K250900 · EvoEndo, Inc. · FDS · Jun 16, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K250900
Device Name
EvoEndo Single-Use Endoscopy System
Applicant
EvoEndo, Inc.
Product Code
FDS · Gastroenterology, Urology
Decision Date
Jun 16, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K250900 · Jun 16, 2025
EvoEndo Single-Use Endoscopy System
EvoEndo, Inc.
Published clinical literature
A clinical literature review was conducted to evaluate the safety of endoscopy procedures in patients under 5 years of age and to assess anatomical considerations for using the EvoEndo Endoscopy System in this smaller patient population. This evidence supports the removal of the age restriction (5 years and above) from the device's indications for use.
Literature review; Pediatric safety; Indication expansion
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical literature review; Literature review
Patients under 5 years of age
Not applicable for this study
Safety of endoscopy procedures and anatomical considerations for use in pediatric patients under 5 years
Indications for Use
The EvoEndo Model LE Gastrocope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The gastroscope is a sterile single-use device and can be inserted orally or transnasally. The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation.
Device Story
System consists of sterile, single-use gastroscope and controller; gastroscope features HD camera at distal end, 4-way steering, and air/water/suction channels; controller processes video/images from gastroscope for display on external medical-grade monitor via HDMI; used in clinical settings by physicians for upper GI endoscopy; provides real-time visualization to assist in diagnosis and treatment; benefits include single-use sterility, eliminating reprocessing requirements and reducing cross-contamination risk.
Clinical Evidence
No new clinical trials performed. Evidence consists of a clinical literature review evaluating safety of endoscopy in patients under 5 years and anatomical considerations. Previous usability testing (benchtop and animal models) and biocompatibility testing (ISO 10993-1) from the predicate device (K213606) are leveraged to support safety and effectiveness.
Technological Characteristics
Single-use gastroscope (3.5mm OD, 2.0mm working channel, 1100mm length) and aluminum controller. Illumination via integrated light source. 4-way steering (Up 210°, Down 90°, Left/Right 100°). EO sterilization. Connectivity via HDMI/USB. Electrical safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Photobiological safety per IEC/TR 62471.
Indications for Use
Indicated for visualization of the upper digestive tract (esophagus, stomach, duodenal bulb) in adult and pediatric patients for observation, diagnosis, and endoscopic treatment. Can be inserted orally or transnasally.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 16, 2025
EvoEndo, Inc.
% Keira Jessop
Regulatory Affairs Consultant
AlvaMed, Inc.
935 Great Plain Avenue
Unit 166
Needham, Massachusetts 02492
Re: K250900
Trade/Device Name: EvoEndo Single-Use Endoscopy System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: Class II
Product Code: FDS, FET
Dated: March 24, 2025
Received: March 25, 2025
Dear Keira Jessop:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250900 - Keira Jessop
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K250900 - Keira Jessop
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# SIVAKAMI VENKATACHALAM -S
for
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250900 | |
| Device Name EvoEndo Single-Use Endoscopy System | |
| Indications for Use (Describe) The EvoEndo Model LE Gastroscope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The gastroscope is a sterile single-use device and can be inserted orally or transnasally. The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K250900 Page 1 of 7
510(k) Summary
K250900
1.1 Name and Address of Submitter
Paul Imaoka
EvoEndo, Inc.
888 East Belvidere Road
Suite 212
Grayslake, IL 60030
(303) 223-7445
pimaoka@evoendo.com
1.2 Correspondent/Primary Contact Person
Keira Jessop, M.S., RAC
Regulatory Affairs Consultant
AlvaMed, Inc.
935 Great Plain Avenue, Unit 166
Needham, MA 02492 USA
kjessop@alvamed.com
Phone: +1 (888) 331-3485 Fax: +1 (617) 249-0955
1.3 Submission Information
Date Summary Prepared: May 7, 2025
Subject Device:
Trade/Device Name: EvoEndo Single-Use Endoscopy System
Manufacturer: EvoEndo, Inc.
Common Name: Gastrocope and accessories, flexible/rigid
Endoscopic video imaging system/
Component, Gastroenterology-Urology
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulation Class: Class II
Product Code: FDS, FET
Review Panel: Gastroenterology/Urology
{6}
K250900 Page 2 of 7
| Predicate Device: | Clearance: | K213606 |
| --- | --- | --- |
| | Trade/Device Name: | EvoEndo Single-Use Endoscopy System |
| | Manufacturer: | EvoEndo, Inc. |
| | Regulation Number: | 21 CFR 876.1500 |
| | Regulation Name: | Endoscope and accessories |
| | Regulation Class | Class II |
## Valid Predicate Discussion
EvoEndo Single-Use Endoscopy System was selected as the valid predicate device used to support the 510(k) submission because it was cleared using well-established methods and this submission is to extend the indicated age from the previously submitted device. The only differences between the subject and predicate device are the indicated age and the shelf life. After conducting a search on the Manufacturer and User Facility Device Experience (MAUDE) Database, Medical Device Reporting (MDR), MedSun Reports Database, Medical Device Safety and CBER Safety & Availability (Biologics) and Recall Database websites it was confirmed that there are no known unmitigated use-related or design safety issues and the predicate device has not been subject to a design-related recall.
Table 1: Valid Predicate Device
| Valid Predicate Device | A - Well established methods | B - Meets or exceeds expected predicate performance | C - Unmitigated use-related or design-related safety issues | D – Associated design-related recall |
| --- | --- | --- | --- | --- |
| EvoEndo Single-Use Endoscopy System | Used relevant methods that were published in the public domain. | History of Safe use, established due to duration of device on the market | No known unmitigated use-related or design related safety issues | No design-related recall identified |
{7}
K250900 Page 3 of 7
## 1.4 Device Description
The EvoEndo Endoscopy System is comprised of two regulated components:
- EvoEndo Model LE Single-Use Gastrocope (hereafter referred to as the EvoEndo Endoscope)
- EvoEndo Controller (hereafter referred to as the Controller)
The EvoEndo Endoscope is a sterile, single-use gastroscope intended to perform oral or transnasal diagnostic endoscopy in adult and pediatric patients. The EvoEndo Endoscope is ethylene oxide (EO) sterilized and is comprised of:
- Handle
- Umbilical Bundle that includes air, water, and suction lines, as well as the video connector
- Endoscope shaft with HD Camera
The Controller of the EvoEndo Endoscopy System translates the images or video captured by the camera at the distal end of the EvoEndo Endoscope to a monitor via an HDMI cable.
## 1.5 Indications for Use
The EvoEndo Model LE Gastrocope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The gastroscope is a sterile single-use device and can be inserted orally or transnasally.
The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation.
## 1.6 Summary of Technological Characteristics
The following table provides an overview of general technological characteristics in comparison to the predicate device.
{8}
K250900 Page 4 of 7
Table 2: Comparison of Predicate Device and Subject Device
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) Number | K250900 | K213606 |
| Sponsor | EvoEndo, Inc | EvoEndo, Inc |
| Device Name | EvoEndo Endoscopy System | EvoEndo Endoscopy System |
| Product Code | FDS, FET | FDS, FET |
| Indications for Use | The EvoEndo Model LE Gastroscope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The gastroscope is a sterile single-use device and can be inserted orally or transnasally.The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation. | The EvoEndo Model LE Gastroscope is intended for the visualization of the upper digestive tract in adults and pediatric patients, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb in patients over the age of five years. The gastroscope is a sterile single-use device and can be inserted orally or transnasally.The EvoEndo Controller is intended for use with an EvoEndo Endoscope for endoscopic diagnosis, treatment, and video observation. |
| Sterilization Method | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
| Shelf Life | 14 months | 6 months |
| Endoscope Outer Diameter | 3.5mm | 3.5mm |
| Working Channel Diameter | 2.0mm | 2.0mm |
| Working Length | 1100mm | 1100mm |
| Field of View | 120° | 120° |
| Steering Method | 4-way | 4-way |
| Maximum Bending Angle | Up 210°Down 90°Left 100°Right 100° | Up 210°Down 90°Left 100°Right 100° |
{9}
K250900 Page 5 of 7
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) Number | K250900 | K213606 |
| Air, Water, Suction (AWS) Functionality | Yes | Yes |
| Video Inputs | Ferric HDMI Cable & USB 3.0 Cable | Ferric HDMI Cable & USB 3.0 Cable |
| Monitor Compatibility | Medical grade monitor (27-inch, 1080p HD resolution recommended) | Medical grade monitor (27-inch, 1080p HD resolution recommended) |
| Illumination Technology | Located in EvoEndo Endoscope | Located in EvoEndo Endoscope |
| Electrical Specifications Input Voltage | 100 – 240 VAC, 50-60 Hz 0.5A | 100 – 240 VAC, 50-60 Hz 0.5A |
| U.S. Power Cord Specifications | 115 VAC Length – 1.5 meters Voltage Rating – 240 VAC Current Rating – 12 A Connector Type – Horizontal | 115 VAC Length – 1.5 meters Voltage Rating – 240 VAC Current Rating – 12 A Connector Type – Horizontal |
| Controller Physical Specifications | | |
| Material | Aluminum | Aluminum |
| Height | 6.0 cm | 6.0 cm |
| Weight | 15.0 cm | 15.0 cm |
| Depth | 21.5 cm | 21.5 cm |
## 1.7 Performance Testing Summary
This submission is specific to the indication extension. There are no changes to the design of the device as compared to K213606. No testing was performed specific to this submission. Please refer to K213606 for performance testing associated with this device. To support the removal of "in patients 5 years and above" from the indication a clinical literature search was conducted.
## Clinical Testing
Published literature was evaluated for the safety of endoscopy procedures in patients under 5 years as well as review anatomical considerations of using the EvoEndo Endoscopy system in the smaller patient population. A risk/benefit evaluation was also completed. The evaluation supported the removal of the specific age of 5 years and above.
## Summary of Previous Biocompatibility Testing:
Cytotoxicity, sensitization, irritation, acute systemic toxicity, and material- mediated pyrogenicity testing was performed per ISO 10993-1 and the FDA Guidance for Industry and
{10}
K250900 Page 6 of 7
FDA Staff, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.'" There were no cytotoxicity, sensitization, irritation, acute systemic toxicity, or material-mediated pyrogenicity concerns associated with the EvoEndo Endoscopy System.
## Summary of Previous Non-Clinical Functional and Performance Testing
Visual, dimensional, functional, and transit simulation verification testing was performed on final sterilized EvoEndo Endoscopes. In functional testing, devices were verified for camera and video control unit functioning; tip deflection; air, water, and suction flow rates; functionality with accessory device; functionality after $360^{\circ}$ bend shaft orientation; functionality with simulated repeated use; and post simulated use functionality. The maximum bending angle of the device in each direction is shown in the table below. The deflection achieved with the EvoEndo Endoscope is consistent with that of the identified predicate device in all directions, indicating substantial equivalence.
Table 3: Deflection of EvoEndo Endoscope
| Deflection Direction | Maximum Bending Angle |
| --- | --- |
| Up | 210° |
| Down | 90° |
| Left | 100° |
| Right | 100° |
In transit simulation testing, test units underwent environmental conditioning, distribution simulation, and packaging testing to confirm the integrity of the sterile barrier and functionality of the device under expected transportation conditions. The EvoEndo Endoscopy System met acceptance criteria set for all verification testing procedures.
Optical performance testing was performed on the EvoEndo Endoscope and compared to the predicate device. The testing evaluated the resolution, depth of field, geometric distortion, image intensity uniformity, and color performance of the endoscope. The test results suggest equivalent performance across all metrics assessed, demonstrating substantial equivalence between the EvoEndo Endoscope and the predicate device.
Photobiological safety testing was performed per FDA recognized consensus standard IEC/TR 62471 First Edition 2006-07: Photobiological safety of lamps and lamp systems. Based on the measured output of the EvoEndo's light source, it was deemed to be below the limits of the exempt risk group.
## Summary of Previous Human Factors / Usability Engineering Testing:
EvoEndo conducted a two-stage usability study of representative intended users. The study assessed the overall usability of the EvoEndo Endoscopy System and the
{11}
K250900 Page 7 of 7
ability for users to perform critical tasks. Participants were asked to perform the entire workflow of the device across a benchtop model session and a representative animal model session with prior training provided. Tests were performed in a simulated clinic-office setting. Throughout the procedure, there were no significant device or user malfunctions or errors that would result in patient harm during an actual procedure.
Summary of Previous Animal and Clinical Data
Animal testing was not required for this device.
## 1.8 Shelf Life and Sterilization
The EvoEndo Endoscope is packaged individually, provided sterile, labeled for single use only, and sterilized using 100% ethylene oxide (EO) gas in a fixed chamber. Validation performed for the sterilization technique met all acceptance criteria. Functional testing confirmed that sterilization did not impact product safety or effectiveness. The predicate device (K213606) was cleared with a shelf life of 6 months. Additional testing done using the protocol described in K213606 was performed to extend the shelf life of the subject device to 14 months.
## 1.9 Previous Electrical Safety and EMC Testing:
The Electrical Safety (ES) and Electromagnetic Compatibility (EMC) testing conformed to IEC 60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-2:2014, and IEC 60601-2-18:2009. All acceptance criteria were met.
## 1.10 Previous Software Testing:
The firmware used in the EvoEndo Controller has been determined to be a moderate level of concern. The software in the controller consisting of 6 embedded firmware items is considered off-the-shelf software and, as such, is documented in accordance with the FDA Guidance "Off- The-Shelf Software Use in Medical Devices" (Sep 27, 2019).
## 1.11 Conclusion
There have been no technological changes between the subject and predicate device. Therefore, previous testing conducted on the predicate device applies to the subject device and supports substantial equivalence of the EvoEndo Endoscopy System subject device to the legally marketed predicate device.
To support the removal of the specific age from the indication, a literature review was conducted. The evaluation supported the removal of the specific age of 5 years and above. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that it can be concluded that the subject device is substantially equivalent with predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.