Suction Valve (MAJ-1443); Air/Water Valve (MAJ-1444)

K250949 · Olympus Medical Systems Corporation · ODC · Jul 16, 2025 · Gastroenterology, Urology

Device Facts

Record IDK250949
Device NameSuction Valve (MAJ-1443); Air/Water Valve (MAJ-1444)
ApplicantOlympus Medical Systems Corporation
Product CodeODC · Gastroenterology, Urology
Decision DateJul 16, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The AIR/WATER VALVE MAJ-1444 and SUCTION VALVE MAJ-1443 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract, as follows. Air/Water Valve (MAJ-1444) - To feed air to remove any fluids or debris adhering to the objective lens. - To feed water for lens washing. - To fill the balloon with sterile water. Suction Valve (MAJ-1443) - To remove any fluids, debris, or air from the patient. - To remove the water from the balloon.

Device Story

Reusable suction and air/water valves designed for Olympus GI ultrasound endoscopes; manual actuation by clinician during endoscopic procedures. Suction Valve (MAJ-1443) attaches to endoscope suction cylinder to remove fluids, debris, or air from patient and deflate balloon. Air/Water Valve (MAJ-1444) attaches to air/water cylinder to clear lens via air/water feed and inflate balloon with sterile water. Devices facilitate visualization and therapeutic access; reusable design requires manual cleaning, disinfection, and autoclaving. Output is mechanical control of endoscope fluid/air channels; benefits include maintained clear visualization and balloon-assisted ultrasound imaging.

Clinical Evidence

Bench testing only. Evaluated endoscope compatibility, suction/flow rates, leakage, depression force, composite durability, and microbiological backflow prevention. All samples met pre-defined acceptance criteria.

Technological Characteristics

Reusable endoscope channel accessories. Manual actuation. Materials compatible with manual cleaning, disinfection, and autoclaving. No electronic components; purely mechanical fluid/air control valves.

Indications for Use

Indicated for use with Olympus ultrasound endoscopes in the gastrointestinal tract to control suction, air/water delivery for lens cleaning, and balloon inflation/deflation.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 16, 2025 Olympus Medical Systems Corporation % Susan Lewandowski Program Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway Center Valley, Pennsylvania 18034 Re: K250949 Trade/Device Name: Suction Valve (MAJ-1443); Air/Water Valve (MAJ-1444) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODC Dated: March 28, 2025 Received: March 28, 2025 Dear Susan Lewandowski: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250949 - Susan Lewandowski Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250949 - Susan Lewandowski Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, SIVAKAMI VENKATACHALAM -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Suction Valve Page 9 of 35 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250949 | ? | | Please provide the device trade name(s). | | ? | | Suction Valve (MAJ-1443); Air/Water Valve (MAJ-1444) | | | | Please provide your Indications for Use below. | | ? | | The AIR/WATER VALVE MAJ-1444 and SUCTION VALVE MAJ-1443 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract, as follows. | | | | Air/Water Valve (MAJ-1444) - To feed air to remove any fluids or debris adhering to the objective lens. - To feed water for lens washing. - To fill the balloon with sterile water. | | | | Suction Valve (MAJ-1443) - To remove any fluids, debris, or air from the patient. - To remove the water from the balloon. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | {4} OLYMPUS K250949 Page 1 of 4 Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 # 510(k) Summary ## 1. General Information Date Prepared: March 28, 2025 510(K) submitter: Olympus Medical Systems Corporation 2951 Ishikawa-cho, Hachioji-shi, Tokyo Japan 192-8507 Contact: Osamu Tamada Product RA Lead GI Endoscope, OMSC Correspondent: Olympus Surgical Technologies of America 800 West Park Drive, Westborough, MA 01581 Primary Contact: Susan Lewandowski Email: susan.lewandowski@olympus.com ## 2. Device Information Device Name: Suction Valve MAJ-1443, Air/Water Valve MAJ-1444 Common Name: Endoscope channel accessory Classification: 876.1500 – Endoscope and accessories Regulatory Class: II Product Code: ODC (endoscope channel accessory) Device Panel: Gastroenterology & Urology ## 3. Predicate Device Information BioGuard EUS Air/Water and Suction Valves K202104 ## 4. Device Description The Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract. The subject devices are reusable. The subject devices are compatible with Olympus endoscopes GF-UCT180 and GF-UE160-AL5. The Suction Valve MAJ-1443 is attached to the suction cylinder of a compatible endoscope to remove any fluids, debris, or air from the patient and to remove water from the balloon. Suction Valve MAJ-1443 has no patient-contacting components and is a reusable device. The Air/Water Valve MAJ-1444 is attached to the air/water cylinder of a compatible endoscope to feed air to remove any fluid or debris adhering to the objective lens, to feed water for lens Page 1 of 4 {5} OLYMPUS K250949 Page 2 of 4 Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 washing, and to fill the balloon with sterile water. MAJ-1444 has indirect patient-contacting components and is a reusable device. The subject device has the same technological characteristics and similar design as the applicable predicate device. # 5. Indications for Use The Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract, as follows: Suction Valve MAJ-1443 - To remove any fluids, debris, or air from the patient - To remove the water from the balloon Air/Water Valve MAJ-1444 - To feed air to remove any fluid or debris adhering to the objective lens - To feed water for lens washing - To fill the balloon with sterile water # 6. Predicate Comparison | Description | Subject Device (SD) Suction Valve MAJ-1443 Air/Water Valve MAJ-1444 | Predicate Device (PD) BioGuard EUS Air/Water and Suction Valves K202104 | Comparison | | --- | --- | --- | --- | | Product Code | ODC | ODC, FDF | SAME The predicate device additionally has the assigned FDA product code FDF (colonoscope and accessories, flexible/rigid) which the subject device does not have. | | Classification | II | II | SAME | | Regulation No. | 21 CFR 876.1500 | 21 CFR 876.1500 | SAME | | Common Name | Endoscope channel accessory | Endoscope channel accessory | SAME | | Regulation Name | Endoscope and accessories | Endoscope and accessories | SAME | | Review Panel | Gastroenterology and Urology | Gastroenterology and Urology | SAME | | Indications for Use | The Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 have been designed for use with an Olympus ultrasound endoscope for the gastrointestinal (GI) tract, as follows: | The BioGuard EUS Suction Valve is intended to be used to control the suction function of an endoscope during a GI endoscopic procedure. | SAME The SD and PD have the same intended use/indications for use but are phrased differently. The SD indications statement specifically details the | Page 2 of 4 {6} K250949 Page 3 of 4 # OLYMPUS Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 | Description | Subject Device (SD) Suction Valve MAJ-1443 Air/Water Valve MAJ-1444 | Predicate Device (PD) BioGuard EUS Air/Water and Suction Valves K202104 | Comparison | | --- | --- | --- | --- | | | Suction Valve MAJ-1443 • To remove any fluids, debris, or air from the patient • To remove the water from the balloon Air/Water Valve MAJ-1444 • To feed air to remove any fluid or debris adhering to the objective lens • To feed water for lens washing • To fill the balloon with sterile water | The BioGuard EUS Air/Water Valve is intended to be used to control the air/water function of an endoscope during a GI endoscopic procedure. | expected use/function of the valves. | | Compatible Endoscopes | Olympus GI ultrasound endoscope (with a balloon channel) | Olympus GI ultrasound endoscope (with a balloon channel) | SAME The PD has been designed to be used with an Olympus GI ultrasound endoscope (with a balloon channel) | | Single Use | No | Yes | DIFFERENT Performance testing was conducted to demonstrate that this difference in technology does not impact the safety or effectiveness of the SD. | | Function | Suction valve • remove any fluids, debris, or air from the patient. • remove the water from the balloon. Air/water valve • feed air to remove any fluid or debris adhering to the objective lens. • feed water for lens washing. • fill the balloon with sterile water. | Suction valve • remove any fluid or debris adhering to the objective lens. • remove the water from the balloon. Air/water valve • feed air to remove any fluid or debris adhering to the objective lens. • feed water for lens washing. • fill the balloon with sterile water. | SAME | | Reprocessing Method | - Manual cleaning - Manual disinfection - Autoclaving | Not applicable as the PD is single use | DIFFERENT Performance testing was conducted to demonstrate | Page 3 of 4 {7} OLYMPUS K250949 Page 4 of 4 Suction Valve MAJ-1443 and Air/Water Valve MAJ-1444 | Description | Subject Device (SD) Suction Valve MAJ-1443 Air/Water Valve MAJ-1444 | Predicate Device (PD) BioGuard EUS Air/Water and Suction Valves K202104 | Comparison | | --- | --- | --- | --- | | | | | that this difference in technology does not impact the safety or effectiveness of the SD. | | Method of application | Manual actuation | Manual actuation | SAME | # 7. Non-Clinical/Clinical Tests Summary and Conclusion Olympus performed bench testing to demonstrate the functionality of the Suction Valve MAJ-1443 and the Air/Water Valve MAJ-1444. Test samples were final, finished devices subjected to the full manufacturing process. The following performance bench tests were conducted. All test samples passed pre-defined acceptance criteria. Suction Valve MAJ-1443 - Endoscope compatibility, Suction rate, Balloon suction rate, Leakage, Depression Force, and Composite Durability Air/Water Valve MAJ-1444 - Endoscope compatibility, Air flow rate, Water flow rate, Balloon water rate, Leakage, Depression Force, Composite Durability, and Microbiological Evaluation of Backflow Prevention # 8. Conclusion Based on the comparison to the predicate device, intended use, technological characteristics, and performance testing, the Suction Valve MAJ-1443 and the Air/Water Valve MAJ-1444 raise no new issues of safety and effectiveness and are substantially equivalent to the predicate devices in terms of safety, efficacy, and performance. Page 4 of 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...