Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution, Dewin One Step medium [with HSA and without HSA])
K251637 · Donnevie Medical Technology (Shanghai) Co. , Ltd. · MQL · Aug 27, 2025 · Obstetrics/Gynecology
Device Facts
Record ID
K251637
Device Name
Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution, Dewin One Step medium [with HSA and without HSA])
Applicant
Donnevie Medical Technology (Shanghai) Co. , Ltd.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Aug 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Dewin Gamete Buffer is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI). Dewin Follicle Flushing Solution is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures. Dewin One Step Medium is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus.
Device Story
Dewin Reproductive Media consists of three solutions: Gamete Buffer, Follicle Flushing Solution, and One Step Medium. Used in assisted reproductive technology (ART) procedures; operated by embryologists/clinicians in laboratory settings. Follicle Flushing Solution facilitates oocyte collection from ovaries; Gamete Buffer enables short-term handling/washing of gametes and ICSI; One Step Medium supports embryo culture from fertilization to blastocyst stage and subsequent uterine transfer. Media contain inorganic salts, energy sources, amino acids, HSA, antioxidants, buffers, EDTA, phenol red, and gentamicin. Solutions are sterile-filtered and stored in glass bottles. Output is the biological environment for gamete/embryo development; clinical decision-making relies on successful embryo development and viability for transfer. Benefits include standardized, controlled media for IVF procedures.
Clinical Evidence
Bench testing only. Performance validated via Mouse Embryo Assay (MEA) per 2021 FDA guidance, demonstrating ≥80% development to expanded blastocyst stage. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity). Sterility and shelf-life validated via real-time aging and USP <71>, <785>, <791>, and <85> standards.
Technological Characteristics
Aqueous solutions containing inorganic salts, energy sources, amino acids, HSA, antioxidants, buffers, EDTA, phenol red, and gentamicin. pH 7.2–7.5; osmolality 260–295 mOsm/kg. Sterile-filtered (0.2 µm). Glass bottle containers sterilized via dry heat (ISO 20857:2010); caps sterilized via electron beam irradiation (ISO 11137-1/2).
Indications for Use
Indicated for human gamete and embryo handling, oocyte collection, and embryo culture to the blastocyst stage in in vitro fertilization procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 26, 2025
DonneVie Medical Technology (Shanghai) Co. Ltd.
Lily Liu
RA Manager
Suite 201, Bld. 1, 138 Xinjun Ring
Minhang District
Shanghai, 201114
CHINA
Re: K251637
Trade/Device Name: Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution; Dewin One Step medium [with HSA and without HSA])
Regulation Number: 21 CFR 884.6180
Regulation Name: Reproductive Media And Supplements
Regulatory Class: Class II
Product Code: MQL
Dated: May 29, 2025
Received: May 29, 2025
Dear Lily Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251637 - Lily Liu
Page 2
Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
{2}
K251637 - Lily Liu
Page 3
(https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251637 | |
| Device Name Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution; Dewin One Step medium [with HSA and without HSA]) | |
| Indications for Use (Describe) | |
| Dewin Gamete Buffer is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI). | |
| Dewin Follicle Flushing Solution is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures. | |
| Dewin One Step Medium is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus. | |
| Type of Use (Select one or both, as applicable) | |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251637
Page 1 of 6
510(k) Summary
K251637
I. SUBMITTER
Applicant: DonneVie Medical Technology (Shanghai) Co. Ltd.
Address: Suite 201, Bld. 1, 138 Xinjun Ring Minhang District, Shanghai, 201114, China
Phone: 86 (21) 347-81568 /86 18221012749
Contact Person: Lily Liu, RA Manager
Email: liulj@dewinivf.com
Date Prepared: July 31, 2025
II. DEVICE
Trade Name: Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution; Dewin One Step medium [with HSA and without HSA])
Common Name: Assisted Reproduction Media
Regulation Name: Reproductive Media and Supplements
Regulation Number: 884.6180
Product Code: MQL (Media, Reproductive)
Regulatory Class: II
III. PREDICATE DEVICE
VitaVitro Fertilization Medium, VitaVitro Gamete Buffer Medium, VitaVitro Flushing Buffer Medium (K200408) from Shenzhen VitaVitro Biotech Co.,Ltd. The predicate device has not been subject to a design-related recall.
IV. DEVICE DESCRIPTION
The DonneVie Dewin Reproductive Media are solutions used in assisted reproductive procedures. The family of Reproductive Media include the following 3 types and their variants:
- Dewin Gamete Buffer [with and without Human Serum Albumin (HSA)];
- Dewin Follicle Flushing Solution;
- Dewin One Step Medium [with and without HSA].
{5}
K251637
All variants contain gentamicin, an antibiotic agent that suppresses bacterial growth. The Gamete Buffer and Follicle Flushing media are offered in $100\mathrm{mL}$ volume, while the One Step Medium is offered in volumes of $25\mathrm{mL}$ and $50\mathrm{mL}$ . The three Dewin Reproductive Media are used together during an assisted reproductive procedure. The Follicle Flushing Solution is used to first flush/wash the Cumulus-Oocyte Complex after it's removed from the ovary; it is then transferred to the Gamete Buffer for cleaning. After the prepared oocyte has been fertilized, it is transferred to the One Step Medium to culture the embryo from fertilization to blastocyst stage.
The primary glass bottle containers are washed, then sterilized and depyrogenated via dry heat. The bottle caps are supplied sterile by electron beam irradiation. The media are sterile filtered through a $0.2\mu \mathrm{m}$ filter in an aseptic isolator at DonneVie Medical Technology Co., Ltd.
# V. INDICATIONS for USE
Dewin Gamete Buffer is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI).
Dewin Follicle Flushing Solution is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures.
Dewin One Step Medium is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus.
# VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A comparison of the intended use and technological features of the subject and predicate devices are described in the tables below:
Table 1 - Dewin Gamete Buffer & Follicle Flushing Solution Comparison to VitaVitro Gamete Buffer Medium & VitaVitro Flushing Buffer Medium
| | Subject device - Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution; Dewin One Step | Predicate device - VitaVitro® Fertilization Medium, VitaVitro® Gamete Buffer Medium, VitaVitro® Flushing Buffer Medium / K200408 | Comparison |
| --- | --- | --- | --- |
{6}
K251637
Page 3 of 6
| | medium [with HSA and without HSA] | | |
| --- | --- | --- | --- |
| Indications for Use | Dewin Gamete Buffer is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI).
Dewin Follicle Flushing Solution is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures.
Dewin One Step Medium is intended for use for culture of embryos from fertilization to the blastocyst stage. This device can be used for transfer of embryos to the uterus. | VitaVitro® Gamete Buffer Medium is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI).
VitaVitro® Flushing Buffer Medium is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures.
VitaVitro® Fertilization Medium is intended for preparation and handling of human gametes and for in vitro fertilization. | The Indications for use statements are same for Gamete Buffer and Follicle Flushing Solution. The difference in language for One Step Medium and Fertilization Medium do not raise different questions of S&E. |
| Primary Ingredients | Inorganic salts, energy sources, amino acids, antibiotic, HSA, antioxidant, buffer, EDTA, phenol red. | Inorganic salts, energy sources, amino acids, antibiotic, HSA, antioxidant, buffer, EDTA, phenol red. | Different: The formulations of the subject and predicate devices are not the same. These differences do not raise different questions of S&E. |
| pH | 7.2 – 7.5 | 7.2 - 7.6 | Different: The subject device and predicate devices have differences in pH specifications. These differences do not raise different questions of S&E. |
| Osmolality | 260-295mOsm/Kg | 260 - 290mOsm/Kg | Different: The subject device and predicate devices have differences in |
{7}
K251637
Page 4 of 6
| | | | osmolality specifications. These differences do not raise different questions of S&E. |
| --- | --- | --- | --- |
| Endotoxin | <0.25EU/mL | <0.25EU/mL | Same |
| MEA | One-cell MEA: ≥80% developed to expanded blastocyst at 96 hours after a 30 minute exposure to Dewin Gamete Buffer.
One-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours after a 30 minute exposure to Dewin Follicle Flushing Solution.
One-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 120 hours for Dewin One Step medium. | 1-cell MEA: ≥80% expanded blastocyst at 96 hours after a 2-hour exposure to media (Gamete Buffer Medium and Flushing Buffer Medium).
1-cell MEA: ≥80% expanded blastocyst at 96 hours after a 24-hour exposure to media (Fertilization media). | Different: The subject device and predicate devices have differences in MEA specifications. These differences do not raise different questions of S&E. |
| Sterilization method | Aseptic filtration. | Aseptic filtration. | Same |
| Storage conditions | 2 - 8°C | 2 - 8°C | Same |
| Shelf-life | Dewin Gamete Buffer – 7 months
Dewin Follicle Flushing Solution – 12 months
Dewin One Step Medium – 4 months. | 6 months. | Different: The subject device and predicate devices have differences in shelf-life. These differences do not raise different questions of S&E. |
As shown in the table above, there are differences in the indications for use statements and technological characteristics of the subject and predicate devices. However, as stated in the table, the differences in indications for use do not raise different questions of safety and effectiveness and the differences in technological characteristics do not raise different questions of safety and effectiveness.
{8}
K251637
Page 5 of 6
# VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
The following studies have been conducted in support of the substantial equivalence to the predicate devices:
- Aseptic filtration and aseptic filling validation, per ISO 13408-1:2008 & A1:2013 and ISO 13408-2:2018.
- For filter challenge test per ISO 13408-2, the solutions used for testing did not contain antimicrobials (gentamicin was excluded from subject media formulation) and were representative of worst-case production conditions.
- Glass bottles were sterilized by dry heat, per 20857:2010/(R)2015 (Revision of ANSI/AAMI/ST63:2002) Sterilization of health care products - Dry heat - Requirements for the development, validation and routine control of a sterilization process for medical devices. Bottle caps were sterilized by radiation, per ISO 11137-1:2006 (including: Amendment 1 (2013) and Amendment 2 (2018)) Sterilization of health care products — Radiation — Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2:2013 - Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose [Including Amendment 1 (2022)].
- Shelf-life testing was conducted to support shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after real time aging:
- Appearance: Clear, particulate free
- pH per USP <791>: 7.2 -7.5
- Osmolarity per USP <785>: 260–295 mOsm/kg
- Endotoxin per USP <85>: < 0.25 EU/mL
- Sterility per USP <71>: No microbial growth
- MEA per the 2021 FDA Guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices:
- One Step Medium: One-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 120 hours for Dewin One Step medium
- Gamete Buffer: “One-cell MEA: ≥80% developed to expanded blastocyst at 96 hours after a 30 minute exposure to Dewin Gamete Buffer
{9}
K251637
Page 6 of 6
- Follicle Flushing Solution: One-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 96 hours after a 30 minute exposure to Dewin Follicle Flushing Solution
> Transportation testing per ASTM D4169-22 and cap/seal leak testing using a method equivalent to USP <1207.2> on transportation-conditioned devices.
Biocompatibility testing was conducted on the patient contacting subject media Dewin Follicle Flushing Solution and Dewin One Step Medium in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process, as follows:
- Cytotoxicity (ISO 10993-5:2009)
- Guinea Pig Maximization Sensitization Test (ISO 10993-10:2021)
- Intracutaneous Reactivity (ISO 10993-23:2021)
The results of the testing conducted support the biocompatibility of the patient contacting subject devices.
## VIII. CONCLUSION
The results of the performance testing described above demonstrate that Dewin Reproductive Media (Dewin Gamete Buffer, Dewin Follicle Flushing Solution and Dewin One Step Medium) are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.