The Cure Catheter Insertion Kit is an intermittent urinary catheter kit indicated for the purpose of bladder drainage for males and females. The urinary catheter kit comes in a variety of configurations and sizes packaged sterile for single-use.
Device Story
Cure Catheter Insertion Kits are sterile, single-use convenience kits designed to facilitate intermittent urinary catheterization; used by patients or clinicians in clinical or home settings. Kits contain various components—underpads, BZK wipes, gloves, urine collection bags, PVP swabsticks, lubricating gel, sterile wipes, and a urinary catheter (or provided without a catheter). Components assist in maintaining a no-touch technique during insertion to prevent contamination. The device functions as a mechanical aid for bladder drainage; no electronic or software components are involved. Benefits include reduced risk of contamination during the catheterization process and improved convenience for the user.
Clinical Evidence
Bench testing only. Performance validated per ISO 20696:2018 (urinary catheters), ISO 11135-1:2014/AMD 1:2018 and ISO 10993-7:2008/AMD:2019 (sterilization), ISO 10993-1:2018 (biocompatibility), ISO 11607-1:2019/11607-2:2019 (sterile packaging), and ASTM F2096-11 (2019) (packaging integrity).
Technological Characteristics
Convenience kit for intermittent catheterization. Components: underpads, BZK wipes, gloves, urine collection bags, PVP swabsticks, lubricating gel, sterile wipes, and urinary catheter. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Shelf life: 3 years. Standards: ISO 20696:2018, ISO 11135-1, ISO 10993-1, ISO 11607-1/2, ASTM F2096-11.
Indications for Use
Indicated for bladder drainage in males and females requiring intermittent urinary catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 6, 2025
ConvaTec Limited
Alexandra Hyde
Regulatory Affairs Manager
First Avenue
Deeside Industrial Park
Deeside, CH5 2NU
UNITED KINGDOM
Re: K250891
Trade/Device Name: Cure Catheter Insertion Kit (K1); Cure Catheter Insertion Kit (K2);
Cure Catheter Insertion Kit (K2-90); Cure Catheter Insertion Kit (K3);
Cure Hydrophilic Catheter Kit (HM12UK);
Cure Hydrophilic Catheter Kit (HM14UK);
Cure Hydrophilic Catheter Kit (HM16UK);
Cure Pocket Catheter Kit (M14UK); Cure Catheter
Closed System Kit (CS8); Cure Catheter Closed System Kit (CS10);
Cure Catheter Closed System Kit (CS12); Cure Catheter
Closed System Kit (CS14); Cure Catheter Closed System Kit (CS14C);
Cure Catheter Closed System Kit (CS16)
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: FCM
Received: May 7, 2025
Dear Alexandra Hyde:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kits have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250891 - Alexandra Hyde
Page 2
in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K250891 - Alexandra Hyde
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen
-S
Jessica K. Nguyen, Ph.D
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250891
Device Name
The Cure Catheter insertion Kits (K1, K2, K2-90, K3, Cure Hydrophilic Catheter Kit (HMxxUK), Cure Pocket Catheter Kit (MxxUK), Cure Catheter Closed System Kit (CSxx)
Indications for Use (Describe)
The Cure Catheter insertion Kit is an intermittent urinary catheter kit indicated for the purpose of bladder drainage for males and females. The urinary catheter kit comes in a variety of configurations and sizes packaged sterile for single-use.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K250891
Page 1 of 3
# 510(k) Summary
| Date Prepared | 05^{th} June 2025 |
| --- | --- |
| Submitter | ConvaTec, Ltd.
First Avenue
Deeside Industrial Park
Deeside
Flintshire CH5 2NU |
| Contact person | Alexandra Hyde
Alexandra.hyde@convatec.com |
| Name of Device | Cure Catheter Insertion Kit |
| Common Name | Tray, catheterization, sterile urethral, with or without catheter (kit) |
| Regulation Name | Urological catheter and accessories |
| Regulation Number | 21 CFR 876.5130 |
| Product Code | FCM |
| Classification | II |
| Primary Predicate Device | Cure Catheter Closed System (K230400) |
| Purpose of Submission | Device modification |
| Device Description | The Cure Catheter Insertion Kits (K1, K2, K2-90, K3), Cure Hydrophilic Catheter Kit (HMxxUK), Cure Pocket Catheter Kit (MxxUK), Cure Catheter Closed System Kit (CSxx)) contain sterile products used during intermittent urinary catheterization. Two types of kit are provided, one where a urinary catheter is included and one without a catheter. The components allow users to select one or more of the products to help prevent contamination of the environment and the user and to maintain a no-touch technique during the insertion of the catheter. Catheters are inserted through the urethra and indicated for the purpose of bladder drainage. |
| Intended Use | The Cure Catheter Insertion Kit is an intermittent urinary catheter kit indicated for the purpose of bladder drainage for males and females. |
| Indications for Use | The Cure Catheter Insertion Kit is an intermittent urinary catheter kit indicated for the purpose of bladder drainage for males and females. The urinary catheter kit comes in a variety of configurations and sizes packaged sterile for single-use. |
| Performance Data | Performance testing for Cure Catheter Insertion Kit was conducted on the subject devices or representative devices per applicable sections of voluntary and FDA consensus standards:
• Sterilization validation per AAMI/ISO 11135-1:2014/AMD 1:2018 and ISO 10993-7:2008/AMD:2019
• Biocompatibility testing according to ISO 10993-1:2018 and FDA Guidance “Use of International Standard ISO 10993-1”
• Sterile packaging in accordance with ISO 11607-1:2019 and ISO 11607-2:2019
• Real time aged shelf-life testing according to ISO 11607-1:2006 with justification to version 2019
• Packaging integrity testing according to ASTM F2096-11 (2019)
• Urinary catheter testing according to ISO 20696:2018 |
| Technological Comparison: | |
{5}
K250891
Page 2 of 3
| | Subject Device | Predicate Device | Substantial Equivalence |
| --- | --- | --- | --- |
| Device | Cure Catheter Insertion Kit | Cure Catheter Closed System | N/A |
| Intended Use | The Cure Catheter Insertion Kit is an intermittent urinary catheter kit indicated for the purpose of bladder drainage for males and females. | The Cure Catheter Closed System is an intermittent urinary catheter indicated for the purpose of bladder drainage for males and females. | Equivalent to predicate. |
| Classification Regulation | 876.5130 – Urological catheter and accessories | 876.5130 – Urological catheter and accessories | Equivalent to predicate. |
| Product Code | FCM | KOD | N/A |
| Technological Features | Convenience kit to aid insertion, components include: Underpads, BZK wipes, gloves, urine collection bags, PVP swabsticks, lubricating gel, sterile wipes and a urinary catheter. | Convenience kit to aid insertion, components include: Underpads, BZK wipes, gloves, urine collection bags, PVP swabsticks, lubricating gel, sterile wipes and a urinary catheter. | Equivalent to predicate. |
| Kit Configurations | Cure Catheter Insertion Kits – contain: underpads, BZK wipes, gloves, collection bags, lubricating gel (K3 does not contain any lubricating gel). Kits provided over-the-counter (OTC),
Cure Hydrophilic Catheter Kit – contains: underpads, BZK wipes, gloves, urine collection bags, and a Cure Hydrophilic Catheter. Kits provided prescription only,
Cure Pocket Catheter Kit – contains: underpads, BZK wipes, gloves, urine collection bags, and a Cure Pocket Catheter. Kits provided prescription only, | Cure Catheter Closed System Kit – contains: underpads, BZK wipes, gloves, urine collection bags, PVP swabsticks, lubricating gel, sterile wipes and Cure Closed System Catheter. | Substantially equivalent. |
{6}
K250891
Page 3 of 3
| | | | Cure Catheter Closed System Kit – contains: underpads, BZK wipes, gloves, urine collection bags, PVP swabsticks, lubricating gel, sterile wipes and Cure Closed System Catheter. Kits provided prescription only. | | |
| --- | --- | --- | --- | --- | --- |
| | | Biocompatibility | All tests were performed in accordance with US FDA General Program Memorandum #G95-1 and Part- 10993-1 of the International Standard Organisation (ISO) Standard (Biological Evaluation of Medical Devices). The test articles were considered biocompatible under the conditions tested. | All tests were performed in accordance with US FDA General Program Memorandum #G95-1 and Part- 10993-1 of the International Standard Organisation (ISO) Standard (Biological Evaluation of Medical Devices). The test articles were considered biocompatible under the conditions tested. | Equivalent to predicate. |
| | | Sterilization | EO, SAL^{-6} | EO, SAL^{-6} | Equivalent to predicate. |
| | | Shelf Life | 3 years | 3 years | Equivalent to predicate. |
| Conclusion | The intended use, indications, technological characteristics and performance data between the subject device, Cure Catheter Insertion Kits and the predicate Cure Catheter Closed System (K230400) has been established as being substantially equivalent. | | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.