Digital Radiographic Imaging Acquisition Software - DR (RiasDR)

K251038 · Shen Zhen Cambridge-Hit Co., Ltd. · LLZ · Aug 6, 2025 · Radiology

Device Facts

Record IDK251038
Device NameDigital Radiographic Imaging Acquisition Software - DR (RiasDR)
ApplicantShen Zhen Cambridge-Hit Co., Ltd.
Product CodeLLZ · Radiology
Decision DateAug 6, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The RiaspDR software directly controls and acquires general radiographic images of human anatomy (excluding fluoroscopic, angiographic, dental and mammographic applications). The RiaspDR software is designed to work with X-ray images from the Mars1417X detector (K210316).

Device Story

RIASPDR is a radiographic imaging acquisition software platform. It interfaces with the Mars1417X digital X-ray detector to control image acquisition and process radiographic data. The software supports image viewing, search, storage, annotation, measurement, and processing. It operates via TCP/IP or Wi-Fi to communicate with the detector and supports DICOM 3.0 for data transmission to PACS. Used in clinical settings, the software allows healthcare providers to acquire and review images to assist in clinical decision-making. The system provides automated measurement tools for length, area, perimeter, and angle, with defined accuracy tolerances.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical software verification and validation testing, which confirmed that the device meets all predetermined acceptance criteria and complies with relevant FDA guidance and international standards.

Technological Characteristics

Software-based radiographic imaging acquisition platform. Connectivity: TCP/IP or Wi-Fi for detector communication; DICOM 3.0 for PACS integration. Measurement accuracy: length ±5%, area ±10%, perimeter ±5%, angle ±0.5°. Operates as a software platform controlling external digital X-ray detectors.

Indications for Use

Indicated for acquisition of general radiographic images of human anatomy. Excludes fluoroscopic, angiographic, dental, and mammographic applications. Designed for use with the Mars1417X detector.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 6, 2025 Shen Zhen Cambridge-hit Co., Ltd. % Fu Field Senior Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue Xili Town, Nanshan District Shenzhen, GD 518000 China Re: K251038 Trade/Device Name: Digital Radiographic Imaging Acquisition Software - DR (RiasDR) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: July 7, 2025 Received: July 7, 2025 Dear Fu Field: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251038 - Fu Field Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K251038 - Fu Field Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K251038 Device Name Digital Radiographic Imaging Acquisition Software - DR (RiasDR) Indications for Use (Describe) The RiaspDR software directly controls and acquires general radiographic images of human anatomy (excluding fluoroscopic, angiographic, dental and mammographic applications). The RiaspDR software is designed to work with X-ray images from the Mars1417X detector (K210316). Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K251038 Shen Zhen Cambridge-hit [510(k)] Application # 510(k) Summary ## 1. Contact Details ### 1.1 Applicant information | Applicant Name | Shen Zhen Cambridge-hit Co., Ltd. | | --- | --- | | Address | 1st Floor, Building A1, Hualianfeng Building, Huaning Road,Dalang Street, Longhua District, Guangdong Province, China. | | Phone No. | +86(755)23226094 | | Contact person | Ming Zhao | | Date Prepared | Apr. 7, 2025 | | E-mail | ming.zhao@iraygroup.com | ### 1.2 Proposed device manufacturer | Applicant Name | iRay Group | | --- | --- | | Address | No. 999, Huanqiao Road, Pudong New Area 201315, Shanghai, China. | ### 1.3 Submission Correspondent | 电话天键 | Shenzhen Joyantech Consulting Co., Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District, Shenzhen, Guangdong, 518000, China | | --- | --- | | Phone No. | +86-755-86069197 | | Contact person | Field Fu | | Contact person's e-mail | field@cefda.com | | Website | http://www.cefda.com | ## 2. Device information | Trade name | Digital Radiographic Imaging Acquisition Software Platform – DR (RIASPDR) | | --- | --- | | Model | Not applicable | | Classification | II | | Classification name | Radiology | | Product code | LLZ | | Regulation No. | 892.2050 | ## 3. Legally Marketed Predicate Device | Trade Name | Econsole1 | | --- | --- | | 510(k) Number | K152172 | {5} Shen Zhen Cambridge-hit [510(k)] Application | Product Code | LLZ | | --- | --- | | Manufacturer | DRTECH Corporation | ## 4. Device Description RIASPDR is Radiographic Imaging Acquisition Software Platform. RIASPDR software directly controls and acquires images from Mars1417X detector(K210316) whose manufacturer is iRay Technology. Furthermore, RIASPDR acquires and processes images. In addition, it complies with DICOM standards and is able to transmit and receive data with the PACS system. ## 5. Device specification The deviation of the length measurement should not exceed 5%; The deviation of the area measurement should not exceed 10%; The deviation of the perimeter measurement should not exceed 5%; The deviation of the angle indication value from the actual value should be within ±0.5°. ## 6. Intended Use/Indication for Use The RiaspDR software directly controls and acquires general radiographic images of human anatomy (excluding fluoroscopic, angiographic, dental and mammographic applications). The RiaspDR software is designed to work with X-ray images from the Mars1417X detector (K210316). ## 7. Substantial Equivalence Comparison | Item | Subject Device | Predicate Device: (K152172) | Comments | | --- | --- | --- | --- | | Regulation number | 892.2050 | 892.2050 | Same | | Classification | II | II | Same | | Product Code | LLZ | LLZ | Same | | Intended use/Indications for use | The RiaspDR software directly controls and acquires general radiographic images of human anatomy (excluding fluoroscopic, angiographic, dental and mammographic applications). The RiaspDR software is designed to work with X-ray images from the Mars1417X detector | The Econsole1 software is indicated for use in general radiographic images of human anatomy (excluding fluoroscopic, angiographic, and mammographic applications). | Same | {6} Shen Zhen Cambridge-hit [510(k)] Application | Item | Subject Device | Predicate Device: (K152172) | Comments | | --- | --- | --- | --- | | | (K210316). | | | | Acquisition devices | Digital X-ray Detector | Digital X-ray Detector | Same | | Software Function | Image viewing; Image search; Image storage; Image annotation; Image measurement; Image processing; | Image viewing; Image search; Image storage; Image annotation; Image measurement; Image processing; Image stitch | Same | | DICOM 3.0 Compatibility | Yes | Yes | Same | | Wireless function | Yes, by TCP/IP or WIFI. The software interacts with the Flat Panel Detector (FPD, that is, Digital X-ray Detector) by calling the driver provided by the manufacturer of the FPD. The data interaction between the FPD and the software can be wired or wireless, depending on the Design of the FPD. | No | Different (note) | # 8. Non-clinical Testing As required by FDA Guidance, Software verification and validation testing were conducted and documentation was provided in this 510(k). Results demonstrated that the predetermined acceptance criteria were met. Software Verification and Validation Testing was performed in accordance with internal requirements, international standards and guidance shown below, the safety and effectiveness of RIASPDR were supported, and the substantial equivalence to the predicate device was demonstrated: --- Content of Premarket Submissions for Device Software Functions, issued on June 14, 2023. ---Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued on September 27, 2023. {7} Shen Zhen Cambridge-hit [510(k)] Application ---Digital Imaging and Communications in Medicine (DICOM) Set. ## 9. Clinical testing Not applicable. ## 10. Conclusions Drawn from Non-Clinical and Clinical Tests Based on device comparison information and non-clinical bench testing, the subject device is substantially equivalent to legally marketed predicate devices (K152172).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...