Medline Microtek C-Flo Bag Decanter, Sterile (2000S); Medline Microtek Bag Decanter II, Sterile (2002S); Medline Microtek Vial Decanter, Sterile (2006S); Medline Microtek Transfer Device, Sterile (2008S)
Applicant
Microtek Medical, LLC
Product Code
LHI · General Hospital
Decision Date
Sep 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Microtek decanters are intended for the aseptic transfer of solutions (eg. IV fluids and medications) from IV containers - Bag decanters are for use in transferring IV fluids and medication from a bag to an IV fluid administration device. - Vial decanters are for use in transferring IV fluids and medications from a vial to an IV fluid administration device. - Transfer devices are for use in transferring IV fluids and medications from a vial or medication container to an IV fluid administration device. Microtek decanters are suitable for adult and pediatric patients. The use of decanters is at the discretion of the healthcare professional.
Device Story
Single-use, sterile medical fluid transfer devices; consist of hollow injection-molded tubes with protective caps and at least one spike for container access. Vial decanter includes integrated splash guard. Used in clinical settings by healthcare professionals to facilitate aseptic fluid transfer between IV containers and administration sets. Operates via spike-and-tube mechanism to create sterile fluid path. Benefits include maintenance of aseptic conditions during medication preparation and administration.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (ISO 10993-1, -4, -5, -10, -11, -23), performance testing (ISO 22413:2021, ISO 8536-4:2019), and sterilization validation (ISO 11135:2014).
Technological Characteristics
Materials: PMMA, HIPS825, ABS. Energy source: None (manual). Form factor: Hollow tubes with spikes, 3" to 8.25" length. Sterilization: Ethylene Oxide (EO) per ISO 11135 (SAL 10^-6). Connectivity: None. Software: None.
Indications for Use
Indicated for aseptic transfer of IV fluids and medications from IV containers (bags or vials) to IV administration devices in adult and pediatric patients.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 16, 2025
Microtek Medical LLC
% Prithul Bom
Most Responsible Person
REGULATORY TECHNOLOGY SERVICES, LLC
1000 Westgate Drive, Suite #510k
Saint Paul, MN 55114
Re: K251676
Trade/Device Name: Medline Microtek C-Flo Bag Decanter, Sterile (2000S); Medline Microtek Bag Decanter II, Sterile (2002S); Medline Microtek Vial Decanter, Sterile (2006S); Medline Microtek Transfer Device, Sterile (2008S)
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular Administration Set
Regulatory Class: Class II
Product Code: LHI
Dated: September 8, 2025
Received: September 8, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251676 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251676 - Prithul Bom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Wolloscheck -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251676
Device Name
Medline Microtek C-Flo Bag Decanter, Sterile (2000S);
Medline Microtek Bag Decanter II, Sterile (2002S);
Medline Microtek Vial Decanter, Sterile (2006S);
Medline Microtek Transfer Device, Sterile (2008S)
Indications for Use (Describe)
Microtek decanters are intended for the aseptic transfer of solutions (eg. IV fluids and medications) from IV containers
- Bag decanters are for use in transferring IV fluids and medication from a bag to an IV fluid administration device.
- Vial decanters are for use in transferring IV fluids and medications from a vial to an IV fluid administration device.
- Transfer devices are for use in transferring IV fluids and medications from a vial or medication container to an IV fluid administration device.
Microtek decanters are suitable for adult and pediatric patients. The use of decanters is at the discretion of the healthcare professional.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
MICROTEK
# K251676 - 510(k) Summary
## Submitter
510(k) submitted by:
Microtek Medical LLC
Three Lakes Drive
Northfield, IL 60093
Establishment Registration Number: 3014342658
Contact person:
Lam Duong
Senior Manager Regulatory Affairs
Phone: (651) 412 6559
Email: LDuong@medline.com
Submission Date: 15 September 2025
## Device
The following trade names are applicable to the devices
- Medline Microtek C-Flo Bag Decanter, Sterile (2000S)
- Medline Microtek Bag Decanter II, Sterile (2002S)
- Medline Microtek Vial Decanter, Sterile (2006S)
- Medline Microtek Transfer Device, Sterile (2008S)
The common or usual names for this type of product:
I.V. Fluid Transfer Set
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular administration set
Product Code: LHI
Classification Panel: General Hospital
Device Classification: Class II
Submission Type: 510(k) Traditional
## Primary Predicate Device
GCMEDICA Transfer Device (K182819)
## Device Description
Microtek decanters and transfer devices are provided sterile and for use in a single procedure only. They consist of a hollow tube with protective removable caps at either end. The injection molded hollow tube has at least one spiked end used to access the source container to initiate fluid transfer. The vial decanter includes a built-in splash guard.
Decanters and transfer devices are designed to transfer medical fluids (e.g.: medications and solutions) between various containers. Below is a summary of the different configurations.
Page 1 of 6
{5}
MICROTEK
| Model Number(s) | Description | Container Configuration | Overall Length (of finished device without caps) | Materials |
| --- | --- | --- | --- | --- |
| 2000S | C-Flo Bag Decanter, Sterile | Bag to an IV fluid administration device | 8.25” | PMMA |
| 2002S | Bag Decanter II, Sterile | Bag to an IV fluid administration device | 8.25” | HIPS825 |
| 2006S | Vial Decanter, Sterile | Vial to an IV fluid administration device | 6” | HIPS825, ABS |
| 2008S | Transfer Device, Sterile | Vial or medication container to an IV fluid administration device | 3” | ABS |
# Indications for Use
Microtek decanters are intended for the aseptic transfer of solutions (eg. IV fluids and medications) from IV containers
- Bag decanters are for use in transferring IV fluids and medication from a bag to an IV fluid administration device.
- Vial decanters are for use in transferring IV fluids and medications from a vial to an IV fluid administration device.
- Transfer devices are for use in transferring IV fluids and medications from a vial or medication container to an IV fluid administration device.
Microtek decanters are suitable for adult and pediatric patients. The use of decanters is at the discretion of the healthcare professional.
# Substantial Equivalence
Microtek decanters are medical fluid transfer devices with the same intended use as the predicate device: GCMEDICA Transfer Device.
There are a variety of medical fluid transfer devices on the market covering a wide range of materials and designs. The use of the spike and tube facilitates a sterile path for the medical fluid to flow through. Decanting or transfer devices help ensure the aseptic transfer of medical fluids from one container to another.
| Device Name | Proposed Device: Microtek decanting and transfer devices | GCMEDICA Transfer Device (various models) | Comparison Analysis |
| --- | --- | --- | --- |
| Regulation Number | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| Product Code | LHI | LHI | Same |
| Premarket Notification | New Device | K182819 | N/A |
Page 2 of 6
{6}
MICROTEK
MICROTEK
| Indications for Use | Microtek decanters are intended for the aseptic transfer of solutions (eg IV fluids and medications) from IV containers.
• Bag decanters are for use in transferring IV fluids and medication from a bag to an IV fluid administration device.
• Vial decanters are for use in transferring IV fluids and medications from a vial to an IV fluid administration device.
• Transfer devices are for use in transferring IV fluids and medications from a vial or medication container to an IV fluid administration device. | • Bag Transfer Devices: Intended for the aseptic dispensing of solutions from IV containers. For use in transferring IV fluids/medication from a bag to an IV fluid administration device.
• Vial Transfer Devices: Intended for the aseptic dispensing of solutions from IV containers. For use in transferring IV fluids/medication from a vial to an IV fluid administration device. | Same |
| --- | --- | --- | --- |
| Intended Patient Population | Microtek decanters are suitable for adult and pediatric patients. The use of decanters is at the discretion of the healthcare professional. | None stated / general population | Similar. Microtek decanters meet EO residual level for adult and pediatric patients, including infants and neonates. |
| Device Materials | 2000S: PMMA
2002S: HIPS825
2006S: HIPS825, ABS
2008S: ABS | GC0652DD, GC0653DD, GC0654DD: ABS, PE
GC0654DT: K Resin, PE | Similar polymer based plastic materials. Biocompatibility testing was performed on subject devices per ISO 10993-1. The differences do not raise any question regarding the device safety or effectiveness. |
| Operating principle | • Aseptic transfer of medical fluids
• Spike and tube to facilitate fluid flow | • Through the use of a spike/stick, facilitates creation of a sterile path
• Decanting device helps ensure an aseptic transfer or removal of fluids/medication | Same |
| Sterilization | EO | EO | Same |
Page 3 of 6
{7}
MICROTEK
| Device components/ features
Spike, tube, splashguard, caps | • Bag Decanters (2000S, 2002S) consists of one spike, tube and caps for ends
• Vial Decanters (2006S) consists of one spike, tube, splashguard, and caps for ends
• Transfer Device(2008S) consists of two spikes and caps for spikes | • GC0654DT and GC0654DD consist of one spike, tube and caps for ends
• GC0653DD consists of one spike, tube, splashguard and caps for ends
• GC0652DD consists of two spikes and caps for spikes | Same |
| --- | --- | --- | --- |
| Length | • 2000S, 2002S: 8.25"
• 2006S: 6"
• 2008S: 3" | • GC0654DT: 9"
• GC0654DD: 9"
• GC0653DD: 6"
• GC0652DD: 3" | No significant differences.
Any slight length differences do not raise any question regarding the device safety or effectiveness. |
| Biocompatibility
Test to ISO 10993-1 and applicable parts of series | • ISO 10993-1:2018
• ISO 10993-4:2017
• ISO 10993-5:2009
• ISO 10993-10:2021
• ISO 10993-11:2017
• ISO 10993-23:2021 | • ISO 10993-1:2009
• ISO 10993-4:2017
• ISO 10993-5:2009
• ISO 10993-10:2010
• ISO 10993-11:2017 | No significant differences.
Biocompatibility testing did not raise any question regarding device safety. |
The proposed Microtek decanters and transfer devices are substantially equivalent in intended use, device characteristics, device design and function in comparison to the predicate devices. As reflected in the Substantial Equivalence Comparison Table, there are no significant differences when comparing the proposed devices to the predicate devices. Functional performance and material biocompatibility testing included within this 510(k) supports the safety and effectiveness of the proposed devices and further verifies predicate equivalence.
## Technological characteristics
Information included in this submission demonstrates that there are no significant differences in technological characteristics between Microtek decanting and transfer devices and the cited predicate devices.
## Sterilization
The Microtek decanters and transfer devices are sterilized by terminal ethylene oxide sterilization. The method used for this validation study is the half cycle approach referenced in Annex B Overkill approach of ISO 11135:2014. Sterility Assurance Level (SAL) = 10⁻⁶
- Ethylene Oxide residuals ≤ 0.21 mg, in accordance with ISO 10993-7 for limited exposure devices.
Page 4 of 6
{8}
MICROTEK
- Ethylene Chlorohydrin levels ≤ 0.45 mg, in accordance with ISO 10993-7 for limited exposure devices.
## Biocompatibility
The biocompatibility evaluation for this device was conducted in accordance with ISO 10993-1:2018, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA.
The following Biocompatibility tests were conducted on the clear bag decanter and a vial decanter. The results of all testing were passing.
| Biocompatibility Test | Standards |
| --- | --- |
| In Vitro Hemocompatibility Testing (Rabbit blood)
Standard practice for Assessment of Hemolytic
Properties of Materials | ISO 10993-4:2017
ASTM F756-17 |
| In Vitro Cytotoxicity Test | ISO 10993-5:2009 |
| Skin Sensitization Test: Guinea Pig Maximization
Test (0.9% Sodium Chloride Injection Extract) | ISO 10993-10:2021 |
| Skin Sensitization Test: Guinea Pig Maximization
Test (Sesame oil Extract) | ISO 10993-10:2021 |
| Acute Systemic Toxicity (0.9% Sodium Chloride
Injection Extract, Intravenous) | ISO 10993-11:2017 |
| Acute Systemic Toxicity (Sesame Oil Extract,
Intraperitoneal) | ISO 10993-11:2017 |
| Pyrogen Test (0.9% Sodium Chloride Injection
Extract, Rabbit) | ISO 10993-11:2017 |
| Intracutaneous Reactivity Test (0.9% Sodium
Chloride Injection Extract and Sesame Oil Extract) | ISO 10993-23:2021 |
## Performance Testing
The Microtek decanters and transfer devices have been tested for the dimensional requirement of closure piercing per ISO 22413:2021 and all other relevant performance requirements in accordance with ISO 8536-4:2019. All performance testing on the subject devices has demonstrated that they meet the necessary safety and efficacy standards, establishing equivalence to the proposed predicate devices.
## Risk Management
Page 5 of 6
{9}
MICROTEK
The design and use risks for the Microtek decanters and transfer devices were evaluated using a process compliant with ISO 14971:2019. After application of risk control, all risks identified in the risk assessment were deemed to be broadly acceptable.
## Conclusion
In accordance with 21 CFR Part 807 and based on the information provided in this premarket notification, Microtek Medical LLC concludes that the Microtek decanters and transfer devices are as safe, as effective and substantially equivalent to predicate devices in terms of intended use, device characteristics, device design, function and biocompatibility as described herein.
Page 6 of 6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.