K251303 · Menovex Medical Technology (Shenzhen) Co., Ltd. · GEX · Nov 4, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251303
Device Name
Tamburo Thulium Fiber Laser System
Applicant
Menovex Medical Technology (Shenzhen) Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications:urology, lithotripsy gastroenterological surgery and gynecological surgery.
Device Story
Tamburo Thulium Fiber Laser System is a 1940 nm thulium fiber laser surgical instrument. System components include laser console, footswitch, power cord, interlock connector, ground wire, protective goggles, and optical delivery fibers. Operated by physicians in clinical settings (urology, gastroenterology, gynecology). Laser energy is transmitted via fiber to the treatment site for soft tissue interaction. Device features include air cooling, aiming beam, and LCD touch screen interface. Output is controlled laser energy for surgical procedures; benefits include precise tissue management (incision, ablation, coagulation).
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including optical, mechanical, and hardware performance verification, biocompatibility (ISO 10993 series), and electrical safety/EMC (IEC 60601 series).
Technological Characteristics
Thulium fiber laser; 1940 nm wavelength; 60W max power; 0.025-6J pulse energy; 0.2-50ms pulse duration; up to 2400 Hz frequency. Air cooling. Optical fiber delivery (150-800μm). Sterilization: EO. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, ISO 10993.
Indications for Use
Indicated for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue in urology, lithotripsy, gastroenterological surgery, and gynecological surgery, with or without an endoscope.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
K242987 — Thulium Fiber Laser (FiberLAZE+, FiberLAZE) · Allengers Global Healthcare Private Limited · Dec 17, 2024
K221306 — SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE SuperPulsed Laser Fibers, and Accessories · Gyrus Acmi, Inc. · Jan 27, 2023
K242293 — Medical Thulium Fiber Laser Systems (UroFiber 60Q) · Rhein Laser Technologies Co., Ltd. · Nov 19, 2024
K183647 — SOLTIVE Laser System(SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVELaser Fibers, and Accessories) · Olympus Surgical Technologies America · Aug 9, 2019
K232568 — IPG Medical Family of Thulium Fiber Lasers, Surgical Fibers, and Accessories · Ipg Medical Corporation · Mar 21, 2024
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 4, 2025
menovex Medical Technology (Shenzhen) Co., Ltd
% Kyra Kang
Official Correspondent
Landlink Healthcare Technology (Shanghai) Co., Ltd.
Room 1308, Baohua International Plaza, West Guangzhong Road
Jingan District
Shanghai,
China
Re: K251303
Trade/Device Name: Tamburo Thulium Fiber Laser System;
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: October 30, 2025
Received: October 30, 2025
Dear Kyra Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251303 - Kyra Kang
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K251303 - Kyra Kang
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2025.11.04
19:43:58 -05'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Tamburo Thulium Fiber Laser System
Page 11 of 55
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251303 | ? |
| Please provide the device trade name(s). | | ? |
| Tamburo Thulium Fiber Laser System | | |
| Please provide your Indications for Use below. | | ? |
| The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications:urology, lithotripsy gastroenterological surgery and gynecological surgery. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251303
# 510(k) Summary
## I. Submitter
menovex Medical Technology (Shenzhen) Co., Ltd
401, 501, Building 13, Biomedicine Innovation Industrial Park, No. 14, Jinhui Road, Jinsha Community, Kengzi Street, Pingshan District, Shenzhen 518118, China
Contact person: Ms. Xueying Qin
Tel.: +86-0755-89454602
E-mail: qxy100@menovex.com
Preparation date: April 27, 2025
Submission Correspondent
Ms. Kyra Kang
Landlink Healthcare Technology (Shanghai) Co., Ltd.
E-mail: kyra.kang@landlink-health.com
## II. Proposed Device
Device Trade Name: Tamburo Thulium Fiber Laser System
Common name: powered laser surgical instrument
Regulation Number: 21 CFR 878.4810
Regulatory Class: Class II
Product code: GEX
Review Panel: General & Plastic Surgery
## III. Predicate Devices
510(k) Number: K183647
Trade name: SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories).
Common name: powered laser surgical instrument
Classification: Class II
Product Code: GEX
Manufacturer: General & Plastic Surgery
Page 1 of 6
{5}
Page 2 of 6
## IV. Device description
Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications:
urology, lithotripsy gastroenterological surgery and gynecological surgery.
The Tamburo is a Thulium fiber laser device emitting at a wavelength of 1940 nm, which is used by physicians as a tool in surgical procedures.
This device consists of a thulium laser, footswitch, power cord, Interlock connector, ground wire and protective goggles.
The thulium laser includes thulium laser, cooling device, aiming beam device, and LCD touch screen.
Tamburo Thulium Laser Fiber:
Tamburo Thulium Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.
This device consists of a fiber, dust cap, ferrule, outer cap and handle.
The fiber includes core, cladding, coating and outer jacket.
Tamburo Thulium Laser Fibers are compatible for use with Tamburo Thulium Laser Systems.
## V. Indications for use
The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy gastroenterological surgery and gynecological surgery.
## VI. Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate devices are listed in below table
Table 5.1 General Comparison
| Characteristics | Proposed device | Predicate device (K183647) | Discussion |
| --- | --- | --- | --- |
| Trade name | Tamburo Thulium Fiber Laser System | SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium | / |
{6}
| | | SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories). | |
| --- | --- | --- | --- |
| Classification Name | Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology | Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology | Same |
| Product Code | GEX | GEX | Same |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Indications for use | The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy gastroenterological surgery and gynecological surgery. | The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. | Same |
| Laser Source | Thulium laser | Thulium laser | Same |
| Wavelength | 1938.9nm(1940 ±20 nm) | 1936.5nm(1940 ±20 nm) | Same |
Page 3 of 6
{7}
| Laser mode | Pulsed / CW | Pulsed | Different 1 |
| --- | --- | --- | --- |
| Divergence angle of terminal laser beam | 15.40° (16°±20%) | 12.96° | Different 2 |
| Aiming beam peak wavelength | 522.55nm (520 ±10nm) | 518.50nm (500-550nm) | Similar |
| Frequency | Up to 2400 Hz | Up to 2400 Hz | Same |
| Pulse Energy | 0.025-6J | 0.025-6J | Same |
| Pulse duration | 0.2-50ms | 0.2-50ms | Same |
| Max average power | 60W | 60W | Same |
| Delivery system | Optical fibers | Optical fibers | Same |
| Cooling system | Air cooling system | Air cooling system | Same |
| Fiber size | 150, 200, 272, 365, 550, 800μm | 150, 200, 365, 550, 940 μm | Similar |
| Fiber Sterilization | EO | EO | Same |
# Analysis 1 Laser mode
Both pulsed and CW modes have been rigorously tested and validated to ensure they meet all applicable regulatory requirements and deliver consistent, reliable performance. The difference in laser mode does not affect the safety or effectiveness.
{8}
Analysis 2 Divergence angle of terminal laser beam
By calculation, the distance of the Tamburo laser is 0.84 times that of Olympus. During clinical use, the fiber directly contacts the tissue. The distance difference can be ignored. The device has been rigorously tested and validated to ensure that it meets all applicable safety and performance standards, demonstrating that the divergence angle does not affect its ability to deliver effective and controlled laser energy for its intended use. Therefore, this difference does not pose any risk to the safety or effectiveness of the device.
## VII. Non-Clinical Testing
The device described in this summary, were tested and demonstrated to be in conformance with the following standards:
### Performance testing :
A comprehensive performance test was conducted on the Tamburo Thulium Fiber Laser System and Tamburo Thulium Laser Fiber to verify the performance of the product in optical, mechanical and Hardware aspects, along with comparative testing against equivalent products to demonstrate substantial equivalence.
### Biocompatibility testing
Biocompatibility of the device was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1" The following testing was conducted:
- ISO 10993-1 Fifth edition 2018-08
- ISO 10993-5 Third edition 2009-06-01
- ISO 10993-10 Fourth edition 2021-11
- ISO 10993-11 Third edition 2017-09
- ISO 10993-23 Software First edition 2021-01
Guidance for Industry and Food and Drug Administration Staff-Content of Premarket Submissions for Device Software Functions
### Electrical Safety and EMC
- IEC 60601-1 Edition 3.2 2020-08
- IEC 60601-1-2 Edition 4.1 2020-09
- IEC 60601-4-2 Edition 1.0 2-2024-03
- IEC 60601-1-6 Edition 3.2 2020-07
- IEC 60601-2-22 Edition 4.0 2019-11
Page 5 of 6
{9}
Page 6 of 6
## Shelf -life
The shelf life of the Tamburo Thulium Fiber Laser System is ensured by the stability evaluation performed in both real time and accelerated conditions, by the stability of laser performance. The performance is maintained after transport simulation and after 7 years aging.
The shelf life of the Tamburo Thulium Laser Fiber is set to 3 years. The sterile barrier performance (with the sterile barrier manufactured at validated worst case manufacturing conditions) is maintained after transport simulation and after 3 years aging (corresponding to the shelf-life of the device).
## VIII. Clinical Testing
No clinical study is included in this submission.
## IX. Conclusion
The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.