K251125 · Tonica Elektronik A/S · OBP · Aug 11, 2025 · Neurology
Device Facts
Record ID
K251125
Device Name
MagVenture TMS Therapy System
Applicant
Tonica Elektronik A/S
Product Code
OBP · Neurology
Decision Date
Aug 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K251125 · Aug 11, 2025
MagVenture TMS Therapy System
Tonica Elektronik A/S
Systematic clinical literature review; Retrospective case series; Open-label clinical studies
The sponsor conducted a systematic literature review of clinical data from the subject device and substantially equivalent devices to support the safety and effectiveness of the MagVenture TMS Therapy System for the treatment of MDD in adolescents (15-21 years).
Major Depressive Disorder; Adolescent; TMS; Literature Review; Retrospective Case Series
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic Literature Review (various studies); Systematic review of prospective, retrospective, controlled, and non-controlled clinical trials and case series; Follow-up/Duration: Varies by study (up to 6 months); Study Period: Up to April 30, 2024
Adolescent patients (15-21 years) with Major Depressive Disorder; Sample Size: 447 total patients across all reviewed studies; 148 patients specifically using the MagVenture TMS Therapy System
Not applicable for this study
HAM-D, PHQ-9, GAD-7, CDRS-R, response and remission rates
Indications for Use
MagVenture TMS Therapy System is indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21).
Device Story
MagVenture TMS Therapy System is a computerized, electromechanical device delivering non-invasive, repetitive transcranial magnetic stimulation (rTMS) to the brain. Input: system settings defined by clinician; Transformation: high-current pulses through an electromagnetic figure-of-eight coil generate focused magnetic fields, inducing electrical currents in the left dorsolateral prefrontal cortex (L-DLPFC) to modulate neuronal activity via axonal depolarization; Output: magnetic pulse trains. Used in inpatient/outpatient clinics by trained clinical professionals. The system includes a stimulator, treatment coils, positioning arm, and head support. Clinicians use the system to deliver standardized stimulation protocols (e.g., 10Hz rTMS or iTBS) to treat MDD. The device assists in reducing depressive symptoms, providing a non-pharmacological adjunct treatment option for adolescents.
Clinical Evidence
Clinical evidence includes a systematic literature review of 19 studies (5 RCTs, 10 open-label, 1 prospective comparative, 1 retrospective comparative, 2 retrospective case series) totaling 447 adolescent patients. Six studies (N=148) specifically utilized the MagVenture TMS system. Primary endpoints included depression symptom reduction (HAM-D, PHQ-9, CDRS-R). Results consistently demonstrated large effect sizes (Hedges' g > 0.8) and significant symptom improvement. The device was well-tolerated; reported adverse events were mild/transient (headache, discomfort, fatigue). No serious adverse events were reported.
Technological Characteristics
Integrated system: MagPro stimulator, figure-of-eight coils (Cool-B65, Cool-B70, Cool D-B80), positioning arm, and head support. Principle: electromagnetic induction of cortical electrical currents. Biphasic waveforms; frequency 0.1-100 PPS; pulse width 290 μs (±5%). Biocompatible materials per ISO 10993-1. Electrical safety/EMC per IEC 60601-1/1-2. Software-controlled GUI for protocol management. Liquid-cooled coils. Connectivity: standalone system.
Indications for Use
Indicated for adolescent patients (age 15-21) with Major Depressive Disorder (MDD) as an adjunct treatment.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 11, 2025
Kirstine Klitgaard Schou, PhD
Senior Medical Affairs Specialist
MagVenture A/S
Lucernemarken 15
Farum, DK-3520
Denmark
Re: K251125
Trade/Device Name: MagVenture TMS Therapy System
Regulation Number: 21 CFR 882.5805
Regulation Name: Repetitive Transcranial Magnetic Stimulation System
Regulatory Class: Class II
Product Code: OBP
Dated: July 11, 2025
Received: July 14, 2025
Dear Dr. Klitgaard Schou:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251125 - Kirstine Klitgaard Schou
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}
K251125 - Kirstine Klitgaard Schou
Page 3
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**PAMELA D. SCOTT-S**
Digitally signed by PAMELA D. SCOTT-S
Date: 2025.08.11 19:31:09
04:00
Pamela D. Scott
Assistant Director
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
MagVenture TMS Therapy System
Page 10 of 44
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251125 | ? |
| Please provide the device trade name(s). | | ? |
| MagVenture TMS Therapy System | | |
| Please provide your Indications for Use below. | | ? |
| MagVenture TMS Therapy System is indicated as an adjunct for the treatment of Major Depressive Disorder in adolescent patients (age 15-21). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Traditional 510(k) Premarket
Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
^{}[]
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
Date Prepared: August 8, 2025
# I. SUBMITTER
Tonica Elektronik A/S
Lucernemarken 15
DK-3520 Farum, Denmark
Tel: +45 4499 1544
Primary Contact: Kirstine Klitgaard Schou, Ph.D.
Senior Medical Affairs Specialist
Phone: +45 6114 6675
E-mail: kks@magventure.com
Secondary Contact: Jan Kjøller
Head of Regulatory Affairs
Phone: +45 2489 9976
E-mail: jk@magventure.com
# II. DEVICE
Device Trade Name: MagVenture TMS Therapy System
Classification Names: Repetitive transcranial magnetic stimulation system
Regulation: 21 CFR 882.5805
Regulatory Class: Class II
Device Panel: Neurology
Product Code: OBP
# III. PRIMARY PREDICATE DEVICE
Predicate Manufacturer: Neuronetics Inc.
Predicate Trade Name: NeuroStar Advanced Therapy System
Predicate 510(k): K231926
Predicate Regulation: 21 CFR 882.5805
Predicate Regulatory Class: Class II
Predicate Device Panel: Neurology
Predicate Product Code: OBP
Page 1 of 14
{5}
Traditional 510(k) Premarket Submission MagVenture TMS Therapy System TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
^{}[]
# IV. PREDICATE DEVICE
| Predicate Manufacturer: | Tonica Elektronik A/S |
| --- | --- |
| Predicate Trade Name: | MagVita TMS Therapy System |
| Predicate 510(k): | K150641, K170114, K171481, K171967, K172667, K173620 |
| Predicate Regulation: | 21 CFR 882.5805 |
| Predicate Regulatory Class: | Class II |
| Predicate Device Panel: | Neurology |
| Predicate Product Code: | OBP |
# V. DEVICE DESCRIPTION
The MagVenture TMS Therapy System is a transcranial magnetic stimulation device. Specifically, it is a computerized, electromechanical medical device that produces and delivers non-invasive magnetic fields to induce electrical currents targeting specific regions of the cerebral cortex.
Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp.
The pulses generate an electric field in the underlying brain tissue. When this field surpasses a specific threshold and aligns appropriately with the brain's neuronal pathways, it induces localized axonal depolarization, leading to neuron activation in the targeted brain region.
The MagVenture TMS Therapy System represents an integrated system comprised of the following components:
- Magnetic Stimulator (MagPro Family)
- Coil for motor threshold determination: C-B60, C-B70
- Treatment Coils: Cool-B65, Cool-B70, Cool D-B80
- Accessories:
- Trolley with mounting for super flexible arm and coil holder arrangement)
- Patient head fixation
- Super flexible arm or Flow Arm for coil fixation
- Isolation transformer
- Cooler Unit
- Caps and Marking accessory (marking plate, pen, ruler) – Beam F3 or 5.5 cm Coil Placement
- Vacuum pump and Vacuum pillow with Pillow Case for patient head fixation (Optional)
- Treatment Chair (Optional)
Page 2 of 14
{6}
Traditional 510(k) Premarket Submission MagVenture TMS Therapy System TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
- Coil Hub (Optional)
- MagVenture TMS Atlas Neuro Navigation System (Optional)
Except for the Beam F3 marking plate, all components have previously received FDA clearance. The MagVenture TMS Therapy System and its technological characteristics remain equivalent to those cleared under K150641, K170114, K171481, K171967, K172667, K173620.
This submission introduces the following modifications:
- Expanded Indications: MagVenture TMS Therapy System is indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21)
- Coil Approval Expansion: Inclusion of Cool D-B80 for MDD in adolescent patients (age 15-21).
- Beam F3 Marking Method: Introduces the Beam F3 marking plate as an alternative to the standard 5.5 cm method for coil positioning in the dorsolateral prefrontal cortex (DLPFC).
Apart from the modifications outlined above, the core design, fundamental operating principles, and safety characteristics remain substantially equivalent to the previously cleared MagVenture TMS Therapy System.
## VI. INDICATIONS FOR USE STATEMENT
MagVenture TMS Therapy System is indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21).
## VII. SUBSTANTIAL EQUIVALENCE DISCUSSION
The following table compares the MagVenture TMS Therapy System to the (primary) predicate device with respect to indications for use, and technological characteristics including principles of operation, design and output stimulation parameters, and forms the basis for the substantial equivalence determination.
The new device is substantially equivalent to the primary predicate device (K231926). They have the same intended use and indications for use. Both devices use the same principles of operation to apply TMS treatment at a defined intensity as repetitive pulse trains delivered as brief rapidly alternating magnetic fields to induce electrical currents to the prefrontal cortex. They have substantially equivalent technological characteristics. Both devices represent integrated systems of components including TMS stimulator with software, electromagnetic figure-of-8-coil for MT determination and treatment, coil fixture and positioning system and head support system. The operational procedures including system setup, patient preparations, motor threshold
Page 3 of 14
{7}
Traditional 510(k) Premarket
Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
determination, coil positioning and patient treatment with predefined treatment stimulation settings are substantially equivalent. Devices target the identical anatomical area of the brain. Minor identified differences in system components and output stimulation paraments do not raise different questions of safety and effectiveness than the primary predicate.
Substantial equivalence between the MagVenture TMS Therapy System and NeuroStar Advanced Therapy System, having identical technological characteristics as described for the respective systems in this 510(k) submission, has previously been demonstrated via 510(k) clearances K150641 and K171481 covered by the predicate device (MagVita TMS Therapy System).
The new device is substantially equivalent to the predicate device. They have the same intended use and identical technological characteristics. The predicate device previously obtained FDA clearance for the treatment of major depressive disorder (MDD) in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode (K150641, K170114, K171481, K171967, K172667, K173620).
Page 4 of 14
{8}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
+
Table-1 Comparison of Technological Characteristics for Major Depressive Disorder (MDD) in adolescent patients (age 15-21) and (Primary) Predicate Device
| Attribute | MagVenture TMS Therapy System (New device) | NeuroStar Advanced Therapy System (Primary predicate device) | MagVita TMS Therapy System (Predicate device) | Comparison |
| --- | --- | --- | --- | --- |
| 510(k) Number | K251125 | K231926 | K150641, K170114, K171481, K171967, K172667, K173620 | Not applicable |
| Manufacturer of the system | Tonica Elektronik A/S | Neuronetics, Inc. | Tonica Elektronik A/S | Not applicable |
| Product Code | OBP | OBP | OBP | Same |
| Indications for Use | MagVenture TMS Therapy System is indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21) | NeuroStar Advanced Therapy is indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (age 15-21) | MagVita TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode | Same intended use and indications for use as primary predicate. Same intended use as predicate. |
| Intended use | Treatment of MDD | Treatment of MDD | Treatment of MDD | Same |
| Intended users | Trained clinical professionals | Trained clinical professionals | Trained clinical professionals | Same |
| Anatomical Site | Left dorsolateral prefrontal cortex (L-DLPFC) | Left dorsolateral prefrontal cortex (L-DLPFC) | Left dorsolateral prefrontal cortex (L-DLPFC) | Same |
| Principle of operation | Generation of varying and focused magnetic fields to induce electric currents in specific areas of the brain to modulate neuronal activity | Generation of varying and focused magnetic fields to induce electric currents in specific areas of the brain to modulate neuronal activity | Generation of varying and focused magnetic fields to induce electric currents in specific areas of the brain to modulate neuronal activity | Same |
| Target Population | Adolescent patients aged 15-21 years | Adolescent patients aged 15-21 years | Adults | Same as primary predicate. Substantial equivalent to predicate |
Page 5 of 14
{9}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
+
| Attribute | MagVenture TMS Therapy System
(New device) | NeuroStar Advanced Therapy System
(Primary predicate device) | MagVita TMS Therapy System
(Predicate device) | Comparison |
| --- | --- | --- | --- | --- |
| **Clinical setting** | Inpatient and outpatient settings including physician’s offices and clinics, hospitals, and general medical/surgical hospitals | Inpatient and outpatient settings including physician’s offices and clinics, hospitals, and general medical/surgical hospitals | Inpatient and outpatient settings including physician’s offices and clinics, hospitals, and general medical/surgical hospitals | Same |
| **Device design & components** | Mobile console for housing of MagPro family stimulator and electronics, built in display, system software with graphic user interface
Coil for MT determination and treatment
Coil fixture and positioning system
Treatment chair
Accessory for marking and locating treatment area
Head Support System
Data management system | Mobile console for housing of stimulator and electronics, display monitor, system software with graphic user interface
Coil for MT determination and treatment
Coil fixture and positioning system
Treatment chair
Accessory for marking and locating treatment area
Head Support System
Data management system | Mobile console for housing of MagPro family stimulator and electronics, built in display, system software with graphic user interface
Coil for MT determination and treatment
Coil fixture and positioning system
Treatment chair
Accessory for marking and locating treatment area
Head Support System
Data management system | Substantially equivalent to primary predicate.
Same as predicate. |
| **Biocompatibility** | Patient-contacting device components use standard materials compliant with ISO 10993-1 that are commonly used in consumer products and medical device applications | Patient-contacting device components use standard materials compliant with ISO 10993-1 that are commonly used in consumer products and medical device applications | Patient-contacting device components use standard materials compliant with ISO 10993-1 that are commonly used in consumer products and medical device applications | Same |
| **Coil Configuration** | Figure-of-eight coil
Air core
Liquid cooling | Figure-of-eight coil
Ferromagnetic Core
Air cooling | Figure-of-eight coil
Air core
Liquid cooling | Substantial equivalent to primary predicate.
Same as predicate. |
Page 6 of 14
{10}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
+
| Attribute | MagVenture TMS Therapy System (New device) | NeuroStar Advanced Therapy System (Primary predicate device) | MagVita TMS Therapy System (Predicate device) | Comparison |
| --- | --- | --- | --- | --- |
| Performance | Waveforms: Biphasic
Frequency: 0.1 - 30 PPS or 0.1 - 100 pulses, depending on model
Preset range of % MT: 0 - 140%
Amplitude Range: 0 - 1.7 SMT (80 - 120% MT) (Standard & iTBS treatment)
Pulse width: 290 μs (± 5%) | Waveforms: Biphasic
Frequency: 1 - 30 PPS. For MT determination: 0.1 - 0.3 PPS
Preset range of % MT: 25 - 140%
Amplitude Range: Standard treatment: 0.22 to 2.08 SMT; iTBS treatment: 0.22 to 1.9 SMT
Pulse width: 185 μs (±10%) | Waveforms: Biphasic
Frequency: 0.1 - 30 PPS or 0.1 - 100 pulses, depending on model
Preset range of % MT: 0 - 140%
Amplitude Range: 0 - 1.7 SMT (80 - 120% MT) (Standard & iTBS treatment)
Pulse width: 290 μs (± 5%) | Same |
| Treatment Protocol | Standard Protocol | | | |
| Stimulation intensity | 120% MT | 120% MT | 120% MT | Same |
| Repetition rate | 10 Hz | 10 Hz | 10 Hz | |
| Train duration | 4 sec | 4 sec | 4 sec | |
| Inter-train Interval | 11 - 26 sec | As low as 11 sec | 11 - 26 sec | |
| Number of pulses/session | 3000 | 3000 | 3000 | |
| Total treatment duration | 19 - 37.5 min | As low as 18.75 min | 19 - 37.5 min | |
| Treatment Protocol | iTBS# | | | |
| Stimulation intensity | 120% MT | 80-120% MT | 120% MT | Substantially equivalent to primary predicate. Same as predicate |
| Repetition rate | 50 Hz | 50 Hz | 50 Hz | |
| Train duration | 2 sec | 2 sec | 2 sec | |
| Inter-train interval | 8 sec | 8 sec | 8 sec | |
| Burst pulses | 3 | 3 | 3 | |
| Bursts | 200 | 200 | 200 | |
| Interpulse interval | 20 msec | 20 msec | 20 msec | |
Page 7 of 14
{11}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
| Attribute | MagVenture TMS Therapy System
(New device) | NeuroStar Advanced Therapy System
(Primary predicate device) | MagVita TMS Therapy System
(Predicate device) | Comparison |
| --- | --- | --- | --- | --- |
| Number of trains | 20 | 20 | 20 | |
| Pulses per session | 600 | 600 | 600 | |
| Total treatment duration | 3 min and 9 sec | 3 min and 20 sec | 3 min and 9 sec | |
| Electrical Safety & Electromagnetic Compatibility | Complies with IEC60601-1 and IEC60601-1-2 | Complies with IEC60601-1 and IEC60601-1-2 | Complies with IEC60601-1 and IEC60601-1-2 | Same |
| Quality & Risk standards | Company complies with ISO 13485:2016 and ISO 14971:2019 | Company complies with ISO 13485:2016 and ISO 14971:2019 | Company complies with ISO 13485:2016 and ISO 14971:2019 | Same |
Page 8 of 14
{12}
Traditional 510(k) Premarket Submission MagVenture TMS Therapy System TONICA ELEKTRONIK A/S
## VIII. PERFORMANCE TESTING SUMMARY
### Non-Clinical Performance Testing and Performance Standards
The MagVenture TMS Therapy System has been tested and complies with the following recognized electrical safety and performance standards:
- IEC 60601-1: General safety and essential performance requirements.
- IEC 60601-1-2: Electromagnetic compatibility compliance.
- IEC TR 60601-4-2: Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- ISO 13485:2016 & ISO 14971:2019: Quality and risk management compliance.
Additionally, computational modeling (SimNIBS v4.0.1) has been used to compare the electric field strength of the MagVenture coils against the predicate device.
### Clinical Performance Data
The primary predicate device (K231926) has been shown to be clinically safe and effective as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients aged 15-21 years.
The new device is substantially equivalent to the primary predicate device. Since the proposed device is sufficiently similar to the NeuroStar TMS Therapy System in terms of indications, device specifications and energy output, the omission of clinical testing is justified by the SE determination and reliance on prior performance testing per indicated relevant consensus standards and FDA Special Controls Guidance for rTMS as performed for the predicate MagVita TMS Therapy System.
To further support the substantial equivalence determination for the new device, however, the applicant performed a systematic clinical literature review based on PICO principles to evaluate the safety and effectiveness of TMS for the treatment of MDD in adolescent patients (15-21 years).
Searches were performed using PubMed, PubMed Central (PMC) and Cochrane Central Register of Controlled Trials databases up to April 30, 2024. Searches included combinations of the following terms: [Transcranial magnetic stimulation], [(r)TMS], [depression], [major depressive disorder], [adolescent], [young adult], [child(ren)], [teenager], [MagVita TMS Therapy], [NeuroStar Advanced Therapy], [MagVenture], [Neuronetics], [Tonica Elektronik], [antidepressant] and [antidepressant therapy]. Defined clinical paper selection criteria included:
- Prospective, retrospective, controlled or non-controlled clinical trials or case series covering ≥ 6 patients addressing depression related endpoints
- Adolescents (15-21 years), male or female
- Patients with clinical diagnosis of major depressive disorder
Page 9 of 14
{13}
Traditional 510(k) Premarket Submission MagVenture TMS Therapy System TONICA ELEKTRONIK A/S
- MagVita/MagVenture TMS System, NeuroStar TMS System or comparable TMS system with Figure-of-eight-coil
- Anatomical location: Left-DLPFC
- rTMS or intermittent Theta Burst Stimulation (iTBS) treatment protocol
- Full text English paper
The systemic review resulted in the selection of a total of five (5) prospective randomized controlled clinical trials (Pan et al., 2020; Chen et al., 2022; Croarkin et al., 2021; Zhao et al., 2023; Zhang M et al., 2024) in which TMS was used as an adjunct for the treatment of MDD, ten (10) open label, single arm studies (Wall, 2011, 2016; Dhami et al., 2019; Zhang T et al., 2019; MacMaster et al., 2019; Sphere et al., 2021; Croarkin et al., 2018, 2016; Lu et al., 2023; Zhang L et al., 2020), one (1) prospective comparative study (Gordon et al, 2022), one (1) retrospective comparative study (Rosenich et al., 2019) and two (2) retrospective case series (Croarkin et al., 2024; Nakano et al., 2023). They cover a total of 447 adolescent patients in which TMS was applied for the treatment of MDD.
The review further identified ten (10) systematic literature reviews (Sigrist et al. 2022; Oberman et al., 2021; Hett et al., 2021; Magavi et al., 2017; Cao et al., 2023; Bejenaru et al., 2022; Majumder et al., 2021; Qiu et al., 2023; Sun et al., 2023; Donaldson et al., 2014). Five (5) of these reviews included a meta-analysis (Sigrist et al. 2022; Cao et al., 2023; Majumder et al., 2021; Qiu et al., 2023; Sun et al., 2023).
Clinical data on the proposed MagVenture TMS Therapy System is reported by the studies of Zhang T et., 2019 (N=42), Rosenich et al., 2019 (N=15), Croarkin et al., 2024 (N=58), Zhang L et al., 2020 (N=13), Nakano et al., 2023 (N=6) and Gordon et al., (N=14), covering a total of 148 adolescent patients (Table 2). These studies further indicate the safety and effectiveness of the new TMS device for the treatment of MDD in adolescent patients aged 15-21 years.
The literature review indicates that TMS is effective as an adjunct for the treatment of MDD in adolescents aged 15-21 years. All studies concluded that TMS is well tolerated and safe for this patient population. Side effects of TMS treatment are similar to those observed in adults. They are typically mild and transient and include headaches, skull discomfort, musculoskeletal discomfort, neck pain, twitching/tingling, mood changes, fatigue, nausea, dizziness anxiety, tingling and tinnitus. No serious adverse events were reported.
The systematic clinical literature review, covering clinical data from the proposed device and substantially equivalent devices, therefore confirms that the use of TMS therapy as an adjunct for the treatment of MDD in adolescent patients aged 15-21 years is safe and effective. Clinical data further demonstrates substantially equivalent treatment effect of TMS therapy as an adjunct to antidepressant therapy over antidepressant therapy alone in reducing depression in adolescents that is consistent within and across the studies.
Page 10 of 14
{14}
Traditional 510(k) Premarket Submission MagVenture TMS Therapy System TONICA ELEKTRONIK A/S
## IX. PERFORMANCE TESTING SUMMARY
The proposed MagVenture TMS Therapy System is substantially equivalent to the primary predicate device. They have the same intended use and indications for use and the same or similar technological characteristics. Identified differences do not raise new or different questions of safety and effectiveness.
Non-clinical performance testing per relevant consensus standards and FDA Special Controls Guidance Document for rTMS System as performed for the predicate device support the conclusion that the performance and safety characteristics of the MagVenture TMS Therapy System are substantially equivalent to the primary predicate device.
The safety and effectiveness of the new device and the substantial equivalence determination is further supported by clinical data retrieved from systematic literature review using defined inclusion criteria and covering the use of the subject device, primary predicate device and comparable devices for the treatment of MDD in adolescent patients aged 15-21 year.
Page 11 of 14
{15}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
+
Table-2. Overview of Clinical Data for the MagVenture TMS Therapy System
| Reference | Study Design and Protocol | Patients | N | Outcome Measures | Clinical effectiveness | | | Clinical safety | HAMD Effect Size Hedges' g * | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Zhang T et al., 2019 | Design: Open label
Groups: rTMS applied in Adolescents, Adults, and Older Adults
TMS Device: MagVenture TMS System, Figure-of-8-coil
Protocol: L-DLPFC 120% MT; 10 Hz ITI: 12s 2,400 pulses; 20 sessions, 4 wks | Patients with mood and anxiety disorders.
Age (years): Adolescents: M = 14.6 S = 2.0 R = 10-17 Gender (M/F): 13/29 | Adolesc: N = 42 Adult: N = 75 | HAM-D | Results: An overview of the mean depression scores from baseline to 2 weeks and treatment end (4 weeks) is provided in the table below: | | | No serious adverse events reported
AEs reported included transient mild headache and musculoskeletal discomfort | 5.49 (@ wk2)
95% CI: [4.275, 6.695]
Large effect size
3.86 (@ wk4)
95% CI: [2.561, 5.164]
Large effect size | |
| | | | | | Active TMS (N = 42) | P | | | | |
| Rosenich et al., 2019 | Design: Retrospective, comparative
Group: rTMS (MagVenture TMS System, Figure-of-8-coil; Neuro-MS/D TMS System, Figure-of-8-coil)
Protocol: Right unilateral (n=8) or bilateral (n=4) DLPFC 110% MT; 1 Hz (R-DLPFC) and/or 10 Hz (L-DLPFC) 900-2,400 pulses 18 sessions, 3/wk for 6 wks. | Patients with treatment resistant MDD
Age (years): M = 20.69 S = 2.55 R = 17-25 Gender (M/F): 8/7 | N = 15 | HAM-D MADRS
Zung Self-Rating Depression Scale | Results: An overview of the mean depression scores from baseline to treatment end (3 weeks) is provided in the table below: | | | No serious adverse events reported
AEs included mild and transient side effects such as discomfort at the treatment site, headache, and tiredness following treatment | -6.48
95% CI: [-8.854, -4.115]
Large effect size | |
| | | | | | Active rTMS (N = 15) | F | P | | | |
| Crourkin et al., 2024 | Design: Retrospective case series
TMS Device: MagVenture TMS | Diagnosed MDD
Age (years): M = 15.8 S = 1.1 R = 12-17 | N = 58 | PHQ-9 GAD-7 | Results: An overview of the mean depression and anxiety scores PHQ-9 and GAD-7, respectively, from baseline to 8 weeks is provided in the table below: | | | No serious adverse events reported
AEs: An overview of the mean depression and anxiety scores PHQ-9 and GAD-7, respectively, from baseline to 8 weeks is provided in the table below: | Not applicable | |
| | | | | | Overall, 32/57 treatment-completers (56.1%) met response criteria and 20/57 (35.1%) met remission criteria. Overall PHQ-9 scores improved by an average of 48.7%, from 15.2 ± SD6.1 to 7.8 ± SD6.1, indicating a large effect size (Cohen's d = 1.10). | | | | | |
Page 12 of 14
{16}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
+
| Reference | Study Design and Protocol | Patients | N | Outcome Measures | Clinical effectiveness | Clinical safety | HAMD Effect Size Hedges' g * |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Therapy System; Figure-of-8-coil
Protocol: R-DLPFC: 120% MT; 1Hz 60 s on/30 s off, 6 trains, 360 pulses total
L-DLPFC: 120% MT; 20 Hz 2 s on/4 s off, 30 trains, 1200 pulses total
Daily sessions (week days) total of 36 sessions | | | | Conclusion: The authors indicated that the observations from the retrospective case series suggest that a bilateral 1 Hz right/20 Hz left DLPFC protocol is safe, tolerable, and effective for both depressive symptoms and anxiety, with large effect sizes seen in this sample of adolescents.
**Active rTMS** (N = 58)
PHQ-9
Baseline mean scores 15.2 ± 6.08 -
8 wk mean scores 7.8 ± 6.05 -1.10
GAD-7
Baseline mean scores 11.8 ± 5.85 -
8 wk mean scores 5.9 ± 5.2 -0.94 | Safety and tolerability were strong, with 57/59 patients completing at least 30 sessions and no dropouts due to adverse effects. | |
| Zhang L et al., 2020 | Design: Open label
Groups: rTMS applied in Adolescents, Adults, and Older Adults
TMS Device: MagVenture TMS System; Figure-of-8-coil
Protocol: L-DLPFC: 120% MT; 10 Hz ITI: 12s; 30 pulses per train
80 stimulation trains per session
20 sessions, 4 wks
R-DLPFC: 120% MT; 1 Hz ITI: 12s; 30 pulses per train
80 stimulation trains per session
20 sessions, 4 wks | Patients with depression and anxiety symptoms
Age (years): M = 15.24 S = 1.61 | N = 13 | HAMD-17 | Results: An overview of the mean depression scores from baseline to 4 weeks is provided in the table below:
Active rTMS P (N = 13)
HAMD
Baseline mean scores 14.31 ± 5.60
4 wk mean scores 3.23 ± 1.74 <0.001
Conclusion: Results indicate that rTMS is a safety and effective add-on therapy for the treatment of depression in adolescents. | No serious adverse events reported.
AE reported was dizziness | 2.69
95% CI: [1.536, 3.839]
Large effect size |
| Nakano et al., 2023 | Design: Case series
TMS Device: MagVenture TMS System, Figure-of-8-coil | Patients diagnosed with MDD
Age (years): R = 16-18
Gender (M/F): 5/1 | N = 6 | PHQ-9 | Results:
Remission: 50% (3/6) (PHQ-9 score < 5)
Response: 33% (2/6) (≥ 50% reduction in PHQ-9 scores)
The Standard Mean Difference (SMD) for PHQ-9 scores was 1.6 at 1-month posttreatment.
Conclusion: These findings support the safe, clinical use of accelerated iTBS as an adjunct for the treatment of MDD in adolescents | No serious adverse events reported.
All patients tolerated treatment well. Reported AEs included headache, minor eye strain, ‘abdominal twitching’, irritability | Not applicable |
Page 13 of 14
{17}
Traditional 510(k) Premarket Submission
MagVenture TMS Therapy System
TONICA ELEKTRONIK A/S
Tonica Elektronik A/S
+
| Reference | Study Design and Protocol | Patients | N | Outcome Measures | Clinical effectiveness | Clinical safety | HAMD Effect Size Hedges' g * |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Protocol: L-DLPFC: iTBS 120% MT 1800 pulses Mean total sessions: 44 up to 1 month | | | | | | |
| Gordon et al., 2022 | Design: Prospective randomized trial TMS Device: MagVenture TMS System Protocol: 120% MT; 20 treatments in 4 weeks L-DLPFC: high frequency rTMS; 10 Hz 45 trains of 4 sec; ITT: 26 sec; 1800 pulses/day R-DLPFC: low frequency rTMS; 1 Hz 1 train of 30 min; 1800 pulses per day | Diagnosed MDD Age (years): M = 16.5 S = 1.16 Gender (M/F): 4/10 | N = 14 | CDRS-R | Results: An overview of the response rates based on CDRS-R scores after 20 treatments and 1-month follow-up is provided in the table below:
Active iTMS (N = 14)
20 treatments
Partial response 7.1% (1/14)
Response 0% (0/14)
1-month
Partial response 28.6% (4/14)
Response 14.2% (2/14)
Partial response (25-50% reduction in CDRS-R); response (≥50 % score reduction in CDRS-R)
CDRS-R score changes from baseline were -14.99 after 20 treatments (P<0.001; [95% CI: -19.77, -10.20]) and -17.62 after 1-month follow-up (P<0.001; [95% CI: -22.54, -12.70]). CDRS-R scores improved significantly across the 20 rTMS treatments, with peak response at 1-month follow-up. Two (14%) adolescents had ≥ 50% score improvement and a further four (29%) demonstrated partial response (between 25 and 50% reduction) by 1-month follow-up. Treatment gains were sustained at 6-month follow-up. There was no significant difference in efficacy between left- and right-sided treatment
Conclusion: It is concluded that adolescents with MDD benefited from rTMS | No serious adverse events reported
Treatment was well-tolerated
Most reported side-effects during treatment were lethargy and drowsiness
Headaches were reported by nine participants | Not applicable |
Significance values for P are expressed in comparison to baseline values, or to control group in case of controlled studies;
Beck Scale for Suicide Ideation (BSI); Hamilton Depression Rating Scale (HAMD); Montgomery-Åsberg Depression Rating Scale (MADRS); Wisconsin Card Sorting Test (WCST)
Continuous Performance Test (CPT); Stroop Color-Word Test (SCWT); Children's Depression Rating Scale-Revised (CDRS-R)
Quick Inventory of Depressive Symptomatology-- Adolescent (17 Item) – Self Report (QIDS- A17-SR); Clinical Global Impression's Scale – Severity (CGI-S)
Beck Depression Inventory (BDI); Quality of Life Enjoyment and Satisfaction Questionnaire (QLES-Q; Brief Psychiatric Rating Scale for Children (BPRS- C)
Children's Global Assessment Scale (CGAS); Suicidal Ideation Questionnaire (SIQ); Patient Health Questionnaire (PHQ-9); 7-item Generalized Anxiety Disorder scale (GAD-7);
Hamilton Anxiety Scale (HAMA)
* Where Hedges' g represents an effect size Small: < 0.2; Medium: 0.5; Large: > 0.8.
Page 14 of 14
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.