Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (VH-6000); Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield (VH-6001); Vasoview Hemopro 3 Power Supply (VH-6010)
K251238 · Maquet Cardiovascular, LLC · GEI · May 30, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251238
Device Name
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (VH-6000); Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield (VH-6001); Vasoview Hemopro 3 Power Supply (VH-6010)
Applicant
Maquet Cardiovascular, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
May 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vasoview Hemopro 3 System is indicated for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or the radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall. The Vasoview Hemopro 3 Power Supply is used in conjunction with the Vasoview Hemopro 3 Endoscopic Vessel Harvesting System.
Device Story
The Vasoview Hemopro 3 is an endoscopic vessel harvesting system used in minimally invasive surgery. It consists of a harvesting cannula with four lumens housing an endoscope, a C-Ring for vessel cradling and lens washing, and a harvesting tool. The harvesting tool, powered by a dedicated DC power supply, utilizes heat and pressure via heating elements in its jaws to cut and cauterize tissue and blood vessels; it does not use radiofrequency energy. Operated by a surgeon under endoscopic visualization, the device allows for precise dissection and hemostasis. The system benefits patients by enabling minimally invasive access for vessel harvesting, reducing the trauma associated with traditional open harvesting techniques. The power supply provides the necessary energy for the tool's heating elements, which are controlled by an activation toggle on the handle.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by bench testing, including mechanical and thermal characterization of the modified C-Ring, durability testing, and extreme use cadaver testing.
Technological Characteristics
Electrosurgical cutting and coagulation accessory. Powered by direct current (DC) via a dedicated power supply. Features a four-lumen cannula, C-Ring, and harvesting tool with three heating elements (two cauterization, one cutting). Insulation protects adjacent tissue. Standards referenced: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, ISO 14971, ISO 11137-3, ASTM D4169-22.
Indications for Use
Indicated for patients undergoing minimally invasive endoscopic surgery for arterial bypass, including saphenous vein or radial artery harvesting for coronary or peripheral artery bypass grafting. Also indicated for blunt dissection of blood vessels, ducts, and tissues in extraperitoneal, subcutaneous extremity, and thoracic spaces, and thoracoscopic exposure/dissection of chest wall structures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (K101274)
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (K233879)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
Maquet Cardiovascular, LLC
Brad Sheals
Sr. Manager, Regulatory Affairs
45 Barbour Pond Road
Wayne, New Jersey 07470
RE: K251238
Trade/Device Name: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (VH-6000); Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield (VH-6001); Vasoview Hemopro 3 Power Supply (VH-6010)
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical cutting and coagulation device and accessories
Regulatory Class: Class II
Product Code: GEI
Dated: April 21, 2025
Received: April 22, 2025
Dear Brad Sheals:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251238 - Brad Sheals
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devicese/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251238 - Brad Sheals
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Jang -S
Digitally signed by
James H. Jang -S
Date: 2025.05.30
17:20:06 -04'00'
James Jang, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251238
Device Name
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield
Vasoview Hemopro 3 Power Supply
Indications for Use (Describe)
The Vasoview Hemopro 3 System is indicated for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or the radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall.
The Vasoview Hemopro 3 Power Supply is used in conjunction with the Vasoview Hemopro 3 Endoscopic Vessel Harvesting System.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
Page 1 of 4
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
K251238
GETINGE
*
# 510(k) Summary - K251238
Submitter: Maquet Cardiovascular, LLC
Address: 45 Barbour Pond Rd
Wayne, NJ 07470
United States
Contact Person: Mr. Brad Sheals, MS
901-286-0318
brad.sheals@getinge.com
Date Prepared: May 30, 2025
Device Trade Name: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield
Vasoview Hemopro 3 Power Supply
Common Name: Electrosurgical cutting and coagulation device accessories
Classification Name: Electrosurgical cutting and coagulation device accessories
Regulation Number: 878.4400
Product Code: GEI
Classification Panel: General and Plastic Surgery
Class: II
| Legally Marketed Predicate Device(s) | 510(k) Number | Product Code |
| --- | --- | --- |
| Vasoview Hemopro 2 Endoscopic Vessel Harvesting System | K101274 | GEI |
| Vasoview Hemopro 3 Endoscopic Vessel Harvesting System | K233879 | GEI |
{5}
Page 2 of 4
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
K251238
GETINGE
# Device Description:
The Vasoview Hemopro 3 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and Vasoview Hemopro 3 Harvesting Tool for cutting and cauterization of vessel branches. The C-Ring with the built-in distal lens washer is independently controlled by a C-Ring Slider on the handle of the Harvesting Cannula. The C-Ring cradles the vessel and delivers the saline that washes the distal tip of the Endoscope.
The Harvesting Tool can be inserted, removed, rotated, extended, and retracted from the main Harvesting Cannula through the Tool Adapter Port.
The Harvesting Tool is powered by direct current only; it does not utilize radiofrequency energy. The Harvesting Tool cuts and cauterizes through a process of heat and pressure. The Harvesting Tool has two curved Jaws. One of the Jaws contains the heating elements for branch cutting and cauterization and spot cautery. There are three heating elements: two cauterization elements and a cutting element between them. Both Jaws have insulation protecting the adjacent tissue. The concave side of the Jaws has a larger buffer of insulation and should, therefore, be positioned toward sensitive tissue during Jaw activation. An area near the tip of the convex side of the Jaw can be used for spot cautery to cauterize tissue contacted by that area of the Jaw.
The Activation Toggle is used to control the Jaws and to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization. This device is intended for specific use with the Vasoview Hemopro 3 Power Supply.
Model numbers applicable to the Vasoview Hemopro 3 System are:
- VH-6000 – Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
- VH-6001 – Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield
- VH-6010 – Vasoview Hemopro 3 Power Supply
# Intended Use/Indications for Use:
The Vasoview Hemopro 3 System is indicated for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or the radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall.
The Vasoview Hemopro 3 Power Supply is used in conjunction with the Vasoview Hemopro 3 Endoscopic Vessel Harvesting System.
# Indications for Use Comparison:
The indications for use for the subject Vasoview Hemopro 3 Endoscopic Vessel Harvesting System are the same as for the predicate devices: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System and the Vasoview Hemopro 2 Endoscopic Vessel Harvesting System.
{6}
Page 3 of 4
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
K251238
GETINGE
# Comparison of Technological Characteristics with the Predicate Devices:
The subject device (Vasoview Hemopro 3 Endoscopic Vessel Harvesting System) has the same intended use and similar technological characteristics as those of the predicate devices:
- The same intended use
- The same operating principles
- The same basic design
- The same sterilization method
- The same sterile packaging material
The subject device incorporates the following modifications:
- Material change to the C-Ring component of the Hemopro 3 Endoscopic Vessel Harvesting System Cannula
The design verification/validation testing demonstrates that the modification to the Hemopro 3 Endoscopic Vessel Harvesting System has similar technological characteristics compared to the predicate devices, or in the case where differences are noted, such differences do not raise new questions of safety or effectiveness. Such conclusions are also supported by the successful design verification and validation conducted and included in the submission.
# Summary of Nonclinical and/or Clinical Tests:
Testing was performed to demonstrate that the material change to the C-Ring component of the Hemopro 3 System's Cannula subassembly does not affect the device's ability to perform as intended. No new safety or effectiveness issues were raised during the testing.
# Bench Testing
The following design verification and validation tests were conducted:
- Mechanical and Thermal Characterization of C-Rings
- C-Ring Temperature Durability and Mechanical Strength
- C-Ring Extreme Use Cadaver Testing
- Technical Report - HP3 C-Ring Change Justifications
The following consensus standards were used or referenced as applicable in the bench testing performed:
- ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
- ANSI/AAMI HE75:2009/(R)2018
- IEC 60601-1-2 Edition 4.1 2020-09 Consolidated Version
- IEC 60601-1-6 Edition 3.2 2020-07 Consolidated Version
- ISO 15223-1 Fourth Edition 2021-07
- ISO 20417 First edition 2021-04 Corrected version 2021-12
- ISO 14971 Third Edition 2019-12
- ISO 17664-2 First edition 2021-02
{7}
Page 4 of 4
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System
K251238
GETINGE
- ISO 11137-3 Second edition 2017-06
- ASTM D4169-22
Clinical Testing
No clinical testing was conducted in support of this submission.
Conclusion:
Based on similarities in the indications for use and technological characteristics and on the results of nonclinical performance tests, the modified Vasoview Hemopro 3 Endoscopic Vessel Harvesting System is found substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.