Ceftobiprole BPR 5 µg Disc

K250885 · Liofilchem, Inc. · JTN · Jun 23, 2025 · Microbiology

Device Facts

Record IDK250885
Device NameCeftobiprole BPR 5 µg Disc
ApplicantLiofilchem, Inc.
Product CodeJTN · Microbiology
Decision DateJun 23, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2
AttributesPCCP

Indications for Use

Liofilchem Antibiotic Discs are a semi-quantitative agar diffusion method for in vitro determination of antimicrobial susceptibility of clinical isolates tested on agar media after overnight incubation. The Ceftobiprole BPR 5 μg Disc is intended to determine susceptibility of Enterobacterales and Staphylococcus aureus to ceftobiprole, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC). Ceftobiprole at concentrations of 5 μg should be interpreted at 16-18 hours of incubation. The Ceftobiprole BPR 5 μg Disc demonstrated acceptable performance with the following organisms: Enterobacterales (Escherichia coli and Klebsiella pneumoniae) Staphylococcus aureus (includes methicillin resistant isolates)

Device Story

Device consists of high-quality paper discs impregnated with 5 µg ceftobiprole; used for semi-quantitative antimicrobial susceptibility testing (AST). Performed in clinical microbiology laboratories; technicians apply discs to inoculated Mueller Hinton agar plates. After 16-20 hours incubation, inhibition zone diameters (mm) are measured manually. Results are categorized as Susceptible, Intermediate, or Resistant based on FDA-recognized interpretive criteria. Output assists clinicians in selecting appropriate antibiotic therapy for bacterial infections. Device is single-use.

Clinical Evidence

No clinical data provided; performance was established via bench testing demonstrating acceptable performance for the specified organisms (Enterobacterales and S. aureus) using the agar diffusion method.

Technological Characteristics

Semi-quantitative agar diffusion disc; 5 µg ceftobiprole concentration; standard laboratory agar media; 16-18 hour incubation requirement.

Indications for Use

Indicated for in vitro antimicrobial susceptibility testing of clinical isolates of Enterobacterales (e.g., E. coli, K. pneumoniae) and Staphylococcus aureus (including MRSA) to ceftobiprole using agar diffusion methods.

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K250885 B Applicant Liofilchem Inc. C Proprietary and Established Names Ceftobiprole BPR 5 µg Disc D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JTN | Class II | 21 CFR 866.1620 - Antimicrobial Susceptibility Test Disc | MI - Microbiology | ## II Submission/Device Overview: ### A Purpose for Submission: In this Traditional 510(k) submission, Liofilchem Inc. seeks the following: 1. Obtain a substantial equivalence determination for ceftobiprole antimicrobial susceptibility test disc. 2. Establish a Pre-Determined Change Control Plan (PCCP) to address future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage. ### B Measurand: Ceftobiprole 5 µg ### C Type of Test: Antimicrobial Susceptibility Test Disc Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} K250885 - Page 2 of 9 ## III Intended Use/Indications for Use: ### A Intended Use(s): See Indications for Use below. ### B Indication(s) for Use: Liofilchem Antibiotic Discs are a semi-quantitative agar diffusion method for *in vitro* determination of antimicrobial susceptibility of clinical isolates tested on agar media after overnight incubation. The Ceftobiprole BPR 5 µg Disc is intended to determine susceptibility of Enterobacterales and *Staphylococcus aureus* to ceftobiprole, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC). Ceftobiprole at concentrations of 5 µg should be interpreted at 16-18 hours of incubation. The Ceftobiprole BPR 5 µg Disc demonstrated acceptable performance with the following organisms: - Enterobacterales (*Escherichia coli* and *Klebsiella pneumoniae*) - *Staphylococcus aureus* (includes methicillin resistant isolates) ### C Special Conditions for Use Statement(s): Rx - For Prescription Use Only Due to the insufficient number of resistant *Staphylococcus aureus* isolates evaluated, the following limitation was included in the device labeling: The ability of the Antibiotic Disc to detect resistance or non-susceptibility to antimicrobials as shown below is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory. Ceftobiprole: *Staphylococcus aureus*. ### D Special Instrument Requirements: Not applicable ## IV Device/System Characteristics: ### A Device Description: The Liofilchem Ceftobiprole BPR 5 µg Disc is made of special high-quality paper impregnated with a predefined concentration of 5 µg of ceftobiprole. Both sides of the disc are labeled with "BPR 5", designating the agent and the drug content. Interpretive category results are determined by determining the inhibition zone diameter (mm) after incubation on an inoculated agar surface. The Liofilchem Ceftobiprole BPR 5 µg Disc is single use only. {3} K250885 - Page 3 of 9 B Principle of Operation: Liofilchem Antibiotic Discs are made of specialized high-quality paper impregnated with a predefined concentration of an antimicrobial agent. The disc is removed from its packaging aseptically and applied to an inoculated agar surface, where the antimicrobial agent diffuses into the agar. After 16-20 hours incubation, the inhibition zone diameter is measured to the nearest millimeter with zone edges read at the point of complete inhibition. To categorize the result, the recognized zone size interpretive criteria for the antimicrobial agent being tested is used. V Substantial Equivalence Information: A Predicate Device Name(s): BD BBL Sensi-Disc Cefiderocol 30ug (FDC-30) B Predicate 510(k) Number(s): K221826 C Comparison with Predicate(s): | Device & Predicate Device(s): | Device: K250885 | Predicate: K221826 | | --- | --- | --- | | Device Trade Name | Ceftobiprole BPR 5 μg Disc | BD BBL Sensi-Disc Cefiderocol 30μg (FDC-30) | | General Device Characteristic Similarities | | | | Test Method | Antimicrobial susceptibility testing using paper disks impregnated with an antimicrobial agent | Same | | Plate Media | Mueller Hinton Agar | Same | | Disc Material | High quality paper impregnated with a predefined concentration of antimicrobial agent | Same | | Inoculation | Isolated colonies from a culture in a suspension equivalent to 0.5 McFarland. Inoculum is applied to agar with swab manually or with rotation plate. | Same | | Reading Method | Measurement of inhibition zone diameter in mm | Same | | Result Interpretation | Categorical interpretation (S/I/R) as determined by the measured zone diameters according to established interpretive criteria for the drug | Same | {4} K250885 - Page 4 of 9 | Device & Predicate Device(s): | Device: K250885 | Predicate: K221826 | | --- | --- | --- | | General Device Characteristic Differences | | | | Intended Use/Indications for Use | Semi-quantitative susceptibility to antimicrobial agents against specified gram-negative and gram-positive organisms | Semi-quantitative susceptibility to antimicrobial agents against specified gram-negative organisms | | Antimicrobial Agent (Concentration) | Ceftobiprole (5μg) | Cefiderocol (30μg) | VI Standards/Guidance Documents Referenced: - CLSI M100-35th ed. Performance Standards for Antimicrobial Susceptibility Testing (January 2025). - CLSI M07-11th ed. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically (January 2018). VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Reproducibility testing of the Liofilchem Ceftobiprole BPR 5 μg Disc was conducted at one internal site using 20 organisms, including 10 gram-negative organisms and 10 gram-positive organisms. Each isolate was tested in triplicate on three separate days across two disc lots using one lot of media (Mueller Hinton agar media). Each test was visually read by three independent readers with results blinded, resulting in 540 data points for evaluation (20 isolates x 3 replicates x 3 users x 3 days x 1 site x 1 media lot = 540 data points). The reproducibility study included the following species: ten Staphylococcus aureus and ten Enterobacterales (4 Enterobacter cloacae, 4 Escherichia coli, and 2 Klebsiella pneumoniae). Reproducibility was calculated as the percent of results which were within ±3 mm difference in zone diameter comparing test results with the modal zone diameter value. The reproducibility performance between disc lots and across readers was 100% and meets the acceptance criteria. Results are shown in Table 1. Table 1: Reproducibility Summary for Ceftobiprole (5 μg) Disc | Between Disc Lots | | | Across Readers | | | | | --- | --- | --- | --- | --- | --- | --- | | Lot #1 | Lot #2 | All Lots | Reader #1 | Reader #2 | Reader #3 | All Readers | | 100% (270/270) | 100% (270/270) | 100% (540/540) | 100% (180/180) | 100% (180/180) | 100% (180/180) | 100% (540/540) | {5} 2. Linearity: Not applicable. 3. Analytical Specificity/Interference: Not applicable. 4. Assay Reportable Range: Not applicable. 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): # Quality Control (QC) Testing: The CLSI-recommended quality control (QC) isolates Escherichia coli ATCC 25922 and Staphylococcus aureus ATCC 25923 were tested using two Ceftobiprole BPR $5\mu \mathrm{g}$ Disc lots and at least 20 replicates per lot per reader. Each test was visually read by three independent readers, resulting in 240 disc data points. Results were compared to reference broth microdilution. The Liofilchem Ceftobiprole BPR $5\mu \mathrm{g}$ Disc QC performance was $>95\%$ and is acceptable. The performance is summarized in Table 2. Table 2: Quality Control Performance of Ceftobiprole BPR $5\mu \mathrm{g}$ Disc | QC Organism | Zone Diameter in millimeter (mm) | | | | --- | --- | --- | --- | | | Range | Liofilchem Lot 1 | Liofilchem Lot 2 | | Escherichia coli ATCC 25922 | 24 | | | | | 25 | | | | | 26 | 18 | 11 | | | 27 | 22 | 28 | | | 28 | 12 | 11 | | | 29 | 8 | 7 | | | 30 | | 3 | | | 31 | | | | | 32 | | | | Staphylococcus aureus ATCC 25923 | 19 | | | | | 20 | | | | | 21 | | | | | 22 | 13 | 10 | | | 23 | 27 | 27 | | | 24 | 20 | 15 | | | 25 | | 8 | | | 26 | | | | | 27 | | | | | 28 | | | ATCC=American Type Culture Collection K250885 - Page 5 of 9 {6} Broth micro dilution (BMD) for QC isolates was performed as per CLSI M07 on every day of clinical study to provide assurance of the data quality obtained from the reference method. The QC performance for the reference BMD method was acceptable at ≥95%. The performance is shown in Table 3 below. Table 3. Broth Microdilution Quality Control Summary for Ceftobiprole | QC Organism | MIC (μg/mL) | Reference Results | | --- | --- | --- | | Escherichia coli ATCC 25922 Expected range: 0.03-0.12 μg/mL | ≤0.016 | | | | 0.032 | 13 | | | 0.064 | 47 | | | 0.125 | | | | ≥0.25 | | | Staphylococcus aureus ATCC 29213 Expected range: 0.12-1 μg/mL | ≤0.064 | | | | 0.125 | | | | 0.25 | 46 | | | 0.5 | 14 | | | 1 | | | | 2 | | ATCC=American Type Culture Collection ## Inoculum Density Check: Colony counts were conducted for all QC and reproducibility isolates, as well as 10% of clinical isolates. All counts were within the expected range. ## Stability A real-time stability study was conducted with the device stored in the storage conditions recommended in the package insert. Three lots of disks were tested throughout the 36-month claimed shelf life, and microbiological and chemical performance results were within the expected range which is acceptable. Stability and storage information is provided in the package insert. 6. Detection Limit: Not applicable. 7. Assay Cut-Off: Not applicable. ## B Comparison Studies: 1. Method Comparison with Predicate Device: The qualitative Liofilchem Ceftobiprole BPR 5 μg Disc results were compared to the qualitative categorical interpretation (S/I/R) based on the minimum inhibitory concentration (MIC) values obtained from reference broth microdilution (BMD) to assess the categorical agreement (CA) and error rates. The study was conducted at one internal testing site. Three K250885 - Page 6 of 9 {7} independent operators participated in reading test results with isolates evenly distributed to mimic multiple testing sites. Testing was performed utilizing Mueller Hinton agar media. The reference BMD method was performed per CLSI M07 guidelines. Isolated colonies from an overnight agar plate were suspended in saline to achieve a 0.5 McFarland standard turbidity (approximately $10^{8}$ CFU/mL). Testing conditions consisted of incubation of the inoculated Mueller Hinton agar plate in an inverted position at $35^{\circ}\mathrm{C} \pm 2^{\circ}\mathrm{C}$ for 16-18 hours. At the end of the appropriate incubation, the Ceftobiprole BPR $5\mu \mathrm{g}$ Disc categorical interpretations as determined by zone edges at the point of complete inhibition (inhibition zone) were compared to categorical interpretations as determined by MIC results obtained with the CLSI reference broth microdilution method. # Clinical: Clinical testing was performed using a total of 557 clinical isolates including 282 Enterobacterales (186 isolates of Klebsiella pneumoniae, 63 isolates of Escherichia coli, and 33 isolates of Enterobacter cloacae), and 275 Staphylococcus aureus. The clinical testing included 216 $(38.8\%)$ contemporary isolates (isolated no longer than 6 months prior to testing) and 341 $(61.2\%)$ stock isolates (isolated more than 6 months prior to testing). # Challenge: Challenge testing was performed at one site using a total of 85 challenge isolates including 35 Enterobacterales (18 isolates of Klebsiella pneumoniae, 11 isolates of Escherichia coli, 6 isolates of Enterobacter cloacae), and 50 Staphylococcus aureus. Performance results for the total 642 clinical and challenge isolates are shown in Table 4. Table 4: Performance of the Ceftobiprole BPR $5\mu \mathrm{g}$ Disc vs. BMD | | Total | CA# | CA% | #S | #I | #R | MIN | MAJ | VMJ | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Enterobacterales [Breakpoints (mm): ≥23 (S), 21-22 (I), ≤20(R)] | | | | | | | | | | | Clinical | 282 | 278 | 98.6 | 83 | 2 | 197 | 3 | 1 | 0 | | Challenge | 35 | 35 | 100 | 3 | 0 | 32 | 0 | 0 | 0 | | Combined | 317 | 313 | 98.7 | 86 | 2 | 229 | 3 | 1 | 0 | | Staphylococcus aureus [Breakpoints (mm), ≥16 (S), 14-15 (I), ≤13 (R)] | | | | | | | | | | | Clinical | 275 | 262 | 95.3 | 263 | 12 | 0 | 7 | 0 | 0 | | Challenge | 50 | 46 | 92.0 | 14 | 36 | 0 | 4 | 0 | 0 | | Combined | 325 | 308 | 94.8 | 277 | 48 | 0 | 11 | 0 | 0 | CA - Category Agreement MIN - minor errors S - Susceptible isolates MAJ - major errors I - Intermediate isolates VMJ - very major errors R - Resistant isolates Category Agreement (CA) is when the device result interpretation agrees exactly with the comparator result interpretation. The performance of the Ceftobiprole BPR $5\mu \mathrm{g}$ Disc as compared to the reference method MIC for Enterobacterales (Table 4) is acceptable with $98.7\%$ CA. There were three minor errors and no very major errors. There was one major error ( $1.1\%$ , 1/88 susceptible isolates), which is acceptable. K250885 - Page 7 of 9 {8} The performance of the Ceftobiprole BPR 5 µg Disc as compared to the reference method for S. aureus (Table 4) is acceptable with 94.8% CA. There were 11 minor errors and no major or very major errors. Testing/Reporting Non-Indicated Species: As required under 511A(b)(2)(C)(ii)(I) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the 'Precautions' section in the device labeling to address testing of non-indicated species: Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved. ## Resistance Isolates: A total of 317 clinical and challenge Enterobacterales isolates and 325 clinical and challenge S. aureus isolates were tested when the Ceftobiprole BPR 5 µg Disc was compared to the interpretive category based on MIC values determined by the reference method. However, an insufficient number of resistant isolates were available for testing for S. aureus. To address the lack of resistant strains encountered during the clinical evaluation, the following limitation was applied to ceftobiprole testing in the device labeling: The ability of the Antibiotic Disc to detect resistance or non-susceptibility to antimicrobials as shown below is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory. Ceftobiprole: Staphylococcus aureus. 2. Matrix Comparison: Not applicable. ## C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable. K250885 - Page 8 of 9 {9} D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: Table 5. FDA-Recognized Interpretive Criteria for Ceftobiprole | Organism | Disk Diffusion (Zone diameters in mm)a | | | | --- | --- | --- | --- | | | Susceptible | Intermediate | Resistant | | Staphylococcus aureus | ≥16 | 14-15 | ≤13 | | Enterobacteralesb | ≥23 | 21-22 | ≤20 | aAccording to the FDA STIC Webpage VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/fda-recognized-antimicrobial-susceptibility-test-interpretive-criteria). The protocol outlined the specific procedures and acceptance criteria that Liofilchem Inc. intends to use to evaluate the Ceftobiprole BPR 5 µg Disc when revised breakpoints for ceftobiprole are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, Liofilchem Inc. will update the Ceftobiprole BPR 5 µg Disc device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation. K250885 - Page 9 of 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...