K250974 · Bio Compression Systems, Inc. · JOW · Dec 12, 2025 · Cardiovascular
Device Facts
Record ID
K250974
Device Name
Bio Arterial Deluxe (IC-BAP-DX)
Applicant
Bio Compression Systems, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Dec 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bio Arterial Deluxe Arterial Blood Flow Enhancement System is intended: • For the improvement of blood circulation in the upper and lower extremities to help prevent and reduce complications of poor circulation • For treatment of patients with intermittent claudication, rest pain, diabetic foot, ischemic neuritis, arterial ulcers, gangrene, poor runoff • As an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes: amputations (minor), angioplasty/stent failure, arteriopathic wounds, graft failure, intermittent claudication, ischemia, night pain, rest pain, small vessel disease, ulcers
Device Story
Pneumatic compression device consisting of pump and inflatable garment; pump cyclically inflates garment chambers in distal-to-proximal sequence to stimulate blood flow. Core components: motor, air compressor, disc valves, micro switch. Motor-driven rotating disc valve directs airflow; micro switch triggers cycle. Controlled via softkeys and LED display. Used in clinical or home settings by patients or caregivers. Provides adjustable pressure, compliance monitoring, and timed treatment. Benefits patient by enhancing arterial blood flow in extremities, aiding management of ischemic conditions and ulcers.
Clinical Evidence
No clinical or animal studies were submitted or relied upon. Substantial equivalence is supported by bench testing, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1 series), EMC, and functional verification (cycle time, pressure accuracy, and mode operation).
Technological Characteristics
Pneumatic compression pump with motor, air compressor, and disc valve assembly. Garments: 200 Denier Nylon Oxford coated with 3 mil Polyurethane. Power: 120 VAC, 60 Hz. Class II, Type BF applied part, IP21 ingress protection. Cycle time: 20s (5s compression, 15s non-compression). Pressure range: 20-150 mmHg (±20% accuracy). Software safety class A (IEC 62304).
Indications for Use
Indicated for patients requiring improved blood circulation in upper/lower extremities to prevent/reduce complications of poor circulation, or treatment of intermittent claudication, rest pain, diabetic foot, ischemic neuritis, arterial ulcers, gangrene, poor runoff, and ischemic disease of lower extremities. Contraindicated for patients with open/freshly healed ulcers or fragile skin in treatment areas, limb infections (cellulitis) without antibiotics, DVT, inflammatory phlebitis, pulmonary embolism, uncontrolled CHF, undesirable venous/lymphatic return, limb sepsis, recent skin grafts, pulmonary edema, or acute thrombophlebitis.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
IC-BAP-DL Bio Arterial Plus Arterial Blood Flow Enhancement System (K131327)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
Bio Compression Systems, Inc.
Marc Somelofski
Vice President of RA/QA
120 West Commercial Ave
Moonachie, New Jersey 07074
Re: K250974
Trade/Device Name: Bio Arterial Deluxe (IC-BAP-DX)
Regulation Number: 21 CFR 870.5800
Regulation Name: Compressible Limb Sleeve
Regulatory Class: Class II
Product Code: JOW
Dated: November 12, 2025
Received: November 12, 2025
Dear Marc Somelofski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250974 - Marc Somelofski
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250974 - Marc Somelofski
Page 3
Sincerely,
Nicole M. Gillette -S
Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Bio Arterial Deluxe
Page 9 of 39
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250974 | ? |
| Please provide the device trade name(s). | | ? |
| Bio Arterial Deluxe (IC-BAP-DX) | | |
| Please provide your Indications for Use below. | | ? |
| The Bio Arterial Deluxe Arterial Blood Flow Enhancement System is intended:
• For the improvement of blood circulation in the upper and lower extremities to help prevent and reduce complications of poor circulation
• For treatment of patients with intermittent claudication, rest pain, diabetic foot, ischemic neuritis, arterial ulcers, gangrene, poor runoff
• As an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes: amputations (minor), angioplasty/stent failure, arteriopathic wounds, graft failure, intermittent claudication, ischemia, night pain, rest pain, small vessel disease, ulcers | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
BIO BIO COMPRESSION SYSTEMS inc.
120 West Commercial Avenue, Moonachie, New Jersey 07074
# Summary
Date Prepared: December 11, 2025
# I. SUBMITTER
Bio Compression Systems, Inc.
120 West Commercial Avenue
Moonachie, NJ 07074, USA
Phone: +1-201-939-0716
Contact Person: Marc L. Somelofski
# II. DEVICE
| Device Name | Bio Arterial Deluxe (IC-BAP-DX) |
| --- | --- |
| Trade/Proprietary Name | Bio Arterial Deluxe (IC-BAP-DX) |
| Common/Usual Name | IC-BAP-DX Bio Arterial Deluxe |
| Classification Name | Compressible Limb Sleeve (21 CFR 870.5800) |
| Product Code | JOW |
| Class | Class II |
# III. PREDICATE DEVICE
IC-BAP-DL Bio Arterial Plus Arterial Blood Flow Enhancement System (K131327).
Reference devices K024019, K072666, K131146, K210417, and K240499 are mentioned.
Summary
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{5}
BIO BIO COMPRESSION SYSTEMS inc.
120 West Commercial Avenue, Moonachie, New Jersey 07074
# IV. DEVICE DESCRIPTION
Bio Compression Systems' Bio Arterial Deluxe is a pneumatic compression device which consists of a pneumatic compression sleeve ("garment") connected to a pneumatic compression pump ("pump"). The pump cyclically inflates the garment's sections ("chambers") in sequence from the distal end toward the trunk of the body. The inflation of the garment compresses the area of the body on which it is worn, stimulating blood flow.
The core components of the pump are the motor, air compressor, disc valves, and micro switch. The air compressor generates air flow into a stationary disc valve. The motor moves a rotating disc valve. The geometry of the disc valve directs air flow and cyclically triggers the micro switch.
The pump is controlled by softkeys and an LED display.
The device uses the Predicate Device's garments.
Summary
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{6}
BIO
BIO
COMPRESSION
SYSTEMS inc.
120 West Commercial Avenue, Moonachie, New Jersey 07074
V. INDICATIONS FOR USE
| | Subject Device | Predicate Device (K131327) | Comparison |
| --- | --- | --- | --- |
| Indications for Use | The Bio Arterial Deluxe Arterial Blood Flow Enhancement System is intended:
• For the improvement of blood circulation in the upper and lower extremities to help prevent and reduce complications of poor circulation
• For treatment of patients with intermittent claudication, rest pain, diabetic foot, ischemic neuritis, arterial ulcers, gangrene, poor runoff
• As an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes: amputations (minor), angioplasty/stent failure, arteriopathic wounds, graft failure, intermittent claudication, ischemia, night pain, rest pain, small vessel disease, ulcers | The Bio Arterial Plus Arterial Blood Flow Enhancement System is intended as an adjunct therapy for patients with ischemic disease of the lower extremities, due to one or more of the following causes: amputations (minor), angioplasty/stent failure, arteriopathic wounds, graft failure, intermittent claudication, ischemia, night pain, rest pain, small vessel disease, ulcers | Equivalent.
The addition of the improvement of blood circulation in the upper and lower extremities to help prevent and reduce complications of poor circulation indication is not a new use; devices in the category with similar characteristics have been cleared for this indication (K024019).
The addition of the treatment of patients with intermittent claudication, rest pain, diabetic foot, ischemic neuritis, arterial ulcers, gangrene, poor runoff indication is not a new use; devices in the category with similar characteristics have been cleared for these indications (K131146, K240499). |
Summary
Page 3
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{7}
BIO
BIO
COMPRESSON
SYSTEMS inc.
120 West Commercial Avenue, Moonachie, New Jersey 07074
| | Subject Device | Predicate Device (K131327) | Comparison |
| --- | --- | --- | --- |
| Contra - indications | Use of this device is contraindicated for patients with any of the following conditions:
• Open or freshly healed ulcers or other wounds or otherwise fragile skin in the arms, core or full leg region
• Infections in the limb, including cellulitis, without appropriate antibiotic coverage
• Suspicion or confirmation of the presence of Deep Vein Thrombosis (DVT)
• Inflammatory phlebitis or episodes of pulmonary embolism
• Congestive Heart Failure (CHF) unless controlled by medication
• Undesirable venous and lymphatic return
• Sepsis in the limb or limbs
• Immediately following skin grafts in or around treatment sites
• Pulmonary edema
• Acute thrombophlebitis | Use of this device is contraindicated for patients with any of the following conditions:
• Infections in the limb, including cellulitis, without appropriate antibiotic coverage
• Suspicion or confirmation of the presence of Deep Vein Thrombosis (DVT)
• Inflammatory phlebitis or episodes of pulmonary embolism
• Congestive Heart Failure (CHF)
• Undesirable venous and lymphatic return
• Sepsis in the limb or limbs
• Immediately following skin grafts in or around treatment sites
• Pulmonary edema
• Acute thrombophlebitis | Equivalent.
Subject device adds open or freshly healed ulcers or other wounds or otherwise fragile skin in the core or full leg region. |
| Prescription/ over-the-counter use | Prescription | Prescription | Identical |
Summary
Page 4
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{8}
BIO BIO COMPRESSION SYSTEMS inc.
120 West Commercial Avenue, Moonachie, New Jersey 07074
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | Subject Device | Predicate Device (K131327) | Comparison |
| --- | --- | --- | --- |
| Operating Principle | Same as Predicate | Same as Predicate | Identical |
| Power Supply Rating | 120 VAC, 60 Hz | 100-240V, 50/60 Hz | Equivalent |
| Electrical Classification | Class II | Class II | Identical |
| Applied Part | Type BF | Type BF | Identical |
| Ingress Protection | IP21 | IP21 | Identical |
| Cycle Time | 20 seconds (compression 5 seconds, non-compression 15 seconds) | 20 seconds (compression 5 seconds, non-compression 15 seconds) | Identical |
| Pressure Range | 20-150 mmHg | 20-150 mmHg | Identical |
| Pressure Accuracy | ± 20% | ± 20% | Identical |
| Software Safety Class (IEC 62304) | A | A | Identical |
| Weight | 3.3 lbs. (1.5 kg) | 7.5 lbs. (3.4 kg) | Similar |
| Dimensions | 4.5" x 12" x 7.34" (114 mm x 304 mm x 186 mm) | 5.5" x 12" x 8" (140 mm x 305 mm x 203 mm) | Similar |
| Compression Sleeves ("Garments") | Uses Predicate's garments and adds arm garment | 200 Denier Nylon Oxford, coated with 3 mil of Polyurethane | Identical or identical material covering an additional anatomic areas (arm). This is the same anatomic areas as other cleared devices in this category with similar technological characteristics and indications (K024019) |
| Features | • Adjustable Pressure
• Compliance/Usage Meter
• Pause
• Timed Treatment | • Adjustable Pressure
• Compliance/Usage Meter | Minor differences in features |
The Subject Device is based upon the Predicate Device and has the same technological characteristics with respect to design, materials used, and construction.
- Subject Device and Predicate Device have the same operating principle and use the same disc valve assembly for operation
Summary
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{9}
BIO BIO COMPRESSION SYSTEMS inc.
120 West Commercial Avenue, Moonachie, New Jersey 07074
- Components used in Subject and Predicate Devices are identical, the same material, or have the same specifications
- The Predicate Device uses the Subject Device's compression garments
- Subject Device and Predicate Device have the same or similar performance specifications
The differences between the Subject Device and Predicate Device exist in similar cleared devices and do not raise any different questions of safety and/or effectiveness.
- The Subject Device has an additional garment model covering an additional anatomic area (arm). This is the same area as other cleared devices in this category with similar technological characteristics and the same indications (K024019).
## VII. PERFORMANCE DATA
The following testing and performance data were provided in support of the substantial equivalence determination.
### Electrical safety and electromagnetic compatibility (EMC)
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012 and A2:2021
- ANSI/AAMI HA60601-1-11:2015-08 and A1:2021
- IEC 60601-1:2005/AMD1:2012 and A2:2020
- IEC 60601-1-2:2014 and A1:2020
- IEC 60601-1-6:2010/AMD1:2013 and A2:2020
- IEC 60601-1-11:2015 and A1:2020
### Predicate Device routine acceptance tests conducted on Subject Device
- Observation of continuous and timed operation
- Pressure testing
- HiPot (dielectric withstand test) testing
### Functional verification and validation testing
- Cycle time verification and validation
- Treatment time verification and validation
- Pressure setting endpoint testing
- Operation to confirm all modes, settings, and mode/setting changes function as intended
Summary
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
{10}
BIO
BIO
COMPRESSION
SYSTEMS INC.
120 West Commercial Avenue, Moonachie, New Jersey 07074
## Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". The software documentation level for this device is basic and there are no cybersecurity risks.
## Clinical/Animal Studies
Clinical/animal studies were not submitted, referenced, or relied on in this premarket notification submission for a determination of substantial equivalence.
## VIII. CONCLUSION
The data included in this submission demonstrates that the Subject Device is substantially equivalent to the legally marketed Predicate Device and performs comparably to the Predicate Device that is currently marketed for the same intended use.
Summary
Page 7
201-939-0716 ■ 800-888-0908 ■ Fax: 201-939-4503
E-Mail BioSystems@biocompression.com ■ www.biocompression.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.