StealthStation S8 Spine Software

K251282 · Medtronic Navigation, Inc. · OLO · Oct 17, 2025 · Neurology

Device Facts

Record IDK251282
Device NameStealthStation S8 Spine Software
ApplicantMedtronic Navigation, Inc.
Product CodeOLO · Neurology
Decision DateOct 17, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K251282 · Oct 17, 2025StealthStation S8 Spine SoftwareMedtronic Navigation, Inc.Published clinical literatureA retrospective clinical evaluation of published literature was conducted to support the safety and effectiveness of the StealthStation Navigation System for use in skeletally mature pediatric (adolescent) patients.Retrospective literature review; Pediatric population; Safety and effectiveness

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective clinical evaluation of published literature; Retrospective literature reviewSkeletally mature pediatric (adolescent) patientsNot applicable for this studySafety and effectiveness

Indications for Use

The StealthStation™ System, with StealthStation™ Spine Software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures in adult and skeletally mature pediatric (adolescent) patients. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy. This can include the following spinal implant procedures in adult patients, such as: • Pedicle Screw Placement • Iliosacral Screw Placement • Interbody Device Placement This can include the following spinal implant procedures in skeletally mature pediatric (adolescent) patients: • Pedicle Screw Placement

Device Story

StealthStation S8 Spine Software v2.1 provides intraoperative surgical navigation; tracks surgical instruments relative to patient anatomy using optical (infra-red) technology. Inputs include pre-operative or intra-operative X-ray images; system registers images to patient anatomy via PointMerge, SurfaceMerge, FluoroMerge, or automatic registration (StealthAiR). Software displays instrument tip position and trajectory on axial, sagittal, coronal, oblique, and 3D renderings; allows surgeon to create/store surgical plans and simulate progression. Used in OR by surgeons to guide instrument placement; provides visual feedback on monitor to assist in pedicle screw, iliosacral screw, and interbody device placement. Benefits include precise anatomical localization and guidance along planned trajectories.

Clinical Evidence

Retrospective clinical evaluation of published literature conducted to assess safety and effectiveness in skeletally mature pediatric (adolescent) patients. Bench testing (positional/trajectory accuracy) previously established under K201189 remains valid; mean positional error ≤ 2.0 mm and mean trajectory error ≤ 2 degrees.

Technological Characteristics

Optical (infra-red) tracking; X-ray based imaging; C++ programming language. Interfaces with O-arm and various 3D C-Arms. Features include PointMerge, SurfaceMerge, FluoroMerge, and StealthAiR automatic registration. Accuracy: mean positional error ≤ 2.0 mm, mean trajectory error ≤ 2 degrees.

Indications for Use

Indicated for adult and skeletally mature pediatric (adolescent) patients undergoing open or percutaneous neurosurgical and orthopedic procedures where stereotactic surgery is appropriate and rigid anatomical structures (spine/pelvis) can be referenced to images. Includes pedicle screw, iliosacral screw, and interbody device placement.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 17, 2025 Medtronic Navigation, Inc. Victoria Baldock Senior Regulatory Affairs Specialist 200 Medtronic Drive Lafayette, Colorado 80026 Re: K251282 Trade/Device Name: StealthStation™ S8 Spine Software v2.1 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 19, 2025 Received: September 19, 2025 Dear Victoria Baldock: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251282 - Victoria Baldock Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251282 - Victoria Baldock Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Tejen D. Soni -S For Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251282 | | | Device Name StealthStation™ S8 Spine Software v2.1 | | | Indications for Use (Describe) The StealthStation™ System, with StealthStation™ Spine Software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures in adult and skeletally mature pediatric (adolescent) patients. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy. This can include the following spinal implant procedures in adult patients, such as: • Pedicle Screw Placement • Iliosacral Screw Placement • Interbody Device Placement This can include the following spinal implant procedures in skeletally mature pediatric (adolescent) patients: • Pedicle Screw Placement | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Page 1 of 5. 510(k) Summary - K251282 April 23, 2025 I. Company: Medtronic Navigation, Inc. 200 Medtronic Drive Lafayette, CO 80026 Telephone Number: (720) 890-3160 Contact: Victoria Baldock Senior Regulatory Affairs Specialist Telephone Number: (423) 863-5907 Email: tori.a.baldock@medtronic.com Kyle Hoefling (Alternate) Regulatory Affairs Director Telephone Number: (760) 207-2432 Email: kyle.d.hoefling@medtronic.com II. Proprietary Trade Name: StealthStation™ S8 Spine Software v2.1 III. Common Name: Orthopedic Stereotaxic Instrument IV. Classification Name: Stereotaxic Instrument (21 CFR 882.4560) V. Classification: Class II VI. Product Code: OLO (Stereotaxic Instrument) VII. Predicate Devices: The legally marketed predicate and reference devices are identified below: | Subject Device | Predicate Device | Reference Device | | --- | --- | --- | | StealthStation™ S8 Spine Software v2.1 | StealthStation™ S8 Spine Software v1.3 | Stryker Spine Guidance Software | | | K201189 | K220593 | | | S.E. May 29, 2020 | S.E. May 27, 2022 | VIII. Device Description: StealthStation S8 Spine Software helps guide surgeons during spine surgical procedures. The subject software works in conjunction with a navigation system, surgical instruments, a referencing system, and computer hardware. Navigation tracks the position of instruments in relation to the surgical anatomy and identifies this position on pre-operative or intraoperative images of the patient. The mouse, keyboard, touchscreen monitor, and footswitch of the {5} StealthStation platforms are used to move through the software workflow. Patient images are displayed by the software from a variety of perspectives (axial, sagittal, coronal, oblique) and 3-dimensional (3D) renderings. During navigation, the system identifies the tip location and trajectory of the tracked instrument on images and models the user has selected to display on the monitor. The surgeon may also create and store one or more surgical plan trajectories before and during surgery and simulate progression along these trajectories. During surgery, the software can display how the actual instrument tip position and trajectory relate to the plan, helping to guide the surgeon along the planned trajectory. IX. Indications for Use: The StealthStation™ System, with StealthStation™ Spine Software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures in adult and skeletally mature pediatric (adolescent) patients. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy. This can include the following spinal implant procedures in adult patients, such as: - Pedicle Screw Placement - Iliosacral Screw Placement - Interbody Device Placement This can include the following spinal implant procedures in skeletally mature pediatric (adolescent) patients: - Pedicle Screw Placement X. Comparison of the Technological Characteristics: A comparison of the technological characteristics of the subject, predicate, and reference devices is provided in the table below. Table 1. Technological Comparison of StealthStation S8 Spine Software v2.1 (Subject), StealthStation S8 Spine Software v1.3 (Predicate), and Stryker Spine Q Guidance Software (Reference) K251282, Page 2 of 5. {6} K251282, Page 3 of 5. | Technological Characteristic | StealthStation S8 Spine Software v2.1 (Subject) | StealthStation S8 Spine Software v1.3 (Predicate, K201189) | Stryker Spine Guidance Software (Reference, K220593) | | --- | --- | --- | --- | | Intended Use/ Indications for Use | The StealthStation™ System, with StealthStation™ Spine Software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures in **adult and skeletally mature pediatric (adolescent) patients**. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy. This can include the following spinal implant procedures in **adult patients**, such as: • Pedicle Screw Placement • Iliosacral Screw Placement • Interbody Device Placement **This can include the following spinal implant procedures in skeletally mature pediatric (adolescent) patients:** • **Pedicle Screw Placement** | The StealthStation™ System, with StealthStation™ Spine Software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy. This can include the following spinal implant procedures, such as: • Pedicle Screw Placement • Iliosacral Screw Placement • Interbody Device Placement | The Stryker Q Guidance System, when used with the Spine Guidance software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in **adult and pediatric (adolescent) patients**. The system is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the spine, pelvis or skull can be identified. The system assists in the positioning of instruments for procedures on the spine and pelvis, including: • Screw Placement in the spine or pelvis | | Product Code | OLO (Orthopedic Stereotaxic Instrument) | OLO (Orthopedic Stereotaxic Instrument) | OLO (Orthopedic Stereotaxic Instrument) | | StealthStation Operating Principle (Tracking Method) | Optical (infra-red) | Optical (infra-red) | Optical (infra-red) | | System Accuracy Requirement | Under representative worst-case Configuration, the StealthStation S8 Spine software v2.1, has demonstrated performance in 3D positional accuracy with a mean positional error of ≤ 2.0 mm and mean trajectory error of ≤ 2 degrees. Mean Accuracy Values (StealthAiR Spine): Positional Error – 1.01 mm Trajectory Error – 0.37 degrees | Under representative worst-case Configuration, the StealthStation S8 Spine software v1.3.0, has demonstrated performance in 3D positional accuracy with a mean positional error of ≤ 2.0 mm and mean trajectory error of ≤ 2 degrees. Mean Accuracy Values (StealthAiR Spine): Positional Error – 1.01 mm Trajectory Error – 0.37 degrees | The system has a mean accuracy of 2 mm for positional displacement and 2° for trajectory angle displacement. Accuracy values apply to tracking in the workspace. | {7} | Technological Characteristic | StealthStation S8 Spine Software v2.1 (Subject) | StealthStation S8 Spine Software v1.3 (Predicate, K201189) | Stryker Spine Guidance Software (Reference, K220593) | | --- | --- | --- | --- | | | Mean Accuracy Values (Overlapping Slices): Positional Error – 0.51 mm Trajectory Error – 0.41 degrees | Mean Accuracy Values (Overlapping Slices): Positional Error – 0.51 mm Trajectory Error – 0.41 degrees | | | Imaging Modalities | X-Ray Based Imaging | X-Ray Based Imaging | N/A | | Patient Registration Features | PointMerge Registration SurfaceMerge Registration FluoroMerge Registration Automatic 2D Image Registration Automatic 3D Image Registration StealthAiR Spine Automatic Registration | PointMerge Registration SurfaceMerge Registration FluoroMerge Registration Automatic 2D Image Registration Automatic 3D Image Registration StealthAiR Spine Automatic Registration | N/A | | Planning Features | Plan Entry and Target Selection 3D Model Building Deformity Planning | Plan Entry and Target Selection 3D Model Building Deformity Planning | N/A | | View (Display) Features | Look Sideways 3D Anatomic Orthogonal Trajectory 1 and 2 Trajectory Guidance Look Ahead Probe's Eye AP and Lateral Synthetic AP and Lateral Maximum Intensity Projection Video Input | Look Sideways 3D Anatomic Orthogonal Trajectory 1 and 2 Trajectory Guidance Look Ahead Probe's Eye AP and Lateral Synthetic AP and Lateral Maximum Intensity Projection Video Input | N/A | | Software Interface (GUI) | Black and gray style with procedure task overview in left menu option and next/back task flow at bottom of the screen. Software controls for images, planning, and instrument management are contained in a right side bar. | Black and gray style with procedure task overview in left menu option and next/back task flow at bottom of the screen. Software controls for images, planning, and instrument management are contained in a right side bar. | N/A | | Programming Language | C++ | C++ | N/A | | Medical Device Interfaces | O-arm Imaging System Ziehm Vision FD Vario 3D C-Arm ISO-C 3D C-Arm Ziehm Vision RFD 3D C-Arm Stealth-Midas MR8 | O-arm Imaging System Ziehm Vision FD Vario 3D C-Arm ISO-C 3D C-Arm Ziehm Vision RFD 3D C-Arm Stealth-Midas MR8 | N/A | # XI. Discussion of Nonclinical Testing: The Verification and Validation activities of the StealthStation S8 Spine Software have not changed since the previous clearance of the device under K201189. A requirements review, risk assessment and clinical summary were performed to demonstrate that the proposed labeling changes met the design requirements, is safe and effective for use in skeletally mature pediatric (adolescent) patients population and is substantially equivalent to the predicate StealthStation S8 Spine Software v1.3 (K201189). K251282, Page 4 of 5. {8} K251282, Page 5 of 5. ## XII. Discussion of Clinical Testing: A retrospective clinical evaluation of published literature was performed to assess the safety and effectiveness of the StealthStation Navigation System for use in the skeletally mature pediatric (adolescent) patient population. ## XIII. Conclusion: The StealthStation S8 Spine Software v2.1 with skeletally mature pediatric (adolescent) indications has been shown through comparison and clinical literature to be substantially equivalent to the identified predicate and reference devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...