Telemetry Monitor (iT30, iT50)

K251508 · Edan Instruments, Inc. · MHX · Aug 6, 2026 · Cardiovascular

Device Facts

Record IDK251508
Device NameTelemetry Monitor (iT30, iT50)
ApplicantEdan Instruments, Inc.
Product CodeMHX · Cardiovascular
Decision DateAug 6, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Indications for Use

The iT30 and iT50 Telemetry Monitor are intended for use on Adult and Pediatric patients (over three years old) to monitor physiological parameters include ECG, SpO2, PR, NIBP (Applicable to iT50 only), and Skin TEMP. The physiological data can be analyzed, alarmed, stored locally on the monitor, and the Edan Central Monitoring System can configure, display and review the physiological parameters from the iT30 and iT50 Telemetry Monitor. The arrhythmia detection and ST Segment analysis of ECG are intended for adult patients. The SpO2 and PR are used under no motion conditions. The NIBP is used under no motion conditions for patients over 12 years old and excluding pregnant women. The iT30 and iT50 Telemetry Monitor must be used by professional trained clinical medical staff in healthcare facilities.

Device Story

Telemetry monitors (iT30, iT50) provide continuous monitoring of ECG, SpO2, PR, NIBP, and skin temperature; data transmitted to Edan Central Monitoring System for display/review. Operated by trained clinical staff in healthcare facilities. Device processes physiological signals to detect arrhythmias and ST-segment abnormalities; triggers alarms for clinical response. NIBP uses oscillometric method (iCUFS algorithm). Power via AA or Li-ion batteries. Output enables real-time patient status assessment and clinical decision-making regarding patient care.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing against consensus standards for ECG (IEC 60601-2-27), NIBP (IEC 80601-2-30), pulse oximetry (ISO 80601-2-61), and clinical thermometers (ISO 80601-2-56).

Technological Characteristics

Battery-powered (AA/Li-ion) telemetry monitor. Parameters: ECG (3/5/6/10-lead), SpO2, PR, NIBP (oscillometric), Skin TEMP. Connectivity: IEEE 802.11 a/b/g/n Wi-Fi. Software-based arrhythmia/ST analysis. Complies with IEC 60601-1, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-30, ISO 80601-2-56, ISO 80601-2-61.

Indications for Use

Indicated for adult and pediatric patients (>3 years) for monitoring ECG, SpO2, PR, NIBP (iT50 only), and skin temperature. Arrhythmia/ST analysis limited to adults. NIBP limited to patients >12 years, excluding pregnant women. SpO2/PR monitoring restricted to no-motion conditions. Must be used by trained clinical staff in healthcare facilities.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} **U.S. FOOD & DRUG** ADMINISTRATION August 6, 2026 Edan Instruments Inc. Tracy Yue #15 Jinhui Road, Jinsha Community, Kengzi Sub-District Pingshan District Shenzhen, Guangdong 518122 China Re: K251508 Trade/Device Name: Telemetry Monitor (iT30, iT50) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, MSX, DPS, DSB, DSI, MLD, DRT, DXN, FLL, DQA Dated: January 9, 2026 Received: January 9, 2026 Dear Tracy Yue: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251508 - Tracy Yue Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251508 - Tracy Yue Page 3 Sincerely, # KIMBERLY N. CROWLEY -S For: Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251508 | ? | | Please provide the device trade name(s). | | ? | | Telemetry Monitor (iT30, iT50) | | | | Please provide your Indications for Use below. | | ? | | The iT30 and iT50 Telemetry Monitor are intended for use on Adult and Pediatric patients (over three years old) to monitor monitor physiological parameters include ECG, SpO2, PR, NIBP (Applicable to iT50 only), and Skin TEMP. The physiological data can be analyzed, alarmed, stored locally on the monitor, and the Edan Central Monitoring System can configure, display and review the physiological parameters from the iT30 and iT50 Telemetry Monitor. The arrhythmia detection and ST Segment analysis of ECG are intended for adult patients. The SpO2 and PR are used under no motion conditions. The NIBP is used under no motion conditions for patients over 12 years old and excluding pregnant women. The iT30 and iT50 Telemetry Monitor must be used by professional trained clinical medical staff in healthcare facilities. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | Telemetry Monitor Page 9 of 72 {4} Traditional 510(k) of Telemetry Monitor K251508 ### 510(K) Summary Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92 # **1. Submitter:** # **Applicant:** Edan Instruments, Inc. #15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District, Shenzhen, 518122 P.R.China. Tel: +86(0755) 26858736 Fax: +86(0755) 26882223 # **Contact person:** Tracy Yue # **Preparing date:** January 09, 2026 # **2. Device name and classification:** # **Trade name:** Telemetry Monitor, Model: iT30, iT50 # **Common/Usual Name:** Telemetry Monitor # **Classification Name:** 21 CFR 870.1025 Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms) # **Regulatory class:** Class II # **Primary product code:** MHX - Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms) # **Subsequent product code:** | Regulation number/Device | Product Code | | --- | --- | | 21 CFR 870.2300 System, network and communication, physiological monitors | MSX | | 21 CFR 868.2375 Electrocardiograph | DPS | | 21 CFR 870.2770 Plethysmograph, impedance | DSB | | 21 CFR 870.2340 Detector and Alarm, Arrhythmia | DSI | | 21 CFR 870.1025 Monitor, ST Segment with Alarm | MLD | | 21 CFR 870.2300 Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm) | DRT | | 21 CFR 870.1130 System, Measurement, Blood-Pressure, Non-Invasive | DXN | | 21 CFR 880.2910 | FLL | 1 / 6 {5} Traditional 510(k) of Telemetry Monitor 510(k) Summary K251508 | Thermometer, Electronic, Clinical | | | --- | --- | | 21 CFR 870.2700 Oximeter | DQA | ### 3. Predicate Device(s): 1) Primary Predicate Device: Shenzhen Mindray Bio-Medical Electronics Co., LTD., Telemetry Monitor TM80 (K193391) 2) Reference Device: Edan Instruments, Inc., Patient Monitor iX10/iX12/iX15 (K232962) ### 4. Device Description: The iT30&iT50 telemetry monitor can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of displaying, storing, and transferring of multiple physiological parameters connecting to Central Monitoring System. It will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate. ### 5. Indication for Use The iT30 and iT50 Telemetry Monitor are intended for use on Adult and Pediatric patients (over three years old) to monitor physiological parameters include ECG, SpO2, PR, NIBP (Applicable to iT50 only), and Skin TEMP. The physiological data can be analyzed, alarmed, stored locally on the monitor, and the Edan Central Monitoring System can configure, display and review the physiological parameters from the iT30 and iT50 Telemetry Monitor. The arrhythmia detection and ST Segment analysis of ECG are intended for adult patients. The SpO2 and PR are used under no motion conditions. The NIBP is used under no motion conditions for patients over 12 years old and excluding pregnant women. The iT30 and iT50 Telemetry Monitor must be used by professional trained clinical medical staff in healthcare facilities. ### 6. Predicate Device Comparison The table below compares the indication for use and key technological feature of the subject devices to the predicate device. | Characteristic | <Subject Device> (iT30, iT50 Telemetry Monitor) | <Predicate Device> (TM80 Telemetry Monitor) | | --- | --- | --- | | Manufacturer/K# | K251508 | K193391 | | **Intended Use/Indications for Use** | | | | Description | The iT30 and iT50 Telemetry Monitor are intended for use on Adult and Pediatric patients | The TM80 telemetry monitor is intended for use on Adult and Pediatric patients over three | 2 / 6 {6} Traditional 510(k) of Telemetry Monitor 510(k) Summary K251508 | | (over three years old) to monitor physiological parameters include ECG, SpO2, PR, NIBP (Applicable to iT50 only), and Skin physiological data can be analyzed, alarmed, stored locally on the monitor, and the Edan Central Monitoring System can display and review the physiological parameters from the iT30 and iT50 Telemetry Monitor. The arrhythmia detection and ST Segment analysis of ECG are intended for adult patients. The SpO2 and PR are used under no motion conditions. The NIBP is used under no motion conditions for patients over 12 years old and excluding pregnant women. The iT30 and iT50 Telemetry Monitor must be used by professional trained clinical medical staff in healthcare facilities. | years old to monitor ECG, SpO2, NIBP and Resp physiological data. The physiological data can be analyzed, alarmed, stored, reviewed locally on the display of the monitor, and the CentralStation can configure and display the physiological parameters from the TM80. The TM80 telemetry monitor is intended for use in professional healthcare facilities under the direct supervision of a licensed healthcare practitioner. | | --- | --- | --- | | **ECG module** | | | | Lead Mode | 3 Electrodes; 5 Electrodes; 6 Electrodes; **10 Electrodes** | 3 Electrodes; 5 Electrodes; 6 Electrodes | | **ST value** | | | | Measurement Range | -2.0 mV to +2.0 mV | -2.0 mV to +2.0 mV | | **Pace** | | | | Pulse Indicator | Amplitude: ±2 mV to ±700 mV Width: 0.1 ms to 2.0 ms Ascending time: 10 μs to 100 μs | Amplitude: ±2 mV to ±700 mV Width: 0.1 ms to 2 ms Rise time: 10 μs to 100 μs | | **HR** | | | | Measurement range | ADU: 15 bpm to 300 bpm PED: 15 bpm to 350 bpm | Adult: 15 bpm to 300 bpm Pediatric: 15 bpm to 350 bpm | | **QT Analysis** | | | | QT/QTc/ΔQTc measurement | QT Range: 200 ms ~ 800 ms QTc Range :200ms ~ 800 ms Δ QTc Range: -600 ms ~ 600 ms | QT Range: 200 ms ~ 800 ms QTc Range :200ms ~ 800 ms Δ QTc Range: -600 ms ~ 600 ms | | **Arrhythmia Analysis** | | | | Arrhythmia analysis | **ASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED** | Asystole, V-Fib/V-Tach, V-Tach, Vent Brady, Extreme Tachy, Extreme Brady, PVCs/min, Pauses/min, R on T, Run PVCs, | 3 / 6 {7} Traditional 510(k) of Telemetry Monitor 510(k) Summary K251508 | | **BEATS, IRR, VBRADY, PNC, PNP** | Couplet, Multiform PVC, PVC, Bigeminy, Trigeminy, Tachy, Brady, Pacer Not Pacing, Pacer Not Capture, Missed Beats, Nonsus V-Tach, Vent Rhythm, Pause, Irr Rhythm, A-Fib | | --- | --- | --- | | **NIBP module (EDAN)** | | | | Technique | Oscillometry | Oscillometry | | Measurement Mode | **Manual, Auto, STAT** | Manual, Automatic, Continuous, Sequence and ABPM | | Measurement Algorithm | iCUFS | The method principle is equivalent to the iCUFS algorithm | | Measurement Range | Adult SYS: 25 mmHg to 290 mmHg DIA: 10 mmHg to 250 mmHg MAP: 15 mmHg to 260 mmHg Pediatric SYS: 25 mmHg to 240 mmHg DIA: 10 mmHg to 200 mmHg MAP: 15 mmHg to 215 mmHg | Adult SYS: 25 mmHg to 290 mmHg DIA: 10 mmHg to 250 mmHg MAP: 15 mmHg to 260 mmHg Pediatric SYS: 25 mmHg to 240 mmHg DIA: 10 mmHg to 200 mmHg MAP: 15 mmHg to 215 mmHg | | **PR from NIBP** | | | | Measurement range | **40 bpm to 240 bpm** | 30 bpm to 300 bpm | | **SpO_{2} module** | | | | Measurement Range | 0% to 100% | Nellcor SpO_{2} 0% to 100% | | Accuracy | Adult/pediatric, non-motion conditions ±2% (70% to 100% SpO_{2}) Undefined (0% to 69% SpO_{2}) | Nellcor SpO_{2} 70-100%: ±2% 0-69%: Not specified | | The subject device supports SpO2 sensors SH1, SH4, SH5, SHD-A, SHD-P, 02.57.225230 and 02.57.225231. | | | | **PR from SpO_{2}** | | | | Measurement Range | 20 bpm to 300 bpm | 20 bpm to 300 bpm | | Accuracy | **± 2 bpm** | 20 bpm to 250 bpm: ±3 bpm 251 bpm to 300 bpm: Not specified | | **Skin Temperature (Skin TEMP) module** | | | | Number of channels | 1 | / | | Measurement Range | 0 °C to 50 °C(32 °F to 122 °F) | | | Measurement Site | Skin | | | The TEMP measurement is a newly-added function and is similar to the reference device K232962 | | | | **Wi-Fi** | | | | Radio Technology | **IEEE 802.11 a/b/g/n** | IEEE 802.11 a/b/g/n/ac | | Frequency Range | **2412 MHz-2472 MHz** | 2412 MHz to 2472 MHz | 4 / 6 {8} Traditional 510(k) of Telemetry Monitor 510(k) Summary K251508 | | **5180 MHz-5240 MHz 5745 MHz-5825 MHz** | 5180 MHz to 5825 MHz | | --- | --- | --- | | **Power Supply** | | | | Power Type | AA batteries Rechargeable Li-ion battery | AA batteries Rechargeable lithium-ion battery | | Charger | Charger stand | Central Charger | As seen in the comparison tables, the subject and predicate device have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness. ## **7. Performance Data:** ### **Non-clinical data:** #### **Electrical safety and electromagnetic compatibility (EMC)** iT30 and iT50 Telemetry Monitors were assessed for conformity with the relevant requirements of the following standards and found to comply: - IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION-Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2014 (Fourth Edition) Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests. #### **Performance testing-Bench** Edan has conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards. - IEC 60601-1-8:2020 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems - IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment - IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers - IEC 80601-2-49:2018 Medical electrical equipment –Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment - ISO 80601-2-56:2017+A1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement 5 / 6 {9} Traditional 510(k) of Telemetry Monitor 510(k) Summary K251508 - ISO 80601-2-61:2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment - ISO 14971:2019 Medical devices - Application of risk management to medical devices - IEEE ANSI USEMCSC C63.27 American National Standard for Evaluation of Wireless Coexistence. **Software Verification and Validation Testing** Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Food and Drug Administration Staff, "Content of Premarket Submissions for Device Software Functions". **Clinical data:** Not applicable. **Summary** The non-clinical performance testing showed that the subject devices are as safe and effective as the predicate device. **8. Conclusion** The bench testing data and software verification and validation demonstrate that iT30&iT50 Telemetry Monitor are substantially equivalent to the predicate device. 6 / 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...