← Product Code [IRP](/productcode/IRP) · K251531

# AIR COMPRESSION BOOTS 1018195,1018196 (K251531)

_Merchsource, LLC · IRP · Oct 30, 2025 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K251531

## Device Facts

- **Applicant:** Merchsource, LLC
- **Product Code:** [IRP](/productcode/IRP.md)
- **Decision Date:** Oct 30, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5650
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The AIR COMPRESSION BOOTS is intended for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas in people who are in good health. The AIR COMPRESSION BOOTS simulates kneading and stroking of tissues by using an inflatable garment.

## Device Story

Device consists of hand controller, air pump, solenoid valve, and inflatable leg garments; operates by pumping air into garment chambers to simulate kneading/stroking of tissues. Used in clinics, hospitals, athlete training, or home environments; operated by user. User selects therapy time (10, 20, 30 mins) and massage mode (Pulse, Gradual, Sequential, Zones, Manual). Output pressure range 120-210mmHg. Provides temporary relief of minor muscle aches/pains and increased circulation. Safety features include manual stop/pause and automatic power-off after 3 minutes.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench performance testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), accelerated lifecycle reliability testing, and software verification/validation.

## Technological Characteristics

Powered inflatable tube massager; 5-chamber leg garment made of polyethylene terephthalate (PET). Microprocessor-controlled air pump and solenoid valve system. Power: 100-240V, 50/60Hz, 36W. Dimensions: 207x145x112mm (controller). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-1. Non-conductive patient contact.

## Regulatory Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

## Predicate Devices

- Rapid Reboot Compression Therapy System ([K182668](/device/K182668.md))
- Air compression therapy system: FO-3001 ([K201982](/device/K201982.md))

## Reference Devices

- The Oligio ([K221989](/device/K221989.md))
- Relaxor Perfect Touch Air Massaging System ([K030437](/device/K030437.md))
- NormaTec Pulse and NormaTec Pulse Pro ([K160608](/device/K160608.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 30, 2025

Merchsource LLC
Shirley Luo
Registered Engineer
7755 Irvine Center Drive SUITE 100
Irvine, California 92618

Re: K251531
Trade/Device Name: AIR COMPRESSION BOOTS 1018195, 1018196
Regulation Number: 21 CFR 890.5650
Regulation Name: Powered inflatable tube massager
Regulatory Class: Class II
Product Code: IRP
Dated: May 19, 2025
Received: May 19, 2025

Dear Shirley Luo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251531 - Shirley Luo
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K251531 - Shirley Luo
Page 3

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251531  |   |
|  Device Name AIR COMPRESSION BOOTS 1018196,1018195  |   |
|  Indications for Use (Describe) The AIR COMPRESSION BOOTS is intended for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas in people who are in good health. The AIR COMPRESSION BOOTS simulates kneading and stroking of tissues by using an inflatable garment.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K251531
510(k) Summary

Prepared on 2025-10-27

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a traditional 510(K) submission with no previous application.

1. Submitter's Information

Sponsor
- Company Name: Merchsource LLC
- Address: 7755 Irvine Center Drive SUITE 100, Irvine, California 92618
- Phone: +1 949 900 6598
- Contact Person (including title): Shirley Luo
- E-mail: shirley.luo@merchsource.com

Application Correspondent
- Company: Merchsource LLC
- Address: 7755 Irvine Center Drive SUITE 100, Irvine, California 92618
- Tel: 86-13512755282
- Email: jianda-lee@foxmail.com

2. Subject Device Information
- Name of Device: AIR COMPRESSION BOOTS 1018196, 1018195
- Common or Usual Name: AIR COMPRESSION BOOTS
- Regulation Name: Powered inflatable tube massager
- Regulation Number: 21 CFR 890.5650
- Regulatory Class: II
- Product Code: IRP

3. Predicate Device
- Predicate Device 1: Rapid Reboot Compression Therapy System, K182668
- Predicate Device 2: Air compression therapy system: FO-3001, K201982

{5}

4. Device Description

4.1 Overall product description

This product operates through an air pump stored internally, sending air into the air bag, inflating and deflating through the alternate work of the solenoid valve, and squeezing the leg to achieve a massage effect.

4.2 Structure description

This product consists of hand controller, air pump, air valve and leg bag.

4.3 The statement of difference between models

The following model numbers are identical in circuitry and electrical, mechanical and physical construction; the only differences are the size and model number for trading purpose. The model 1018196 is for the 26-inch size and model 1018195 is for the 32-inch size.

5. Indications for Use

The AIR COMPRESSION BOOTS is intended for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas in people who are in good health. The AIR COMPRESSION BOOTS simulates kneading and stroking of tissues by using an inflatable garment.

{6}

# 6. Comparison of Technological Characteristics with the Predicate Device

|  Elements of Comparison | Subject Device | Predicate Device I | Predicate Device 2 | Verdict  |
| --- | --- | --- | --- | --- |
|  Company | Merchsource LLC | Rapid Reboot Recovery Products, LLC | FOSHAN HONGFENG CO., LTD. |   |
|  Trade Name | AIR COMPRESSION BOOTS | Rapid Reboot Compression Therapy System | Air compression therapy system: FO-3001 | --  |
|  Regulation Name | Powered inflatable tube massager | Powered inflatable tube massager | Powered inflatable tube massager | Same  |
|  510(k) Number | K251531 | K182668 | K201982 | --  |
|  Product Code | IRP | IRP | IRP | Same  |
|  Regulation Number | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | Same  |
|  Regulatory Class | II | II | II | Same  |
|  Review Panel | Physical Medicine | Physical Medicine | Physical Medicine | Same  |
|  Rx or OTC | OTC | OTC | OTC |   |
|  Indications for Use | The AIR COMPRESSION BOOTS is intended for the temporary relief of minor muscle aches and pains and for temporary increase in blood circulation to the treated areas in people who are in good health. The AIR COMPRESSION BOOTS simulates kneading and stroking of tissues by using an inflatable garment. | The Rapid Reboot Compression Therapy System is intended for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Rapid Reboot Compression Therapy System simulates kneading and stroking of tissues by using an inflatable garment. | Intended for home to temporarily relieve minor muscle aches and/or pains, promote blood circulation in treated areas. | Same  |

3

{7}

|  Elements of Comparison | Subject Device | Predicate Device I | Predicate Device 2 | Verdict  |
| --- | --- | --- | --- | --- |
|  Treatment Time | User determines therapy time. Choose from 10, 20, or 30 minute session time, with option to add additional 10 minutes to any therapy time. | User determines therapy time. Choose from 10, 20, or 30 minute session time, with option to add additional 10 minutes to any therapy time. | 1 - 60mins | Same  |
|  Treatment Area | Leg | Leg (including of foot, calf, knee, upper leg).Hip (including of hips, upper leg, glutes, lower lack).Arm (including of entire arm, shoulder, upper chest and back) | Leg | Same  |
|  Environment of Use | Clinics, hospital, athlete training, and home environments | Clinics, hospital, athlete training, and home environments | Not provided in the 510k Summary | Same  |
|  Modes | 5 massage modes
Scroll adjustment working mode, from pulse (default)--gradient
5 massage modes:
Pulse(default): All zones compress and decompress. | 2 modes: "A" mode inflates and deflates chambers from bottom up (distal to proximal), out at a time. "B" mode also inflates chambers from bottom up, but maintains pressure in lower chambers as it works its way to top. Then all | 6 chambers:
(Normal Mode):
Chamber ① inflating till setup pressure or for 2 seconds, then hold air for 2 seconds, start deflating;
chamber ② start like chamber ① | Similar Note 1  |

{8}

|  Elements of Comparison | Subject Device | Predicate Device I | Predicate Device 2 | Verdict  |
| --- | --- | --- | --- | --- |
|   | Gradual: Zones compress sequentially, then all decompress. Sequence: Zones compress and decompress sequentially Zones: Select up to 5 zones for targeted pulse massage. Manual: Select up to 5 zones and manually adjust compression. | chambers release pressure at same time once all chambers have sequentially inflated. | way till chamber ⑥, then rest 3 seconds, then restart chamber ① ②③④⑤⑥ again.

(Squeeze Mode):
chamber inflating till set up pressure or for 28 seconds, then hold the pressure, chamber ② inflating, till setup pressure or for 28 seconds, then chamber ① hold pressure in same time, then chamber ③ start inflating, same way after chamber ⑥ till Chamber ①②③④⑤⑥ deflating in same time for 3 seconds. Then repeat. (Double Wave Mode): chamber ①② inflating till setup pressure or for 40 seconds, hold air for 2 seconds, then start deflating.

Chamber ⑤④ start inflating till setup pressure or for 40 seconds, hold air for 2 seconds, then deflating, same way for chamber ③④ till chamber ⑤⑥, pause for 3 seconds. Then repeat. D (Whole Squeeze Mode): chamber ①②③④⑤⑥ inflating at the same time till setup pressure or for 90 seconds, then deflating in the same time for 3 seconds. Then repeat. E (Combined B + C): sequential squeeze + double wave F (Combined A + C + D): normal + double wave + whole squeeze. 4 and 8 chambers same as 6 chambers above. |   |
|  Output Pressure Range | 120-210mmHg | 0-200 mmHg | 30-110mmHg | Similar Note 2  |
|  Power Source | 100-240V~, 50/60Hz | 110VAC, 60Hz | 100 - 127V/220 - 240V, 50/60Hz | Similar Note 3  |
|  Power | 36W | 30W | 65 W | Similar Note 3  |

{9}

|  Elements of Comparison | Subject Device | Predicate Device I | Predicate Device 2 | Verdict  |
| --- | --- | --- | --- | --- |
|  Number of Chambers | 5 | 4 | 4, 6, 8 | Similar Note 4  |
|  Appearance and Size of Control Unit | 207*145*112mm | 10" x 6.5" x 5" | 220*190*113mm | Similar Note 5  |
|  Appearance and Size of Boots | 32inch:787*1266mm
26inch:787*1108mm | X-Short: 14" x 41"
Short: 14" x 43"
Medium: 14" x 45"
Long: 14" x 48"
X-Long: 14" x 52" | M:91*65cm
L:100*74cm
XL:110*70cm
(Overlapping) | Similar Note 5  |
|  Housing Materials | Molded ABS enclosure | Molded ABS enclosure | Molded ABS enclosure | Same  |

{10}

|  Elements of Comparison | Subject Device | Predicate Device I | Predicate Device 2 | Verdict  |
| --- | --- | --- | --- | --- |
|  Compression Applicator Garments Boots/Sleeves /Pants Material | Polyethylene terephthalate (PET) | Nylon with a Polyurethane laminate | Not mentioned in the 510k summary | Different Note 6  |
|  Safety Feature | a. Quick press the power button during a massage to begin Decompression.
b. The massager will automatically power off after 3 minutes, with boots fully decompressed. | Button on display allows users to stop or pause therapy session at any time | Power button on main unit allows user to stop therapy session at | Same  |
|  Mode of Compression | Sequential | Sequential | Sequential | Same  |
|  Software/Firmware/ Microprocessor Control | Microprocessor | Microprocessor | Microprocessor | Same  |
|  Technology | Compressor and valve system which sequentially inflates cells of inflatable chambers. | Compressor and valve system which sequentially inflates cells of inflatable chambers. | Compressor and valve system which sequentially inflates inflatable chambers. | Same  |
|  Standards | IEC 60601-1-2
IEC 60601-1
IEC 60601-1-11
ISO 10993-1 | ANSI - AAMI
ES 60601 - 1
IEC 60601-1-2
IEC 60601-1-11
ISO 10993-5
ISO 10993-10 | IEC 60601 - 1
IEC 60601-1-2
IEC 60601-1-11
ISO 10993-5
ISO 10993-10
ISO 10993-12 | Same  |
|  Housing Materials And Construction | Molded ABS enclosure | Molded ABS enclosure | Molded ABS enclosure | Same  |
|  Patient Contact | Non-conductive attachments | Non-conductive attachments | Non-conductive attachments | Same  |
|  Cycle Time | Pulse Mode: 25-30 sec
Gradual Mode: 25-30 sec
Sequential Mode: 20-30 sec
Zones Mode: 25-30 sec
Manual Mode: 4-4.5 min | 1 min 20 sec | Not provided in the 510k Summary | Different Note 7  |

{11}

Comparison in Detail(s):

Note 1:
The Modes of subject device are similar to those of the cited predicate devices. The subject device have also been demonstrated to comply with the applicable requirements of IEC 60601-1, IEC 60601-1-2, IEC60601-1-11 as well as relevant performance testing. Therefore, this difference does not impact the safety and effectiveness of the device.

Note 2:
The Output Pressure Range of the subject device is similar to that of Predicate Device 1. The subject device has also been demonstrated to comply with the applicable requirements of IEC 60601-1, IEC 60601-1-2, IEC60601-1-11 as well as relevant performance testing. Therefore, this difference does not impact the safety and effectiveness of the device.

Note 3:
The Power Source and Power specifications of the subject device are similar to those of the cited predicate devices. The subject device has been demonstrated to comply with the applicable requirements of IEC 60601-1, IEC 60601-1-2, IEC60601-1-11, as well as relevant performance testing. Therefore, this difference does not impact the safety and effectiveness of the device.

Note 4:
The Number of Chambers in the subject device is similar to that of the cited predicate devices. The subject device has been demonstrated to comply with the applicable requirements of IEC 60601-1, IEC 60601-1-2, IEC60601-1-11, as well as relevant performance testing. Therefore, this difference does not impact the safety and effectiveness of the device.

Note 5:
The Appearance and Size of the Control Unit and Boots of the subject device is similar to those of the cited predicate devices. The subject device has been demonstrated to comply with the applicable requirements of IEC 60601-1, IEC 60601-1-2, IEC60601-1-11, as well as relevant performance testing. Therefore, these differences do not impact the safety and effectiveness of the device.

Note 6:
The compression boots of the subject device are made of polyethylene terephthalate (PET). Although this material differs from those of the cited predicate devices, PET is included in Attachment G, Section B of the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". PET has a documented history of safe use in legally marketed medical devices that contact intact skin (e.g., "The Oligio", K221989). None of the exclusions listed in Attachment G, Section C apply. Therefore, this material

8

{12}

difference does not impact safety and effectiveness.

9

Note 7:

The cycle times for the subject device differ from that of Predicate Device 1. The cycle time for Predicate Device 2 is not provided in its 510(k) Summary. The subject device's manual mode has a longer cycle time, while all other therapeutic modes operate with shorter cycles times. The subject device has been demonstrated to comply with the applicable requirements of IEC 60601-1, IEC 60601-1-2, IEC60601-1-11, as well as relevant performance testing. The cycle times of the subject device are also comparable to other Class II powered inflatable tube massager devices cleared under 21 CFR 890.5650 with product code IRP such as the "Relaxor Perfect Touch Air Massaging System", (K030437) and "NormaTec Pulse and NormaTec Pulse Pro" (K160608), which have similar intended uses and technology. Since FDA has previously cleared devices under the same regulation and product code with comparable intended use, technology, and cycle times without identifying safety or effectiveness concerns, this difference does not impact the safety and effectiveness of the subject device.

7. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

1) Electrical safety and electromagnetic compatibility

The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

IEC 60601-1 Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.

IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

2) Software verification and validation testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's 2023 Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". The software for this device was developed and reviewed under an Enhanced Documentation Level.

3) Cybersecurity

The subject device does not have any external interfaces. According to FDA's guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", no further cybersecurity evaluation is needed.

{13}

1C

4) Bench Testing

Bench Performance and Safety Testing Report for AIR COMPRESSION BOOTS.
Accelerated Lifecycle Reliability Test Report for AIR COMPRESSION BOOTS.
Post-Reliability Performance Verification Test Report for AIR COMPRESSION BOOTS.

5) Clinical Testing

Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.

8. Conclusion

The subject device has the same indications for use, core technological characteristics and principles of operation as the cited predicate devices. The technological differences between the subject and predicate devices do not impact safety and effectiveness. The performance data demonstrates that the subject device performs comparably to the predicate devices. Thus the subject device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K251531](https://fda.innolitics.com/device/K251531)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
