K251485 · Contego Medical, Inc. · QEW · Jun 13, 2025 · Cardiovascular
Device Facts
Record ID
K251485
Device Name
Excipio LV Prime Thrombectomy Device
Applicant
Contego Medical, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Jun 13, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Excipio LV Prime Thrombectomy Device is indicated for the non-surgical removal of emboli and thrombi from peripheral blood vessels.
Device Story
The Excipio LV Prime Thrombectomy Device is a mechanical thrombus displacement catheter used in conjunction with an aspiration catheter to remove emboli and thrombi from peripheral blood vessels. The device features a nitinol braided basket at the distal end, which is controlled via an activation wire connected to a proximal handle. The operator expands the basket to match the target vessel diameter (6-16 mm) under fluoroscopic guidance; the basket wires are inherently radiopaque. The device is used in a clinical setting by a physician. By mechanically displacing thrombus while applying vacuum aspiration (-28 in Hg), the device facilitates clot removal, potentially restoring blood flow and benefiting the patient by resolving vascular occlusions.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through visual inspection, corrosion testing, dimensional inspection, compatibility testing, kink resistance, clot removal testing, torsional strength, tensile testing, simulated use, and radial force. Biocompatibility was established per ISO 10993-1 by leveraging data from the predicate device.
Technological Characteristics
Materials: Nitinol, Polyimide, Pebax, PET, Stainless Steel, Loctite, Nylon, ABS, Dymax. Principle: Mechanical thrombus displacement via expandable nitinol basket. Energy: Manual actuation. Dimensions: 16 mm max expanded basket diameter, 150 cm working length, 0.035" guidewire compatibility. Connectivity: None. Sterilization: Ethylene Oxide (EO). Single-use.
Indications for Use
Indicated for non-surgical removal of emboli and thrombi from peripheral blood vessels in patients requiring mechanical thrombectomy.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 13, 2025
Contego Medical, Inc.
Nitin Mehta
Director, Regulatory Affairs
3801 Lake Boone Trail
Suite 100
Raleigh, North Carolina 27607
Re: K251485
Trade/Device Name: Excipio LV Prime Thrombectomy Device
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy Catheter
Regulatory Class: Class II
Product Code: QEW
Dated: May 13, 2025
Received: May 14, 2025
Dear Nitin Mehta:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251485 - Nitin Mehta
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251485 - Nitin Mehta
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
GREGORY W.
O'CONNELL -S
Digitally signed by GREGORY
W. O'CONNELL -S
Date: 2025.06.13 08:11:54
-04'00"
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251485 | |
| Device Name Excipio LV Prime Thrombectomy Device | |
| Indications for Use (Describe) The Excipio LV Prime Thrombectomy Device is indicated for the nonsurgical removal of emboli and thrombi from peripheral blood vessels. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Contego Medical
INTEGRATED EMBOLIC PROTECTION
K251485
Page 1 of 5
# 510(K) SUMMARY
510(k) Summary [as required per 21 CFR 807.92]
510(k) K251485
| Date Prepared | June 12, 2025 |
| --- | --- |
| Applicant: | Contego Medical, Inc.
3801 Lake Boone Trail, Suite 100
Raleigh, NC 27607 |
| FDA RegistrationNumber | 3011471056 |
| Contact Person: | Mr. Nitin Mehta
Contego Medical, Inc.
3801 Lake Boone Trail, Suite 100
Raleigh, NC 27607
Phone: + 1 805 813 7897
Email: nmehta@contegomedical.com |
| Proprietary Name: | Excipio LV Prime Thrombectomy Device |
| Common Name: | Embolectomy Aspiration Device |
| Device Classification: | Class II per 21 CFR §870.5150 |
| Classification Name: | Catheter, Embolectomy |
| Product Code: | QEW |
| Predicate Device: | Excipio LV Thrombectomy Device (K223897) |
| Reference Devices: | Excipio SV Thrombectomy Device (K221204) |
## Intended Use/Indications for Use:
The Excipio LV Prime Thrombectomy Device is indicated for the non-surgical removal of emboli and thrombi from peripheral blood vessels.
## Device Description:
The Excipio LV Prime Thrombectomy Device consists of a Thrombectomy Catheter to mechanically displace thrombus when used with an aspiration catheter. The device will be sold as a sterile, single use device.
## Thrombectomy Catheter:
The Thrombectomy Catheter is a mechanical thrombus displacement device with a nitinol braided component (basket) at the distal end that can be opened and closed via an activation wire that attaches to the distal end of the braid and attaches to a proximal handle. The operator can open the braided component to a diameter that best matches the target vessel (6-16 mm in diameter). The thickness of the basket wires enhances the radiopacity to facilitate visualization under fluoroscopy without the need for additional
{5}
Kontego Medical
INTEGRATED EMBOLIC PROTECTION
K251485
Page 2 of 5
radiopaque markers.
## Comparison with Predicate Devices:
A comparison of the Excipio LV Prime Thrombectomy Device (subject device) with the predicate devices shows that the technological characteristics of the subject device are the same, or similar, including the indication for use/intended use, principle of operation, design, sterilization method, and packaging.
There is no change of intended use or fundamental technology between the subject device and predicates. In addition, the subject device treats vessels within the range of the predicate devices. The Excipio LV Prime device performs thrombectomy using the same method as the predicate and reference devices.
A tabular comparison of the technological characteristics between the subject and predicate devices is provided below.
{6}
K251485
Page 3 of 5
Contego Medical
| Device Name | Excipio LV (Prime) Thrombectomy Device | Excipio LV Thrombectomy Device | Excipio SV Thrombectomy Device |
| --- | --- | --- | --- |
| Device | Subject Device | Predicate Device | Reference Device |
| Manufacturer | Contego Medical | Contego Medical | Contego Medical |
| 510(k) Number | TBD | K223897 | K221204 |
| Class/Product Code | II/QEW | II/QEW | II/QEW |
| Device Classification Description | Peripheral mechanical thrombectomy with aspiration | Peripheral mechanical thrombectomy with aspiration | Peripheral mechanical thrombectomy with aspiration |
| Indications for Use/Intended Use | Non-surgical removal of emboli and thrombi from peripheral blood vessels | Non-surgical removal of emboli and thrombi from peripheral blood vessels | Non-surgical removal of soft emboli and thrombi from peripheral blood vessels |
| Principle of operation - Thrombectomy mechanism | NiTi Basket | NiTi Basket | NiTi Basket |
| Vessel diameters to be treated | 6 – 16 mm | 10 – 25 mm | 4 – 8 mm |
| Handle – Mechanism of action | Ergonomic handle | Ergonomic handle | Ergonomic handle |
| Basket Material | Nitinol | Nitinol | Nitinol |
| Basket diameter (fully expanded) | 16 mm | 25 mm | 8 mm |
| Basket Braid working length | 60 mm ± 5mm | 82 mm minimum | 34 ± 2mm |
| Basket Length (closed) | 7 cm | 10 cm | 4.4 cm |
| Basket (max crossing profile) | 0.120" (3.05mm) | 0.160" (4.06mm) | 0.066" (1.68mm) |
| Materials of Construction | Nitinol, Polyimide, Pebax, PET, SS, Loctite, Nylon, ABS, Dymax (high level) | Nitinol, Polyimide, Pebax, PET, SS, Loctite, Nylon, ABS, Dymax (high level) | Nitinol, Polyimide, Pebax, PET, SS, Loctite, Nylon, ABS, Dymax (high level) |
| Thrombectomy device working length | 150 cm | 125 cm | 165 cm |
| Guidewire compatibility | 0.035" | 0.035" | 0.014" |
| Radiopaque markers | No radiopaque markers. Basket is radiopaque itself. | No radiopaque markers. Basket is radiopaque itself. | Proximal and distal to basket and on basket circumference |
| Recommended Aspiration Catheter diameter | 9.1 Fr inside diameter (minimum) | 12.6 Fr inside diameter minimum | 6.3 Fr inside diameter minimum |
| Aspiration Source | Vacuum (-28 in Hg) | Vacuum (-28 in Hg) | Vacuum (-28 in Hg) |
| Shelf Life | 2 years | 2 years | 2 years |
| Sterilization | EO, Single Use device | EO, Single Use device | EO, Single Use device |
{7}
K251485
Page 4 of 5
# Non-Clinical Testing/Performance Data:
In accordance with Contego's Risk Analysis process, non-clinical laboratory testing was performed on the Excipio LV Prime Thrombectomy Device to support the substantial equivalence of Excipio LV Prime device.
The following testing/assessments were performed:
- Visual Inspection
- Corrosion Testing
- Dimensional Inspection
- Compatibility Testing
- Kink Resistance
- Clot removal testing
- Torsional Strength
- Tensile Testing
- Simulated Use
- Radial Force
The in vitro bench tests demonstrated that the Excipio LV Prime Thrombectomy Device met all acceptance criteria, functioned as intended and has a performance profile that is similar to the predicate devices.
# Biocompatibility
Biocompatibility evaluation of the Excipio LV Prime Thrombectomy Device was performed. The following aspects of biocompatibility tests were leveraged from the predicate device:
- Cytotoxicity
- Sensitization
- Intracutaneous Irritation
- Acute System Toxicity
- Material Mediated Pyrogenicity
- Hemolysis
- Complement Activation
- Thrombogenicity
The results from the evaluation performed showed the Excipio LV Prime Thrombectomy Device to be biocompatible and meets requirements per ISO 10993-1.
{8}
Kontego Medical
INTEGRATED EMBOLIC PROTECTION
K251485
Page 5 of 5
# Sterilization and Packaging
Due to the limited nature of the device change, packaging testing was leveraged, and sterilization was assessed through sterilization adoption.
# Conclusion:
The Excipio LV Prime Thrombectomy Device has the same intended use and the same or similar technological characteristics such as design, sterilization method, and operating principles as Contego's own legally marketed predicate and reference devices. None of the differences between the subject device and predicates raise any new questions of safety or effectiveness.
The conclusions drawn from the nonclinical tests demonstrate that the Excipio LV Prime Thrombectomy Device is substantially equivalent to the predicate and reference devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.