← Product Code [OWB](/productcode/OWB) · K251520

# Cios Alpha; Cios Flow (K251520)

_Siemens Medical Solutions USA, Inc. · OWB · Jul 9, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K251520

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [OWB](/productcode/OWB.md)
- **Decision Date:** Jul 9, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures. The patient population may include pediatric patients. The Cios Flow is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

## Device Story

Mobile fluoroscopic C-arm X-ray system for surgical environments; consists of C-arm with X-ray source and flat panel detector, and mobile display trolley. Captures X-ray images of anatomical structures; supports orthopedic and vascular procedures. Operated by trained healthcare professionals in clinical settings. System includes updated software (VA31A), collimator, and optional Trixell Pixium 3131SOD IGZO flat panel detector. Imaging data processed via FLC imaging system with high-performance graphics card. Provides real-time visualization to assist clinicians in surgical guidance and decision-making; benefits patients through improved intraoperative imaging capabilities.

## Clinical Evidence

Clinical cadaver study conducted to evaluate subjective image quality of the new IGZO flat panel detector versus the predicate a-Si detector across four anatomical regions in an ortho-trauma setting. Results demonstrated non-inferiority (equal or better) of the IGZO detector. Additional verification and validation testing performed on software and hardware components.

## Technological Characteristics

Mobile C-arm X-ray system; X-ray source and flat panel detector (Trixell Pixium 3131SOD with IGZO technology). Connectivity via data network; power via main outlet and Eaton UPS 5P 850i G2. Software version VA31A. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-43, 60601-2-54, and DICOM standards.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Cios Alpha ([K210055](/device/K210055.md))
- Cios Flow ([K203504](/device/K203504.md))
- Cios Alpha ([K132094](/device/K132094.md))

## Reference Devices

- Ziehm Vision RFD ([K231692](/device/K231692.md))
- CIARTIC Move ([K233748](/device/K233748.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

July 9, 2025

Siemens Medical Solutions USA, Inc
% Patricia Jones
Regulatory Affairs Professional
40 Liberty Boulevard
MALVERN, PA 19355

Re: K251520
Trade/Device Name: Cios Alpha; Cios Flow
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-Intensified Fluoroscopic X-Ray System
Regulatory Class: Class II
Product Code: OWB, JAA, OXO
Dated: May 16, 2025
Received: May 16, 2025

Dear Patricia Jones:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251520 - Patricia Jones
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}

K251520 - Patricia Jones
Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251520  |   |
|  Device Name Cios Alpha; CiosFlow  |   |
|  Indications for Use (Describe) The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures. The patient population may include pediatric patients. The Cios Flow is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

SIEMENS Healthineers
K251520

510(k) Summary: Cios Alpha
Cios Flow

Company: Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard, 65
Malvern, PA 19355

Date Prepared: May 16, 2025

1. General Information:
Importer / Distributor:
Siemens Medical Systems USA, Inc.
40 Liberty Boulevard, 65
Malvern, PA 19355
Establishment Registration Number: 2240869

Manufacturing Site:
Roentgenstrasse 19 -21
95478 Kemnath, Germany
Establishment Registration Number: 3002466018

2. Contact Person:
Ms. Patricia D. Jones
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Phone: (678) 575-8832
Email: patricia.jones@siemens-Healthineers.com

3. Device Name and Classification:

Trade Name: Cios Alpha
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-Ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Codes: OWB
Subsequent Product Codes: JAA, OXO

Trade Name: Cios Flow
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-Ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Regulatory Class: Class II
Product Code: OWB

Page 1 of 11

{5}

SIEMENS Healthineers

Subsequent Product Codes: JAA, OXO

## 4. Legally Marketed Primary Predicate Device

Trade Name: Cios Alpha
510(k) Clearance: K210055
Clearance Date: February 05, 2021
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Code: OWB
Subsequent Product Codes: JAA, OXO
Total Product Life Cycle: All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any applicable issues.

Trade Name: Cios Flow
510(k) Clearance: K203504
Clearance Date: December 22, 2020
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Code: OWB
Subsequent Product Codes: JAA, OXO
Total Product Life Cycle: All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any applicable issues.

## Secondary Predicate Device:

Trade Name: Cios Alpha
510(k) Clearance: K132094
Clearance Date: March 11, 2014
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Code: OWB
Subsequent Product Codes: JAA, OXO
Total Product Life Cycle: All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues.

Page 2 of 11

{6}

SIEMENS Healthineers

Reference Devices

Trade Name: Ziehm Vision RFD
510(k) Clearance: K231692
Clearance Date: November 20, 2023
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Code: OWB
Subsequent Product Codes: JAA,OXO

Trade Name: CIARTIC Move
510(k) Clearance: K233748
Clearance Date: March 15, 2024
Classification Name: Image-intensified fluoroscopic X-ray System
Common Name: Interventional Fluoroscopic X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Code: OWB
Subsequent Product Codes: JAA, OXO

5. Device Description:

The Cios Alpha and Cios Flow (VA31A) mobile fluoroscopic C-arm X-ray System is designed for the surgical environment. The Cios Alpha and Cios Flow provide comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:

a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The C-arm can be angulated in both planes and lifted vertically, shifted to the side, and moved forward/backward by an operator.
b) The second unit is the image display station with a movable trolley for the image processing and storage system, image display, and documentation. Both units are connected with a cable.

The main unit is connected to the main power outlet, and the trolley is connected to a data network.

The following modifications were made to the predicate device Cios Alpha and Cios Flow. Siemens Medical Solutions USA, Inc. submits this Bundled Traditional 510(k) to request clearance for Subject Devices Cios Alpha and Cios Flow (VA31A) for the following device modifications made to the Predicates Device Cios Alpha and Cios Flow (VA30).

Page 3 of 11

{7}

SIEMENS Healthineers

This 510(k) submission, Subject Devices "Cios Alpha" and "Cios Flow" with software version VA31A, will support the following categories of modifications made to the Subject Devices in comparison to the Predicate Devices:

## Modified Software:

Table 1: Overview of Software/Hardware Modifications supported by Software Version VA31A

|  Subject Devices Cios Alpha and Cios Flow (VA31A) Systems Modifications |   | Predicate Devices Cios Alpha (K210055) VA30 / Cios Alpha (K132094) VA10 Cios Flow (K203504) VA30  |
| --- | --- | --- |
|  1. | Software updated from VA30 to VA31A to support the following software features: | VA30 System Software  |
|   |  A. Updated InstantLink with Extended NXS Interface | InstantLink with NXS Interface  |
|  2. | Updated Collimator | Collimator  |
|  3. | New optional flat detector Trixell Pixium 3131SOD with IGZO (Indium Gallium Zinc Oxide) technology | Cios Alpha (K132094) VA10  |
|   |   |  Flat detector: aSi detector in Secondary Predicate Cios Alpha VA10  |
|  4. | Updated FLC imaging system with new PC hardware | FLC imaging system PC  |
|   |  Updated the High Performance Graphic Card on the Apphost PC | Apphost PC  |
|  5. | Updated Eaton UPS 5P 850i G2 as successor of UPS 5P 850i due to obsolescence | Eaton UPS 5P 850i  |
|  6. | The Cios Alpha is also known as “Cios Alpha.neo” The Cios Flow is also known as Cios Flow.neo | Trade name: Cios Alpha Trade name: Cios Flow  |
|  |   |   |

## 6. Indications for Use:

The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures. The patient population may include pediatric patients.

The Cios Flow is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include, but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures. The patient population may include pediatric patients.

Table 2: Predicate Device Comparable Properties for Subject Device Modifications:

|  Predicate Device Name and Manufacturer | 510(k) Number | Clearance Date | Comparable Properties  |
| --- | --- | --- | --- |
|  Primary Predicate | K210055 | 02/05/2021 | • Indications for Use • Software Version VA30 • Collimator • FLC Imaging PC • Apphost PC  |
|  Cios Alpha Siemens  |   |   |   |

Page 4 of 11

{8}

SIEMENS Healthineers

|  Predicate Device Name and Manufacturer | 510(k) Number | Clearance Date | Comparable Properties  |
| --- | --- | --- | --- |
|   |  |  | • Eaton UPS  |
|  Cios Flow Siemens | K203504 | 12/22/2020 | • Indications for Use
• Software Version VA30
• Collimator
• FLC Imaging PC
• Apphost PC
• Eaton UPS  |
|  **Secondary Predicate Device** | K132094 | 03/11/2014 | • Flat Panel Detector (Amorphous Silicon)  |
|  Cios Alpha (VA10) Siemens  |   |   |   |
|  **Reference Devices** | K231692 | 11/20/2023 | • Flat Panel Detector (IGZO technology)  |
|  Ziehm Vision RFD Ziehm  |   |   |   |
|  CIARTIC Move Siemens | K233748 | 03/15/2024 | • Collimator  |

7. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Cios Alpha and Cios Flow (VA31A) Systems are substantially equivalent to the commercially available Siemens Cios Alpha (VA30), cleared with K210055 and Cios Flow (VA30) cleared with K203504. Indication for use remains unchanged. The technology and design of the Cios Alpha and Cios Flow are based on the predicate Cios Alpha and Cios Flow (VA30). Technological differences between the Subject Devices and the Predicate Devices are provided in Table 3 below for all modifications.

Page 5 of 11

{9}

SIEMENS Healthineers

Table 4: Summary of Comparison of Technological Characteristics

|  Modifications | Subject Devices
Cios Alpha / Cios Flow (VA31A) | Predicate Devices
Cios Alpha (K210055) VA30 /
Cios Alpha (K132094) VA10
Cios Flow (K203504) VA30 | Comparison
Results  |
| --- | --- | --- | --- |
|  New System Software/
Hardware Changes | 1. Software updated from VA30 to VA31A to support the following software features: | VA30 System Software | Comparable:
System software VA31A was updated to support modifications 1A-5. Software modifications conform to “Guidance for the Content of Premarket Submissions for Device Software Functions”. All software documentation meets “Basic Level” requirements.

Bench tests were conducted and found acceptable and did not raise any new issues of safety or effectiveness.

All software validation data demonstrate that the Subject devices are as safe and effective when compared to the Predicate Devices, that is currently marketed for the same intended use. All test results met all acceptance criteria.  |
|   |  A. Updated InstantLink with Extended NXS Interface | InstantLink with NXS Interface  |   |
|   |  2. Updated Collimator | Collimator  |   |
|   |  3. New optional flat detector Trixell Pixium 3131SOD with IGZO (Indium Gallium Zinc Oxide) technology | Cios Alpha (K132094) VA10  |   |
|   |   |  Flat detector: aSi detector in Secondary Predicate Cios Alpha VA10  |   |
|   |  4. Updated FLC imaging system with new PC hardware | FLC imaging system PC  |   |
|   |  Added a High-Performance Graphic Card to the Apphost PC | Apphost PC  |   |
|   |  5. Updated Eaton UPS 5P 850i G2 as successor of UPS 5P 850i due to obsolescence | Eaton UPS 5P 850i  |   |
|  6. The Cios Alpha is also known as “Cios Alpha.neo”.
The Cios Flow is also known as “Cios Flow.neo”. | Trade name: Cios Alpha
Trade name: Cios Flow | The Cios Alpha and Cios Flow will also be known known as “Cios Alpha.neo” and “Cios Flow.neo”  |   |
|  The changes of the proposed modified device Cios Alpha / Cios Flow (VA31A) described in the table do not change the fundamental control mechanism, operating principle, energy type, or intended use found on predicate devices and supports substantially equivalents to the predicate devices Cios Alpha (K210055) VA30 / Cios Alpha (K132094) VA10 and Cios Flow (K203504) in accordance with its labeling.  |   |   |   |

Page 6 of 11

{10}

SIEMENS Healthineers

# 8. Nonclinical Performance Testing:

During product development, non-clinical tests were conducted for the Cios Alpha and Cios Flow (VA31A). The following non-clinical testing was conducted on the presented modifications: Updated software from VA30 to VA31A; updated InstantLink with NXS interface; updated collimator; optional flat panel detector; new PCs for imaging FLC and high-performance graphic card for Apphost; and an updated Eaton UPS power supply. Software functional, verification, and System validation testing were conducted with passing results.

The Cios Alpha and Cios Flow (VA31A) were certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance, and Electromagnetic Compatibility:

The modifications described in this Premarket Notification are supported by verification and validation testing.

|  Recog-nition # | Product Area | Title of Standard | Reference Number and Date | Standards Development Organization  |
| --- | --- | --- | --- | --- |
|  19-49 | General II (ES/EMC) | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied, see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section. | 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION | IEC  |
|  19-46 | General II (ES/EMC) | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)] | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | ANSI AAMI  |
|  19-36 | General II (ES/EMC) | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | IEC  |
|  12-336 | Radiology | Medical Electrical Equipment: Part 1: 1. Collateral Standard: General requirements for radiation protection in diagnostic x-ray equipment | 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION | IEC  |
|  5-132 | General I (QS/RM) | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability | 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION | IEC  |
|  12-273 | Radiology | Safety of laser products - Part 1: Equipment classification and | 60825-1:2014 | IEC  |

Page 7 of 11

{11}

SIEMENS
Healthineers

|  Recognition # | Product Area | Title of Standard | Reference Number and Date | Standards Development Organization  |
| --- | --- | --- | --- | --- |
|   |  | Requirements | (recognized 2007) |   |
|  13-79 | Software/Informatics | Medical Device Software – Software Life Cycle Processes | 62304:2015 | IEC  |
|  12-309 | Radiology | Medical electrical equipment - Part 2-28: Particular requirements for basic safety and essential performance of X-ray tube assemblies for medical diagnosis | 60601-2-28 Edition 3.0 2017-06 | IEC  |
|  12-351 | Radiology | Medical electrical equipment - Part 2-43: Particular requirements for basic safety and essential performance of X-ray equipment for interventional procedures | 60601-2-43 Edition 3.0 2022-12 | IEC  |
|  12-348 | Radiology | Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy | 60601-2-54 Edition 2.0 2022-09 | IEC  |
|  5-125 | General I (QS/RM) | Medical devices: Application of Risk management to medical devices | 14971:2019 | ISO  |
|  5-129 | General I (QS/RM) | Medical devices - Part 1: Application of usability engineering to medical devices | 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION | IEC  |
|  12-352 | Radiology | Digital Imaging and Communications in Medicine (DICOM) Set | PS 3.1 - 3.20 2023e | NEMA  |
|  19-19 | General II (ES/EMC) | Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems | TR 60601-4-2 Edition 1.0 2016-05 | IEC  |
|  5-134 | General I (QS/RM) | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements | 15223-1 Fourth edition 2021-07 | ISO  |
|  14-579 | Sterility | Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices. | 17664-2 First edition 2021-02 | ISO  |
|  5-135 | General I (QS/RM) | Medical devices - Information to be supplied by the manufacturer | 20417 First edition 2021-04 | ISO  |

Page 8 of 11

{12}

SIEMENS
Healthineers

|  Recog-nition # | Product Area | Title of Standard | Reference Number and Date | Standards Development Organization  |
| --- | --- | --- | --- | --- |
|   |  |  | Corrected version 2021-12 |   |
|  5-137 | General I (QS/RM) | Graphical symbols for electrical equipment in medical practice [Including: Corrigendum 1 (2023)] | TR 60878 Ed. 4.0 2022-11 | IEC  |
|  12-341 | Radiology | Medical electrical equipment - Medical image display systems - Part 1: Evaluation methods | 62563-1 Edition 1.2 2021-07 CONSOLIDATED VERSION | IEC  |
|  12-344 | Radiology | Medical electrical equipment - Medical image display systems - Part 2: Acceptance and constancy tests for medical image displays | 62563-2 Edition 1.0 2021-11 | IEC  |
|  13-38 | Software/Infor matics | Application of risk management for IT-networks incorporating medical devices - Part 1: Roles, responsibilities and activities | 80001-1 Edition 1.0 2010-10 | IEC  |
|  12-364 | Radiology | Evaluation and routine testing in medical imaging departments - Part 3-8: Acceptance and constancy tests - Imaging performance of X-ray equipment for radiography and radioscopy | 61223-3-8 Edition 1.0 2024-03 | IEC  |
|  13-83 | Software/Infor matics | Principles for medical device security - Risk management. | TIR57:2016 | AAMI  |
|  13-104 | Software/Infor matics | Standard for Safety, Software Cybersecurity for Network-Connectable Products, Part 2-1: Particular Requirements for Network Connectable Components of Healthcare and Wellness Systems | 2900-2-1 First Edition 2017 | UL ANSI  |
|  13-96 | Software/Infor matics | Standard for Safety, Standard for Software Cybersecurity Network- Connectable Products, Part 1: General Requirements | 2900-1 First Edition 2017 | UL ANSI  |
|  13-131 | Software/Infor matics | Standard for medical device security - Security risk management for device manufacturers | SW96:2023 | ANSI AAMI  |
|  19-48 | General II (ES/EMC) | American National Standard for Evaluation of Wireless Coexistence | USEMCSC C63.27-2021 | IEEE ANSI  |

Page 9 of 11

{13}

SIEMENS Healthineers

|  Recog-nition # | Product Area | Title of Standard | Reference Number and Date | Standards Development Organization  |
| --- | --- | --- | --- | --- |
|  19-22 | General II (ES/EMC) | Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems. | TIR69:2017/(R20 20) | AAMI  |

## Clinical Cadaver Report:

Summary of performed pre-market clinical development activities for Cios Flow VA31 and Cios Alpha VA31 new optional IGZO flat panel detector is as follows: In comparison to the a-Si detector, the IGZO detector can be regarded as non-inferior (equal or better) concerning the subjective image quality for four anatomical regions that have been investigated in the ortho-trauma setting.

## Verification and Validation:

Software Documentation for a Basic documentation level per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for "Guidance for the Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with medical devices containing software. Non-clinical tests were conducted on Cios Alpha and Cios Flow software (VA31A) during product development.

The Risk analysis was completed, and risk control was implemented to mitigate identified hazards. The testing results show that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

Cios Alpha and Cios Flow software (VA31A) were tested and found to be substantially equivalent for intended users, uses, and use environments through the design control verification and validation process. Customer employees are adequately trained in the use of this equipment.

Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse, or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1-2010 falls on the hospital. Cybersecurity information is provided in the Software Section, is the required cybersecurity information.

## Summary:

Performance tests and a Clinical Evaluation were conducted to test the functionality of the Cios Alpha and Cios Flow software (VA31A) systems. These tests have been performed to assess the functionality of the subject device. The results of all

Page 10 of 11

{14}

SIEMENS Healthineers

conducted testing were found acceptable and did not raise any new issues of safety or effectiveness.

9. General Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device safely and effectively.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practices, and all equipment is subject to final performance testing. Furthermore, the operators are healthcare professionals who are trained and responsible for evaluating the post-processing of X-ray images.

10. Conclusion as to Substantial Equivalence:
The predicate devices were cleared based on non-clinical supportive information. Non-clinical test results demonstrate that the Cios Alpha and Cios Flow software (VA31A) acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data, Clinical Cadaver Report, and software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Devices that are currently marketed for the same intended use and Indications for use.

Page 11 of 11

---

**Source:** [https://fda.innolitics.com/device/K251520](https://fda.innolitics.com/device/K251520)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
