VersaCoat Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
Device Story
VersaCoat Nerve Protector is an absorbable implant functioning as a gelatinous interface between injured nerves and surrounding tissues. It consists of a hollow, cylindrical hydrogel pellet (alginate and hyaluronic acid) and an aqueous citrate wetting solution. The clinician hydrates the pellet with the solution using a provided dual-syringe system to create a viscous, flowable, tissue-adherent gel. The gel is applied to the nerve site to provide a non-constricting, protective encasement. By acting as a physical barrier, it keeps damaged tissues separated during healing and provides a gliding surface, facilitating normal axonal survival and growth. It is intended for single use in a clinical setting.
Clinical Evidence
No human clinical data. Evidence includes bench performance testing (visual, dimensional, weight, gel integrity, handling, migration, coverage, cadaveric evaluation), biocompatibility testing (ISO 10993 series), and an animal study in rats. The animal study evaluated neurological, macroscopic, and histological outcomes, demonstrating the device safely provides an interface for injured nerves, permitting axonal survival and growth comparable to the predicate.
Technological Characteristics
Materials: Alginate and hyaluronic acid hydrogel; aqueous citrate wetting solution. Form: Hollow cylindrical pellet. Energy: None (passive implant). Sterilization: Radiation (SAL 10^-6). Biocompatibility: Tested per ISO 10993-3, 5, 6, 10, and 11. Endotoxin: < 20 EU/device.
Indications for Use
Indicated for management of peripheral nerve injuries without substantial loss of nerve tissue.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 22, 2025
Alafair Biosciences
Sarah Mayes, Ph.D.
Chief Scientific Officer
6101 W Courtyard Drive, Suite 1-225
Austin, Texas 78730
Re: K251505
Trade/Device Name: VersaCoat Nerve Protector (VTP-44G2); VersaCoat Nerve Protector (VTP-12G1)
Regulation Number: 21 CFR 882.5275
Regulation Name: Nerve Cuff
Regulatory Class: Class II
Product Code: JXI
Dated: May 15, 2025
Received: May 16, 2025
Dear Dr. Mayes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}
K251505 - Sarah Mayes, Ph.D.
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251505 - Sarah Mayes, Ph.D.
Page 3
Sincerely,
Adam D. Pierce -S
Digitally signed by
Adam D. Pierce -S
Date: 2025.07.22
15:01:56 -04'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
VersaCoat Nerve Protector
Page 8 of 35
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251505 | ? |
| Please provide the device trade name(s). | | ? |
| VersaCoat Nerve Protector (VTP-44G2);
VersaCoat Nerve Protector (VTP-12G1) | | |
| Please provide your Indications for Use below. | | ? |
| VersaCoat Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251505 Page 1 of 3
# 510(k) Summary
Prepared on: July 17, 2025
| Contact Details | |
| --- | --- |
| Submitted by: | Alafair Biosciences, Inc.
6101 W Courtyard Drive Ste. 1-225
Austin, TX 78730 |
| Contact: | Dr. Sarah Mayes
Alafair Biosciences Inc.
6101 W Courtyard Dr Ste 1-225
Austin TX 78730 United States
Phone: 512-739-9510
Email: info@alafairbiosciences.com |
| Device Name | |
| Product Name | VersaCoat Nerve Protector |
| Common Name | Cuff, Nerve |
| Classification number | 21 CFR 882.5275 |
| Product Code | JXI |
| Legally Marketed Predicate Device | |
| VersaWrap Nerve Protector (K232029) | |
| Device Description Summary | |
| VersaCoat Nerve Protector is an absorbable implant (device) designed to function as a gelatinous interface between an injured nerve and surrounding tissues, providing a non-constricting, protective encasement for injured nerves. VersaCoat Nerve Protector consists of a Hydrogel Pellet and Wetting Solution. The Pellet is a hollow, cylindrical hydrogel of alginate and hyaluronic acid. The Pellet is easy to handle and is designed for quick hydration with the Wetting Solution. The Pellet is supplied sterile, non-pyrogenic, for single use, in double peel pouches. The Wetting Solution hydrates the Pellet, rendering a viscous, flowable, tissue-adherent gel. The Wetting Solution is aqueous citrate and is provided in a dropper packaged in a double peel pouch. The Wetting Solution is sterile, non-pyrogenic, and is intended for single use only. A sterile container (dual syringe system) is provided and may be used to facilitate hydration of the Pellet with Wetting Solution. | |
| Indications for Use | |
| VersaCoat Nerve Protector is indicated for the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue. | |
| Principle of Operation | |
| The mechanism of action of VersaCoat Nerve Protector is to manage nerve injuries by providing a gliding surface and by keeping damaged tissues physically separated during healing | |
| Comparative Technology Characteristics | |
| A comparison of the subject device and the predicate device demonstrates equivalent technological characteristics. Equivalence is based upon intended use, indications for use, materials of construction, operating principle, fundamental scientific technology, and performance. When implanted, the subject and predicate device are identical. Minor technological characteristic differences do not raise new questions of safety and effectiveness. Minor differences include the geometry, or the form of hydrogel from a flat sheet to a hollow, cylindrical pellet. This change is to allow end users an easier ability to apply the device as a gel. This change is not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device because the difference | |
{5}
K251505 Page 2 of 3
does not affect the safety and effectiveness of the device when used as labeled, and does not describe a new disease, condition, or patient population. The risk-based assessment did not identify any new risks or significantly modified existing risks. Therefore, the subject device is substantially equivalent to the predicate device.
| Functional and Safety Testing |
| --- |
| To verify that device design met functional and performance requirements, representative samples of the device underwent nonclinical testing. These data demonstrate substantially equivalent safety and effectiveness to the predicate. |
# Non-Clinical Tests Summary & Conclusions
The following tests were performed to support substantial equivalence.
Performance Testing, including:
- Visual inspection
- Dimensional and weight measurements
- Gel integrity (including homogeneity, cohesivity, and viscosity)
- Handling
- Migration testing
- Coverage testing
- Cadaveric evaluation
Animal Testing, including:
- In the animal study conducted, rats underwent various peripheral nerve injuries that received either the subject or predicate device. There were no procedure related complications or device related premature deaths in this study, at all follow-up timepoints. The safety and efficacy of the subject device were evaluated by neurological assessment, and macroscopic and histological evaluation. These studies demonstrate that the subject device can safely provide an interface between injured nerve and surrounding tissues, permitting normal axonal survival and growth in a variety of clinically relevant nerve injury models, performing similar to the predicate device.
Biocompatibility Testing, including:
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Irritation - Intracutaneous Reactivity (ISO10993-10)
- Acute Systemic Toxicity (ISO10993-11)
- Pyrogenicity (ISO 10993-11)
- Genotoxicity (ISO 10993-3)
- Subchronic toxicity (13 weeks, ISO 10993-11 and ISO 10993-6)
- Muscle implantation toxicity/irritation (ISO 10993-6)
Sterilization:
Sterilization of the VersaCoat Nerve Protector was completed using radiation. Sterilization validation was conducted to verify the subject device meets SAL of 10-6.
Pyrogenicity (endotoxin):
Bacterial endotoxin testing was conducted to verify the subject device meets the requirement of $< 20$ EU/device.
{6}
K251505 Page 3 of 3
| Shelf life / packaging: Packaging integrity and stability were tested to verify the packaging meets product requirements and maintains its integrity during conditions normally encountered during sterilization, shipping, and storage over the proposed shelf-life. |
| --- |
| Conclusion |
| No new questions of safety or effectiveness were identified during device testing; therefore, the VersaCoat Nerve Protector device is considered substantially equivalent to the predicate device(s) in terms of safety and effectiveness. |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.