OTS 25-L (100-5)

K251310 · Ots Medical , Ltd. · MBI · Jul 22, 2025 · Orthopedic

Device Facts

Record IDK251310
Device NameOTS 25-L (100-5)
ApplicantOts Medical , Ltd.
Product CodeMBI · Orthopedic
Decision DateJul 22, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The OTS 25-L All-Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, both may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the healing period, the suture anchor system stabilizes the damaged soft tissue.

Device Story

OTS 25-L is an all-suture anchor for soft-tissue fixation; expandable push-in design; provided preloaded on disposable inserter. Device components: braided polyester suture sleeve; stainless-steel inserter, drill bit, and drill guide; polypropylene drill guide obturator. Used in orthopedic surgery (arthroscopic or open) by surgeons to reattach ligaments, tendons, or joint capsules to bone. The anchor is placed into bone; suture strands are used to secure soft tissue. The subject device allows the extension component above the bone surface to be optionally packed with autologous bone marrow to augment soft tissue healing. Device stabilizes damaged tissue during the healing period, supported by postoperative immobilization. Benefits include secure tissue-to-bone fixation.

Clinical Evidence

No clinical testing performed. GLP animal study conducted on 20 animals (8 at 6 weeks, 12 at 16 weeks) comparing subject device (with autologous bone marrow) to predicate (K233429) in a Bankart surgical model. Assessments included blood collection, body weight, lameness, CT imaging, macroscopic analysis, microCT, histopathology, and biomechanical testing. Results demonstrated equivalence in tissue healing and attachment strength.

Technological Characteristics

All-suture anchor; braided polyester sleeve; stainless-steel inserter, drill bit, and drill guide; polypropylene drill guide obturator. Expandable push-in design. Sterile, single-use. Optional augmentation with autologous bone marrow.

Indications for Use

Indicated for patients requiring soft tissue-to-bone reattachment in orthopedic surgical procedures, including ligaments, tendons, or joint capsules, via arthroscopic or open surgery.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 22, 2025 OTS Medical Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103 Re: K251310 Trade/Device Name: OTS 25-L (100-5) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: July 2, 2025 Received: July 2, 2025 Dear Ms. Hogan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251310 - Janice Hogan Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251310 - Janice Hogan Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} OTS 25-L Page 8 of 41 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251310 | ? | | Please provide the device trade name(s). | | ? | | OTS 25-L (100-5) | | | | Please provide your Indications for Use below. | | ? | | The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The OTS 25-L All-Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, both may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the healing period, the suture anchor system stabilizes the damaged soft tissue. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} | 510(k) #: K251310 | 510(k) Summary | Prepared on: 2025-07-02 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | OTS Medical Ltd. | | | Applicant Address | 94 Yigal Alon St., Floor 31 Tel Aviv 6789139 Israel | | | Applicant Contact Telephone | +972-54-6171726 | | | Applicant Contact | Dr. Dekel Assaf | | | Applicant Contact Email | dekel.assaf@gmail.com | | | Correspondent Name | Hogan Lovells US LLP | | | Correspondent Address | 1735 Market Street, Floor 23 Philadelphia PA 19103 United States | | | Correspondent Contact Telephone | 1 267 675 4611 | | | Correspondent Contact | Ms. Janice Hogan | | | Correspondent Contact Email | janice.hogan@hoganlovells.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | OTS 25 -L (10 05 ) | | | Common Name | Smooth or threaded metallic bone fixation fastener | | | Classification Name | Fastener, Fixation, Nondegradable, Soft Tissue | | | Regulation Number | 888 .304 0 | | | Product Code(s) | MBI | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K241234 | OTS 25-L | MBI | | K151068 | Cayenne Medical's Surelock W Suture Anchor | MBI | | Device Description Summary | | 21 CFR 807.92(a)(4) | | OTS 25-L is an all-suture anchor. It is a soft-tissue fixation device with an expandable push-in design, provided preloaded on a disposable inserter. The device is constructed of a flat suture sleeve that is interlaced longitudinally along its central width by two #2 suture strands. The flat suture sleeve and the double loaded suture strands are folded back on themselves at the distal end of the disposable inserter. The device consists of the following components and accessories: non-absorbable, single use, suture anchor is made of a braided polyester sleeve; inserter; drill bit; and drill guide are made of stainless steel; and drill guide obturator is made of polypropylene. The disposable inserter has a stainless-steel shaft supporting a forked shaped tip, with a handle, and is provided sterile, for single use. | | | {5} | Intended Use/Indications for Use | 21 CFR 807.92(a)(5) | | --- | --- | | The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The OTS 25-L All-Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, both may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the healing period, the suture anchor system stabilizes the damaged soft tissue. | | | Indications for Use Comparison | 21 CFR 807.92(a)(5) | | The subject OTS 25-L has the same intended use and indications, technological characteristics, and principles of operation as the previously cleared predicates OTS Medical’s OTS 25-L. The only change to the device is a labeling change to allow use of the device when packed with autologous bone marrow. | | | Technological Comparison | 21 CFR 807.92(a)(6) | | The subject OTS 25-L and predicate cleared OTS 25-L device (K241234) are identical in all technological characteristics. The only difference between the subject OTS 25-L and the cleared OTS 25-L (K241234) predicate is a change to the labeling which allows the extension component of the anchor above the bone surface to be optionally filled with autologous bone marrow to allow the porous, braided polyester to act an interface for augmentation of the soft tissue. | | | Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) | | GLP animal testing was performed on 8 animals at 6 weeks (3 control, 5 investigational) and 12 animals at 16 weeks (4 control, 8 investigational). The control device consisted of the cleared OTS 25 anchor (K233429) and the investigational device was the subject OTS 25-L anchor packed with autologous bone marrow/bone fragments. Animals underwent a Bankart surgical procedure involving unilateral exposure of the glenohumeral joint, a bone lesion creation, and OTS test articles implantation. In vivo assessments performed included blood collection, body weight, lameness, CT imaging (16 weeks only). Ex-vivo assessments included macroscopic analysis, microCT analysis, histopathology (organs and shoulder samples), and biomechanical testing (6 weeks only). No clinical testing was performed to establish equivalence. The results of the test demonstrated the investigational device was at least equivalent to the cleared predicate in terms of tissue healing and strength of attachment. | |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...