← Product Code [ODG](/productcode/ODG) · K251542

# InstaFAN (K251542)

_Onepass Medical , Ltd. · ODG · Jul 18, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K251542

## Device Facts

- **Applicant:** Onepass Medical , Ltd.
- **Product Code:** [ODG](/productcode/ODG.md)
- **Decision Date:** Jul 18, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The InstaFAN device is intended for sampling targeted submucosal and extramural lesions within or adjacent to the gastrointestinal tract through the accessory channel of a curvilinear echo-endoscope. Use only with EUS endoscopes.

## Device Story

InstaFAN is an endoscopic ultrasound-guided fine needle aspiration/biopsy device; collects multiple samples from different locations in a single pass. Device comprises central 22-gauge needle and four peripheral 25-gauge needles; single handle design with two moving parts. Distal handle section activates central needle; proximal section deploys all five needles simultaneously. Used in clinical settings by physicians; inserted through accessory channel of curvilinear echo-endoscope. Samples acquired by penetrating lesions while applying suction via syringe. Output is tissue/aspiration sample for diagnostic evaluation. Benefits include multi-site sampling efficiency in single procedure.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (tensile, elasticity, bend strength, connection force, insertion force, needle orientation) and animal studies. All results met safety and performance criteria.

## Technological Characteristics

Endoscopic ultrasound-guided biopsy device; 22-gauge central needle and four 25-gauge peripheral needles. Mechanical actuation via dual-section handle. Compatible with curvilinear echo-endoscopes. Non-powered, manual operation.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Acquire Endoscopic Ultrasound Fine Needle Biopsy (FNB) Device ([K160845](/device/K160845.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

July 18, 2025

OnePass Medical Ltd.
% Danielle Besal
Principal Consultant
MRC Global
9085 East Mineral Circle
Centennial, Colorado 80112

Re: K251542
Trade/Device Name: InstaFAN
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: ODG, FCG
Dated: May 20, 2025
Received: May 20, 2025

Dear Danielle Besal:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251542 - Danielle Besal
Page 2 of 3

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K251542 - Danielle Besal
Page 3 of 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

SIVAKAMI VENKATACHALAM -S

for

Shanil P. Haugen, Ph.D.

Assistant Director

DHT3A: Division of Renal, Gastrointestinal,

Obesity and Transplant Devices

OHT3: Office of Gastrorenal, ObGyn,

General Hospital and Urology Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

InstaFAN
Page 9 of 36

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251542 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  InstaFAN  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The InstaFAN device is intended for sampling targeted submucosal and extramural lesions within or adjacent to the gastrointestinal tract through the accessory channel of a curvilinear echo-endoscope. Use only with EUS endoscopes.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

K251542 Page 1 of 2

|  510(k) #: | 510(k) Summary | Prepared on: 2025-05-20  |
| --- | --- | --- |
|  Contact Details |   | 21 CFR 807.92(a)(1)  |
|  Applicant Name | OnePass Medical Ltd. |   |
|  Applicant Address | 1st The Four Street Katzrin Industrial Park Katzrin 1290000 Israel |   |
|  Applicant Contact Telephone | 972.54.6670181 |   |
|  Applicant Contact | Jacob Ben Arie |   |
|  Applicant Contact Email | jacob@onepassmedical.com |   |
|  Correspondent Name | MRC Global |   |
|  Correspondent Address | 9085 East Mineral Circle Centennial CO 80112 United States |   |
|  Correspondent Contact Telephone | 9018278670 |   |
|  Correspondent Contact | Danielle Besal |   |
|  Correspondent Contact Email | danielle.besal@askmrcglobal.com |   |
|  Device Name |   | 21 CFR 807.92(a)(2)  |
|  Device Trade Name | InstaFAN |   |
|  Common Name | Endoscope and accessories |   |
|  Classification Name | Endoscopic Ultrasound System, Gastroenterology-Urology |   |
|  Regulation Number | 876.1500 |   |
|  Product Code(s) | ODG, FCG |   |
|  Legally Marketed Predicate Devices |   | 21 CFR 807.92(a)(3)  |
|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
|  K160845 | Acquire Endoscopic Ultrasound Fine Needle Biopsy (FNB) Device | ODG  |
|  Device Description Summary |   | 21 CFR 807.92(a)(4)  |
|  InstaFAN is an endoscopic ultrasound-guided fine needle aspiration/fine needle biopsy device that collects several samples from different locations in a single pass. The InstaFAN can be coupled to the biopsy channel of a Curvilinear Array (CLA) Echoendoscope and delivered to the gastrointestinal tract. The InstaFAN device is comprised of a central single 22-gauge needle and four peripheral 25-gauge needles. The device has a single handle design that contains two moving parts. The distal section of the handle controls the central needle, which can be activated first on its own or together with the proximal section of the handle to deploy all five needles simultaneously. The needles are used to acquire samples from lesions within and adjacent to the digestive system's major lumens that can be identified and targeted using the echoendoscope. Aspiration/biopsy samples are obtained by penetrating the lesion with the needle(s) while applying suction with the syringe.  |   |   |
|  Intended Use/Indications for Use |   | 21 CFR 807.92(a)(5)  |

{5}

K251542 Page 2 of 2

The InstaFAN device is intended for sampling targeted submucosal and extramural lesions within or adjacent to the gastrointestinal tract through the accessory channel of a curvilinear echo-endoscope. Use only with EUS endoscopes.

|  Indications for Use Comparison | 21 CFR 807.92(a)(5)  |
| --- | --- |

The subject device includes a statement about use with EUS endoscopes; however, this is a clerical difference for enhanced clarity that does not impact the indications for use.

|  Technological Comparison | 21 CFR 807.92(a)(6)  |
| --- | --- |

The subject InstaFAN device uses the same fundamental technology as the predicate device. The safety and effectiveness of the subject device has been verified and validated and performance was shown to be substantially equivalent. Thus, it can be concluded that the subject InstaFAN device does not raise different questions about safety and effectiveness.

|  Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b)  |
| --- | --- |

Performance testing was conducted on the InstaFAN device to verify the specifications of the device design. Testing included verification of the performance of the needles (tensile, elasticity, bend strength) as well as that of the overall device (EUS connection force, bend strength, force required to move needles, insertion force, and needle orientation). Results of the testing showed the device met its specifications. Animal studies were also conducted on the InstaFAN device and results met the safety and performance criteria to demonstrate substantial equivalence to the predicate.

Clinical testing was not required for the subject device; therefore, this section is not applicable.

The InstaFAN device uses the same fundamental technology as the predicate device. The differences between the InstaFAN device and its predicate device do not introduce a new intended use and do not raise different questions of safety and effectiveness. Verification and validation testing demonstrated that the device performs as intended. From the results of nonclinical testing, OnePass Medical concludes that the InstaFAN device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K251542](https://fda.innolitics.com/device/K251542)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).