K251614 · Dentsply Sirona, Inc. · KCT · Feb 6, 2026 · General Hospital
Device Facts
Record ID
K251614
Device Name
Stainless Steel Surgical Kits
Applicant
Dentsply Sirona, Inc.
Product Code
KCT · General Hospital
Decision Date
Feb 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Stainless Steel Surgical Kits are intended for organizing, sterilizing and storing of drills, instruments, and prosthetic accessories. The Stainless Steel Surgical Kits are not intended to maintain sterility and are to be used in conjunction with a legally marketed, validated sterilization pouch.
Device Story
Reusable stainless steel surgical kits designed to organize, store, and facilitate sterilization of dental drills, instruments, and prosthetic accessories. Kits consist of a base, lid, and silicone holders; some models include a removable insert or bath for soiled items. Used in dental clinical settings by practitioners. Kits are provided non-sterile and must be cleaned and steam-sterilized (pre-vacuum, 132°C for 4 minutes) by the end-user before and after each procedure. Kits are placed inside a legally marketed, validated sterilization pouch for the sterilization process. The device ensures instruments are held securely during transport and sterilization, supporting efficient clinical workflows and instrument maintenance.
Clinical Evidence
Bench testing only. Validated per ANSI/AAMI ST98:2022 for cleaning and ISO 17665:2024 for sterilization. Results confirmed SAL of ≤ 10^-6, no visible soil, and residual protein ≤ 6.4 μg/cm². Service life validated for 50 reprocessing cycles. Biocompatibility (cytotoxicity) per ISO 10993-5:2009 passed.
Indicated for organizing, sterilizing, and storing dental drills, instruments, and prosthetic accessories in a clinical setting. Not intended to maintain sterility; must be used with a validated sterilization pouch.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 6, 2026
Dentsply Sirona Inc.
Laura Sobrin
Sr. Technical RA Manager
221 West Philadelphia Street, Suite 60W
York, Pennsylvania 17401
Re: K251614
Trade/Device Name: Stainless Steel Surgical Kits
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: KCT
Dated: January 5, 2026
Received: January 5, 2026
Dear Laura Sobrin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251614 - Laura Sobrin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251614 - Laura Sobrin
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN A. ANISKO -S
Digitally signed by
STEPHEN A. ANISKO -S
Date: 2026.02.06
12:40:51 -05'00'
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251614
Device Name
Stainless Steel Surgical Kits
Indications for Use (Describe)
Stainless Steel Surgical Kits
- The Stainless Steel Surgical Kits are intended for organizing, sterilizing and storing of drills, instruments, and prosthetic accessories.
- The Stainless Steel Surgical Kits are not intended to maintain sterility and are to be used in conjunction with a legally marketed, validated sterilization pouch.
Sterilization parameters
- Pre-vacuum Steam at 132 °C (270 °F) for 4 minutes with a 20 minute dry time.
- Do not exceed the worst-case validated maximum load:
| Item Number | Product Name | Dimensions (inch) | Max Load (gr) | Vent to Volume Ratio (inch-1) | Kit type |
| --- | --- | --- | --- | --- | --- |
| MK-0020 | Bone compression kit | 3.94x3.46x1.86 | 318.77 | 0.15 | Closed |
| MK-0056 | Scan post Int. Hexagon connection Kit | 2.60x2.83x1.02 | 143.59 | 0.29 | Closed |
| MK-0057 | Scan post Conical connection Kit | 2.60x2.83x1.02 | 146.32 | 0.29 | Closed |
| MK-0059 | Bone profiler Int. Hex. kit | 2.79x1.18x1.77 | 79.26 | 10.18 | Open |
| MK-0060 | Bone profiler coni. con. kit | 2.79x1.18x1.77 | 81.00 | 10.18 | Open |
| MK-0065 | Scan post organizer box | 2.60x2.83x1.02 | 146.32 | 0.29 | Closed |
| MK-BC101 | Drill stopper kit for C1 implant, SP | 2.60x2.83x1.02 | 144.87 | 0.29 | Closed |
| MK-BS001 | Drill stopper kit for Seven implant, SP | 2.60x2.83x1.02 | 144.64 | 0.29 | Closed |
| MK-CDS08 | Drill stopper kit for 8mm length C1 implant procedure | 1.97x2.83x1.02 | 117.27 | 0.26 | Closed |
| MK-CDS10 | Drill stopper kit for 10mm length C1 implant procedure | 1.97x2.83x1.02 | 116.70 | 0.26 | Closed |
| MK-CDS11 | Drill stopper kit for 11.5mm length C1 implant procedure | 1.97x2.83x1.02 | 116.28 | 0.26 | Closed |
| MK-CDS13 | Drill stopper kit for 13mm length C1 implant procedure | 1.97x2.83x1.02 | 115.86 | 0.26 | Closed |
| MK-LDSK | Lance Drill stoppers kit | 4.22x3.96x1.20 | 282.17 | 0.20 | Closed |
| MK-SDS06 | Drill stopper kit for 6mm length Seven implant procedure | 1.97x2.83x1.02 | 118.43 | 0.26 | Closed |
| MK-SDS08 | Drill stopper kit for 8mm length Seven implant procedure | 1.97x2.83x1.02 | 117.19 | 0.26 | Closed |
| MK-SDS10 | Drill stopper kit for 10mm length Seven implant procedure | 1.97x2.83x1.02 | 116.64 | 0.26 | Closed |
| MK-SDS11 | Drill stopper kit for 11.5mm length Seven implant procedure | 1.97x2.83x1.02 | 116.22 | 0.26 | Closed |
| MK-SDS13 | Drill stopper kit for 13mm length Seven implant procedure | 1.97x2.83x1.02 | 115.81 | 0.26 | Closed |
{4}
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{5}
510(k) SUMMARY for Stainless Steel Surgical Kits, K251614
1. **Submitter Information:**
Dentsply Sirona Inc.
221 West Philadelphia Street
Suite 60W
York, PA 17401
Contact Person: Laura Sobrin
Telephone Number: 717-849-4434
Email: laura.sobrin@dentsplysirona.com
Date Prepared: February 4, 2026
2. **Device Name:**
- Proprietary Name: Stainless Steel Surgical Kits
- Classification Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
- CFR Number: 21 CFR 880.6850
- Device Class: Class II
- Product Code: KCT
3. **Predicate and Reference Devices:**
The predicate and reference devices identified are the following:
| Type of Device | Device Name | 510(k) | Company Name |
| --- | --- | --- | --- |
| Predicate | Surgical and Prosthetic Trays | K220791 | Dentsply Sirona Inc. |
| Reference | Washtrays | K231144 | Dentsply Sirona Inc. |
| Reference | Biotech Dental Kits | K212068 | Biotech Dental, SAS |
4. **Device Description:**
The Stainless Steel Surgical Kits are reusable containers that are used to store and organize drills, instruments, and prosthetic accessories needed before, during and after the implantation and the prosthetic procedure.
The Stainless Steel Surgical Kits are also designed to hold the drills, instruments, and prosthetic accessories during the sterilization process. The kits are intended to be used in conjunction with a legally marketed, validated sterilization pouch that is intended to maintain sterility of the items placed inside the kits.
The proposed Stainless Steel Surgical Kits must be cleaned and sterilized before the first use and after each use according to the Instructions for Use (IFU) included with each kit.
Page 1 of 9
{6}
The proposed Stainless Steel Surgical Kits can be divided into two types of kits: Open and Closed Kits.
Open kits consist of a lid and base with holders. Open kits cannot be closed with a lid to cover the items, but contain a lid that is always open and can be moved to a horizontal position. Open kits also do not contain an insert; the instrument holders are integrated on the base of the kit. In addition, in Open kits, the base is open and the holders can be seen through the bottom of the base.
By contrast, Closed kits consist of a lid, removable insert with holders, and base. Closed kits have a lid that can be closed. The lid has perforations that allows for steam penetration. In addition, one of the proposed Closed kits also contains a removable bath. The bath is used to hold soiled items after use during the surgical procedure.
The lid, base, insert, and bath of these kits are made of stainless steel. Silicone holders hold the kit components in each surgical kit. The role of the holders is to properly hold the items during transportation, sterilization and before use by the dentist.
## 5. Intended Use and Indications for Use
### Stainless Steel Surgical Kits
- The Stainless Steel Surgical Kits are intended for organizing, sterilizing and storing of drills, instruments, and prosthetic accessories.
- The Stainless Steel Surgical Kits are not intended to maintain sterility and are to be used in conjunction with a legally marketed, validated sterilization pouch.
### Sterilization parameters
- Pre-vacuum Steam at 132 °C (270 °F) for 4 minutes with a 20 minute dry time.
- Do not exceed the worst-case validated maximum load:
| Item Number | Product Name | Dimensions (inch) | Max Load (gr) | Vent to Volume Ratio (inch-1) | Kit type |
| --- | --- | --- | --- | --- | --- |
| MK-0020 | Bone compression kit | 3.94x3.46x1.86 | 318.77 | 0.15 | Closed |
| MK-0056 | Scan post Int. Hexagon connection Kit | 2.60x2.83x1.02 | 143.59 | 0.29 | Closed |
| MK-0057 | Scan post Conical connection Kit | 2.60x2.83x1.02 | 146.32 | 0.29 | Closed |
| MK-0059 | Bone profiler Int. Hex. kit | 2.79x1.18x1.77 | 79.26 | 10.18 | Open |
| MK-0060 | Bone profiler coni. con. kit | 2.79x1.18x1.77 | 81.00 | 10.18 | Open |
| MK-0065 | Scan post organizer box | 2.60x2.83x1.02 | 146.32 | 0.29 | Closed |
| MK-BC101 | Drill stopper kit for C1 implant, SP | 2.60x2.83x1.02 | 144.87 | 0.29 | Closed |
| MK-BS001 | Drill stopper kit for Seven implant, SP | 2.60x2.83x1.02 | 144.64 | 0.29 | Closed |
| MK-CDS08 | Drill stopper kit for 8mm length C1 implant procedure | 1.97x2.83x1.02 | 117.27 | 0.26 | Closed |
| MK-CDS10 | Drill stopper kit for 10mm length C1 implant procedure | 1.97x2.83x1.02 | 116.70 | 0.26 | Closed |
| MK-CDS11 | Drill stopper kit for 11.5mm length C1 implant procedure | 1.97x2.83x1.02 | 116.28 | 0.26 | Closed |
| MK-CDS13 | Drill stopper kit for 13mm length C1 implant procedure | 1.97x2.83x1.02 | 115.86 | 0.26 | Closed |
| MK-LDSK | Lance Drill stoppers kit | 4.22x3.96x1.20 | 282.17 | 0.20 | Closed |
| MK-SDS06 | Drill stopper kit for 6mm length Seven implant procedure | 1.97x2.83x1.02 | 118.43 | 0.26 | Closed |
| MK-SDS08 | Drill stopper kit for 8mm length Seven implant procedure | 1.97x2.83x1.02 | 117.19 | 0.26 | Closed |
| MK-SDS10 | Drill stopper kit for 10mm length Seven implant | 1.97x2.83x1.02 | 116.64 | 0.26 | Closed |
Page 2 of 9
{7}
| Item Number | Product Name | Dimensions (inch) | Max Load (gr) | Vent to Volume Ratio (inch-1) | Kit type |
| --- | --- | --- | --- | --- | --- |
| | procedure | | | | |
| MK-SDS11 | Drill stopper kit for 11.5mm length Seven implant procedure | 1.97x2.83x1.02 | 116.22 | 0.26 | Closed |
| MK-SDS13 | Drill stopper kit for 13mm length Seven implant procedure | 1.97x2.83x1.02 | 115.81 | 0.26 | Closed |
# 6. Comparison of Technological Characteristics
The proposed Surgical Kits have the same intended use, similar Indications for Use, and incorporate the same fundamental technology as the predicate device, Surgical and Prosthetic Trays (K220791), with similar general design and functionality.
The proposed and predicate (K220791) devices share the same sterilization method (pre-vacuum) and sterilization parameters. Both devices are intended to be used with an FDA cleared sterilization single pouch.
Both the proposed and predicate devices are reusable. The difference in reuse life is based on the number of cycles validated (50 cycles) for the proposed kits when compared to the cycles validated on the predicate device (200 cycles).
An overview of the similarities and differences between the proposed, predicate, and reference devices is given in Table 6.1.
{8}
Table 6.1: Similarities and Differences between the proposed, predicate, and reference devices
| Element | Proposed Device Stainless Steel Surgical Kits | Predicate Device Surgical and Prosthetic Trays (K220791) | Reference Device Washtrays (K231144) | Reference Device Biotech Dental Kits 2-piece metal trays (K212068) | Comparison |
| --- | --- | --- | --- | --- | --- |
| Manufacturer | MIS Implants Technologies Ltd. (part of Dentsply Sirona) | Dentsply Implants Manufacturing GmbH (part of Dentsply Sirona) | Dentsply Implants Manufacturing GmbH (part of Dentsply Sirona) | Biotech Dental, SAS | Same corporation for the predicate and reference (K231144) device |
| Intended Use | Intended for organizing, sterilizing and storing of drills, instruments, and prosthetic accessories | Intended for organizing, sterilizing and storing of instruments | Intended for organizing, cleaning, sterilizing, and storing of instruments | Intended to allow sterilization of the group medical devices | Same general intended use as the predicate device. |
| Product code | KCT | KCT | KCT | KCT | Same |
| Regulation # | 21 CFR 880.6850 | 21 CFR 880.6850 | 21 CFR 880.6850 | 21 CFR 880.6850 | Same |
| General design and Components | Closed kit: Perforated stainless steel box (lid and base) with removable stainless steel insert, on which holders made of silicone are assembled. One kit contains a removable stainless steel bath. Open Kit: Stainless steel kit (lid and base) without an insert. The base integrates the silicone holders. | Perforated plastic box (lid and base) with removable plastic insert, on which holders made of silicone are assembled. The base also holds instruments. | Perforated stainless steel box (lid and base) which incorporates a small stainless steel basket, and an anodized aluminum insert with large stainless steel basket underneath. The insert contains thermoplastic holders with stainless steel springs, and the small basket contains a silicone mat. | 2 piece stainless steel tray with silicone holders | Similar general design and components. The proposed (Closed) Stainless Steel Surgical Kits and the predicate device (K220791) have the same components: lid, base, insert and holders. One of the proposed Stainless Steel Surgical Kits contains an additional bath component. For Closed kits, the lid and base have ventilation apertures (i.e. perforations) and the insert is a perforated plate mounted on the base, which is similar to the |
| | | | | | material. The base is a perforated plate mounted on the base, which is similar to the |
{9}
| Element | Proposed Device Stainless Steel Surgical Kits | Predicate Device Surgical and Prosthetic Trays (K220791) | Reference Device Washtrays (K231144) | Reference Device Biotech Dental Kits 2-piece metal trays (K212068) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | | | predicate device (K220791). The holders are assembled in the insert, and they hold the items in place.
The difference is that predicate device (K220791) allows for placement of items in the base, whereas the proposed Stainless Steel Surgical Kits hold all items within the insert only.
The proposed Open kits are similar to the reference device K212068 in design concept and materials.
The reference device K231144 also shares similar materials compared to the proposed Closed kits, which is also a perforated kit. |
| Dimensions (inch) | 1.97 x 2.83 x 1.02
2.60 x 2.83 x 1.02
2.79 x 1.18 x 1.77
3.94 x 3.46 x 1.86
4.22 x 3.96 x 1.20 | 5.4 x 3.9 x 2.4
7.3 x 5.5 x 2.4 | 10.83 x 6.93 x 2.32 | 1.57 x 1.97 x 0.79
(40 mm x 50 mm x 20 mm) | Similar.
The dimensions of the proposed Stainless Steel Surgical Kits are within the range of the predicate and reference devices. |
Page 5 of 9
{10}
| Element | Proposed Device Stainless Steel Surgical Kits | Predicate Device Surgical and Prosthetic Trays (K220791) | Reference Device Washtrays (K231144) | Reference Device Biotech Dental Kits 2-piece metal trays (K212068) | Comparison |
| --- | --- | --- | --- | --- | --- |
| Raw materials | Lid, Base, Insert, and Bath: AISI 304L Holders: Colored Silicone | Base, Lid and Insert: PPSU polyphenylsulfone - Radel R-5000 Holders - Colored Silicone | • Metal Grid Base: Stainless steel 1.4301
• Metal grid base feet: Silicone
• Large and small basket: Stainless steel 1.4303
• Lid with Handle: Stainless Steel 1.4310
• Silicone mat (inside the small basket): Colored Silicone
• Instrument Holders (within the Overlay): PEEK, with Stainless steel springs
• Overlay: Anodized aluminum alloy | Base and Lid: Stainless Steel AISI 304L Holders: Medical grade silicone | Similar
Same material as the predicate device for the silicone holders.
Similar stainless steel materials as the reference devices. |
| Air permeance | Allows moist heat (steam) penetration to achieve sterilization | Allows moist heat (steam) penetration to achieve sterilization | Allows moist heat (steam) penetration to achieve sterilization | Allows moist heat (steam) penetration to achieve sterilization | Same |
| Maximum number of instruments | 5 - 20 instruments | 22 - 56 instruments | 38 - 76 instruments | 3 instruments | Similar and within range of the predicate and reference devices, combined.
The maximum number of instruments that can be loaded in the proposed Stainless Steel Surgical Kits is lower than that of the predicate device (K220791). |
{11}
| Element | Proposed Device Stainless Steel Surgical Kits | Predicate Device Surgical and Prosthetic Trays (K220791) | Reference Device Washtrays (K231144) | Reference Device Biotech Dental Kits 2-piece metal trays (K212068) | Comparison |
| --- | --- | --- | --- | --- | --- |
| Mass of maximum load (g) | 79.26–318.77 | 271.9 - 563.5 | 1107.3 | 39.4 | Similar and within the range of the predicate and reference devices, combined. The weight difference of the mass of maximum load of the proposed Stainless Steel Surgical Kits does not affect the sterilization process and sterility assurance level (SAL) of 10-6as demonstrated by the sterilization validation. |
| Vent to Volume Ratio (in-1) | 0.15 – 10.18 | 0.033 – 0.046 | 2.54 | 0.9500 cm2/cm3(2.413 in-1) | Similar or higher when compared to the predicate and reference devices. The difference in the vent to volume ratio (in-1) does not affect the sterilization process and sterility assurance level (SAL) of 10-6as demonstrated by the sterilization validation. |
| Sterility | Provided non-sterile (to be steam sterilized by the end user) | Provided non-sterile (to be steam sterilized by the end user) | Provided non-sterile (to be steam sterilized by the end user) | Provided non-sterile (to be steam sterilized by the end user) | Same |
| Sterilization Method | Dynamic air removal (pre-vacuum) | Dynamic air removal (pre-vacuum) | Dynamic air removal (pre-vacuum) | Dynamic air removal (pre-vacuum) | Same |
| Sterilization Parameters | Dynamic air removal (pre-vacuum) at 132 °C (270 | Dynamic air removal (pre-vacuum) at 132 °C (270 | Dynamic air removal (pre-vacuum) at 132 °C (270 | Dynamic air removal (pre-vacuum) at 132 °C (270 | Same |
| Sterilization Time (min) | 1000 | 1000 | 1000 | 1000 | Same |
{12}
| Element | Proposed Device Stainless Steel Surgical Kits | Predicate Device Surgical and Prosthetic Trays (K220791) | Reference Device Washtrays (K231144) | Reference Device Biotech Dental Kits 2-piece metal trays (K212068) | Comparison |
| --- | --- | --- | --- | --- | --- |
| | °F) for 4 minutes with a 20 minute dry time. | °F) for 4 minutes with a 20 minute dry time. | °F) for 4 minutes with a 20 minute dry time. | °F) for 4 minutes with a 20 minute dry time. | |
| Sterile barrier | FDA cleared sterilization pouch (single pouch) | FDA cleared sterilization pouch (single pouch) | FDA cleared sterilization pouch (single pouch) | FDA cleared sterilization pouch (double pouch) | Same as the predicate and reference device (K231144). |
| Reusable | Yes | Yes | Yes | Yes | Same |
Page 8 of 9
{13}
# 7. Non-Clinical Performance Data
The proposed Stainless Steel Surgical Kits are reusable devices provided non-sterile which need to be end user cleaned and sterilized. Cleaning and sterilization of the proposed trays were validated according to the FDA guidance document, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".
Non-clinical testing data submitted, referenced, or relied upon, including acceptance criteria and set specifications in test methodology and standards, are summarized below:
| Test Name | Test Methodology | Purpose | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| Cleaning (Manual Pre-Cleaning and Automated Cleaning) validation | ANSI/AAMI ST98:2022 | To validate that the trays can be cleaned as specified on labeling | No visible soil. Residual protein: ≤ 6.4 μg/cm2Total Organic Carbon: ≤ 12 μg/cm2 | Pass |
| Sterilization cycle validation and drying time validation | • ISO 17665:2024 • AAMI TIR 12:2020/(R) 2023 | To validate that the trays can be sterilized via moist-heat sterilization as specified on labeling (132°C for 4 minutes)To validate that the drying time parameters are suitable | Sterility assurance level (SAL) of ≤ 10-6No visible moisture ≤ 1% weight gain of packaging and absorbable materials | Pass |
| Service Life validation | • ANSI/AAMI ST98:2022 • Internal Test - visual and functional tests | To confirm that the trays can withstand simulated use and up to 50 reprocessing cycles | Internal visual and functional acceptance criteria must be met.Total Organic Carbon: ≤ 12 μg/cm2 | Pass |
| Transportation validation | • ASTM D4332-22 • ASTM D4169-23e1 • Internal Test - visual and functional tests | To confirm that the packaging protects the device during simulated distribution | Internal visual and functional acceptance criteria must be met. | Pass |
| Cytotoxicity testing | ISO 10993-5:2009/(R)2022 | To confirm that no cytotoxic substances are released after 50 or 51 reprocessing cyclesTo confirm that no cytotoxic residuals of detergent remain on the device after 8 simulated use cycles | Inhibition of cell proliferation must be at or below 30% compared to untreated cultures | Pass |
# 8. Conclusion
The conclusions drawn from the non-clinical test data and evaluation support that the proposed Stainless Steel Surgical Kits are as safe, as effective and perform as well as the legally marketed predicate device Dentsply Sirona Surgical Trays (K220791).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.