K250815 · Hubly, Inc. · HBE · Nov 6, 2025 · Neurology
Device Facts
Record ID
K250815
Device Name
Hubly Drill (H100)
Applicant
Hubly, Inc.
Product Code
HBE · Neurology
Decision Date
Nov 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hubly Auto-Stop Drill (H100, H200, H201, H0270, H0271, H0272) is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone. The Hubly Auto-Stop Drill (H100) is a single-use, sterile, disposable device intended for use on adult patients during laminectomy and laminotomy procedures for drilling holes through the lamina.
Device Story
Battery-powered, single-use, sterile, disposable drill; used in OR, ER, or bedside for cranial burr holes or laminectomy/laminotomy. Physician-operated; one-handed trigger activation. Features tapered stainless steel bit for mechanical plunge prevention; LED indicator (green: sufficient force; red: auto-stop triggered). Auto-stop feature detects bone breakthrough to halt rotation. Provides visual depth guidance via 5mm and 10mm markings. Facilitates bone excision; enhances safety by preventing dural damage during drilling.
Clinical Evidence
No clinical studies performed. Evidence includes bench testing (battery performance, depth penetration, trigger performance, EMC/electrical safety), cadaver studies (safety validation for laminectomy/laminotomy), and a summative usability study. Literature review characterized lamina/dura properties. Bench testing confirmed performance on sawbones and bovine bone models was equivalent to cranial bone performance.
Technological Characteristics
Battery-powered, single-use, disposable drill. Bit material: 316 Stainless Steel (ASTM F899). Features: mechanical plunge prevention (tapered bit), auto-stop sensing, LED status indicators. Standards: IEC 60601-1 (safety), IEC 60601-1-2/60601-4-2 (EMC). Software: Class B (IEC 62304), Basic Documentation level.
Indications for Use
Indicated for adult patients undergoing neurosurgical procedures for drilling cranial bone, and specifically for laminectomy and laminotomy procedures for drilling holes through the lamina.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 6, 2025
Hubly, Inc.
Julie Byars
Quality Lead
815 Ogden Ave.
Lisle, Illinois 60532
Re: K250815
Trade/Device Name: Hubly Drill (H100)
Regulation Number: 21 CFR 882.4310
Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories
Regulatory Class: Class II
Product Code: HBE
Dated: October 3, 2025
Received: October 8, 2025
Dear Julie Byars:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}
K250815 - Julie Byars
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250815 - Julie Byars
Page 3
Sincerely,
Adam D. Pierce -S
Digitally signed by Adam D. Pierce -S
Date: 2025.11.06
17:02:01 -05'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250815
Device Name
Hubly Auto-Stop Drill (H100)
Indications for Use (Describe)
The Hubly Auto-Stop Drill (H100) is a single-use, sterile, disposable device intended for use on adult patients during laminectomy and laminotomy procedures for drilling holes through the lamina.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) Summary
K250815
A. Device Information:
| Category | Comments |
| --- | --- |
| Sponsor: | Hubly, Inc.
Casey Qadir
815 Ogden Ave.
Lisle, IL 60532
(844) 482-5942 |
| Correspondent Contact Information: | Julie Byars
815 Ogden Ave.
Lisle, IL 60532
901-848-4010 |
| Device Common Name: | Powered Cranial Drill |
| Device Regulation & Name: | 21 CFR 882.4310 - Powered simple cranial drills, burrs, trephines and their accessories |
| Classification & Product Code:
510(k) Number: | Class II, HBE
K250815 |
| Device Proprietary Name: | Hubly Auto-Stop Drill |
Predicate Device Information:
| Predicate Device: | Hubly Electric Drill |
| --- | --- |
| Predicate Device Manufacturer: | Hubly, Inc |
| Predicate Device Common Name: | Powered Cranial Drill |
| Predicate Device Premarket Notification #: | K230619 |
| Predicate Device Classification & Name: | 21 CFR 882.4310 - Powered simple cranial drills, burrs, trephines and their accessories |
| Predicate Device Classification & Product Code: | Class II, HBE |
B. Date Summary Prepared
November 6, 2025
C. Description of Device
The Hubly Drill is a battery powered drill. It is single-use disposable and provided ethylene oxide sterilized. The drill is designed to create optimal holes in bone. For trephination procedures the Hubly Drill is intended to drill cranial bone and create burr holes in the emergency room, at the bedside, or in the operating room. For laminectomy and laminotomy procedures the drill is intended to be used in the operating room to drill holes through the lamina to facilitate excision of the lamina.
Page 1 of 5
{5}
510(k) Summary
K250815
The drill can be used one-handed using either hand. It is trigger-activated after removal of the battery pull tab. The drill has a single speed and turns off when the trigger is released. The drill bit has depth indicators at 5 and 10mm depth, which the physician may use to visually gauge depth of penetration while drilling. The drill features mechanical plunge prevention with a tapered stainless steel drill bit. The drill may be reactivated any number of times using the trigger if the physician desires.
The drill also has an auto-stop feature which detects when the bit breaks through the bone and immediately stops the drill. The device has an LED indicator which indicates to the user (green) when they are applying enough force and (red) when the drill stops.
# D. Indications for Use
The Hubly Auto-Stop Drill (H100, H200, H201, H0270, H0271, H0272) is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
The Hubly Auto-Stop Drill (H100) is a single-use, sterile, disposable device intended for use on adult patients during laminectomy and laminotomy procedures for drilling holes through the lamina.
# E. Comparison of Technological Characteristics
| | Application Device
Hubly Drill – K250815 | Predicate Device
Hubly Electric Drill – K230619 | Impact on Substantial
Equivalence |
| --- | --- | --- | --- |
| Company | Hubly, Inc. | Hubly, Inc. | N/A |
| Regulation Number | 21 CFR 882.4310 | 21 CFR 882.4310 | Identical |
| Product Code | HBE – Class II | HBE – Class II | Identical |
| Intended Use & Indications for use | The Hubly Auto-Stop Drill (H100, H200, H201, H0270, H0271, H0272) is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
The Hubly Auto-Stop Drill (H100) is a single-use, sterile, disposable device intended for use on adult patients during laminectomy and laminotomy procedures for drilling holes through the lamina. | The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone. | Minor Impact – Addition of laminectomy and laminotomy indication.
No impact on intended use, safety or effectiveness. |
| Technology – THE PROPOSED DRILL IS THE SAME AS THE PREDICATE, WITH AN ADDED INDICATION. | | | |
{6}
510(k) Summary
K250815
| Power Source | Battery Power (removable) | Battery Power (removable) | Identical |
| --- | --- | --- | --- |
| Automatic stop | Auto-stop | Auto-stop | Identical |
| Mechanical plunge prevention | Tapered bit | Tapered bit | Identical |
| Bit Material | Stainless Steel (316) | Stainless Steel (316) | Identical |
| Drill activation | Trigger – On/Off | Trigger – On/Off | Identical |
| Drill rotation | Forward | Forward | Identical |
| Flute characteristics | 2 Straight flutes | 2 Straight flutes | Identical |
F. Discussion of Non-Clinical Performance Testing
The following performance data were provided in support of the substantial equivalence determination.
BIOCOMPATIBILITY TESTING
The drill bit is the only patient-contacting material used in the Hubly device. 316 Stainless Steel certified to ASTM F899 was selected for its known biocompatibility with human tissues and fluids. A biological risk assessment was conducted for the device. It considers raw material, manufacturing processing, and full endpoint biocompatibility testing performed, and determines full compliance of the device with ISO 10993-1.
ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY (EMC)
Electrical safety and EMC testing were conducted on the Hubly device. The system complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 and 60601-4-2 standards for EMC.
SOFTWARE VERIFICATION TESTING
Software verification was conducted on the firmware responsible for the Hubly Drill's automatic stopping feature. The device has an equivalent expected safety profile compared to the standard of care without utilizing software mitigations. The software implements additional safety features compared to the current standard of care.
The Hubly Drill software is initially classified as a Class B software system according to IEC 62304 Section 4.3a based on the possible hazards to which the software component can contribute. For this reason, the resulting possible harm due to failure of software is considered non-serious. Based on the FDA Guidance document "Guidance for the Content of Premarket Submissions for Device Software Functions" the documentation level for the Hubly Drilling System has been determined to be "Basic Documentation".
BENCH TESTING
The following benchtop performance tests were conducted. Results demonstrated that the drill meets all design specifications and requirements.
- Performance – battery/continuous use
Page 3 of 5
{7}
510(k) Summary
K250815
- Depth of Penetration, bone thickness variation
- Bit securement
- Trigger performance
- Basic Safety and Essential Performance
- EMC Testing (Basic Safety, Essential Performance, Electromagnetic disturbances)
Additional performance testing was completed using sawbones absolute sheets as representative lamina bone to demonstrate that the drill performance using cranial bone is equivalent to the drill performance on sawbones absolute sheets. Drill penetration depth was evaluated on the sawbones sheets with thicknesses from 2mm-5mm. A representative dural model was evaluated for damage at each of the thicknesses. Testing was performed to characterize the force needed to activate the auto-stop feature during drilling as well as the magnitude of the current spike needed to trigger the stopping signal.
Testing was performed to characterize the performance of the Hubly Drills ability to drill into the lamina, and to evaluate the thickness and range of angles required for drilling.
Alternate bone testing was performed to verify that there was not a significant difference in the Hubly Drilling System penetration depth performance between bovine and cranial bone.
A cadaver study was conducted to validate that the Hubly Drilling System is as safe as the standard of care for the laminectomy and laminotomy indication.
A second cadaver study was conducted to validate that the Hubly Drilling System is as safe as the standard of care and to further characterize the safety of the drill performance in the lamina.
A summative usability study was conducted to confirm that users can effectively perform critical tasks for the Hubly Drilling System for laminectomy and laminotomy procedures and that the use-related risks are at an acceptable level.
A literature review was conducted to characterize the material properties and disease state conditions of the human lamina and spinal dura. The report identified the density, thickness, hardness, elastic modulus, and angles of human lamina.
## G. Discussion of Clinical Performance Testing
ANIMAL STUDY
Animal testing was not performed.
CLINICAL STUDIES
Clinical testing was not performed.
## H. Conclusion
Page 4 of 5
{8}
510(k) Summary
K250815
The performance and design validation testing conducted on the Hubly Drill demonstrates that it performs comparably to the predicate device for the added indication for laminectomy and laminotomy procedures. All hardware and software verification and validation demonstrate that the Hubly Drill will perform as safe and effective as the predicate.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.