc.
% Jennifer Day
Regulatory Correspondent
Prime Path Medtech
1321 Upland Dr
Suite 6792
Houston, Texas 77043
Re: K251724
Trade/Device Name: Remi Impression Material
Regulation Number: 21 CFR 872.3660
Regulation Name: Impression Material
Regulatory Class: Class II
Product Code: SHI
Dated: June 4, 2025
Received: June 4, 2025
Dear Jennifer Day:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251724 - Jennifer Day
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251724 - Jennifer Day
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Remi Impression Material
Page 7 of 44
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251724 | ? |
| Please provide the device trade name(s). | | ? |
| Remi Impression Material | | |
| Please provide your Indications for Use below. | | ? |
| The Remi Impression Material is intended to be placed on a preformed impression tray and used to
reproduce the structure of a patient’s teeth and gums to fabricate patient matched nightguards/
mouthguards. | | |
| Please select the types of uses (select one or both, as
applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
# 510(k) Summary (K251724)
A summary of 510(k) substantial equivalence information in accordance with the requirements of 21 CFR 807.92.
Submitter: GrindGuard Inc DBA Remi
Company Contact Person: Oscar Adelman
Address: 155 Jackson St, 1206, San Francisco CA 94111
Phone: 917-634-1180
Email: oscar@shopremi.com
Submission Correspondent: Jennifer Day, Regulatory Affairs Consultant
Address: 539 W Commerce St Suite 5446
Dallas TX 75208-1953
Phone: 833 812 5885
Email: jday@primepathmedtech.com
Date Prepared: 02/24/2026
Proprietary Name: Remi Impression Material
Common Name: Impression Material
Product Code: SHI
Device Classification: Class II
Predicate Device: Accusil Dental Impression Material (K201483)
Reference Device: Zhengzhou Huaer Impression Material (putty) (K201483)
## Device Description:
The Remi Impression Material is an addition-cure silicone impression material (Type 0 putty consistency) composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry.
## Indications for Use:
The Remi Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums to fabricate patient matched night guards/mouthguards.
{5}
# Comparison to Predicate Devices:
Table 1. Predicate Comparison
| Specification | Subject Device: Remi Impression Material | Predicate Device: Accusil Putty | Reference: Zhengzhou Huaer |
| --- | --- | --- | --- |
| 510k | K251724 | K213244 | K201483 |
| Product Code | SHI | ELW | ELW |
| Classification Panel | Dental | Dental | Dental |
| Device Class | Class II | Class II | Class II |
| Classification | 21 CFR 872.3660 | 21 CFR 872.3660 | 21 CFR 872.3660 |
| Putty Type | Type 0 | Type 0-type 3 | Type 0 |
| Material | Vinyl polysiloxane | Vinyl polysiloxane | Vinyl polysiloxane |
| Mixing Ratio | 1:1 | 1:1 | 1:1 |
| Anatomical Sites | Mouth/Dentition | Worn on maxillary or mandibular teeth | Mouth/Dentition |
| Sterile | Non-Sterile | Non-Sterile | Non-Sterile |
| Fixed/Removable | Removable | Removable | Removable |
| Rx/OTC | OTC | Rx | Rx |
{6}
| Specification | Subject Device: Remi Impression Material | Predicate Device: Accusil Putty | Reference: Zhengzhou Huaer |
| --- | --- | --- | --- |
| Indication for Use | The Remi Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums to fabricate patient matched nightguards/mouthguards | Accusil dental impression material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. | Used for all crown, bridge, and orthodontic impression techniques |
| Material Type | The Impression Material is a kind of addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry. It consists of base and catalyst, in which putty base contains Vinyl polysiloxane, Methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil and silicon. Catalyst mainly contains vinyl polysiloxane, platinum catalyst, dimeticone, white oil and silicon. The product is provided non-sterile | Accusil light, heavy, mono, and bite registration: Mixture of vinyl terminated polydimethylsiloxanes and filler materials with platinum catalyst and SIH capped polysiloxane Accusil putty: Mixture of vinyl terminated polydimethylsiloxanes and filler materials with platinum catalyst and SIH capped polysiloxane plus softener | The Impression Material is a kind of addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry. It consists of base and catalyst, in which putty base contains Vinyl polysiloxane, Methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil and silicon. Catalyst mainly contains vinyl polysiloxane, platinum catalyst, dimeticone, white oil and silicon. The product is provided non-sterile |
| Working/Processing Time | Up to 240 sec | 40-150 sec | Up to 240 sec |
{7}
| Specification | Subject Device: Remi Impression Material | Predicate Device: Accusil Putty | Reference: Zhengzhou Huaer |
| --- | --- | --- | --- |
| Setting time/Time in the mouth | 60 to 120 | 60-300 sec | 60 to 120 |
| Hardness | 58 +/- 2 | 63-70 Shore A | 58 +/- 2 |
| Working humidity | Unknown | 50% | Unknown |
| Dimensional accuracy | >99% | 99.9%-99.2% | >99% |
| Stability (linear dimensional change) | .69 | <0.8% typical | .69 |
| Consistency | Type 0
ISO 4823 | Type 0-type 3
ISO 4823 | Type 0
ISO 4823 |
| Chemical Description | Room temperature vulcanizing 2 components silicone | Room temperature vulcanizing 2 components silicone | Room temperature vulcanizing 2 components silicone |
{8}
| Specification | Subject Device: Remi Impression Material | Predicate Device: Accusil Putty | Reference: Zhengzhou Huaer |
| --- | --- | --- | --- |
| Package | Bite registration Impression Trays (2) and 4 small jars; 2 of base and 2 of catalyst. | Accusil bite registration, heavy, light, and monophase come in 2 x 50 ml double cartridges Monophase regular also comes in a 5:1 380ml double cartridges. Accusil putty comes in plastic jars of 150ml and 300ml for both base and catalyst. | Bite registration Impression Trays (2) and 4 small jars; 2 of base and 2 of catalyst. |
| Method of manipulation | Preformed Impression tray | Preformed Impression tray | Preformed Impression tray |
{9}
| Specification | Subject Device: Remi Impression Material | Predicate Device: Accusil Putty | Reference: Zhengzhou Huaer |
| --- | --- | --- | --- |
| Device Description | The Remi Impression Material is an addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry. It consists of base and catalyst, in which putty base contains Vinyl polysiloxane, Methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil and silicon. The catalyst mainly contains vinyl polysiloxane, platinum catalyst, dimeticone, white oil and silicon. The product is provided non-sterile. | Accusil Dental Impression Materials are designed for dental applications to define and reproduce the structure of a patient’s teeth and gums for producing crowns, bridges, occlusal models and dental implant restorative devices. Base and catalyst components are mixed in an equal ratio 1:1, placed into an impression tray and inserted into the patient’s mouth. The material will conform to the patient’s dentition and when set will produce a reproduction of the patient’s teeth and occlusion. | The Impression Material is a kind of addition-cure rubber impression material composed of vinyl polysiloxane and various fillers, with neutral smell and applicable to impression in dentistry. It consists of base and catalyst, in which putty base contains Vinyl polysiloxane, Methylhydrogensiloxane dimethylsiloxane, dimeticone, white oil and silicon. Catalyst mainly contains vinyl polysiloxane, platinum catalyst, dimeticone, white oil and silicon. The product is provided non-sterile |
| Testing | ISO 4823 & ISO 10993 | ISO 4823 & ISO 10993 | ISO 4823 & ISO 10993 |
{10}
Intended Use:
The device has the same intended use as the predicate and reference device but includes a change in indications from prescription to over-the-counter (OTC) use. The change in indications comparing to the predicate device, Accusil (K213244) and reference device, Zhengzhou Huaer Impression (putty) (K201483) is supported by the clinical validation study and human factor/usability engineering performance data summarized below.
Technological Features:
The Technological Features (i.e., design, material, chemical composition, principle of operation, energy source, etc.) of the Remi Impression Material is identical to the predicate device. There is no difference between the subject and predicate device thus displaying the substantial equivalence of the subject device.
Performance Testing
There are no formulation/technological changes to the Remi Impression Material subject device vs. the legally marketed Zhengzhou Huaer Impression Material (putty) (K201483) reference device.
Human Factors Testing:
The Remi OTC Impression Kit Human Factors Engineering/Usability Study validation study was designed to demonstrate that untrained lay adults can safely and effectively use the kit at home relying solely on the provided labeling. The study employed an all-comers, consecutive enrollment design in which participants first self-selected using the kit's Instructions for Use, then were independently confirmed as suitable candidates by a physician investigator. Critical tasks evaluated included self-selection, tray sizing, putty mixing, tray loading and seating, maintaining a bite for five minutes, tray removal, and packaging and return of impressions. No professional training or real-time coaching was provided during impression-taking, consistent with the intended OTC home-use environment.
All 35 enrolled participants correctly self-selected as appropriate candidates (100%), with no instances of incorrect self-inclusion or self-exclusion. A minor labeling clarification was identified when 5 participants (14.3%) initially misinterpreted a question about restorative dental work; this was proactively addressed by updating the contraindication language in the IFU. Critical tasks were successfully completed across the study population, with only one subject deviating from the intended tray-use workflow — a recoverable error that was self-corrected without harm. No adverse events occurred throughout the study, and all residual risks were classified as Acceptable or ALARP per the Use-Related Risk Analysis.
{11}
# Study Participant Outcomes
| Outcome | n | % |
| --- | --- | --- |
| Enrollment | | |
| Total participants enrolled | 35 | 100% |
| Self-Selection | | |
| Correctly self-selected as suitable (Category 3) | 35 | 100% |
| Incorrectly self-excluded — intended users (Category 1) | 0 | 0% |
| Incorrectly self-included — non-users (Category 2) | 0 | 0% |
| Experienced confusion on self-selection question (restorative dental work) | 5 | 14.3% |
| Critical Task Performance | | |
| Successfully completed all critical tasks | 35 | 100% |
| Experienced minor usability-related issues (non-safety) | 6 | 17.1% |
| Tray-use workflow deviation (self-corrected, no harm) | 1 | 2.9% |
| Submitted impressions of insufficient quality (lost to follow-up) | 2 | 5.7% |
| Safety | | |
| Adverse events | 0 | 0% |
| Critical-task failures resulting in harm | 0 | 0% |
ALARP = As Low As Reasonably Practicable. IFU = Instructions for Use.
# Clinical Study
This clinical study evaluated whether at-home dental impressions taken by laypersons using an over-the-counter impression kit produce nightguards of equivalent fit and quality to those made from clinician-collected impressions. The prospective, single-arm, blinded study enrolled 30 subjects across two clinical sites, two investigators, and one independent dentist reviewer. Each subject provided both a self-taken and a dentist-taken impression, and two nightguards were fabricated from each pair — with both subjects and clinicians blinded to the impression source during all fit evaluations. Qualitative evaluation and Quantitative (over 9 points) dimensional analysis were performed on all impressions, 3D models, and nightguards comparing the impressions taken by the dentist to the study subjects.
The results were strongly positive across all endpoints. Quantitatively, the mean dimensional difference between at-home and in-office impressions was only $-0.001\mathrm{mm}$, with $95\%$ limits of agreement ranging from $-0.255\mathrm{mm}$ to $+0.253\mathrm{mm}$ — well within the pre-defined clinical
{12}
tolerance of ±0.30 mm. A Pearson correlation coefficient of 0.99998 confirmed agreement between the two methods. Qualitatively, physicians rated 100% of all nightguards as "A: Excellent Fit," regardless of impression source. Subject ratings were slightly more variable but remained positive, with no nightguard receiving an "F" (unacceptable) rating. No adverse events were reported throughout the study.
The study concluded that OTC impression kits are a clinically acceptable, safe, and effective means for fabricating custom nightguards comparable in quality to those produced from professional impressions — offering a practical alternative for individuals facing cost or access barriers to traditional dental care.
{13}
Patient Enrollment and Outcomes
| Category | n |
| --- | --- |
| Enrollment & Completion | |
| Total enrolled | 35 |
| Lost to follow-up (failed to submit impressions) | 2 |
| Included in impression analysis | 33 |
| Lost to follow-up (did not return for visit 2) | 3 |
| Included in nightguard analysis | 30 |
| Demographics | |
| Female | 26 (74%) |
| Male | 9 (26%) |
| Mean age (range) | 43.7 ± 14.0 years (18–71) |
| Nightguard Fit Ratings (60 total guards evaluated) | |
| A: Excellent Fit | 54 (90%) |
| B: Good Fit (minor cosmetic issues) | 3 (5%) |
| C: Good Fit (minor, correctable performance issues) | 3 (5%) |
| F: Would not fit / not perform as intended | 0 (0%) |
| Physician Ratings | |
| Physician ratings — subject impression guards | 30/30 A (100%) |
| Physician ratings — dentist impression guards | 30/30 A (100%) |
| Safety | |
| Adverse events | 0 |
# Conclusion
Based on comparison of indications for use, user population, mechanical, clinical and technological features, the Remi Impression Material is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.