K251673 · X9, Inc. · SGH · Oct 17, 2025 · Radiology
Device Facts
Record ID
K251673
Device Name
X9 Ultrasound System
Applicant
X9, Inc.
Product Code
SGH · Radiology
Decision Date
Oct 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Vascular access vessel presence detection
—
Access Presence Sensitivity >= 75%
94.5%
—
—
ML Model Validation: 378 B-mode ultrasound images from 63 participants.
3 (independent sonographers)
Vascular access vessel localization
—
Average lateral error <= 3mm
0.492 mm
—
—
ML Model Validation: 378 B-mode ultrasound images from 63 participants.
3 (independent sonographers)
Vascular access vessel segmentation
—
Dice Similarity Coefficient >= 75%
81.3%
—
—
ML Model Validation: 378 B-mode ultrasound images from 63 participants.
3 (independent sonographers)
Indications for Use
The X9 Ultrasound System is intended for ultrasound imaging of validated vascular accesses for hemodialysis during the preliminary stage of a cannulation procedure, prior to inserting the needle. The device is intended to be used by qualified healthcare professionals in a medical setting. The device is not intended to be used to diagnose the position of the vascular access without confirming position per standard of care. The X9 Ultrasound System is not for use in pediatric patients.
Device Story
X9 Ultrasound System is a handheld ultrasound device for guiding cannulation of arteriovenous fistulas and grafts (AVF/G). System consists of a handpiece with an ultrasound transducer and software running on a user-supplied computer. The user positions the handpiece on the patient's limb; the software GUI provides visual feedback on alignment with the AVF/G. The device uses a machine learning model to detect the presence of the access vessel and calculate its lateral position. The output guides the clinician in positioning the handpiece to facilitate standard-of-care cannulation. Used in medical settings by qualified healthcare professionals. Benefits include improved efficiency and guidance during the preliminary stage of hemodialysis cannulation procedures.
Clinical Evidence
No clinical testing was conducted. Non-clinical ML model validation was performed using 378 B-mode ultrasound images from 63 adult participants (18-75+ years) with AVF/G. Ground truth was established by three independent sonographers. Results: Access Presence Sensitivity was 94.5% (criteria ≥75%); average lateral error was 0.492 mm (criteria ≤3mm); average Dice Similarity Coefficient was 81.3% (criteria ≥75%).
Technological Characteristics
Handheld ultrasound transducer probe connected via cable to a user-supplied computer. B-mode imaging. Software-based GUI for alignment guidance. Machine learning model for vessel detection and localization. Tested for acoustic, electrical, electromagnetic, mechanical, and thermal safety. Cybersecurity and software verification testing performed.
Indications for Use
Indicated for ultrasound imaging of validated vascular accesses for hemodialysis in adult patients during the preliminary stage of a cannulation procedure, prior to needle insertion. Not for use in pediatric patients.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
BD Prevue™ II Peripheral Vascular Access System (K211193)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 17, 2025
X9, Inc.
c/o Melissa Viotti
Quality and Regulatory Consultant
10 Research Parkway, Ste 350
Wallingford, Connecticut 06492
Re: K251673
Trade/Device Name: X9 Ultrasound System
Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system
Regulatory Class: Class II
Product Code: SGH, IYO, ITX, QIH
Dated: September 15, 2025
Received: September 17, 2025
Dear Melissa Viotti:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251673 - Melissa Viotti
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K251673 - Melissa Viotti
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gema Gonzalez -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251673 | |
| Device Name X9 Ultrasound System | |
| Indications for Use (Describe) The X9 Ultrasound System is intended for ultrasound imaging of validated vascular accesses for hemodialysis during the preliminary stage of a cannulation procedure, prior to inserting the needle. The device is intended to be used by qualified healthcare professionals in a medical setting. The device is not intended to be used to diagnose the position of the vascular access without confirming position per standard of care. The X9 Ultrasound System is not for use in pediatric patients. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
Traditional 510(k) X9 Ultrasound System
# 1. Contact Details
Applicant Name: X9, Inc.
Applicant Address: 10 Research Parkway, STE 350, Wallingford, CT 06492
Applicant Contact Telephone: (650) 472-0307
Applicant Contact: Earl “Eb” Bright
Applicant Contact Email: ebright@x9med.com
Date Summary Prepared: October 17, 2025
# 2. Device Name
Device Trade Name: X9 Ultrasound System
Common Name: Ultrasonic System
Classification Name: Ultrasonic pulsed echo imaging system
Regulation Number: 892.1560
Product Code: SGH, IYO
Classification Name: Diagnostic ultrasonic transducer
Regulation Number: 892.1570
Product Code: ITX
Classification Name: Medical image management and processing system
Regulation Number: 892.2050
Product Code: QIH
# 3. Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name | Product Code |
| --- | --- | --- |
| K211193 | BD Prevue™ II Peripheral Vascular Access System | IYO – ITX – LLZ |
# 4. Device Description Summary
The X9 Ultrasound System is intended for guidance in accessing arteriovenous fistulas and grafts (AVF/G). The X9 Ultrasound System consists of a Handpiece and Software. The Handpiece is a handheld device with an ultrasound transducer and is attached via a cable to a user-supplied Computer. The Handpiece is covered by a compatible probe cover and is placed on the patient’s limb and positioned (translated/rotated) by the user to align with the AVF/G. The Handpiece contains a physical button for the user to activate/deactivate the system and an alignment marking to assist the user with positioning the Handpiece over the vessel. The software, which runs on a standard operating system platform installed on the Computer, provides the Graphical User Interface (GUI). The GUI will indicate when the Handpiece is aligned with the AVF/G. The information given from the system is intended to guide the user so that they can efficiently proceed with the standard of care assessment prior to cannulation.
K251673
{5}
510(k) Summary
Traditional 510(k) X9 Ultrasound System
# 5. Intended Use / Indications for Use
The X9 Ultrasound System is intended for ultrasound imaging of validated vascular accesses for hemodialysis during the preliminary stage of a cannulation procedure, prior to inserting the needle. The device is intended to be used by qualified healthcare professionals in a medical setting. The device is not intended to be used to diagnose the position of the vascular access without confirming position per standard of care. The X9 Ultrasound System is not for use in pediatric patients.
# 6. Indications for Use Comparison
The BD Prevue™ II Peripheral Vascular Access System is intended for diagnostic ultrasound imaging of the human body performed by appropriately trained healthcare professionals in a medical setting. Specific clinical applications include:
- Pediatric
- Peripheral Vessel and Vascular Access
Typical examinations performed using the BD Prevue™ II System include:
| Imaging Applications | Exam Type (Adult and Pediatric) |
| --- | --- |
| Vascular | Assessment of vessels in the extremities and neck leading to or coming from the heart, superficial veins in the arms and legs, and vessel mapping. Assessment of superficial thoracic vessels. |
| Vascular Access | Guidance for PICC, CVC, dialysis catheter, port, PIV, midline, arterial line placement, access to fistula and grafts, and general vein and artery access |
The X9 Ultrasound System's indications for use are encompassed by the broader indications of the predicate device. The X9 Ultrasound System is not for use in pediatric patients.
# 7. Technological Comparison
A comparison of the X9 Ultrasound System (subject device) to the BD Prevue™ II Peripheral Vascular Access System (predicate device) demonstrated that any differences between the subject device and predicate device do not constitute a new intended use, and any differences in technological characteristics do not raise new questions of safety and effectiveness. Both subject and predicate devices are intended for use in a hospital/clinic medical setting by a trained clinician. Both systems consist of B-mode imaging ultrasound probes and utilize software analysis during use.
# 8. Clinical Performance Data
No clinical testing was conducted in support of the X9 Ultrasound System, as the intended use, indications, and technology are equivalent to those of the predicate device. The non-clinical
K251673
{6}
510(k) Summary
Traditional 510(k) X9 Ultrasound System
testing summarized in this submission supports the substantial equivalence of this device to the predicate with respect to safety and effectiveness.
# 9. Non-Clinical Test Summary
X9, Inc. has conducted extensive verification and validation testing of the X9 Ultrasound System. The subject device was tested to ensure that it can provide all the capabilities necessary to operate safely and effectively. Acceptance criteria have been established to ensure that the subject device performs in a manner that is substantially equivalent to the cited predicate device. Testing was conducted to verify the safety and performance requirements of the subject device, and the test results support substantial equivalence to the predicate device. The following nonclinical tests were performed on the subject X9 Ultrasound System for a determination of substantial equivalence.
- Acoustic Safety Testing
- Electrical Safety Testing
- Electromagnetic Compatibility Testing
- Mechanical and Thermal Safety Testing
- Software Verification Testing
- Cybersecurity Testing
- Bench Performance Testing
- User Needs Testing
- Ship Testing
- Machine Learning Validation Testing
A Machine Learning Model Validation was performed to confirm that the ML model correctly determined if an access vessel was visible and if the average lateral error between the ML Model and the access vessel True Location was acceptable. For the ML Model Validation, the X9 Ultrasound System was used to collect 378 B-mode ultrasound images on 63 participants known to have an arteriovenous fistula or graft (AVF/G). Participants had their access vessel scanned using the ultrasound handpiece and a laptop running an application capable of saving scanned data. From each participant scan dataset, images at three timepoints were selected in accordance with the approved protocol and analyzed by three independent sonographers to establish ground truth. The training dataset was collected at dialysis clinics not included in the ML model validation and was used exclusively for model training. Separate test datasets were reserved for performance validation against expert-annotated ground truth. ML model validation clinic locations were selected to obtain a sample of participants and included 31 females, 32 males, 34 Black/African American, 27 White, 1 Asian, 1 Unknown/Not Reported, 4 Hispanic or Latino, and 59 Not Hispanic or Latino. The participant age ranged from 18 to 75+ years old. Fifteen (15) participants presented with AV Grafts and 48 with AV Fistulas.
The reviewers used their clinical expertise to determine if an access vessel was visible in each image or not. A subset of images called the "Access Present Subset" was created, containing all images where the majority of independent reviewers determined that an access vessel was
K251673
{7}
510(k) Summary
Traditional 510(k) X9 Ultrasound System
present in the image. The Access Presence Sensitivity percentage was calculated from the number of images where the ML Model determined that an access vessel is present divided by the total number of images in the subset. The "true location" of the access vessel was established by calculating the average of the locations annotated by each of the independent reviewers. The lateral error of the ML Model was calculated as the difference between the True Location of the access vessel and the ML model lateral access vessel location.
The Machine Learning Model (and therefore the X9 Ultrasound System) met the ML Model Validation acceptance criteria for the primary analyses. The calculated Access Presence Sensitivity was 94.5%, which exceeded the acceptance criteria of greater than or equal to 75%. The average lateral error between the ML Model and the access vessel True Location was 0.492 mm, which met the acceptance criterion of 3mm maximum. The average Dice Similarity Coefficient (DSC) across all reviewers was 81.3%, which meets the acceptance criterion of 75% minimum.
## 10. Conclusion
The subject device met all design specifications and non-clinical testing provided support of the substantial equivalence determination.
K251673
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.