Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118

K251048 · STERIS Corporation · JOJ · May 2, 2025 · General Hospital

Device Facts

Record IDK251048
Device NameRapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118
ApplicantSTERIS Corporation
Product CodeJOJ · General Hospital
Decision DateMay 2, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The RAPICIDE™ PA Test Strips are chemical indicators used after the disinfection cycle to ensure that the RAPICIDE™ PA High-Level Disinfectant solution is met the minimum recommended concentration of 850ppm peracetic acid. The RAPICIDE PA Test Strips are designed for use by trained personnel who reprocess endoscopes and their accessories in facilities where endoscopies are performed.

Device Story

Chemical indicator test strip; measures peracetic acid (PAA) concentration in high-level disinfectant solutions. Used by trained personnel in endoscopy facilities post-disinfection cycle. Input: disinfectant solution sample. Principle: colorimetric reaction on test strip pad. Output: visual color change; solid black indicates pass (>850 ppm PAA); dark grey, blue grey, violet grey, light grey, or white indicates failure (≤850 ppm PAA). Assists healthcare providers in verifying disinfectant efficacy; ensures endoscope reprocessing meets safety standards; prevents use of sub-therapeutic disinfectant concentrations.

Clinical Evidence

Bench testing only. Performance validated through dynamic range, comparative sensitivity/specificity, analytic specificity (contaminants and other germicides), shelf life, and in-use (open bottle) stability testing. All tests met acceptance criteria.

Technological Characteristics

Chemical indicator strip; chromatography paper pad; non-sterile; 18-month shelf life; 4-month open-bottle stability. Measures PAA concentration via colorimetric reaction. No patient/user contact. No electronic components or software.

Indications for Use

Indicated for trained personnel reprocessing endoscopes and accessories to verify that RAPICIDE PA High-Level Disinfectant solution maintains a minimum concentration of 850ppm peracetic acid.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION May 2, 2025 STERIS Corporation Steven Deline Sr. Manager, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060 Re: K251048 Trade/Device Name: Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118 Regulation Number: 21 CFR 880.2800(b) Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: April 2, 2025 Received: April 3, 2025 Dear Steven Deline: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251048 - Steven Deline Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251048 - Steven Deline Page 3 Sincerely, Stephen A. Anisko -S Digitally signed by Stephen A. Anisko -S Date: 2025.05.02 16:26:51 -04'00' for: Christopher K. Dugard, M.S. Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251048 | | | Device Name Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118 | | | Indications for Use (Describe) | | | The RAPICIDE™ PA Test Strips are chemical indicators used after the disinfection cycle to ensure that the RAPICIDE™ PA High-Level Disinfectant solution is met the minimum recommended concentration of 850ppm peracetic acid. The RAPICIDE PA Test Strips are designed for use by trained personnel who reprocess endoscopes and their accessories in facilities where endoscopies are performed. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} STERIS® ^{}[] # K251048 510(k) Summary For Rapicide PA High-Level Disinfection Test Strips STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198 Contact: Steve Deline, MS RAS Sr. Manager, Regulatory Affairs Telephone: 952-767-7386 e-mail: steve_deline@steris.com Summary Date: May 1, 2025 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 K251048 May 1, 2025 Page 1 of 4 {5} STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Rapicide PA High-Level Disinfectant Test Strips 1. Device Name Trade Name: Rapicide PA High-Level Disinfectant Test Strips Model Ref # ML02-0118 Device Class: Class II Common/usual Name: Chemical Indicator Classification Name: Physical/Chemical sterilization process Classification Number: indicator 21 CFR 880.2800 Product Code: JOJ 2. Predicate Device K152394 Rapicide PA High-Level Disinfectant Test Strips | Table 1. Device Comparison Table | | | | | --- | --- | --- | --- | | Feature | Predicate Device K152394 | Rapicide PA HLD Test Strips (Proposed device) | Comparison | | Indications for Use | The RAPICIDE™ PA Test Strips are chemical indicators used after the disinfection cycle to ensure that the RAPICIDE™ PA High-Level Disinfectant solution is met the minimum recommended concentration of 850ppm peracetic acid. The RAPICIDE PA Test Strips are designed for use by trained personnel who reprocess endoscopes and their accessories in facilities where endoscopies are performed. | The RAPICIDE™ PA Test Strips are chemical indicators used after the disinfection cycle to ensure that the RAPICIDE™ PA High-Level Disinfectant solution is met the minimum recommended concentration of 850ppm peracetic acid. The RAPICIDE PA Test Strips are designed for use by trained personnel who reprocess endoscopes and their accessories in facilities where endoscopies are performed. | Identical | | Minimum Recommended Concentration | Measures the disinfectant use solution concentration above 850ppm PAA | Measure the disinfectant use solution concentration above 850ppm PAA | Identical | | Endpoint Specification “Pass” | Solid Black > 850 ppm PA | Solid Black > 850 ppm PA | Identical | | Performance | • Dynamic Range • Comparative sensitivity and specificity • Analytic Specificity | • Dynamic Range • Comparative sensitivity and specificity • Analytic Specificity | Identical | | Test Strip Components | • Chromatography paper supplier • Adhesive type • Test Strip dimensions / Pad material | • Chromatography paper supplier • Adhesive type • Test Strip dimensions / Pad material | Similar | | Packaging Configuration | • Aluminum vial • 50 Test Strips per vial • 2 Vials per box | • Polypropylene vial • 100 Test Strips per Vial • 2 Vials per box | Similar | K251048 May 1, 2025 {6} STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Rapicide PA High-Level Disinfectant Test Strips | Table 1. Device Comparison Table | | | | | --- | --- | --- | --- | | Feature | Predicate Device K152394 | Rapicide PA HLD Test Strips (Proposed device) | Comparison | | Direct or indirect body contact | No direct or indirect contact with patients or users | No direct or indirect contact with patients or users | Identical | | Sterile | Non-Sterile | Non-Sterile | Identical | | Shelf Life | 18 Months | 18 Months | Identical | | Open Bottle Shelf Life | 4 months | 4 months | Identical | | Biocompatibility | No patient or user contact | No patient of user contact | Identical | The Subject device is similar to the predicate device. It has the same intended use, scientific technology, and principles of operation. The subject device has the same endpoint specification, the same Minimum Recommended Concentration (MRC), and stability profile. The Subject device has a different vial package material and increased the number of strips per vial. Additionally, the Subject device has increased the strip length, modified the adhesive type, and uses a different chromatography paper supplier. ## 3. Description of Device Rapicide PA High-Level Disinfectant Test Strip has the ability to measure the disinfectant use solution concentration above 850 ppm Peracetic Acid (PAA). This is the minimum recommended concentration (MRC) of PAA for Rapicide PA high-level disinfectant. If the solution is at or below MRC, the test strip pad will indicate a failure by turning dark grey, blue grey, violet grey, light grey, or white (no change in color). A passing result will be indicated by the test strip pad turning solid black after contact with solution at concentrations of 1100 ppm ± 20 ppm PAA. ## 4. Intended Use/Indications for Use The RAPICIDE™ PA Test Strips are chemical indicators used after the disinfection cycle to ensure that the RAPICIDE™ PA High-Level Disinfectant solution is met the minimum recommended concentration of 850 ppm peracetic acid. The RAPICIDE PA Test Strips are designed for use by trained personnel who reprocess endoscopes and their accessories in facilities where endoscopies are performed. K251048 May 1, 2025 Page 3 of 4 {7} STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Rapicide PA High-Level Disinfectant Test Strips 5. Summary of Nonclinical Tests The Subject Device has the same intended use and technological characteristics that do not raise different questions of safety and effectiveness as compared to the predicate device. Testing to assess and demonstrate performance of the subject device is summarized below. | Test | Conclusion | | --- | --- | | Dynamic Range | Met Acceptance Criteria | | Comparative Sensitivity and Specificity | Met Acceptance Criteria | | Analytic Specificity – Contaminants | Met Acceptance Criteria | | Analytic Specificity – Other Germicides | Met Acceptance Criteria | | Shelf Life | Met Acceptance Criteria | | In-Use (Open Bottle) Stability | Met Acceptance Criteria | 6. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed device and is as safe, as effective and performs at least as well as the predicate device K152394, Class II (21 CFR 880.2800), product code JOJ. K251048 May 1, 2025 Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...