Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII). Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes(MII).
Device Story
Ultra-Fast Vitri and Ultra-Fast Warm are media sets for cryopreservation of mature (MII) oocytes. Ultra-Fast Vitri uses Equilibration Solution (ES) and Vitrification Solution (VS) containing ethylene glycol, DMSO, trehalose, and hydroxypropyl cellulose to replace water in oocytes prior to liquid nitrogen storage. Ultra-Fast Warm uses Thawing Solution (TS) to remove cryoprotectants. Used in clinical ART laboratories by embryologists. The device does not contact the patient, only the oocytes. The ultra-fast protocol reduces steps and time compared to conventional methods, potentially improving workflow efficiency while maintaining oocyte viability, survival, and clinical pregnancy outcomes.
Clinical Evidence
Literature-based clinical study comparing conventional vs. ultra-fast protocols using 1,077 mature oocytes (n=519 conventional; n=558 ultra-fast). Results showed comparable outcomes: survival rate (100.0% vs 90.9%), clinical pregnancy rate (65.2% vs 54.3%), and live birth rate (56.5% vs 52.2%). Bench testing included MEA, pH, osmolality, endotoxin, and sterility (USP <71>).
Technological Characteristics
Media based on Medium 199 with HEPES buffer, ethylene glycol, DMSO, trehalose, hydroxypropyl cellulose, and Gentamicin. Aseptic filtration for media; radiation sterilization for vials. Storage at 2–8°C. Shelf life 12 months. No electronic or software components.
Indications for Use
Indicated for preparation, vitrification, and storage of mature (MII) oocytes in assisted reproductive technology (ART) procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 26, 2025
Kitazato Corporation
% Mei Li
Quality and Regulatory Affairs Consultant
Emergo Global Consulting, LLC
2500 Bee Cave Road Building 1, Suite 300
Austin, Texas 78746
Re: K251305
Trade/Device Name: Ultra-Fast Vitri; Ultra-Fast Warm
Regulation Number: 21 CFR 884.6180
Regulation Name: Reproductive Media And Supplements
Regulatory Class: Class II
Product Code: MQL
Dated: March 27, 2025
Received: July 29, 2025
Dear Mei Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251305 - Mei Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific
{2}
K251305 - Mei Li
Page 3
regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251305 | |
| Device Name Ultra-Fast Vitri; Ultra-Fast Warm | |
| Indications for Use (Describe) Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII). Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes(MII). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251305
Page 1 of 5
510(k) Summary – K251305
Ultra-Fast Vitri; Ultra-Fast Warm
1. Submission Sponsor
Kitazato Corporation
100-10 Yanagishima,
Fuji Shizuoka, 416-0932
Japan
Ms. Kyoko Izumi
Quality Assurance Manager
Phone Number: (81) 545 65 7122
2. Submission Correspondent
Emergo by UL
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327-9997
Email: LST.US.EmergoFDASubmissions@ul.com
Contact: Mei Li
Title: Quality and Regulatory Affairs Consultant
3. Date Prepared
August 26, 2025
4. Device Identification
Trade/Proprietary Name: Ultra-Fast Vitri; Ultra-Fast Warm
Common Name: Vitrification and warming media for vitrified oocytes
Regulation Name: Reproductive media and supplements
Regulation Number: 884.6180
Product Code: MQL, Reproductive Media and Supplements
Class: Class II
Classification Panel: Obstetrics/Gynecology
5. Legally Marketed Predicate Device(s)
Device name: Vitrification Kit and Thawing Kit
510(k) number: K171748
Page 1 of 5
{5}
K251305
Page 2 of 5
Manufacturer: Kitazato Corporation
The predicate device has not been subject to a design-related recall.
## 6. Device Description
The Ultra-Fast Vitri and Ultra-Fast Warm is composed of a set of three media to vitrify and warm oocytes for assisted reproductive technology (ART) procedures.
The Ultra-Fast Vitri includes two components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol and dimethyl sulfoxide. There are two vitrification procedures to choose from. During the vitrification process, oocytes are first exposed to ES then in VS within several minutes. Using this methodology, permeating cryoprotectants can replace water in the oocytes prior to vitrification and storage in liquid nitrogen. The Ultra-Fast Vitri comes prepackaged with 1.5 mL vial or 4 mL vial of ES, three 1.5 mL vials or three 4 mL vials of VS.
Ultra-Fast Warm is composed of one media used for warming and removing cryoprotectants from vitrified oocytes. It is composed of Thawing Solution (TS). The Ultra-Fast Warm comes pre-packaged with four 4.0 ml vials of TS.
All the media in the Ultra-Fast Vitri and Ultra-Fast Warm contain Gentamicin. The media undergoes aseptic filtration, while the vials are sterilized by radiation.
## 7. Indication for Use Statement
Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).
Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes (MII).
## 8. Substantial Equivalence Discussion
The following table compares the Ultra-Fast Vitri and Ultra-Fast Warm to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any different questions of safety or effectiveness as compared to the predicate device.
Comparison of Characteristics
| Feature | Subject Device: Ultra-Fast Vitri and Ultra-Fast Warm | Predicate: Vitrification Kit and Thawing Kit (K171748) |
| --- | --- | --- |
| Device Classification | Class II
Media, Reproductive | Class II
Media, Reproductive |
| Product Code | MQL | MQL, MQK |
| Regulation No. | 21 CFR 884.6180 | 21 CFR 884.6180 |
| Manufacturer | Kitazato Corporation | Kitazato Corporation |
| Intended Use | Reproductive media and supplements | Reproductive media and supplements |
017
Page 2 of 5
{6}
K251305
Page 3 of 5
| Feature | Subject Device: Ultra-Fast Vitri and Ultra-Fast Warm | Predicate: Vitrification Kit and Thawing Kit (K171748) |
| --- | --- | --- |
| Indications for Use | Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII). | The Vitrification Kit is indicated for use in the preparation, vitrification and storage of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos. |
| | Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes (MII). | The Thawing Kit is indicated for use in the preparation and thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos. |
| | Neither Ultra-Fast Vitri not Ultra-Fast Warm makes direct contact with the patient, only with the oocyte. | This product does not make direct contact with the patient, only with the gametes/embryos: Oocyte, PN through Blastocyst. |
| Embryo Stage | Ultra-Fast Vitri: Oocyte | Vitrification kit: Oocyte, PN through Blastocyst |
| | Ultra-Fast Warm: Oocyte | Thawing kit: Oocyte, PN through Blastocyst |
| Principle of Operation | Ultra-Fast Vitri: Provides users with the ability to cryopreserve supernumerary oocytes created during the in vitro fertilization procedure | Vitrification kit: Provides users with the ability to cryopreserve supernumerary oocytes or embryos created during the in vitro fertilization procedure |
| | Ultra-Fast Warm: Provides users with the ability to warm vitrified oocytes for use at a future point in time | Thawing kit: Provides users with the ability to warm vitrified oocytes or embryos for use at a future point in time |
| Vitrification Formulation | In a Medium 199 HEPES buffered Medium | In a Medium 199 HEPES buffered Medium |
| | Ethylene glycol | Ethylene glycol |
| | DMSO | DMSO |
| | Trehalose | Trehalose |
| | Hydroxypropyl Cellulose (HPC) | Hydroxypropyl Cellulose (HPC) |
| | Gentamicin | Gentamicin |
| Vitrification Protocol | Procedures provided for both Repro and Standard plates. Faster overall procedure with fewer steps. | Procedure provided for Repro plate. More steps and time to the procedure. |
| Vitrification Steps | 2 Procedures, each procedure with 2 Steps. The first step consists of 1 min equilibration in ES. The second step consists of 1 min vitrification in VS. | 2 Steps. The first step consists of a 15-minute equilibration involving BS and ES. The second step consists of 1-1.5 min vitrification involving VS. |
017
Page 3 of 5
{7}
K251305
Page 4 of 5
| Feature | Subject Device: Ultra-Fast Vitri and Ultra-Fast Warm | | Predicate: Vitrification Kit and Thawing Kit (K171748) |
| --- | --- | --- | --- |
| | The difference between the procedure using repro plates vs standard dishes relates to the volumes used (procedure on standard dish uses drops, this dish does not have wells) and the plate/dish in which the procedure is carried out. | | |
| Vitrification Media Component | Ultra-Fast Vitri devices can be composed on the following based on the reference number: | | Basic Solution (BS) 1.5mL x 1 Vial, Equilibration Solution (ES) 1.5mL x 1 Vial, and Vitrification Solution (VS) 1.5mL x 2 Vials |
| | VT601USUF (91268) | 1×1.5mL vial ES 3×1.5mL vial VS | |
| | VT601USUF-4 (91269) | 1×4mL vial ES 3×4mL vial VS | |
| | VT601USUF-ESx4 (91260) | 4×1.5mL vial ES | |
| | VT601USUF-VSx4 (91261) | 4×1.5mL vial VS | |
| | VT601USUF-ES4x4 (91266) | 4×4mL vial ES | |
| | VT601USUF-VS4x4 (91267) | 4×4mL vial VS | |
| Thawing Formulation | In a Medium 199 HEPES buffered Medium | | In a Medium 199 HEPES buffered Medium |
| | Hydroxypropyl Cellulose (HPC) (v/v) | | Hydroxypropyl Cellulose (HPC) (v/v) |
| | Gentamicin | | Gentamicin |
| | Trehalose | | Trehalose |
| Thawing Protocol | Faster overall procedure with fewer steps. | | More steps and time to the procedure. |
| Thawing Steps | 1 Step consisting of placing the oocyte 1 min in TS. | | Multiple Steps involving three different solutions (TS, DS and WS). |
| Thawing Media Component | TS solution 4.0mL x 4 Vials | | Thawing Solution (TS)4.0mL x 2 Vials Diluent Solution (DS) 4.0mL x 1 Vial Washing Solution (WS)4.0mLx1 Vial |
| Packaging | Solutions are packed in plastic vials. 4 vials are packed in a cardboard box. | | Solutions are packed in plastic vials. 4 vials are packed in a cardboard box. |
| Sterile | Solutions sterilized using aseptic processing techniques through filtration. Vial containers are sterilized via radiation | | Solutions sterilized using aseptic processing techniques through filtration Vial containers are sterilized via radiation |
| Endotoxin | Endotoxin by LAL methodology <0.25 EU/mL (LAL) | | Endotoxin by LAL methodology <0.25 EU/mL (LAL) |
| Mouse Embryo Assay | >80% one-cell 96 hours | | >80% one-cell 96 hours |
| Sterility Testing | Passes USP <71> | | Passes USP <71> |
| pH Test | 7.20 – 7.60 | | 7.20 – 7.60 |
| Biocompatibility | Passes | | Passes |
017
Page 4 of 5
{8}
K251305
Page 5 of 5
| Feature | Subject Device: Ultra-Fast Vitri and Ultra-Fast Warm | Predicate: Vitrification Kit and Thawing Kit (K171748) |
| --- | --- | --- |
| Storage | 2 – 8°C | 2 – 8°C |
| Shelf Life | 12 months | 12 months |
## 9. Non-Clinical Performance Data
Bench testing included the following:
- Color/Appearance
- pH Testing
- Endotoxin testing
- Osmolality Testing
- Sterility Testing
- Gentamicin Test
- Initial Media Dispensing Validation
To demonstrate safety and effectiveness of Ultra-Fast Vitri and Ultra-Fast Warm and to show substantial equivalence to the predicate device, the Mouse Embryo Assay (MEA) was conducted using the subject device with revised protocols to align with the ultra-fast procedures. The Ultra-Fast Vitri and Ultra-Fast Warm passed the testing in accordance with FDA guidance shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- Mouse Embryo Assay (MEA) – Passed
- Mouse Embryo Assay for Assisted Reproduction Technology Devices, issued January 5, 2021
## 10. Clinical Performance Data
Clinical performance data to support the safety and performance of the subject device was included in this submission. The referenced literature used Kitazato’s vitrification and warming solutions (K171748 and K160864) and compared the effectiveness of vitrification and warming using the conventional protocol to the ultra-fast protocols of the subject device (i.e., vitrification: 1 minute in equilibration solution and 1 minutes in vitrification solutions; thawing: 1 minute in thawing solution at 37°C). The study was conducted using 1,077 mature oocytes (n=519 for the conventional vitrification and warming protocols and n=558 for the ultra-fast protocols). The results from the study showed that oocyte survival rate after vitrification and thawing (100.0% vs. 90.9%), clinical pregnancy rate (65.2% vs. 54.3%), and live birth rate (56.5% vs. 52.2%) obtained with the ultra-fast vitrification and warming and with the conventional vitrification and warming protocols, respectively, were comparable.
## 11. Statement of Substantial Equivalence
The results of the performance testing described above demonstrate that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
017
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.