K251596 · William Cook Europe Aps · DQX · Nov 9, 2025 · Cardiovascular
Device Facts
Record ID
K251596
Device Name
Lunderquist Extra Stiff Wire Guide
Applicant
William Cook Europe Aps
Product Code
DQX · Cardiovascular
Decision Date
Nov 9, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Intended Use
The Lunderquist Extra Stiff Wire Guides are intended to facilitate catheterization and/or placement of devices during vascular interventional procedures. The Lunderquist Extra Stiff Wire Guides are intended for use in the aorta, including its access vessels and major adjacent vessels. The Lunderquist Extra Stiff Wire Guides are auxiliary devices and hence not primary interventional devices themselves. Instead, they are intended to facilitate vascular access and/or delivery of over-the-wire medical devices during endovascular procedures, including TEVAR/EVAR procedures. The Lunderquist Extra Stiff Wire Guides are both long and stiff; features that make them suitable for facilitating endovascular procedures that involve placement of large devices e.g., stent grafts in the aorta. The wire guides are not intended for coronary or neurovascular use.
Device Story
Auxiliary guidewire used to facilitate vascular access and delivery of over-the-wire medical devices; specifically designed for endovascular procedures involving large devices like stent grafts in the aorta. Device consists of 0.035-inch PTFE-coated stainless steel wire with a mandril and flexible coil tip. Available in various lengths (90-300 cm) and configurations (straight, curved, double curved, extended double curved). Used by physicians in clinical settings during vascular interventions. Provides structural support and trackability for primary interventional devices. Benefits patient by enabling successful placement of large endovascular implants.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
0.035-inch PTFE-coated stainless steel wire guide; mandril with 11 cm tapered end and 4 or 7 cm flexible tip; some models include inner gold coil. Supplied sterile via ethylene oxide in Tyvek peel pouch. Single-use.
Indications for Use
Indicated for patients undergoing vascular interventional procedures in the aorta, access vessels, and major adjacent vessels to facilitate catheterization and delivery of over-the-wire medical devices (e.g., stent grafts for TEVAR/EVAR). Contraindicated for coronary or neurovascular use.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K180919 — GORE Tri-Lumen Catheter · W. L. Gore and Associates, Inc. · Oct 5, 2018
K220137 — Lunderquist Extra Stiff Wire Guide · William Cook Europe Aps · Feb 17, 2022
K232570 — Steerant Super Stiff Guidewire · Medtronic Vascular · Oct 24, 2023
K251385 — InQwire Super Stiff Guide Wire (IQSS32180J3); InQwire Super Stiff Guide Wire (IQSS35F260S); InQwire Super Stiff Guide Wire (IQSS35F150S); InQwire Super Stiff Guide Wire (IQSS35F100S); InQwire Super Stiff Guide Wire (IQSS35F80S); InQwire Super Stiff Guide Wire (IQSS35F180S) · Merit Medical Ireland, Ltd. · Jan 21, 2026
K151244 — Pre-Formed Guidewire · Lake Region Medical · Jun 11, 2015
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 9, 2025
Anette Hulstrom
Specialist, Submissions, Regulatory Affairs
William Cook Europe ApS
Sandet 6
Bjaeverskov, 4632
Denmark
Re: K251596
Trade/Device Name: Lunderquist Extra Stiff Wire Guide
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: DQX
Dated: October 8, 2025
Received: October 10, 2025
Dear Anette Hulstrom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251596 - Anette Hulstrom
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251596 - Anette Hulstrom
Page 3
Sincerely,
Jenny R. Katsnelson -S
Digitally signed by Jenny R. Katsnelson -S
Date: 2025.11.09 11:46:24 -05'00'
for Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251596 | |
| Device Name Lunderquist Extra Stiff Wire Guide | |
| Indications for Use (Describe) The Lunderquist Extra Stiff Wire Guides are intended to facilitate catheterization and/or placement of devices during vascular interventional procedures. The Lunderquist Extra Stiff Wire Guides are intended for use in the aorta, including its access vessels and major adjacent vessels. The Lunderquist Extra Stiff Wire Guides are auxiliary devices and hence not primary interventional devices themselves. Instead, they are intended to facilitate vascular access and/or delivery of over-the-wire medical devices during endovascular procedures, including TEVAR/EVAR procedures. The Lunderquist Extra Stiff Wire Guides are both long and stiff; features that make them suitable for facilitating endovascular procedures that involve placement of large devices e.g., stent grafts in the aorta. The wire guides are not intended for coronary or neurovascular use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251596
Page 1 of 3
COOK
MEDICAL
WILLIAM COOK EUROPE AFS
SANDET 6, 4632 BJAEVERSKOV
DENMARK
PHONE: +45 56 86 86 86
FAX: +45 56 86 86 96
CVR NO: 83 74 23 13
WWW.COOKMEDICAL.COM
Special 510(k) Summary: K251596
Lunderquist Extra Stiff Wire Guide
Date prepared: 13 May 2025
Submitted By
William Cook Europe ApS
Sandet 6
4632 Bjaeverskov, Denmark
Contact: Mie Dyrholm
Phone Number: +45 56 86 84 92
Email: Mie.Dyrholm@CookMedical.com
Device Information
Trade Name: Lunderquist® Extra Stiff Wire Guide
Common Name: Guidewire
Classification Name: Catheter guide wire
Regulation: 21 CFR 870.1330
Product Code: DQX
Classification/Review Panel: Class II / Cardiovascular
Predicate Device
The subject Lunderquist® Extra Stiff Wire Guide is substantially equivalent to the predicate Lunderquist® Extra Stiff Wire Guide (K220137, cleared 17 Feb 2022).
Device Description
The Lunderquist® Extra Stiff Wire Guides are 0.035 inch PTFE coated stainless steel wire guides consisting of a mandril and a 15 cm or 18 cm long coil. The mandril has a tapered end with a length of 11 cm and an additional 4 or 7 cm flexible tip. The Lunderquist wire guides are available in the following configurations:
{5}
K251596
Page 2 of 3
Straight Lunderquist wire guides (TSMG-/-LES) are available in lengths of 90, 145, 180, 260, and 300 cm and either a 4 cm or 7 cm flexible tip. For the 260 cm and 300 cm lengths, the flexible tip includes an inner gold coil.
Curved Lunderquist wire guides (TSCMG-/-LES) are available in lengths of 90, 145, 180, 260, and 300 cm and either a 4 cm or 7 cm flexible tip. For the 260 cm and 300 cm lengths, the flexible tip includes an inner gold coil. The J-curve radius for the curved wire guides is either 3 mm or 7 mm.
Double Curved Lunderquist wire guides (TSCMG-/-LESDC) are available in lengths of 260 and 300 cm with a 4 cm flexible tip that includes an inner gold coil. The Double Curved Lunderquist wire guides have a large curve/J-curve radius of 75/15 mm.
Extended Double Curved Lunderquist wire guides (TSCMG-/-E-LESDC) are available in lengths of 260 and 300 cm with a 4 cm flexible tip that includes an inner gold coil. The Extended Double Curved Lunderquist wire guides have an extended large curve/J-curve radius of 55/15 mm.
## Indications for Use
The Lunderquist Extra Stiff Wire Guides are intended to facilitate catheterization and/or placement of devices during vascular interventional procedures.
The Lunderquist Extra Stiff Wire Guides are intended for use in the aorta, including its access vessels and major adjacent vessels.
The Lunderquist Extra Stiff Wire Guides are auxiliary devices and hence not primary interventional devices themselves. Instead, they are intended to facilitate vascular access and/or delivery of over-the-wire medical devices during endovascular procedures, including TEVAR/EVAR procedures.
The Lunderquist Extra Stiff Wire Guides are both long and stiff; features that make them suitable for facilitating endovascular procedures that involve placement of large devices e.g., stent grafts in the aorta. The wire guides are not intended for coronary or neurovascular use.
## Comparison to Predicate Device:
The subject device is substantially equivalent to the predicate Lunderquist Extra Stiff Wire Guide cleared under K220137. The principles of operation, design, materials of construction, manufacturing processes, packaging, sterilization process, and basic technological characteristics of the subject device are identical to the predicate device. The device labeling has been updated in alignment with the risk documentation, post market data and global harmonization of the labeling. The labeling updates the indications for use statement, specifically removing the vena cava and specifying the major vessels in alignment with postmarket data.
{6}
K251596
Page 3 of 3
# Technological Characteristics
The technological characteristics remain the same as for the predicate device.
The Lunderquist® Extra Stiff Wire Guides are 0.035 inch PTFE coated stainless steel wire guides consisting of a mandril and a 15 or 18 cm long coil. The mandril has a tapered end with a length of 11 cm and an additional 4 or 7 cm flexible tip. The device is supplied as single use in a Tyvek® peel pouch, sterilized by ethylene oxide gas.
# Performance Testing
No changes to the design, manufacturing, sterilization, or principles of operation have been introduced with the subject device. Therefore, no performance testing was warranted in support of this Special 510(k).
# Conclusion
The subject device is substantially equivalent to the predicate The Lunderquist Extra Stiff Wire Guide cleared under K220137. The design and technological characteristics of the subject device are identical to the predicate device. The changes proposed in this submission are to device labeling in alignment with risk documentation and in an effort to gain global harmonization of the labeling.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.