The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs. The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Personal light-based hair reduction device; uses Intense Pulsed Light (IPL) from a Quartz glass Xenon lamp to reduce hair growth. Models feature optional semiconductor cooling (Peltier effect) via sapphire or other cooling panels. Input: user-selected intensity; skin tone sensor detects skin color to determine required intensity; skin contact sensor ensures safety. Operation: device emits light pulses only when fully applied to skin; supports manual and automatic continuous flashing modes. Used in home environment by end-user. Output: light pulses to hair follicles. Benefits: non-invasive hair reduction with minimal pain. Healthcare provider does not operate; device is OTC.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety/EMC (IEC 60601-1, -1-2, -1-6, -1-11, -2-57, -2-83), photobiological safety (IEC 62471), and software verification/validation.
Technological Characteristics
Materials: body-contacting components compliant with ISO 10993. Energy: Xenon Arc Flashlamp (510-1200nm). Cooling: semiconductor Peltier effect. Form factor: straight-panel or gun-shaped. Connectivity: standalone, powered by external adapter. Software: microprocessor-controlled with skin tone and contact sensors. Sterilization: not required.
Indications for Use
Indicated for OTC removal of unwanted hair (e.g., underarms, facial hair below chin line, legs) and permanent reduction in hair regrowth in individuals with Fitzpatrick Skin Phototypes I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Ice Cooling IPL Home Use Hair Removal Device (K241106)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 14, 2025
Shenzhen Mlay Intelligent Technology Co., Ltd.
Libin Liu
General Manager
201, 301, Building 28, Cuigang Industrial Zone 2, Huaide
Community, Fuyong Street, Baoan District, Shenzhen City,
Shenzhen, 518103
China
Re: K251176
Trade/Device Name: IPL Hair Removal Device (Model(s): T14B, T16B, T19B, T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: April 15, 2025
Received: April 16, 2025
Dear Libin Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}
K251176 - Libin Liu
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K251176 - Libin Liu
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.08.14
15:36:15 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251176 | |
| Device Name IPL Hair Removal Device (Model(s): T14B, T16B, T19B, T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C) | |
| Indications for Use (Describe) The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs. The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
K251176 - 510(k) Summary
"510(k) Summary" as required by 21 CFR Part 807.92.
1. Submitter Information
Shenzhen Mlay Intelligent Technology Co., Ltd.
201, 301, Building 28, Cuigang Industrial Zone 2, Huaide Community, Fuyong Street, Baoan District, Shenzhen City, Guangdong Province, China.
Post Code: 518103
LIU LIBIN
General Manager
Tel.: +86 18344147065
Email: wuxiaotian@mlays.com
Date Prepared: August 8, 2025
2. Device Information
Trade name: IPL Hair Removal Device
Model(s): T14B, T16B, T19B, T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C
Common Name: Light Based Over-The-Counter Hair Removal
Regulation Class: Class II
Product Code: OHT
Review Panel: General & Plastic Surgery
Regulation Number: 21CFR 878.4810
Regulation Description: Laser surgical instrument for use in general and plastic surgery
Submission Type: Traditional 510(k)
3. Predicate Device
| Manufacturer | Predicate Device | 510k Number | Decision Date |
| --- | --- | --- | --- |
| Shenzhen Mlay Intelligent Technology Co., Ltd. | Ice Cooling IPL Home Use Hair Removal Device (Model(s):T10B, T10C, T15A, T17A, T18A, T14A, T16A, T19A) | K241106 | April 22, 2024 |
| Shenzhen Fansizhe Science and Technology Co., Ltd. | Intense Pulsed Light (IPL) System, model: T023K, T023A, T023B, T023C, T023D, T023E, T021K, T021A, T001A, T001B, T001M, T001N, T011C, T016K | K223928 | March 28, 2023 |
| Shenzhen Ulike Smart Electronics Co., Ltd. | Ice Cooling IPL Hair Removal Device | K242039 | October 25, 2024 |
| Shenzhen Ulike | IPL Hair Removal Device | K230122 | January 17, |
{6}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
| Smart Electronics Co., Ltd. | | | 2023 |
| --- | --- | --- | --- |
# 4. Device Description
The IPL Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user.
The device provides hair reduction using Intense Pulsed Light technology(suitable for model T21A, T21B and T22A, T22B). The device provides hair reduction using Intense Pulsed Light technology and cooling technology (suitable for model T14B, T16B, T19B, T15B, T17C, T18B, T21C, T21D, T25B, T25C).
The Intense Pulsed Light technology works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Quartz glass Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the skin of the treatment area, the device will not emit light pulses; If the device is properly and fully applied to the skin of the treatment area, the device can emit light pulses in as quickly as 0.5 seconds. In automatic mode, it supports continuous flashing and automatic light emission.
In auto-recognition skin color mode, the skin tone sensor can detect and identify the color of skin, and determine the required intensity based on the recognized skin color. Make sure the skin tone sensor is in full contact with the skin. If a valid skin color is detected, the corresponding energy level is displayed. If it is not in full contact with the skin, the energy level is 0 and no light pulses are emitted.
The cooling technology based on the temperature difference electrical phenomenon through the semiconductor cooling chip inside the IPL main device and uses the principle of the Peltier effect to achieve the purpose of cooling function. The cooling panel is located around the light-emitting window (suitable for model T14B, T16B, T19B) and does not affect the irradiated area (spot size) of the light outlet; The cooling panel is constructed with sapphire, (suitable for model T15B, T17C, T18B, T21C, T21D, T25B, T25CB) and does not affect the irradiated area (spot size) of the light outlet.
The device is available in two designs: straight-panel and gun-shaped, both featuring a compact and lightweight form factor. Moreover, The enterprise has reserved an ample quantity of lamp heads to ensure maintenance accessibility and end-user convenience.
# 5. Indications for Use
The IPL Hair Removal Device is an over-the-counter device indicated for the removal
{7}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs.
The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
## 6. Comparison of Technological Characteristics With the Predicate Device
The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate device.
{8}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
The IPL Hair Removal Device is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance:
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| K Number | Pending | K241106 | K223928 | K242039 | K230122 | / |
| Trade name | IPL Hair Removal Device, Model(s): T14B, T16B, T19B, T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C | Ice Cooling IPL Home Use Hair Removal; Device Model(s): T10B, T10C, T15A, T17A, T18A, T14A, T16A, T19A | Intense Pulsed Light (IPL) System, model: T023K, T023A, T023B, T023C, T023D, T023E, T021K, T021A, T001A, T001B, T001M, T001N, T011C, T016K | Ice Cooling IPL Hair Removal Device | IPL Hair Removal Device | / |
| Manufacturer | Shenzhen Mlay Intelligent Technology Co., Ltd. | Shenzhen Mlay Intelligent Technology Co., Ltd. | Shenzhen Fansizhe Science and Technology Co., Ltd | Shenzhen Ulike Smart Electronics Co., Ltd. | Shenzhen Ulike Smart Electronics Co., Ltd. | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHT | OHT | OHT | OHT | OHT | Same |
| Device classification | Class II | Class II | Class II | Class II | Class II | Same |
| Location for use | OTC | OTC | OTC | OTC | OTC | Same |
{9}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| Applicable skin | Fitzpatrick Skin Phototypes I-V | Fitzpatrick Skin Phototypes I-V | Unknown | Fitzpatrick Skin Types I-V | Fitzpatrick Skin Types I-V | Same |
| Treatment area | Small areas such as underarm and facial hair below the chin line and large areas such as legs. | Large areas(e.g. legs) and small areas(e.g.underarm, facial hair below the chin line) | Unknown | large areas (e.g. arms, legs, chest) and small areas (e.g. lip) | large areas (e.g. arms, legs, chest) and small areas (e.g. lip) | Same |
| Power source | An external power supply | An external power supply | an external power supply | Supplied by external adapter | Supplied by external adapter | Same |
| Power supply | T14B,T16B,T19B: AC 100-240V, 50/60Hz, 2.5A MAX T15B, T17C, T18B, T21A, T21B, T21C, T21D, T22A, T22B, T25B, T25C: AC 100-240V, 50/60Hz, 1.5A | T10B, T10C, T15A, T17A, T18A: 48W(12V 4A) T14A, T16A, T19A: input AC100-240V; 50/60Hz; 2.5A | Unknown | 100-240V~, 50/60Hz | 100-240V~, 50/60Hz | SE NOTE 1 |
| Dimension | T14B:123.8*49.4*176 .1mm T16B:122.0*46.8*175 .0mm T19B:137.0*59.0*197 | T10B: 186.8*59.7*42.0 mm T10C:186.8*59.7*42. 0 mm T15A: | 116*217.8*42mm for T023K, T023A, T023B, T023C, T023D and T023E, 73.2*81.1*202.2mm | 206.73mm*68.68mm* 54.29mm | 60mm*38mm*170mm | SE NOTE 1 |
5 / 13
{10}
Shenzhen Mlay Intelligent Technology Co., Ltd.
040_510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| | .0mm
T15B:163.0*66.0*43.0mm
T17C:169.5*80.1*52.3mm
T18B:171.0*65.2*45.5mm
T21A:153.6*82.5*50.3mm
T21B:153.6*82.5*50.3mm
T21C:153.6*82.5*50.3mm
T21D:153.6*82.5*50.3mm
T22A:135.8*84.6*49.5mm
T22B:135.8*84.6*49.5mm
T25B:164.0*63.5*41.0mm
T25C:164.0*63.5*41.0mm | 163.0*66.0*43.0 mm
T17A:
169.5*80.1*52.3 mm
T18A:
171.0*65.2*45.5 mm
T14A:
123.8*49.4*176.1 mm
T16A:
122.0*46.8*175.0 mm
T19A:
137.0*59.0*197.0 mm | for T021K and T021A
182*78*151mm for
T001A, T001B,
T001M
and T001N,
211*138*60mm for
T011C,
90*44*225mm for
T016K | | | |
| Sterilization | Not required | Not required | Not required | Not required | Not required | Same |
6 / 13
{11}
Shenzhen Mlay Intelligent Technology Co., Ltd.
040_510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| Wavelength range | 510~1200nm | 510~1200nm | 510~1200nm | 550-1200mm | 560-1200nm | SE NOTE 2 |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Same |
| Light Source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same |
| Pulsing control | Finger switch | Finger switch | Finger switch | Finger switch | Finger switch | Same |
| Delivery device | Direct illumination tissue | Direct illumination tissue | Direct illumination tissue | Direct illumination to tissue | Direct illumination to tissue | Same |
| Max. Fluence (J/cm²) | 5.90J/cm² | 5.23J/cm² | 5.75J/cm² | 6.41J/c m² | 7.2J/cm² | SE NOTE 2 |
| Spot size | 3.1cm², 3.5cm², 3.9cm² | 3.5cm², 3.9cm² | 3.0cm², 3.3cm², 3.6cm², 4.0cm² | 3.9cm² | 3.3cm² | SE NOTE 2 |
| Pulse duration | 0.4~5.5ms | 0.5~8.0ms | 4 ~ 12ms | 0.93ms~3.50ms Single pulse Double pulse Triple pulse | 1.15-6.2ms | SE NOTE 2 |
| Software / Firmware Microprocessor Control? | Yes | Yes | Yes | Yes | Yes | Same |
| Availability of skin sensors (that used to identify the | Presence | Presence | Presence | Presence | Presence | Same |
7 / 13
{12}
Shenzhen Mlay Intelligent Technology Co., Ltd.
040_510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| presence or absence of skin areas) | | | | | | |
| Cooling function | Presence | Presence | Presence | Presence | Presence | Same |
| Skin pigmentation sensor | Detect appropriate skin tones | No | Detect appropriate skin tones | Detect appropriate skin tones | No | SE NOTE 3 |
| Indication for use / intended use | The IPL Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs. The device is also indicated for the permanent reduction in hair regrowth, | The Ice Cooling IPL Home Use Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs. The device is also indicated for the permanent reduction in hair regrowth, | The Intense Pulsed Light (IPL) System is an over - the - counter device intended for the removal of unwanted body hair. | Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the | IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the | Same |
8 / 13
{13}
Shenzhen Mlay Intelligent Technology Co., Ltd.
040_510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| | defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | | number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. | completion of a treatment regime. | |
| Electrical safety statement & EMC safety statement | IEC60601-1:2005+AMD1:2012+AMD2:20 20 CVS
IEC60601-1-6:2010+AMD1:2013+AMD2: 2020 CSV
IEC60601-1-11:2015+AMD1:2020 CVS
IEC60601-1-2:2014/AMD1:2020
IEC TS 60601-4-2:2024 | IEC60601-1:2005+AMD1:2012+AMD2:20 20 CVS
IEC60601-1-6:2010+AMD1:2013+AMD2: 2020 CSV
IEC60601-1-11:2015+AMD1:2020 CVS
IEC60601-1-2:2014/AMD1:2020 | ANSI AAMI ES60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57
IEC 60601-2-83 | ANSI AAMI ES60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57
IEC 60601-2-83 | SE NOTE 4 |
| Photobiological statement | IEC60601-2-57:2011
IEC60601-2-83:2019+ | IEC60601-2-57:2011
IEC60601-2-83:2019+ | IEC 62471 | IEC 62471 | IEC 62471 | SE NOTE 4 |
9 / 13
{14}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
| Comparison Elements | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| | AMD1:2022 CSV
IEC62471:2006 &
IEC62471:2008 | AMD1:2022 CSV
IEC62471:2006 | | | | |
| Materials statement | ISO10993-1:2018
ISO10993-5:2009
ISO10993-10:2021
ISO10993-23:2021 | ISO10993-1:2018
ISO10993-5:2009
ISO10993-10:2021
ISO10993-23:2021 | ISO 10993-5
ISO 10993-10 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10 | SE
NOTE 4 |
## Comparison in details:
NOTE 1: Although the "power supply", "Product appearance", "Dimensions", "Weight", "Environment for operation" and "Environment for storage" of subject device is a little different from the predicate devices, it will not affect the main function and the intended use of the subject device as they all also comply with IEC60601-1 requirements. Besides, the subtle change of the physical characteristics will not affect the critical functions or normal use.
NOTE 2: Although the "Wavelength range", "Spot size", "Energy density", and "pulse duration" of subject device is a little different from the predicate devices; they are very similar, and it will not affect the main function and the intended use of the subject device as they all also comply with IEC60601-1, IEC60601-1-6, IEC60601-11 requirements, so these parameters' differences will not raise any safety or effectiveness issue. Also, the "Wavelength range", "Spot size", "Energy density", and "pulse duration" of the subject device is different from the predicate, but its value is within the range of other predicate devices. so this parameter's difference will not raise any safety or effectiveness issue.
NOTE 3: Although there is a difference of the skin pigmentation sensor between the subject device and predicate device 1, but the predicate device 2 and 3 also have the skin pigmentation sensor to detect appropriate skin tones, so such difference would not raise safety or effectiveness issue.
10 / 13
{15}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
NOTE 4: According to the FDA regulation classification and device characteristics, the subject device belongs to the OHT product code, the regulation recommends that such products comply with the IEC62471 standard requirements. In addition, the relevant tests shall be selected to prove the safety and effectiveness of the subject device according to its design characteristics and features. The battery of testing was performed to, and passed, including:
1) The subject device is powered by an external device and belongs to an active medical device, so the device needs to be tested for 'electrical safety and electromagnetic compatibility' in accordance with the requirements of the standard IEC60601-1 and IEC60601-1-2;
2) And the subject device belongs to a medical electrical equipment, so the device needs to be tested for the 'General requirements for basic safety and essential performance - Collateral standard: Usability' in accordance with the requirements of the standard IEC60601-1-6;
3) And the subject device belongs to a household medical device, so the device needs to be tested for 'the requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment' in accordance with the requirements of the standard IEC60601-1-11;
4) And the subject device is a household light source (non-laser) equipment, so the device needs to be tested for 'the particular requirements for the basic safety and essential performance of such devices' in accordance with the requirements of the standard IEC60601-2-57 and/or IEC60601-2-83.
11 / 13
{16}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
## 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process", as recognized by FDA. The battery of testing was performed to, and passed, including:
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation
## 2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005 MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2:2014/AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC60601-1-6:2010/AMD1:2013/AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-11 Medical Electrical Equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral Standard: Requirements for medical electrical equipment and medical electrical equipment and medical electrical systems used in the home healthcare environment (IEC 60601-1-11:2015 MOD) [Including Amendment1 (2021)]
- IEC 60601-2-57:2011 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic diagnostic monitoring and cosmetic/aesthetic use
- IEC 60601-2-83:2019/AMD1:2022 Medical Electrical Equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
## 3) Eye Safety
- IEC 62471:2006 Photobiological safety of lamps and lamp systems
12 / 13
{17}
Shenzhen Mlay Intelligent Technology Co., Ltd. 040_510(k) Summary
## 4) Software Verification and Validation
In this 510(k) submission, the software documentation are the basic documentation. System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## Summary
Based on the above performance as documented in this application, the subject device IPL Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device.
## 8. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL Hair Removal Device is to be concluded substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.