K251131 · Spine Wave, Inc. · NKB · Jun 6, 2025 · Orthopedic
Device Facts
Record ID
K251131
Device Name
Annex® 2 Adjacent Level System
Applicant
Spine Wave, Inc.
Product Code
NKB · Orthopedic
Decision Date
Jun 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Intended Use
When used with pedicle screw fixation systems, the Annex® 2 Adjacent Level System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). The Annex® 2 Adjacent Level System can be linked to 4.75 mm, 5.5 mm, and/or 6.0 mm diameter rods of the following systems: CapSure®, Sniper® or Salvo® Spine Systems.
Device Story
The Annex® 2 Adjacent Level System is a spinal fixation implant set used to extend existing pedicle screw constructs. It consists of non-sterile, single-use locking screws, adjacent level devices, and connectors. The device is used by surgeons in the operating room to link to 4.75 mm, 5.5 mm, or 6.0 mm diameter rods from the CapSure®, Sniper®, or Salvo® Spine Systems. By connecting to existing hardware, it provides rigid immobilization and stabilization of spinal segments, facilitating fusion and treating spinal instabilities or deformities. The system is provided in various sizes and shapes to accommodate patient anatomy and existing hardware positioning.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including static axial compression bending, static torsion, and dynamic axial compression bending testing conducted in accordance with ASTM F1717.
Technological Characteristics
Materials: Titanium (ASTM F67), titanium alloy (Ti-6Al-4V ELI, ASTM F136), and cobalt-chromium (CoCr, ASTM F1537). Components include locking screws, adjacent level devices, and connectors. Mechanical testing performed per ASTM F1717. Non-sterile, single-use.
Indications for Use
Indicated for skeletally mature patients requiring immobilization and stabilization of thoracic, lumbar, and sacral spinal segments as an adjunct to fusion for degenerative disc disease, spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, or pseudarthrosis.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K132403 — ANNEXTM ADJACENT LEVEL SYSTEM · Spine Wave, Inc. · Jan 10, 2014
K170654 — USS System · Synthes USA Products, LLC · Mar 29, 2017
K182974 — NuVasive Reline System · Nu Vasive, Incorporated · Feb 13, 2019
K231443 — Mini-OSTEO Pedicle Fixation System · Osteomed Indústria E Comércio DE Implantes Ltda · Nov 21, 2023
K170126 — NuVasive® Reline® 4.5-5.0 System · Nu Vasive, Incorporated · Mar 9, 2017
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 6, 2025
Spine Wave, Inc.
Ronald Smith
Executive Vice President - Quality, Regulatory and Clinical Affairs
Three Enterprise Drive
Suite 210
Shelton, Connecticut 06484
Re: K251131
Trade/Device Name: Annex® 2 Adjacent Level System
Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral Pedicle Screw System
Regulatory Class: Class II
Product Code: NKB
Dated: April 11, 2025
Received: April 11, 2025
Dear Ronald Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251131 - Ronald Smith
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251131 - Ronald Smith
Page 3
Sincerely,
EILEEN
CADEL-S
for
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251131 | |
| Device Name Annex® 2 Adjacent Level System | |
| Indications for Use (Describe) When used with pedicle screw fixation systems, the Annex® 2 Adjacent Level System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). | |
| The Annex® 2 Adjacent Level System can be linked to 4.75 mm, 5.5 mm, and/or 6.0 mm diameter rods of the following systems: CapSure®, Sniper® or Salvo® Spine Systems. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251131 - Page 1 of 2
| 510(k) #: K251131 | 510(k) Summary | Prepared on: 2025-04-22 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Spine Wave, Inc. | |
| Applicant Address | Three Enterprise Drive Suite 210 Shelton CT 06484 United States | |
| Applicant Contact Telephone | 203-712-1846 | |
| Applicant Contact | Mr. Ronald Smith | |
| Applicant Contact Email | rsmith@spinewave.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Annex® 2 Adjacent Level System | |
| Common Name | Thoracolumbosacral pedicle screw system | |
| Classification Name | Thoracolumbosacral Pedicle Screw System | |
| Regulation Number | 888.3070 | |
| Product Code(s) | NKB | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K132403 | Annex® Adjacent Level System | MNH |
| K240685 | Salvo® Spine System | NKB |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Annex® 2 Adjacent Level System consists of a selection of non-sterile, single-use locking screws, adjacent level devices, and connectors for attachment to an existing pedicle spinal fixation construct to extend a rigid spinal construct. The Annex® 2 Adjacent Level System components are provided in a variety of sizes and shapes to accommodate variations in anatomy and spacing/positioning of existing screw and rod hardware. The implant components of this system are manufactured from titanium (ASTM F67)), titanium alloy (Ti-6Al-4V ELI (ASTM F136)), and cobalt-chromium (CoCr (ASTM F1537)). | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| When used with pedicle screw fixation systems, the Annex® 2 Adjacent Level System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). The Annex® 2 Adjacent Level System can be linked to 4.75 mm, 5.5 mm, and/or 6.0 mm diameter rods of the following systems: CapSure®, Sniper® or Salvo® Spine Systems. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The Annex® 2 Adjacent Level System is substantially equivalent to the predicate devices. The Annex® 2 Adjacent Level System has the | | |
{5}
K251131 - Page 2 of 2
| same intended use as the primary predicate. Specifically, the devices are used in thoracolumbosacral revision surgery to extend an existing pedicle spinal fixation instrumentation contruct. Therefore, the subject Annex® 2 Adjacent Level System is substantially equivalent to its predicate in intended use, and no new issues of safety and effectiveness are raised. |
| --- |
| Technological Comparison 21 CFR 807.92(a)(6) |
| The Annex® 2 Adjacent Level System has technological characteristics equivalent to those of the predicate devices, including design, and material composition. |
| Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) |
| Non-clinical testing was performed on the Annex® 2 Adjacent Level System to support substantial equivalence to legally marketed predicate devices. Static axial compression bending, static torsion testing and dynamic axial compression bending testing per ASTM F1717 was performed. |
| Clinical testing is not applicable. |
| Evaluation of the subject device performance data as compared to legally marketed predicate devices has found that the Annex® 2 Adjacent Level System is substantially equivalent. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.