K251123 · Taewoong Medical Co., Ltd. · FGE · Dec 29, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K251123
Device Name
Niti-S Biliary Stent; Niti-S Biliary Slim M Stent
Applicant
Taewoong Medical Co., Ltd.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Dec 29, 2025
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Niti-S Biliary Stent is indicated for the palliation of malignant strictures in the biliary tree. The Niti-S Biliary Slim M Stent is indicated for the palliation of malignant strictures in the biliary tree.
Device Story
Niti-S Biliary Stent and Niti-S Biliary Slim M Stent are implantable, self-expanding Nitinol wire stents; used for palliation of malignant biliary strictures. Stents are pre-loaded into a disposable, flexible, co-axial tube-type delivery system. Operated by physicians via endoscopic or percutaneous (Biliary Stent only) access under fluoroscopic or endoscopic guidance. Deployment occurs by pulling the outer sheath of the delivery system, allowing the stent to expand at the target site. Device provides mechanical patency to the biliary tree; benefits include symptom relief from malignant obstruction. No electrical components or software.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included packaging strength, dye penetration, sterility verification, delivery/withdrawal/deployment forces, deployment accuracy, SDS bonding strength, crossing profile, and radiopacity evaluation.
Technological Characteristics
Self-expanding Nitinol wire stent; hook-type design. Delivery system: co-axial tube, disposable, flexible. Materials identical to predicates (stent) and reference device (SDS). Sterilization: Ethylene Oxide (EO). No electrical components or software. Dimensions: 6 Fr or 8 Fr profiles; 50 cm or 180 cm lengths.
Indications for Use
Indicated for patients with malignant strictures in the biliary tree requiring palliation.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 29, 2025
Taewoong Medical Co., Ltd.
% Matthew Krueger
Principal Consultant
Biologics Consulting Group
100 Daingerfield Road
Suite 400
Alexandria, Virginia 22314
Re: K251123
Trade/Device Name: Niti-S Biliary Stent; Niti-S Biliary Slim M Stent
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter And Accessories
Regulatory Class: Class II
Product Code: FGE
Dated: December 1, 2025
Received: December 1, 2025
Dear Matthew Krueger:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
The OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the device's labeling:
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K251123 - Matthew Krueger
Page 2
1. The safety and effectiveness of this device for use in the vascular system has not been established.
Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule").
{3}
K251123 - Matthew Krueger
Page 3
The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
GLENN B. BELL -S
for Michael J. Hoffmann
Director
OHT3: Gastrorenal, ObGyn, General Hospital, and Urology
Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251123
Device Name
Niti-S Biliary Stent ;
Niti-S Biliary Slim M Stent
Indications for Use (Describe)
Niti-S Biliary Stent:
The Niti-S Biliary Stent is indicated for the palliation of malignant strictures in the biliary tree.
Niti-S Biliary Slim M Stent:
The Niti-S Biliary Slim M Stent is indicated for the palliation of malignant strictures in the biliary tree.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
K251123 - Page 1/4
K251123 510(k) Summary
Taewoong Medical
1. SUBMITTER INFORMATION
Applicant: Taewoong Medical Co., Ltd.
Contact: Yongjin Jeff Kim
Phone: +82 70 4649 1543
Email: jinjeff@stent.net
Address: 14, Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi-do, 10022, Republic of Korea
2. CORRESPONDENT INFORMATION
Contact: Matthew Krueger
Title: Principal Consultant
Firm: Biologics Consulting Group
100 Daingerfield Rd, Suite 101
Alexandria, VA 22314
Phone: (667) 352-2578
Email: mkrueger@biologicsconsulting.com
3. DATE PREPARED: DECEMBER 16, 2025
4. DEVICE INFORMATION
Device Name: Niti-S Biliary Stent & Niti-S Biliary Slim M Stent
Common Name: Biliary catheter and accessories
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary catheter and accessories
Product Code: FGE
Regulatory Class: Class II
5. PREDICATE DEVICE INFORMATION
Primary Predicate Device Name: Niti-S Biliary Slim M Stent
510(k) Number: K221071
Manufacturer: Taewoong Medical Co., Ltd.
Page 1 of 4
{6}
K251123 - Page 2/4
K251123 510(k) Summary
Taewoong Medical
Secondary Predicate Device Name: Niti-S Biliary Stent
510(k) Number: K073667
Manufacturer: Taewoong Medical Co., Ltd.
Reference Device Name: Esophageal TTS Stent
510(k) Number: K240522
Manufacturer: Taewoong Medical Co., Ltd.
Additional Reference Device Name: Niti-S Duodenal Comfort Stent, Niti-S Colonic Comfort Stent
510(k) Number: K250663
Manufacturer: Taewoong Medical Co., Ltd.
## 6. DEVICE DESCRIPTION
The Niti-S Biliary Stent consists of an implantable metallic stent and a disposable, flexible Stent Delivery System for placement of the stent. The stent is a flexible and expandable tubular device made of Nitinol wire. Two Nitinol wires are woven in a hook-type design. The stent design is identical to the predicate device cleared in K073667.
The Stent Delivery System is a disposable system for the delivery and deployment of the stent at the target position.
## 7. INDICATIONS FOR USE
The Niti-S Biliary Stent is indicated for the palliation of malignant strictures in the biliary tree.
The Niti-S Biliary Slim M Stent is indicated for the palliation of malignant strictures in the biliary tree.
## 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Indications for Use are for the Niti-S Biliary Stent and Niti-S Biliary Slim M Stent are unchanged from the cleared predicate devises (K073667 and K221071). The purpose of this 510(k) is to align the design of the subject stent delivery systems (SDSs) with those of the Esophageal TTS Stent (K240522). These changes included slight component modifications and some material changes. The principle of operation of the updated Niti-S Biliary Stent and Niti-S Biliary Slim M Stent SDSs remains unchanged. Fluoroscopic and/or endoscopic guidance is still used for SDS positioning and the stent deployment procedure is the same overall.
Page 2 of 4
{7}
K251123 - Page 3/4
K251123 510(k) Summary
Taewoong Medical
Table 1: Niti-S Biliary Stent and Niti-S Biliary Slim M Stent SDS Technological Comparison to Reference Device
| | Subject Device Niti-S Biliary Stent | Subject Device- Niti-S Biliary Slim M Stent | Reference Device Esophageal TTS Stent SDS K240522 |
| --- | --- | --- | --- |
| Trade/Device Name | Niti-S Biliary Stent | Niti-S Biliary Slim M Stent | Esophageal TTS Stent |
| Regulation Number | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 878.3610 |
| Expansion method | The self-expanding Nitinol stent is pre-loaded into the distal part of the delivery system and is expanded in the body by pulling the outer sheath of the delivery device. | The stent is loaded into the distal part of the delivery device and expanded in the body by pulling the outer sheath of the delivery device. | The self-expanding Nitinol stent is pre-loaded into the distal part of the delivery system and is expanded in the body by pulling the outer sheath of the delivery device. |
| Method of introduction | Endoscopic, Percutaneous | Endoscopic | Endoscopic |
| Sterility | EO Sterilization | EO Sterilization | EO Sterilization |
| Delivery system photo | Co-axial tube type (endoscopic) | Co-axial tube type | Co-axial tube type |
| Delivery system photo | Co-axial tube type (percutaneous) | Not applicable. | Not applicable. |
| Delivery system length | 180 cm (endoscopic), 50 cm (percutaneous) | 180 cm (endoscopic) | 180 cm (endoscopic), 220 cm (endoscopic) |
| Delivery system profile | 8 Fr (2.7 mm) | 6 Fr (2 mm) | 10.5 Fr (3.5 mm) |
| Guidewire (in inches) | 0.035 in | 0.025 in | 0.035 in (0.89 mm) 0.038 in (0.97 mm) |
{8}
K251123 - Page 4/4
K251123 510(k) Summary
Taewoong Medical
## 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
### Biocompatibility Testing
The stent materials of the subject devices are identical to those of the Niti-S Biliary Stent cleared in K073667 and the Niti-S Biliary Slim M Stent cleared in K221071, including its formulation, processing, and sterilization, with no additional chemicals (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents) introduced.
Similarly, the SDS material of the subject device is identical to the Esophageal TTS Stent cleared in K240522 by Taewoong Medical, including its formulation, processing, and sterilization, with no additional chemicals added.
Therefore, additional biocompatibility testing for the Stent Delivery System is not required.
### Electrical Safety
Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type.
### Electromagnetic Compatibility (EMC)
Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type.
### Software
Not applicable. The subject device contains no software.
### Performance Testing
Additional performance testing was conducted to demonstrate that the device meets its design requirements and performs as intended. The performance tests include:
- Packaging strength and dye penetration
- Sterility verification
- Delivery Force & Withdrawal Force
- Deployment Force
- Deployment Accuracy
- SDS Bonding Strength
- Crossing Profile
- Radiopacity evaluation
## 10. CONCLUSION
The results of the performance testing described above demonstrate that the Niti-S Biliary Stent and the Niti-S Biliary Slim M Stent perform equivalently as compared to the predicate and reference devices and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.